Page 21

micro, milli, units, or auto ranging. To change the field, use the RIGHT arrow key to move the cursor under “(Auto)”. Use the UP and DOWN arrow keys to select the desired range.

Concentrations (%HCl, %NaOH, %H2SO4) All concentrations are displayed as percent by weight. Setting a measurement for concentration automatically activates specialized temperature compensation for that particular material. The compensation setting is ignored for that channel.

The OK/NEXT key is used to accept the entry for channel A primary and move to the next menu for setting channel “a” secondary. Press the OK/NEXT key a third and fourth time to set the measurement types for channel B primary and channel “b” secondary, respectively.

SELECTING A MEASUREMENT TYPE

When the last menu level is reached (after setting channel “b” secondary), the following message is displayed:

To set or change a measurement type for each of the four measurements:

Save Changes Yes

Press the MEASURE MODE key and the display will show the measurement type assigned to channel A primary. The display may appear as:

Pressing the OK/NEXT key will save the changes and exit the menus. The UP and DOWN arrow keys can be used to change the “Yes” to “No”. Pressing the OK/NEXT key with “No” will discard the changes and exit the menus.

A = Ω-cm (AUTO) This menu indicates that channel A primary measurement is set for resistivity (Ω-cm) with auto ranging. The section “Ω -cm” is the first field to be changed and will be flashed as long as the cursor is under it. Pressing the UP arrow key will change the “Ω -cm” to “S/cm”.

A = S/cm (AUTO) Use the UP and DOWN arrow keys to select the desired measurement type. For conductivity, both temperature compensated and uncompensated measurement are available. Uncompensated readings are needed to meet pharmaceutical USP 23 requirements. Select units of “S/cm” for compensated measurement or “s/cm U” for uncompensated measurement. Using both primary and secondary parameters, both compensated and uncompensated measurements are available simultaneously. In normal operation, uncompensated measure-ments are identified by a flashing cursor under the units, for example, “µs”. Normal compensated readings are identified by “µS”. The field on the right half side of the display is the range field. Some measurements can be set for a fixed range, others can be set only for auto ranging. For example, conductivity can be set for

Chapter 4

CELL CONSTANTS The calibration of each measurement is defined by a set of constants known as cell constants. There are two cell constants for each measurement: a Multiplier Factor and an Adder Factor. They are used to derive an accurate measurement from the sensor’s output signal. As an example, the output of a conductivity sensor can be represented by the following equation: R =x/M + A Where: R = resistivity value x = output from cell M = multiplier factor A = adder factor Example: for a typical conductivity sensor the multiplier (M) is 0.1 and the adder (A) is 0. If the sensor output is 120,000 ohms, then the actual resistivity of the solution measured is 1.2 Mohmcm and is calculated as follows: R = x/M + A R = 120,000/0.1 + 0 R = 1,200,000

Making Measurements

15

Page 22

R = 1.2 Mohm-cm NOTE: The adder is zero for conductivity, resistivity and ORP sensors. The cell constants can be modified via the menus. For more information see ENTERING/EDITING CELL CONSTANTS in Chapter 9.

CONDUCTIVITY TEMPERATURE COMPENSATION

Cation Compensation Power industry applications for cation conductivity measurements with acidic samples are accurately compensated with this setting. It takes into account the effects of temperature on the dissociation of pure water with the presence of very dilute acids. See ASTM Standard D6504 for recommendations on this measurement. Semiconductor acid etch rinse operations are also more accurately monitored using this setting.

Conductivity, resistivity and TDS measurements can be compensated for temperature. Compensation will adjust the measurement to give an equivalent reading of the solution at 25°C. For example, the resistivity of ultrapure water at 25°C is18.18 MΩ-cm. The resistivity of ultrapure water at 30°C is 14.08 MΩ-cm. By compensating the resistivity reading, the value of pure water will continue to read 18.18 MΩ-cm.

