Page 31

When programming the analog output scaling limits with TDS units, the following prefixes should be used (parts per million is the basis):

Aout2 signal = B Press the OK/NEXT key.

m parts per billion

Use the arrow keys to set the minimum value to 1.000 M ohms. The display will appear as follows:

_ parts per million K parts per thousand After setting the desired value, press the OK/NEXT key to accept minimum value and proceed to the next menu.

Aout2 Min=1.000M Press the OK/NEXT key. Use the arrow keys to set the maximum value to 20.00 M ohms. The display will appear as follows:

Aout1 Max=1.000_ This menu is used to set the measurement value that will correspond to an output of 20mA. Repeat the process as described above. Press the OK/NEXT key to accept the maximum value and proceed to the next menu.

1:If err set Max Using an arrow key, select max or min as the fail safe mode for the output signal on overange or failure conditions. The output will go to 20mA if set for max; to 4mA (or 0mA if re-calibrated) if set for min. The meter will ask if changes should be saved.

Save Changes Yes Press the OK/NEXT key to save the changes and return to the display of measurement data.

Example: Setup an Analog Output Set analog output #2 with the following parameters: 1.

Assigned to channel B primary measurement

2.

Minimum value of 1.000 M ohms

3.

Maximum value of 20.00 M ohms

Aout2 Max=20.00M Press the OK/NEXT key.

Save Changes Yes Press the OK/NEXT key to save changes and return the display of measurement data

ANALOG OUTPUT CALIBRATION The analog output signals have been factory calibrated to specifications and normally do not require any further adjustment. To match other equipment, they may be re-calibrated in a two step process where 4mA and 20mA levels are adjusted. A current meter is connected in series with the output. The arrow keys are then used to adjust the current output for the appropriate level (4mA, then 20mA). The 4mA end of the signal can be calibrated down to zero mA, where needed. With a 500 ohm resistor across the output terminals, a 0-10 VDC signal may also be obtained.

Press the OUTPUTS key. The display will show:

For NIST traceability, the outputs can be calibrated with any NIST-calibrated current meter.

Output: Analog

Procedure: Connect the meter in series with the output signal. Press the CAL key and the display will show:

Press the OK/NEXT key. Use the arrow keys to select analog output #2 and assign measurement B to this output. The display will appear as follows:

Chapter 7

Using Analog Outputs

Calibrate Sensor

25

Page 32

Use the UP and DOWN keys to change the display until the “Analog” option is displayed:

Calibrate Analog Press the OK/NEXT key to proceed to the next menu.

Cal Analog Ch 1 Use the UP and DOWN keys to select the desired output channel number (1 or 2). Press the OK/NEXT key to proceed to the next menu.

1:Adj

4mA=10641

Use the UP and DOWN keys to adjust the output current for a 4mA level (as measured by the current meter). The number displayed is an arbitrary value proportional to the analog output signal, near 10,000 for 4 mA, near 0.000 for 0 mA and near 55,000 for 20 mA. Adjusting a more significant digit of this number will change the output signal level faster. Adjust a less significant digit for closer resolution. Press the OK/NEXT key when done adjusting the 4mA or 0mA output. The next menu is for adjusting the 20mA output.

1:Adj

20mA=54091

Adjust the 20mA level in a similar manner. Press the OK/NEXT key when done adjusting the 20mA. The meter will ask if changes should be saved.

Save Changes Yes Press the OK/NEXT key to save the changes and return to the display measurement data.

26

Using Analog Outputs

Chapter 8

Page 33

CHAPTER 8: METER CALIBRATION OVERVIEW The 200pH/CR meter is factory calibrated within specifications. It is not necessary to perform meter re-calibration unless extreme conditions cause out of specification operation shown by Verification (see CALIBRATION VERIFICATION section). Periodic verification/re-calibration may also be necessary to meet Q.A. requirements. ALWAYS perform a verification before considering a meter calibration. If within specification there is no need to perform a meter calibration. CAUTION: Do not perform a meter calibration unless a Thornton calibrator or precision decade box is connected. Factory calibration cannot be restored in the field. The 200pH/CR meter is re-calibrated for conductivity/resistivity by installing a known resistance value in place of the cell and using the calibration menus to complete the process. The meter is designed for calibration at six points for conductivity/resistivity and two points of temperature (on each channel) as shown in Table 8.1. The second column of the table indicates the calibrators that are applicable for a region of measurement. Note that a single conductivity/resistivity calibration covers both Channels A and B at one time but separate temperature calibrations are required on each channel. Calibration Point (ohms)

Calibrator Part No.

