Page 41

The next field selects either the multiplier, M, or the adder, A, factor.

A Cell M=.10000_ Use the RIGHT arrow key to move the cursor to the number field. Use the arrow keys to modify the number desired. The last position in the menu is a unit prefix which should be left at “_” (blank); although µ (micro), m (milli), K (Kilo) and M (mega) are also available. After setting the desired value, the RIGHT arrow key can be used to move the cursor back to the channel/type field (1st field) so that the other constants can be entered and edited. The changes are recorded when you proceed to a different constant.

The Multiplier (slope or span) is an indication of the sensitivity of the sensor to changes in pH. It has a nominal value near 1 and is recalculated after every 2-point calibration. Reduction in this value is usually due to aging, coating or hot caustic attack of the glass measuring membrane of the sensor. A sensor with Multiplier value less than 0.80 should be replaced soon. The Multiplier "M" can be viewed at any time in the Edit Sensor Cal menu. The value must not be changed on this menu (except at startup) or the sensor will have to be recalibrated.

Pressing the OK/NEXT key will accept all set values. The meter will ask if changes should be saved.

Save Changes Yes Press the OK/NEXT key to save the changes and return the display of measurement data.

PH SENSOR DIAGNOSTICS The Adder and Multiplier constants described above can provide valuable preventive maintenance information about the sensors. The Adder factor (zero, asymmetry or standardize offset) is an indication of how far the sensor has drifted from the nominal zero starting point, in pH units. It is recalculated after ever calibration. Drift in this value is usually due to aging or contamination of the reference electrode portion of the sensor. An offset of more than ±2.5 pH units or a more rapid change in this value is an indication that the sensor should be replaced soon. The Adder factor "A" can be viewed at any time in the Edit Sensor Cal menu. However, the value must not be changed in this menu or the sensor will have to be recalibrated. Because the 200pH/CR meter is auto-ranging, when the Adder Factor is very small, it may be displayed in milli-pH units with an "m" prefix following the numeric value. In this case the value is very near the nominal zero offset.

Chapter 9

Sensor Calibration

35

Page 42

CHAPTER 10: SECURITY/LOCKOUT SECURITY FEATURES

Old Pass = 00000

All menus, or only specific ones, can be protected from unauthorized use by using the lockout feature. When enabled, the menu(s) will be inaccessible unless the proper five digit password is entered. Each menu key on the front panel can be locked out individually. The functions that can be locked are: 1.

MEASURE MODE: locks the measure mode key

2.

SETPOINTS: locks the setpoint key.

3.

RELAYS: locks the relays key.

4.

OUTPUTS: locks the output key.

5.

CALIBRATE: locks the calibrate key.

6.

MENUS: locks the menu key.

7.

DISPLAY: locks the arrow keys such that the display mode for measurements cannot be changed.

After setting the current password, press the OK/NEXT key. If the current password is not properly set, then the following message will be displayed before exiting the menus:

Invalid Password If the password was properly entered, then the meter will request the new password:

New Pass = 00000 After setting the new password, press the OK/NEXT key. The meter will ask if changes should be saved.

The password can be changed at any time, although the current password must be entered before the change is allowed. The password and lockout state is retained through a power-down sequence and a system reset. All units shipped from the factory are set with the password of “00000”. It is suggested that this password be changed to another five digit sequence before using the lockout feature.

CHANGING THE PASSWORD

Save Changes Yes Press the OK/NEXT key to save the changes and return to the display of measurement data.

ENABLING THE LOCKOUT To enable the lockout, press the MENUS key and the following menu will appear:

Menus use arrows Press the UP key until the “Set/Clr Lockout” menu is displayed.

To change the password, press the MENUS key and the following menu will appear:

Set/Clr Lockout

Menus use arrows

Press the OK/NEXT key to access this menu. The next menu will request the current password.

Press the UP arrow key until the “Change Password” menu is displayed.

Password = 00000

Change Password Press the OK/NEXT key to access this menu. The next menu will request the current password.

36

Security/Lockout

Chapter 10

Page 43

After setting the password, press the OK/NEXT key. If the correct password has been entered, then the following menu will allow the enabling or disabling of the lockout feature.