Ammonia / ETA Compensation

The 200pH/CR can use one of nine different compensation methods: standard, linear, cation, alcohol, Light 84, 50% glycol, 100% glycol or none. Channels A and B can be programmed with different compensation methods.

Alcohol Compensation

For pharmaceutical USP <645> conductivity measurements where uncompensated measurement is required, select None for compensation. For simultaneous uncompensated and compensated measurements, see SELECTING A MEASUREMENT TYPE earlier in this chapter.

Standard Compensation The standard compensation method includes compensation for non-linear high purity effects as well as conventional neutral salt impurities and conforms to ASTM standards D1125 and D5391.

Linear Compensation

This compensation provides for the temperature characteristic of a 75% solution of isopropyl alcohol in pure water used for some rinsing operations in semiconductor manufacture. Compensated measurements using this solution may go well above 18 Mohm-cm.

Light 84 Compensation This compensation matches the earlier high purity water research results of Dr. T.S. Light published in 1984. It is provided only for use by institutions that have standardized on that work. For all other pure water and general purpose applications, Standard Compensation (using the 1994 high purity water research of Thornton & Light) is recommended.

Glycol Compensation

The raw resistance measurement is compensated by multiplication with a factor expressed as a “% per °C” (deviation from 25°C). The range is 0 99%/°C with a default value of 2%/°C.

16

Power industry applications for specific or direct conductivity on samples with ammonia or ETA (ethanolamine) are accurately compensated with this setting. It takes into account the effects of temperature on the dissociation of pure water with these very dilute bases.

This compensation matches the characteristics of 50% and 100% ethylene glycol in deionized water cooling/antifreeze solutions.

Making Measurements

Chapter 4

Page 23

Setting conductivity compensation Enabling the compensating feature is done via the menus as follows: Press the MENUS key and the following menu will appear:

compensated by displaying a flashing underline cursor at the units character. The following display shows that channel A is compensated and channel B is not compensated (the character “µ” has a blinking underline cursor).

A1.76µS

B2.11µS

Menus use arrows Press the UP arrow key until “Compensation” menu is displayed.

Compensation Press the OK/NEXT key to access this menu.

A: Comp = Standard The cursor will be initially placed under channel. Use the UP and Down arrow keys to change the channel if desired. Use the RIGHT arrow key to move the cursor to the method field. Use the UP and DOWN arrow keys to scroll through the list of available methods: “Standard”, “Linear”, “Cation”, “Alcohol”, "Light84" or “None”. Selecting “None” disables the compensation feature. Press the OK/NEXT key to accept the selection. If “Linear” is selected then another menu will be displayed where the linear value can be entered. This menu will appear as (the actual value may be different):

A: Comp = 2.000%/°C Adjust the numerical field as desired using the arrow keys and press the OK/NEXT key when done. After setting the compensation state for channel A, repeat for channel B. Press OK/NEXT key when done. The meter will ask if changes should be saved.

Save Changes Yes Press OK/NEXT key to save changes and return to the display of measurement data.

Compensation Indication When displaying measurements, the 200pH/CR will indicate that a measurement is NOT

Chapter 4 pH TEMPERATURE COMPENSATION The 200pH/CR Instrument provides two types of pH temperature compensation, described in following sections. If a temperature sensor is not included in a particular pH probe, a manual temperature setting or the temperature from the other channel sensor may be used instead. See Chapter 11. With ORP measurement, temperature is not measured or compensated.

Conventional pH Electrode Temperature Compensation Electrode Temperature Compensation (Nernst Response) is provided in most pH instruments. All pH electrodes produce a millivolt signal with gain proportional to the absolute temperature. Electrode temperature compensation normalizes that variable millivolt output to give pH values. The default and normal operating setting of the instrument has this compensation active. Some specialized measurements may need to disable it by accessing the "Compensation" setting via the MENUS key. Compensation for each channel may be individually turned on or off.

pH Solution Temperature Compensation Solution Temperature Compensation allows settings for compensation of the variable ionization of pure waters. The change is so small in more conductive waters that it is usually ignored, but for high purity water, it is significant. It is used in addition to the conventional (Nernst) compensation, described above, which is normally active. Solution temperature compensation is used primarily with power plant and other pure water samples less than 30 µS/cm conductivity. It