Approx. Meas. Range

Short

1864-09

0-2K ohms

Open

1864-12

>200K ohms

500K

1864-05

>200K ohms

50K

1864-06

20K-200K ohms

5K

1864-07

2K-20K ohms

500

1864-08

0-2K ohms

A Temp 1K

1864-06,-08

0 °C

A Temp 1.4K

1864-05,-07

104 °C

B Temp 1K

1864-06,-08

0 °C

B Temp 1.4K

1864-05,-07

104 °C

Standards for verification or calibration of conductivity and temperature are available as six Thornton NIST-Traceable calibrators (second column) which include one resistance and one temperature value in each unit. They plug directly into patch cables in place of the sensor. Full Range Calibration Kit 1865-07 includes all six calibrators in Table 8.1. High Resistivity Kit 186505 includes 1864-05,-06,-12. Low Resistivity Kit 1865-06 includes 1865-07,-08,-09. If measurements will always be within a certain range, only that range requires calibration. For example, if measurement will always be >200K ohms, as with ultrapure water, only the >200K ohm range requires calibration. CAUTION: Calibrators are for calibrating conductivity/resistivity measurements only. Do not connect a calibrator to a patch cord wired for pH/ORP. If decade boxes are used, connect per Figure 14.10, shown for channel A, and similarly for channel B (temperature only). NOTE: To display the actual value of a calibration device, use the Calibration Verification menu described in the next section. WARNING: INSTALLING A CALIBRATION DEVICE ON A CHANNEL MAY TRIGGER ALARM STATES, RELAY ACTION AND PROCESS UPSET.

CALIBRATION VERIFICATION The calibration verification menu can be used to quickly confirm the meter’s performance. A calibration/verification device is installed on a channel and this menu is used to display the actual resistance value. The verification feature can be accessed from either the CALIBRATE or MENUS keys to allow the calibration menu to be locked yet still allow operators to verify calibration. Press the MENUS key and the following menu will appear:

Menus use arrows

Table 8.1 Calibration Points

Chapter 8

Meter Calibration

27

Page 34

Press the DOWN arrow key until the “Verify Calibrate” menu is displayed.

Calibrate Verify Press the OK/NEXT key to access this menu.

Cal Meter Ch A Use the UP and DOWN keys to select channel B if desired (for temperature). Press the OK/NEXT key to proceed to the next menu. Step 3: Select the signal to calibrate:

Verify Cal: Ch A

Cal A #1: R 500K

Use the UP arrow key to select the desired channel. Press the OK/NEXT key when set. The meter will now display the actual values of the calibrator. A typical display may appear as:

A10.04MΩ 1.003KΩ The first number is the measured value of the conductivity/resistivity input and the second number is the measured value of the temperature input. Compare these numbers with the values printed on the label of the verifier/calibrator as well as on the Certificate of Accuracy supplied with it. Values should be within ±0.5%. Press any key to end this menu. Other calibration/verification values may be checked in the same manner. If within specification, do not perform a calibration.

CALIBRATION PROCEDURE The resistance calibration sequence should use the "short" and/or "open" calibrators first, whenever they are used. Other calibrators may be used in any order. Step 1: Select the Meter Calibration

Use the UP and DOWN keys to select the point to be calibrated. The choices are: “R500K", "R50K", "R5K", "R500", “Temp”, "Volts", "Short", or “Open”. Press the OK/NEXT key when ready to proceed to the next menu. For the following example "Temp" is selected for calibration.

Cal A #1: Temp Step 4: Enter the calibrator value:

A Temp =1.0000K This menu is used to enter the value of the calibrator (printed on the label as well as the Certificate of Accuracy supplied with the calibrator). The display will show a nominal value but the actual value of the calibrator must be entered. Use the UP and DOWN keys to change the digit. Use the RIGHT and LEFT arrow keys to move the cursor to the next digit. The last character on the display is the units prefix, which appears after the value on the calibrator. For Reference:

Press the CALIBRATE key and the display will show:

Calibrate Sensor Use the UP and DOWN keys to change the display until it reads:

Calibrate Meter

The available prefixes are “µ” = micro, “m” = milli, “k” = kilo, “M” = mega and “_” = one. The numerical entry is multiplied by the value of the prefix as follows: “µ” (micro) = multiply value by 0.000001 (10-6). “m” (milli) = multiply value by 0.001 (10-3). “_” (blank) = multiply value by 1.

Press the OK/NEXT key to proceed to the next menu.

