200pH/CR pH, ORP, Conductivity & Resistivity Instrument Instruction Manual
Manual64 pagesPDF 443 KB
Instruction manual for the Thornton 200pH/CR two-channel pH, ORP, conductivity and resistivity instrument: installation, setup and calibration.
Filed with · 1 model
Also filed with these models on the old site
Summary
What is in it
The instruction manual for the Thornton 200pH/CR, an analytical and process control instrument that measures pH, ORP, conductivity and resistivity on two channels. It covers unpacking, panel, wall and pipe mounting, power, relay, output and sensor wiring, the display, keypad and menus, measurement types and temperature compensation, setpoints, relays, analog outputs, meter and sensor calibration, password lockout, printer output, troubleshooting, service, technical illustrations and specifications.
Its parts
- Getting Started
- Installing the 200pH/CR
- Using the 200pH/CR
- Making Measurements
- Using Setpoints
- Using Relays
- Using Analog Outputs
- Meter Calibration
- Sensor Calibration
- Security/Lockout
- Other Functions
- Troubleshooting
- Service
- Technical Illustrations
- Specifications
From the manual
Key facts from this manual
Each line names its page in the PDF.
Instruction manual (part number 84383) for the Thornton 200pH/CR, a two-channel instrument for pH, ORP, conductivity and resistivity. The site files it under the 200pH, but the manual itself covers the 200pH/CR models 6520-1, 6520-2, 6522-1, 6522-2, 6542-1 and 6542-2. It explains mounting, power, relay, output and sensor wiring, diagnostics, troubleshooting, fuse replacement and spare parts, and gives the specifications.
- Device
- 200pH/CR: two-channel analyzer reading pH, ORP, two- and four-electrode conductivity or resistivity, two signals per sensor p. 7
- Measurements
- pH, ORP, resistivity, conductivity, °C, °F, total dissolved solids, %rejection, difference, ratio, %HCl, %NaOH, %H2SO4 p. 7
- Ranges
- -1 to 15 pH, -1400 to +1500 mV, -40° to 200°C; 0.1 constant 2-E cell 0.02 to 1200 µS/cm; 4-E cell 5 to 500,000 µS/cm p. 61
- Accuracy
- Conductivity/resistivity ±0.5% of reading up to 18 Mohm-cm, ±0.25°C; pH/ORP ±0.03 pH, ±2 mV, ±0.3°C p. 61
- Relays
- 2 SPDT relays, 5 amp resistive up to 30 VDC or 250 VAC; optional 2 AC-only solid state, 1.5 amp at 250 VAC, 10 mA minimum; four setpoints p. 61
- Outputs
- Two optional powered, isolated 4-20 mA outputs (0-20 mA possible), 500 ohm maximum, ±0.05 mA; RS232 to 50 feet or RS422 to 4,000 feet, up to 19.2 k baud p. 61
- Power
- 90-130 VAC or 180-250 VAC, 50-60Hz, 12 Watts maximum, or 12-30 VDC at 300 mA steady state (600 mA inrush at 24 VDC) p. 62
- Model numbers
- 6520-1 and 6520-2: 2 SPDT relays, no analog outputs; 6522-1 and 6522-2: 2 SPDT, 2 outputs; 6542-1 and 6542-2: 2 SPDT plus 2 solid state, 2 outputs; -1 is 110 VAC, -2 is 220 VAC, all also 24 VDC p. 62
- Fuse
- 115 VAC units: 1/8 amp time-delayed (Type 2AG, Littlefuse #228.125); 230 VAC units: 1/16 amp (Littlefuse #218.063) p. 49, 50
- Display and enclosure
- 16 character backlit LCD, 11 tactile keys, ABS-PC alloy, IP65 front panel seal, 3.78 x 7.56 in (96 x 192 mm) DIN cutout, 1.9 lbs (0.9 kg) p. 61
- Operating temperature
- -10° to 55°C (14 to 131°F); storage -40° to 70°C; 0 to 95% RH non-condensing p. 61
- Sensor patch cords
- 1XXX-67 series cords; 200 feet (61 m) maximum for 2-E, pH and ORP sensors, 50 feet (15 m) for 4-E sensors p. 11, 61
- Terminal wire size
- 26 AWG to 14 AWG solid or stranded, up to 12 AWG stranded only p. 10
- Spare parts
- Display assembly 06235; panel mounting kit 02181; 9-pin terminal strip 22617; 12-pin 22619; 14-pin 22613 p. 50
What this manual walks you through
- Step-by-step installation and setup sequence p. 8
- Panel, wall and pipe mounting p. 9
- Wiring line power and relays, and setting 115 VAC, 230 VAC or 24 VDC input p. 9, 10
- Wiring the serial port, analog outputs and sensor patch cords p. 11, 12
- Running the built-in self tests p. 47
- Troubleshooting blank display, wrong readings and serial problems p. 48
- Replacing the fuse p. 49
- Shortening a sensor patch cord p. 49
Good to know
- Relays always de-energize on loss of power, whatever their state setting, so control systems need fail-safe logic. p. 2
- Two incompatible sensors may share one instrument only if plastic tanks or piping isolate them electrically; otherwise use two instruments. p. 7
- If the input voltage jumpers are changed, the unit must be relabeled and the fuse changed to the proper rating. p. 10
- The +24 VDC input is not fused inside the meter, and AC connections must be removed when it is used. p. 10
- The ROM self test does not work in software versions up to 3.3; ignore its failure messages. p. 47
Questions this manual answers
What do asterisks on the 200pH/CR display mean?
The measurement could not be made or is over-range, for example from a disconnected or miswired sensor, no sample, or a failed sensor. p. 47
How do I run the 200pH/CR on 24 VDC?
Remove the AC connections, move jumper W6 to the 24V position, and wire the supply to PS+ and PS- on terminal block TB3. p. 10
What fuse does the 200pH/CR use?
A 1/8 amp time-delayed fuse in 115 VAC units and a 1/16 amp time-delayed fuse in 230 VAC units, on the circuit board near the transformer. p. 49
Which 200pH/CR model has four relays?
The 6542-1 (110 VAC) and 6542-2 (220 VAC), with 2 SPDT and 2 AC-only solid state relays. p. 62
How do I clear the Check Setup message on a 200pH/CR?
Enter and exit any menu. The message appears when the meter finds its setup data changed or corrupted. p. 48
The whole text · 7 parts