Making Measurements

17

Page 24

references the pH of pure water to 25°C. All other applications should leave the solution temperature coefficient (STC) set to its default value of zero. For pure makeup water or boiling water reactor samples, the STC should be set to 0.016 pH/°C. For ammonia, phosphate and/or amine-treated samples the STC should be set to 0.033 pH/°C. The appropriate setting for other pure water compositions may be determined by developing temperature vs. pH data for the particular sample with the STC set to zero. The negative slope of this data becomes the STC value. Because Solution Temperature Compensation is unique to the process sample and is different for buffer solutions, it is not active during calibration. The 200pH/CR ignores the STC during calibration. For this reason, the reading in a buffer solution after calibration may not be the exact value entered. To read and verify the exact buffer value, temporarily set the STC to zero. The Solution Temperature Coefficient setting is accessed via the MENUS key / Spec pH Function / STC=0.000 pH/C for each channel.

pH Isopotential Point

Setting 50/60 Hz Operation To set the appropriate filter, press the MENUS key and the following menu will appear:

Menus use arrows Press the DOWN arrow key until the “Set Frequency” menu is displayed.

Set Frequency Press the OK/NEXT key to access this menu.

Frequency = 60 Use the UP or DOWN arrow keys to set the desired frequency. Press the OK/NEXT key when done. The meter will ask if changes should be saved.

Save Changes Yes Press the OK/NEXT key to save the changes and return to the display of measurement.

The isopotential Point, IP, is left at the default setting of 7.0 for all Thornton and other conventional pH sensors. Special purpose pH electrodes with zero potential at values other than 7 pH will have this identified in their instruction manuals. A different IP setting will allow proper temperature compensation of these special electrodes. The IP setting is accessed via MENUS key / Spec pH Function / IP=7.000 pH for each channel.

AC POWER FREQUENCY The 200pH/CR meter was designed to reduce fluctuations in measurements by eliminating noise pickup from the AC power line. The meter can be set to filter either 50Hz or 60Hz power from the measurements. Factory settings are 60 Hz for 115 VAC models and 50 Hz for 230 VAC models.

18

Making Measurements

Chapter 4

Page 25

CHAPTER 5: USING SETPOINTS OVERVIEW A setpoint is a limit or alarm point applied to a measurement. A setpoint can be programmed as a high limit, a low limit or a USP limit. (A USP setpoint is a high alarm used for pharmaceutical water monitoring, described later in this chapter.) When the measurement value is higher than a high setpoint, or lower than a low setpoint, a setpoint error condition exists. The meter will indicate this condition by blinking the measurement on the display. The 200pH/CR can also be programmed to control a relay upon this error condition. Refer to Chapter 6 for more information on relays. Four setpoints are available and can be assigned to any of the four measurements (A, a, B and b). More than one setpoint can be assigned to the same measurement and more than one setpoint can activate the same relay. The following parameters can be programmed for setpoint operation: 1.

The assigned signal: which signal (A, a, B, or b) is monitored by the setpoint.

2.

The setpoint value: the measurement value that triggers the setpoint error condition, with prefix.

3.

The setpoint type: High, Low, Off or USP.

4.

The assigned relay: which relay will be controlled when a setpoint error occurs (this is optional).

5.

Operation on overrange: Whether or not the setpoint will be active when the measurement is outside its range, yes or no.

SETPOINT SIGNAL The setpoint signal is the measurement that will be monitored by the setpoint. The signal can be any of the measurements: 1.

A - Channel A primary

2.

a - Channel a secondary

3.

B - Channel B primary

4.

b - Channel b secondary

Chapter 5

SETPOINT VALUE The setpoint value is the limit that will trigger a setpoint error condition. This number is entered as a four digit number with a prefix. The prefixes are: “µ” (micro) (10-6).