“K” (kilo) = multiply value by 1,000 (103) “M” (Mega = multiply value by 1,000,000 (106)

Step 2: Select the channel

28

Meter Calibration

Chapter 8

Page 35

When the value of the calibrator is entered, press the OK/NEXT key when ready to proceed to the next menu. Step 5: Install the calibrator on the selected channel:

Enter the exact value of the 2nd calibrator and press the OK/NEXT key when ready to proceed. Step 9: Install the second calibrator: As with the 1st calibration point, the meter will now display the measured value of the calibrator as follows:

A= 1.0003KT Ω OK?

A=1.4011KTΩ OK?

The instrument will display the value of the calibrator as measured by the meter. it is important to wait for this displayed measurement to stabilize. The reading may fluctuate slightly but the changes should be no more than +1 digit (least significant). For example, in the display above the reading may fluctuate from 1.0002K to 1.0004KΩ. It is recommended that you wait at least 30 seconds to be certain that the measurement has settled. To discontinue the calibration, press any of the menu keys. To proceed with the calibration process (after the readings have stabilized), press the OK/NEXT key.

When the reading has stabilized, press the OK/NEXT key. Step 10: Perform the calibration (point #2): The calibration process may take a few seconds to complete. When the instrument is done, the following message will be displayed:

Done, press OK Press the OK/NEXT key. The meter will ask if the changes should be saved.

Save Changes Yes

Step 6: Perform the calibration: The calibration process may take a few seconds to complete. When the instrument is done, the following message will be displayed.

Done, press OK Step 7: Proceed to the 2nd calibration point: Press the OK/NEXT key to complete the calibration of the first point. The meter will display the following menu:

Do point #2 No

Press the OK/NEXT key to save the changes and return to the display of measurement data. The calibration process is complete for the temperature circuit of one channel. Repeat this process for the other channel temperature, if used, and for the conductivity/resistivity points. Reminder: due to entered cell constants and temperature compensation, the meter will only read the exact connected calibrator value in the “Verify” mode.

Special Temperature Calibration

Press the OK/NEXT key to save the new calibration factor and return to the measurement mode. Otherwise, use the UP and DOWN arrow keys to change the “No” to “Yes” and press the OK/NEXT key to perform calibration for a 2nd point. The meter will display the following menu, requesting the value of the 2nd calibration point:

NOTE: If using 2_8 series Dot Two cells with Ni-Fe 500 ohm RTD, substitute a decade box with 500 ohms for 0°C and 735 ohms for 100°C in place of the calibrators in the previous procedure.

Step 8: Enter the calibrator value (point #2)

A Temp = 1.4000K

Chapter 8

Meter Calibration

29

Page 36

Voltage Calibration for pH and ORP It is generally not necessary to calibrate the voltage range which is used for pH measurement. pH sensor calibration is performed frequently in NIST traceable pH buffer solutions and will override any voltage calibration. (The pH scale is defined by response in NIST standard reference materials.) The voltage input is used with ORP sensors. To calibrate the voltage range, use a stable voltage source and precision voltmeter and connect as given in Figure 14.10. Follow the procedure above, selecting "Volts" as the signal to be calibrated. A two-point calibration should be done near -400 and +400 mV. A 200pH/CR meter voltage calibration cannot be done at zero millivolts.

30

Meter Calibration

Chapter 8

Page 37

CHAPTER 9: SENSOR CALIBRATION CONDUCTIVITY/RESISTIVITY CELL CONSTANTS Best performance is normally obtained by entering the factory-documented cell constants in the Entering/Editing Cell Constants sections below. Cell constants are provided on the label of the sensors and on their certificates of calibration. However, if the sensor’s performance is not acceptable then it can be calibrated. It is important to ensure that the meter is properly calibrated and that a very accurate standard solution is available before performing a sensor calibration described at the end of this chapter.

Entering/Editing Conductivity/ Resistivity Cell Constants Upon the installation of a conductivity cell, the cell constants must be entered into the meter. Channel A has two cell constant types: one for the resistivity/conductivity sensor and one for the temperature sensor. These constants are referred to as “A Cell” and “A Temp”, respectively. Each of these types has a Multiplier and an Adder cell constant. Channel B has similar constants (“B Cell” and “B Temp”). Only the Multiplier factor is used for conductivity/resistivity and temperature. The Adder factors are left at zero. The procedure to enter and edit constants is as follows: Press the MENUS key and the following menu will appear:

Menus use arrows Press the UP arrow key until “Edit Sensor Cal” menu is displayed.

Edit Sensor Cal Press the OK/NEXT key to access this menu

A Cell M=.10000_ Chapter 9

The cursor is initially under the channel/type field. Use the UP arrow key to select the desired channel/type (either “A Cell”, “A Temp”, “B Cell”, “B Temp”). The next field selects either the Multiplier, M, or the Adder, A, factor.

