Specification sheet for the Thornton 200pH two-channel pH/ORP measurement system, with its pH and ORP sensors and pHure Sensor.
Covers · 1 model
The models it covers
Summary
What is in it
The specification sheet for the Thornton 200pH, a two-channel instrument that measures pH plus temperature and/or ORP for pure water, industrial process and wastewater use. It lists key features and applications (reverse osmosis, power plant cycle chemistry, wastewater neutralization), explains temperature compensation, and gives full specifications for ranges, outputs, relays, accuracy, environment, enclosure and power. Later pages cover the detachable pH and ORP electrodes, housings and preamps with part numbers, and the 363-21X pHure Sensor for pure water.
Its parts
- 200pH: pH and ORP Measurement System
- 200pH Specifications
- 200pH: pH and ORP Sensors
- 200pH: Detachable pH and ORP Sensors
- 363-21X pHure Sensor Specifications
From the sheet · read page by page
Key facts from this sheet
Each line names its page in the PDF.
Eight-page Mettler-Toledo Thornton spec sheet, dated 2005, for the 200pH two-channel pH/ORP measurement system. It gives the instrument's ranges, outputs, relays, power and enclosure, explains its temperature compensation and typical uses, and describes the detachable pH and ORP sensors, the pHure Sensor for pure water, and ordering information.
- Device
- Two-channel instrument measuring pH plus temperature and/or ORP p. 2
- Ranges
- pH -1.00 to 15.00; ORP ± 1250 mV; temperature -40 to 140 °C (-40 to 284 °F) p. 3
- Accuracy
- ± 0.02 pH, ± 2 mV, ± 0.3 °C; resolution 0.01 pH, 0.1 mV, 0.1 °C p. 3
- Analog and serial outputs
- Two optional powered, isolated 4-20 mA outputs (0-20 mA possible), 500 ohm maximum; RS232 to 50 ft or RS422 to 4000 ft, up to 19.2 kbaud p. 3
- Relays
- Standard 2 SPDT mechanical, 5 amp at 250 VAC or 30 VDC; optional 2 more AC-only solid state SPST, 1.5 amp at 250 VAC; four setpoints p. 3
- Power
- 90-130 VAC or 180-250 VAC, 50-60 Hz, 12W maximum, or nominal 12-30 VDC at 300 mA (600 mA inrush) p. 3
- Inputs
- pH and/or ORP electrodes with a Thornton preamp and Pt1000 RTD; preamps accept VP, BNC or S8 electrode connections p. 2, 3
- Display and keypad
- 16 character backlit LCD showing one or two measurements; 11 tactile keys p. 3
- Enclosure
- ABS-PC alloy, 3.78 x 7.56 in (96 x 192 mm) DIN panel cutout, NEMA 4X, IP65, 1.9 lb (0.9 kg); wall or pipe mount with accessory back cover p. 3
- Operating temperature
- -10 to 55 °C (14 to 131 °F); storage -40 to 70 °C; 0 to 95% RH non-condensing p. 3
- Sensor cable
- Patch cord up to 200 ft (61 m); preamp cable up to 33 ft (10 m) p. 4
- Temperature compensation
- Nernst electrode compensation plus adjustable solution compensation for high purity water, referenced to 25°C p. 2
- Memory
- Configuration kept in non-volatile memory without batteries on power loss p. 3
- Pure water sensor
- 363-21X pHure Sensor: 316 SS, glass and silicone rubber wetted parts, 1/4 in NPTF, 50-150 mL/min sample flow, above 0.8 µS/cm for highest accuracy p. 6
What this sheet walks you through
- Typical uses in reverse osmosis, power plant chemistry and wastewater p. 2
- How temperature compensation works for pure water p. 2
- Choosing an electrode, housing and preamp for a detachable sensor p. 4
- Measuring pH in pure, low conductivity water with the pHure Sensor p. 6
- Installing the pHure Sensor with the right clearances p. 7
- Ordering the instrument, patch cords and mounting accessories p. 8
Good to know
- 24 VDC power must be isolated from earth ground and from other instruments. p. 3
- The analog outputs are not for use with externally powered circuits. p. 3
- Serial output needs external isolation. p. 3
- The pHure electrode and flow housing must stand upright. p. 7
- Low conductivity pH must be measured on a side-stream sample in a closed metal flow-through chamber draining to open drain. p. 6
Questions this sheet answers
What power supply does the Thornton 200pH use?
90-130 VAC or 180-250 VAC at 50-60 Hz, or a nominal 12-30 VDC supply; DC power must be isolated from earth ground. p. 3
Can the 200pH measure ORP as well as pH?
Yes. Each of its two channels can measure pH (-1.00 to 15.00) or ORP (± 1250 mV). p. 2, 3
How many relays does the 200pH have?
Two mechanical SPDT relays as standard, plus two optional AC-only solid state relays, driven by four setpoints. p. 3
What pH electrodes work with the 200pH?
Its detachable preamplifiers accept virtually any pH electrode with a VP, BNC or S8 connection. p. 2
How far can the sensor be from the 200pH?
Up to 200 ft (61 m) of patch cord between the sensor and the instrument. p. 2