Enable Lockout N If “N” is selected then the lockout is disabled and the meter will exit the menus. If “Y” is selected the next menu will allow each menu to be individually locked out.

Lock Measure

N

The first field lists the functions to be locked. Use the UP and DOWN arrow keys to access the list. The choices are: MEASURE, SETPOINT, RELAYS, OUTPUTS, CALIBRATE, MENUS, DISPLAY. When the desired function is selected, move the cursor to the next field to change the state. A “Y” indicates that this function is locked and an “N” indicates that the function is not locked. use the LEFT arrow key to move the cursor back to the function field to select another key for lockout. When done setting all the functions, press the OK/NEXT key. The meter will ask if changes should be saved.

Save Changes Yes Press the OK/NEXT key to save the changes and return to the display of measurement data. NOTE: If the lockout feature is used, it is recommended that the MENUS key also be locked to prevent any indirect changes to the meter's setup.

ACCESSING A LOCKED MENU If the lockout feature is enabled and a key is pressed that is also locked, the following message will be displayed:

Password = 00000 If the proper password is entered, then access to that menu will be allowed. This menu key will operate as normal. When the menu is exited, the lockout will be re-enabled.

Chapter 10

Security/Lockout

37

Page 44

CHAPTER 11: OTHER FUNCTIONS AVERAGING The 200pH/CR has various levels of measurement averaging or damping, each for specific applications. The options are: low, medium, high and special. Each channel can be assigned an averaging level. The assigned averaging will apply to both the primary and secondary measurements of that channel. Low averaging is useful in applications that require a fast response to changes in the system. Medium and high averaging help to reduce display fluctuations. The special averaging setting is recommended for most conductivity/resistivity applications. This method provides the most reduction in display fluctuations, while retaining fast response. Special (Spec) averaging is self-adjusting. If a large change in the measurement is detected, then the meter will respond immediately to the change (does not use any averaging). Small changes to the measurement (i.e., system noise less than 1%) will be highly averaged. If measurement noise can exceed 1%, 0.15 pH or 15 mV, then special averaging should not be used.

Use the UP and DOWN arrow keys to change the averaging level. Press the OK/NEXT key when done. The next menu will show the averaging level for channel B. Press the OK/NEXT key after setting channel B. The meter will ask if changes should be saved.

Save Changes Yes Press the OK/NEXT key to save the changes and return to the display of measurement data.

SYSTEM RESET CAUTION: A system reset will set all operational parameters to their default conditions and may require extensive reprogramming. 1.

A Primary Measurement Mode: Conductivity (fixed at µS/cm)

2.

A Secondary Measurement Mode: Temperature (DegC)

3.

To set the averaging, press the MENUS key and the following menu will appear:

B Primary Measurement Mode: Conductivity (fixed at µS/cm)

4.

B Secondary Measurement: Temperature (DegC)

Menus use arrows

5.

Display Mode: Mode #1 (A primary and B primary)

Press the UP arrow key until the “Set Averaging” menu is displayed.

6.

Setpoints (all): Off, value = 0, no relay assigned, active on over-range

7.

Relays (all): Delay = 0, hysteresis = 0, state = normal

8.

Serial Port: Data output off.

9.

Analog Outputs: no signal assigned, min = 0, max = 0

Set Averaging Press the OK/NEXT key to access this menu.

A: Average = High The cursor is initially under the channel field. Use the UP and DOWN keys to change the channel if desired. Use the RIGHT arrow key to move the cursor to the level field. The display will appear as:

A: Average = High

38

10. Manual temperature: None 11. Compensation Method: Standard method.

Other Functions

Chapter 11

Page 45

12. Cell Constants: resistivity multiplier = 0.1, temperature multiplier = 1.0, all adders = 0.

Press the UP arrow key until the “Temp Source” menu is displayed.

Temp Source

13. Auto Display Scroll: Off. A system reset will not change the password, lockout state, meter calibration, analog output calibration, or line power frequency. To reset the meter, press the MENUS key and the following menu will appear:

Menus use arrows Press the UP arrow key until “System Reset” is displayed.

Press the OK/NEXT key to access this menu.