A Cell M=.10000_ Use the RIGHT arrow key to move the cursor to the number field. Use the arrow keys to enter the cell constants from the sensor label. The last position in the menu is a unit prefix which should be left at “_” (blank or units); although µ (micro), m (milli), K (Kilo) and M (mega) are also available. After setting the desired value, the RIGHT arrow key can be used to move the cursor back to the channel/type field (1st field) so that the other constants can be entered and edited. The changes are recorded when you proceed to a different constant. Pressing the OK/NEXT key will accept all set values. The meter will ask if changes should be saved.

Save Changes Yes Press the OK/NEXT key to save the changes and return the display to measurement mode.

CONDUCTIVITY/RESISTIVITY SENSOR CALIBRATION Conductivity sensors can be calibrated with either a one-point or two-point calibration process. With a two-point process, the sensor must be put in two different solutions (or temperature values). Under most circumstances only a 1 point calibration process is recommended. The solution must be known with very high accuracy (either resistivity, conductivity, °C or °F) before the sensor can be calibrated. Calibration is accomplished by entering the known solution value into the meter and instructing the meter to perform a calibration.

Sensor Calibration

31

Page 38

When a one-point calibration is performed, the meter will compute a new sensor Multiplier factor. With a two-point calibration, the meter will compute new Multiplier and Adder factors. Conductivity/resistivity sensor calibration cannot be done if the measurement is set for untemperature compensated conductivity (see SELECTING A MEASUREMENT TYPE in Chapter 4). If necessary, temporarily set the instrument for compensated measurement for the calibration. (Calibration can be done, if desired, with Compensation set to “None”, accessed by the MENUS key) NOTE: The 240-501 cell with 50/cm cell constant uses a 262.4 ohm @ 25°C thermistor. With this sensor, use the factory temperature multiplier constant provided on the cell leadwire label. No further temperature calibration is possible. (Conductivity calibration may be performed in the normal manner.)

Conductivity/ Resistivity Calibration Procedure Press the CALIBRATE key and the display will show:

Calibrate Sensor Press the OK/NEXT key to proceed to the next menu.

Enter the known value. Be certain to set the proper units prefix. In this example, the prefix is “M” for Megohms. Press the OK/NEXT key when done. The meter will now display the reading of the signal to be calibrated. For example, a typical display for calibrating resistivity may appear as:

A= 18.15MΩ

OK?

When the reading has stabilized, press the OK/NEXT key to perform the calibration. The calibration process may take a few seconds to complete. When the instrument is done, the following message will be displayed:

Done, press OK Press the OK/NEXT key. The meter will display the menu for performing the second calibration point.

Do point #2 No If doing a two-point calibration, then use the UP key to change the “No” to “Yes”. Press the OK/NEXT key to perform the 2nd calibration point or to exit the menus. NOTE: Performing a sensor calibration will modify the cell constants shown in the "Edit Sensor Cal" menu such that they will not match the constants written on the cell label.

Cal Sensor Ch A Use the UP arrow key to select the desired channel. Press the OK/NEXT key when set.

Cal Ch A #1: Res Use the UP arrow key to select the measurement to be calibrated. The choices are: Res, Con, pH, Volts, °C and °F. Press the OK/NEXT key when set. The next menu will allow the known value of the signal (or solution) to be entered.

A Res

32

= 18.180 M

Sensor Calibration

Chapter 9

Page 39

PH/ORP SENSOR CALIBRATION Overview Periodic sensor calibration is necessary because pH sensors have some variation in output which can change over time, especially under harsh or contaminating process conditions. The frequency of calibration must be determined by experience with a particular application. For example, begin with daily calibration, then extend to weekly, then to monthly, bi-monthly, etc. as allowed by the drift experienced and process accuracy requirements. Such a sequence is also appropriate since new sensors will show some initial drift as they acclimate to the process. The accuracy of calibration is determined by the accuracy and traceability of the standard buffer solutions used. At initial sensor installation, acceptable performance can usually be obtained by entering the factory-documented sensor constants from the sensor label as described. This allows startup without standard buffer solutions. However, if accuracy must be optimized and for subsequent calibration, the regular procedures must be followed. Calibration is accomplished by entering the known value into the instrument and instructing it to perform the calibration. When a one-point calibration is performed, the instrument computes a new Adder (standardize offset or zero) constant. With a two-point calibration, the instrument computes new Adder and Multiplier (slope or span) constants. It should be noted that a non-functional sensor can always be calibrated at one point. Response to a changing process or a second calibration point are needed to ensure the sensor is functioning. For ORP measurement, no sensor calibration is recommended although it is possible. It is recommended that the instrument read in absolute millivolts established by factory meter calibration. The Adder constant is left at 0 and the Multiplier at 1.0. Standard ORP solutions are available for checking operation of sensors, however, their tolerance under process conditions is generally too wide to be useful for calibration.