= multiply value by

0.000001

“m” (milli) = multiply value by 0.001 (10-3). “-” (blank) = multiply value by 1 “K” (kilo) = multiply value by 1,000 (103). “M” (Mega) = multiply value by 1,000,000 (106). For example, a setpoint value of 18.18M is equivalent to 18,180,000. When programming the setpoint values with TDS units, the following prefixes should be used (parts per million is the basis): m = parts per billion _ = parts per million K = parts per thousand

SETPOINT TYPE The setpoint type can be “High”, “Low”, "USP" or “Off”. A setpoint error condition occurs when the measurement is above the high or USP limit or below the low limit.

ASSIGNED RELAY The assigned relay will change state according to the setpoint condition. When a setpoint error condition exists, the assigned relay will change state. If the relay is normally closed, then it will be opened. Default settings after a system reset: Setpoints are disabled, no signals assigned, no relays assigned and values are zero.

PROGRAMMING A SETPOINT Using Setpoints

19

Page 26

Press the SETPOINTS key to access the first setpoint menu. This menu is used to assign a measurement to each of the four setpoints.

SP1 use Relay#_

SP1 on signal: A

After selecting the desired setpoint number, move the cursor to the relay field. The choices are: 1 - 4 and “_”. The “_” indicates that there are not any relays assigned to the setpoint.

The cursor is under the setpoint number field. Use the UP and DOWN arrow keys to select the desired setpoint number (1 to 4). Use the RIGHT arrow key to move the cursor to the signal field. Then use the UP and DOWN arrow keys to select the desired signal. The signal can be set to “-” which indicates that the setpoint is not assigned (disabled). To change the assigned signal of another setpoint, move the cursor back to the setpoint number. Change the setpoint number, then move the cursor back to the assigned signal field.

SP1=.0000 _ off Press the OK/NEXT key when done to proceed to the next menu. The cursor is initially under the setpoint number field. Select the desired number then move the cursor to the numerical field. Set the desired value and prefix. Examples of numerical entries:

Desired Value

Menu Entry

10.34

10.34_

1,456

1.456K

18,180,000

18.18M

567,456

567.5K

0.003

.0030_ or 3.000m

.000000055

.0550µ

NOTE: Some models of the 200pH/CR are equipped with only 2 relays and will allow settings of only #1 or #2. Press the OK/NEXT key when done to proceed to the next menu.

SP1 over-range Yes This menu selects whether the setpoints will be active when the measurement is out of range, which could be due to process upset, loss of fluid at the sensor, disconnected sensor leads, etc. Use the arrow keys to select Yes or No for each setpoint. Press the OK/NEXT key when done. The meter will ask if changes should be saved.

Save Changes Yes Press the OK/NEXT key to save the changes and return to the display of measurement data.

Example: Set Up a Setpoint Program setpoint #2 with the following conditions: 1.

Assigned to channel A secondary signal (“a”).

2.

A value of 18,200,000 (18.2 M Ω-cm)

3.

Set as a high limit

4.

Use relay #2

5.

Disabled when out of range

Move the cursor over to the setpoint type field. To turn the setpoint on, select High, Low or USP. Press the OK/NEXT key when done to proceed to the next menu. The next menu is used to assign a relay to the setpoint. This is optional. The menu will appear as:

20

Using Setpoints

Chapter 5

Page 27

Press the SETPOINTS key Use the arrow keys to select setpoint #2 and signal “a”. The display will appear as follows:

The 200pH/CR instrument has the USP 23 table in memory and automatically determines the conductivity limit based on the measured temperature.

Use the arrow keys to set the value at 18.20M and the state to high. The display will appear as follows:

The USP setpoint value set in the 200pH/CR is the percentage safety margin below the USP limits to activate the setpoint. For example, the USP table conductivity limit at 15°C is 1.0 µS/cm. If the setpoint value is set at 40% then the setpoint will activate whenever the conductivity goes above 0.6 µS/cm at 15°C.