A: Use this chan The cursor is initially under the channel field. Use the UP or DOWN arrow keys to change the channel if desired. Use the RIGHT arrow key to move the cursor to the source selection field. The display will appear as follows:

A: Use this chan

System Reset Press the OK/NEXT key to access this menu.

Reset Unit? Yes Press the OK/NEXT key to perform the reset. The meter will display a confirmation message for three seconds then exit the menus.

Unit is Reset

Selections in this menu include "Use this chan", "Use other chan" and "Fixed= XX.XX°C". If the last is chosen, using the down arrow key, the display appears as:

A: Fixed =25.00°C Set the desired temperature value. Press the OK/NEXT key when done. The meter will ask if changes should be changed.

Save Changes Yes SETTING THE TEMPERAT URE SOURCE The temperature used for display and compensation may come directly from the sensor, may come from the other channel sensor or may be set to a fixed manual value. These features are useful when a cell does not have a temperature sensor built-in, is slow responding, or if it is desirable to compensate a measurement based on a fixed temperature. NOTE: When “Use other chan” is selected, it is also necessary to enter the temperature Multiplier calibration factor for the sensor into both channels, as described in Chapter 9. To set the temperature source, press the MENUS key and the following menu will appear:

Menus use arrows Chapter 11

Press the OK/NEXT key to save the changes and return to the display of measurement data

SENDING DATA TO A PRINTER OR COMPUTER With measurements other than conductivity with 0.1/cm cell constants, an external isolator for the digital communications signal is strongly recommended to prevent ground loop problems. The 200pH/CR can be set to output measurement data automatically to a printer or computer at a fixed time interval. The time interval can be set from 1 second up to 255 seconds. The data is transmitted as a string of ASCII characters, terminated with a carriage return character. All four measurements are contained in the string. To use this feature, the baud rate, parity, and output timer must be set as follows:

Other Functions

39

Page 46

Use the arrow keys to set the desired time interval in seconds.

Setting the Baud Rate and Parity Press the MENUS key and the following menu will appear:

NOTE: Entering a value greater than 255 seconds will automatically set the timer interval to 255 seconds.

Menus use arrows

Press OK/NEXT when done. The meter will ask if changes should be saved.

Press the UP arrow key until the “Set Serial Port” menu is displayed.

Save Changes Yes

Set Serial Port

Press the OK/NEXT key to save the changes and return to the display of measurement data.

Press the OK/NEXT key to access this menu. A typical menu may appear as:

Baud=9600 P=Even

For details on communications output, refer to Manual 84364.

The cursor is initially under the baud rate setting. Press the UP or DOWN arrow keys to change the baud rate. Use the RIGHT arrow key to move the cursor to the parity field. The parity setting can be switched between even parity and no parity. Press the OK/NEXT key when done. The meter will ask if changes should be saved.

Save Changes Yes Press the OK/NEXT key to save the changes and return to the display of measurement data.

Setting the Data Output Timer Press the OUTPUTS key.

Output: Analog Press the UP-arrow key until “Serial” is displayed. Press the OK/NEXT key to access this menu.

Output off >001s Press the UP arrow key to toggle the serial output from “Off” to “On”. Setting the serial out feature to “On” enables the automatic data output. Use the RIGHT arrow key to move the cursor to the time field.

Output On >001s 40

Other Functions

Chapter 11

Page 47

CHAPTER 12: TROUBLESHOOTING OFF-LINE SELF-DIAGNOSTICS A number of diagnostic and self test functions are available via the menus. The following functions can be tested: 1.

ROM: this test is non-functional in software versions up to 3.3 and failure indications should be ignored.

2.

RAM: the data memory is tested for reading and writing.

3.

AOUT1: analog output channel #1 is tested by setting the output current from 0mA to 20mA in 1mA steps.

The cursor is under the first field which indicates the test to be conducted. The choices are: ROM, RAM, AOUT1, AOUT2, A/D, COMM, NVRAM, DISPLAY, KEYPAD or exit. Use the RIGHT arrow key to select the desired test. Press the OK/NEXT key to perform the test. Select “Exit” to exit this menu. The test results are displayed with the following messages (for example, the ROM test):

ROM: Passed Ok or

4.