most convenient to do a grab sample calibration on a routine basis. A two-point buffer calibration may be performed at less frequent intervals to provide full adjustment to sensor response. Buffer calibration requires removal of the sensor from the process and immersing it in standard buffer solution. It may be done at one or two points and gives the most direct traceability to the standard(s). The sensor should be rinsed well with deionized or distilled water before immersing in each standard. With the sensor removed from the process during buffer calibration, analog outputs and alarm relays are uncontrolled. Provide external means to disable any alarm or control action that could disrupt the process. Grab sample calibration allows the sensor to remain in the process but is limited to a one-point procedure. A sample is removed and measured by a portable pH system that has been previously calibrated in standard buffer solutions. The process pH must be stable enough that there is high confidence that a sample will be representative of the on-line value. For pure water samples (<20 µS/cm conductivity), the "grab sample" should be a flowing side stream not exposed to air before measurement, to minimize carbon dioxide contamination. The steps for grab sample calibration are as follows: 1.

Record the displayed pH at the time the grab sample is taken.

2.

Measure the grab sample pH.

3.

Calculate the difference of the grab sample pH minus the recorded process pH, retaining the sign (positive or negative).

4.

Add the difference value to the current displayed process pH and immediately enter the result as the calibration standard value in the procedure.

Example: 1.

The 200pH/CR instrument reads 6.50 pH when the grab sample is taken.

2.

The grab sample measures 6.20 pH on the portable system.

3.

The difference is - 0.30 pH.

Two methods are used for pH sensor calibration: buffer and grab sample. In some situations, it is

Chapter 9

Sensor Calibration

33

Page 40

4.

The current 200pH/CR instrument reading is 7.00 pH so 6.70 pH is the value entered for one-point calibration.

Press the OK/NEXT key. The meter will display the menu for performing the second calibration point.

Do point #2 No pH Calibration Procedure Press the CALIBRATE key and the display will show:

Calibrate Sensor Press the OK/NEXT key to proceed to the next menu.

If doing a two-point calibration, use the UP key to change the “No” to “Yes”. Press the OK/NEXT key to perform the 2nd point of calibration similar to the first. For a single point calibration, press OK/NEXT with “No” in the display above to exit the menus. NOTE: Performing a sensor calibration will modify the cell constants shown in the "Edit Sensor Cal" menu such that they will not match the constants written on the sensor label.

Cal Sensor Ch A Use the UP arrow key to select the desired channel. Press the OK/NEXT key when set.

Cal Ch A #1: pH Use the UP arrow key to select the measurement to be calibrated. The choices are: pH, Volts, Res, Con, °C and °F. Press the OK/NEXT key when set. The next menu will allow the known value of the signal (or solution) to be entered.

A pH

= 4.000 _

Using the arrow keys, change the display to the solution value. For buffer solutions, enter the exact pH value corresponding to its temperature. Press the OK/NEXT key. The meter will now display the uncalibrated reading to check for stability only. The actual value of the calibration will appear only after completion of the procedure. For example, a typical display may appear as:

A= 3.925

Upon the installation of a pH sensor, its calibration constants are entered into the 200pH/CR, especially when buffer solutions are not available. Channel A has two constant types: one for the pH sensor and one for the temperature sensor. These constants are referred to as “A Cell” and “A Temp”, respectively. Each of these types has a Multiplier and Adder cell constant. Channel B has similar constants (“B Cell” and “B Temp”). Only the Multiplier factor is used for temperature. Its Adder factor is left at zero. The procedure to enter and edit the constants is as follows: Press the MENUS key and the following menu will appear:

Menus use arrows Press the UP arrow key until “Edit Sensor Cal ” menu is displayed.

OK?

Edit Sensor Cal

When the reading has stabilized, press the OK/NEXT key to perform the calibration. The calibration process may take a number of seconds to complete. When the instrument is done, the following message will be displayed:

Done, press OK

34

Entering/Editing pH Calibration Constants

Press the OK/NEXT key to access this menu

A Cell M=1.0000_ The cursor is initially under the channel/type field. Use the UP arrow key to select the desired channel/type (either “A Cell”, “A Temp”, “B Cell”, “B Temp”).

Sensor Calibration

Chapter 9