SP2=18.20 M High

USP <645> Stage 1 Conductivity Limits as a Function of Temperature

SP2 on signal: a Press the OK/NEXT key:

Press the OK/NEXT key: Use the arrow keys to set the relay number to 2. The display will appear as follows:

SP2 use Relay #2 Press the OK/NEXT key: Use the arrow keys to set the over-range to No, resulting in the following display:

SP2 over-range No Press the OK/NEXT key. The meter will ask if changes should be saved.

Save Changes Yes Press the OK/NEXT key to save the changes and return to the display of measurement data.

USP SETPOINT The USP type setpoint is a high alarm used for pharmaceutical water monitoring with nontemperature compensated conductivity measurements. USP (United States Pharmacopoeia) section <645> requires that nontemperature compensated conductivity of pharmaceutical waters must be below a limit from its table based on the temperature of the sample. In other words, USP requirements are to temperature compensate the limit rather than the measurement.

Chapter 5

Temperature (°C) 0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 100

Conductivity Limit (µS/cm) 0.6 0.8 0.9 1.0 1.1 1.3 1.4 1.5 1.7 1.8 1.9 2.1 2.2 2.4 2.5 2.7 2.7 2.7 2.7 2.9 3.1

To configure a USP setpoint, use the previous procedure but select USP instead of High or Low. When selected, the display appears as:

SP1=0.000 % USP Using the arrow keys, enter the percent safety margin below the USP conductivity limit desired.

Using Setpoints

21

Page 28

CHAPTER 6: USING RELAYS DESCRIPTION value is 10% then the measurement must fall below 90 before the relay is deactivated.

The 200pH/CR is equipped with up to four relays. Each relay can be programmed to activate when a setpoint is exceeded (defined as a setpoint error condition). The programmable parameters for a relay are:

RELAY STATE

1.

Delay Time: up to 999 seconds

2.

Hysteresis Value: up to 99%

3.

State: normal or inverted

ELECTRICAL CONNECTIONS For units equipped with two relays, each one has a common connection, a normally open connection and a normally closed connection. Units with 3rd and 4th solid state relays have only a common connection and a normally open connection as shown in Table 2.1.

The relay can be programmed for normal or inverted operation. When the relay is in the inverted state, the relay operation is reversed. When there is no setpoint error condition the relay is activated. The normally open contacts are closed. WARNING: RELAYS WILL ALWAYS DEENERGIZE ON LOSS OF POWER, EQUIVALENT TO NORMAL STATE, REGARDLESS OF RELAY STATE SETTING. HOWEVER, THE SETTING IS RETAINED ON RESTORATION OF POWER. Default settings after the system reset: 1.

Relay is disabled.

2.

Delay is 0 seconds.

3.

Hysteresis is 0%.

4.

Relay state is normal.

DELAY TIME Delay time is the length of time that the setpoint must be exceeded continuously (in a setpoint error condition) before activating the relay. When the setpoint error condition occurs, the delay timer is started. If during the delay time the setpoint error condition no longer exists, the delay timer is reset and the relay will not be activated. The maximum delay time is 999 seconds (16 minutes and 39 seconds).

HYSTERESIS The hysteresis value is entered as a percentage of the setpoint value. For a high setpoint, the measurement must fall more than this percentage point below the setpoint value before the relay is deactivated. With a low setpoint, the measurement must rise at least this percentage above the setpoint value before the relay is deactivated. For example: a high setpoint is set at 100 and the measurement is currently above this value so the setpoint error condition exists. If the hysteresis

22

PROGRAMMING A RELAY To enable or modify a relay: Press the RELAYS key. The first relay menu is used to set the delay time.

R1 Delay= 000 sec In this menu, the cursor is initially under the relay number. Use the UP and DOWN arrow keys to select the desired relay number (1 to 4). Use the RIGHT arrow key to move the cursor to the delay time field. Then use the UP and DOWN arrow keys to set the delay time (000 to 999 seconds). To change the delay time of another relay, move the cursor back to the relay number. Change the relay number, then move the cursor back to the relay time field. Press the OK/NEXT key when done to proceed to the next menu.