AOUT2: analog output channel #2 is tested by setting the output current from 0mA to 20mA in 1mAn steps.

ROM: Failed Ok

5.

A/D: the analog to digital converter circuit (used for making measurements) is tested for functionality.

ON-LINE DIAGNOSTICS

6.

COMM: the communication port is tested for its ability to receive and transmit data. A jumper wire is connected from the transmit line to the receive line before conducting the test.

7.

8.

NVRAM: the non-volatile memory is tested for functionality. This memory is used to hold setup information in case of a power down condition (or low line voltage). DISPLAY: the display is tested by writing various patterns. This is a visual test conducted by the operator.

To perform any of these tests, press the MENUS key and the following menu will appear:

Menus use arrows Press the UP arrow key until the “Diagnostic” menu is displayed.

Error, Over-range & Sensor Error Indications A measurement that cannot be properly measured or computed or is over-range is considered to be in an error state. The display will show asterisk characters (“*”) in place of the allocated digits. An example:

A *****µS *****°C In this display, measurements could not be measured or computed properly. This condition could be caused by a disconnected or mis-wired sensor, no sample at the sensor, too high or too low sample conductivity or resistivity for the sensor/instrument to measure, sensor failure, etc. If the temperature display gives asterisks indicating a problem with that measurement then any temperature compensated measurement will also display asterisks, even if it is otherwise OK.

Diagnostic Menu Press the OK/NEXT key to access this menu.

Test? Chapter 12

ROM Troubleshooting

41

Page 48

“Check Setup” Message The 200pH/CR constantly checks the condition of the setup data (setpoints, compensation methods, etc.). If the meter detects an unauthorized change in this data (i.e., corrupted data) then a message will be displayed every few seconds as follows:

Check Setup This message can be cleared by entering and exiting any menu.

TROUBLESHOOTING Problem

Possible Cause

1.

Display is blank

• no power to unit • blown fuse • loose display cable • display contrast potentiometer needs adjustment • display cable improperly connected • circuit board failure

2.

Wrong readings

• sensor improperly installed • incorrect measurement prefix entered • incorrect cell constant entered • meter improperly calibrated • temperature compensation is incorrectly set or disabled • sensor patch cord is defective • defective sensor • circuit board failure

3.

Keypad not functioning

• keypad cable connector loose or broken • defective keypad

4.

Negative readings in %rejection

•% rejection is calculated on the wrong channel

5.

Readings fluctuate too much

• wrong line power frequency selected • cells and/or cables installed too close to equipment that generates high levels of electrical noise

6.

Data not sent out to serial port

• serial port wiring is wrong • wrong baud rate and/or parity • automatic data output not enabled • data output timer set too high

RECOVERY PROCEDURE

2.

Perform a system reset as outlined in SYSTEM RESET in Chapter 11.

3.

Re-enter the cell constants as shown in ENTER/EDITING CELL CONSTANTS in Chapter 9. Set the desired measurement mode and prefix as shown in SELECTING A MEASUREMENT TYPE in Chapter 4.

4.

Re-calibrate the meter as outlined in Chapter 8.

5.

Check that the proper compensation method is selected.

If the meter is unable to make valid measurements, then the following procedure may help to resolve the problem. 1.

42

Check sensor patch cord wiring. Look for loose wires at the terminal block or incorrect wiring.

Troubleshooting

Chapter 12

Page 49

CHAPTER 13: SERVICE FUSE REPLACEMENT The 200pH/CR is protected from accidental voltage overloading, short circuits, and related damage by a 1/8 amp time-delayed fuse (for 115 VAC units only). The 230 VAC unit uses a 1/16 amp timedelayed fuse. The fuse is located on the printed circuit board (PCB) inside of the case. WARNING: FOR CONTINUED PROTECTION AGAINST RISK OF FIRE, REPLACE ONLY WITH FUSE OF THE SPECIFIED TYPE AND CURRENT RATING. To replace the fuse: 1.

Disconnect all power to the 200pH/CR unit before proceeding.

2.

Remove the two screws from the center of the rear panel.

3.

Slowly pull the rear panel assembly out of the unit, no more than 1”.

4.

Disconnect the two ribbon cables connecting the case to the PCB.