Using Relays

Chapter 6

Page 29

R1 Hyster = 00%

Press the OK/NEXT key. The meter will now ask if changes should be saved.

Select the desired relay number and enter the hysteresis value (00 to 99%).

Save Changes Yes

Press the OK/NEXT key when done to proceed to the next menu.

Press the OK/NEXT key to save the changes and return to the display of measurement data.

R1 State = Normal Select the desired relay number, then use the DOWN arrow key to select either Normal or Invert. Press the OK/NEXT key when done. The meter will ask if changes should be saved.

Save Changes Yes Press the OK/NEXT key to save the changes and return to the display of measurement data.

Example: Setup a Relay Program relay #2 with the following conditions: 1.

Delay of 60 seconds.

2.

A hysteresis of 10%.

3.

Inverted state.

Press the RELAYS key. Use the arrow keys to select relay #2 and set a delay time of “060”. The display will appear as follows:

R2 Delay = 060 sec Press the OK/NEXT key. Use the arrow keys to set the hysteresis value to “10%”. The display will appear as follows: Press the OK/NEXT key.

R2 Hyster = 10% Use the arrow keys to set the state to inverted. The display will appear as follows:

R2 State = Invert

Chapter 6

Using Relays

23

Page 30

CHAPTER 7: USING ANALOG OUTPUTS DESCRIPTION An analog output is an isolated current signal that is proportional to any measurement. The two 200pH/CR analog outputs have a minimum value of 4mA and a maximum value of 20mA (the signal can be re-calibrated to 0-20mA when needed, see ANALOG OUTPUT CALIBRATION later in this chapter). Each output can be scaled to a range of a measurement signal. To use analog outputs, the following parameters must be programmed: 1.

Assigned Signal - The analog output will be proportional to the value of the assigned signal. Any of the four measurements (A, a B, b) can be assigned to the output.

2.

Minimum Value - This is the measurement reading that will correspond to an output of 4mA.

3.

Maximum Value - This is the measurement reading that will correspond to an output of 20mA.

The analog outputs can be programmed for their minimum and maximum values. This programming process is independent from any measurement. See PROGRAMMING THE ANALOG OUTPUTS later in the chapter for details.

PROGRAMMING THE ANALOG OUTPUTS To set up an analog output channel, press the OUTPUTS key:

Output: Analog Press the OK/NEXT key to access this menu. The next menu is used to assign a measurement signal to the output. The choices are: A, a, B, b, and _. The selection of “_” is used to disable the output by not assigning a signal to the output. The output will remain at 4mA when it is disabled. The menu may appear as:

Aout1 signal = A In this menu the cursor is initially under the output number. Use the UP or DOWN keys to select the desired output number ( 1 or 2). Press the RIGHT arrow key to move the cursor under the assigned field.

Aout1 signal = A Use the UP or DOWN arrow keys to change the assigned measurement. Press the OK/NEXT key to accept it and proceed to the next menu.

ELECTRICAL CONNECTIONS Connections to the analog output signals are made at terminal block TB2. Each analog output channel has a signal line (labeled AO1+ or AO2+) and a return line (labeled AO1- or AO2-). Table 2.4 shows the electrical connections. The minimum can be re-calibrated down very close to zero mA when needed. See ANALOG OUTPUT CALIBRATION later in this chapter. The analog outputs are isolated from line power, sensors and earth ground. Each analog output channel can drive a resistive load up to 500 ohms.

24

Aout1 Min=0.000_ This menu is used to set the measurement value that will correspond to a 4mA output. Set the desired value. The last position in this menu is the prefix which can be one of the following: “µ” (micro) (10-6).

= multiply value by

0.000001

“m” (milli) = multiply value by 0.001 (10-3). “_” (blank) = multiply value by 1. “K” (kilo) = multiply value by 1,000 (103) “M” (Mega = multiply value by 1,000,000 (106)

Using Analog Outputs

Chapter 7