5.

The fuse is located near the transformer. Remove the old fuse and replace it with one with the same rating as indicated above.

REDUCING 200PH/CR PATCH CORD LENGTH 200pH/CR patch cords are available in a variety of standard lengths. Occasionally it is necessary to reduce standard cord lengths to accommodate system design. The following procedure outlines how to terminate the end of the cable to assure accurate system operation. The 200pH/CR patch cords include two (2) drain (bare) wires. It is essential that these wires never make contact with each other. Place insulating tubing over these wires. CAUTION: If the wires are touching, it will cause the readings to be inaccurate. Be sure that the wires never make contact with each other. Tools required: Cable cutters, wire strippers, insulating tubing, soldering iron & solder, wire markers (optional). Procedure: 1.

Measure the cable from the end connector to the desired length and cut.

2.

Strip outer jacket and shield 4 inches from the end.

3.

Cut at the jacket the orange, yellow and all white wires that are outside the inner shield. DO NOT CUT the drain (bare) or the wires enclosed inside the inner shield.

6.

Position the PCB near the case and connect the two cables to the PCB. Make sure each cable is properly seated and oriented.

7.

Gently push the rear panel assembly back into the case. Make sure the four mounted posts align with the holes in the assembly.

8.

When the assembly is properly seated, reinstall the two mounting screws.

4.

Strip the inner shield all the way to the jacket.

5.

Strip all leads 1/4 inch and tin the inner shield.

9.

Reconnect power to the meter.

6.

Slide clear insulating tubing over the inner drain (bare) wire. Tuck the insulating tubing under the jacket. THIS WIRE MUST NEVER TOUCH THE OUTER SHIELD OR OUTER DRAIN WIRE.

7.

Place wire markers on leads if desired.

8.

Wire the cable to the 200pH/CR as indicated. See Figure 14.8 and Table 2.5.

Service

43

Chapter 13

Page 50

SPARE PARTS LIST Description

Recommended QTY Part No.

1. Fuse: For 115VAC (Type 2AG, 1/8 Amp SB, Littlefuse® #228.125) For 230VAC (Type 2AG, 1/16 Amp SB, Littlefuse® #218.063)

1 1

35088 35091

2. Display Assembly

1

06235

3. Power Selection Jumper(s): For 115VAC For 230VAC

2 1

25242 25242

4. Panel Mounting Kit (gasket, screws, nuts)

-

02181

5. 9-Pin Plug in Terminal Strip

2

22617

6. 12-Pin Plug in Terminal Strip

1

22619

7. 14-Pin Plug in Terminal Strip

1

22613

ACCESSORIES Description

Part #

Complete Conductivity Calibrator Kit (includes 1864-05, -06, -07, -08, -09, -12) High Resistivity/Low Conductivity Calibrator Kit (includes 1864-05, -06, -12) Low Resistivity/High Conductivity Calibrator Kit (includes 1864-07, -08, -09) Calibrator 500k? & 1400? (104°C) Calibrator 50K? & 1000? (0°C) Calibrator 5K? & 1400? (104°C) Calibrator 500? & 1000? (0°C) Calibrator, 0? (short) & 1097? (25°C) Calibrator, infinite ? (open) & 1097? (25°C)

1865-07 1865-05 1865-06 1864-05 1864-06 1864-07 1864-08 1864-09 1864-12

Back Cover for IP65 Rating Pipe Mounting Kit for 1-1/2 to 4” pipe (requires back cover, above) Ferrite Bead noise suppression kit (one for each sensor cable to meet CE class B requirements)

1000-62 1000-63 02183

Patch Cord, 1 ft Patch Cord, 5 ft Patch Cord, 10 ft Patch Cord, 25 ft Patch Cord, 50 ft Patch Cord, 100 ft Patch Cord, 150 ft Patch Cord, 200 ft Manual, Communications (RS232, RS422) pH/ORP BNC Preamp – for any pH/ORP electrode with BNC connector pH/ORP K9 Preamp – for any pH/ORP electrode with Ingold K9 connector

1001-67 1005-67 1010-67 1025-67 1050-67 1110-67 1115-67 1120-67 84364 1000-77 1000-85

44

Service