Page 11

Output Connections

Sensor Patch Cord Connections

Connections for all outputs are made to terminal block TB4. The serial port can be configured as an RS232 port (shown in Table 2.2.) or an RS422 port (shown in Table 2.3). An external isolator for digital communications is strongly recommended to prevent ground loop problems.

The sensors are connected to plug-in terminal blocks TB2 and TB3. Each channel has nine terminals for the sensors plus earth ground. Patch cords, 1XXX-67 series, for the 200pH/CR have a connector on one end and tinned leads on the other end. Table 2.5. shows the wiring pattern for each type of sensor.

TB2 Label

RS232 Function

DGND

Ground

TXD+

Not Used

TXD-

Transmit Data

RXD+

Not Used

RXD

Receive Data

CAUTION: Wiring for each type of sensor is different. Miswiring patch cords may damage sensors and will void all warranties. Verify wire colors in table 2.5. To meet CE electromagnetic compatibility requirements for emissions Class B, install 02183 ferrite suppression kits on each sensor patch cord close to the instrument.

Table 2.2: RS232 Connections

TB2 Label

RS422 Function

DGND

Ground

TXD+

Transmit Data Positive

TXD-

Transmit Data Negative

RXD+

Receive Data Positive

RXD-

Receive Data Negative

NOTE: For ultrapure water conductivity measurements with all-plastic piping, especially semiconductor wet benches, it is recommended to connect a jumper from one of the three earth ground terminals to sensor ground terminal SIG6(6) on TB3 for Channel A. This will prevent possible instrument damage due to electrostatic charges that build up in plastic piping systems. Do not use a ground jumper with higher conductivity water, with metal piping or with pH or ORP measurements. Terminal

Conductivity/ Resistivity pH/ORP

EARTH GND*

SHIELD

SHIELD

+5V (9)

-

BLUE

I/O (8)

-

-

SIG7(7)

BLUE

WHT/BLUE

SIG6(6)

BLACK

BLACK

SIG5(5)

RED

RED

SIG4(4)

GREEN

GREEN

Table 2.3: RS422 Connections Each analog output has + and – connections. Analog outputs are self-powered with a maximum load resistance of 500 ohms. CAUTION: Do not connect analog outputs to circuits supplying power. CAUTION: Do not connect analog output cable shield(s) to the adjacent DGND terminals. Connect shield(s) only to one of the earth ground terminals next to AC line power. TB2 Label

Analog Output Function

SIG3(3)

WHITE

WHITE

SIG2(2)

CLEAR

CLEAR

AO2-

Output 2 (-)

SIG1(1)

WHT/BLUE

-

AO2+

Output 2 (+)

AO1-

Output 1 (-)

* To any of three earth ground terminals on middle of back panel.

AO1+

Output 1 (+)

Table 2.5: Sensor Patch Cord Connections

Table 2.4: Analog Output Connections

Chapter 2

Installation

5

Page 12

Sensor Models/Types 240-, 243-

2-Electrode Conductivity

244-

4-Electrode Conductivity

363- pH or ORP

Alternative 3-Lead Conductivity Sensor Connections Tinned-lead cells with 1000 Pt or 500 Ni-Fe RTDs may be used with the 200pH/CR. These sensors, with integral leadwire and no connector, enable the lead to pass through small openings. The sensor/instrument separation is limited to less than 50 feet (20 feet recommended). Somewhat lower accuracy may result. Connections are given in Table 2.6. Jumpers should be 22 gauge to match the conductor size in the cable, for secure terminal connections. Terminals

Three-lead Conductivity Sensor Wire

+5V (9)

-

I/O (8)

-

SIG7(7)

Jumper to SIG(6)

SIG6(6)

Jumper to SIG(5)

SIG5(5)

CLEAR (shield)

SIG4(4)

WHITE

SIG3(3)

Jumper to SIG(1)

SIG2(2)

-

SIG1(1)

RED

Table 2.6: Alternative Sensor Connections With 2_8 Series Dot Two sensors there is no label with factory-supplied precision cell constants. Only nominal values are entered into the 200pH/CR. With all tinned-lead sensors, when meter calibration is desired, the cell must be disconnected and a patch cord installed in its place to accept a calibrator.

6

Installation

Chapter 2

Page 13

CHAPTER 3: USING THE 200pH/CR APPLYING POWER TO THE 200PH/CR

THE DISPLAY

After applying power to the meter, the display will show an introduction message for three seconds and then begin making measurements. This message shows the model number and the software version number as follows:

65XX

A7.76pH

Ver X.X

While the message is being displayed the instrument is performing self diagnostics. Various circuits are tested during this process and any failure will be noted with a message. The diagnostics can be repeated at any time via the menus. The default measurement display is the primary readings from the sensors on channel A and B as shown below:

A1.76µS

The 200pH/CR uses a 1 line by 16 character alphanumeric display to convey all measurement and setup information. This instrument will display one or two measurements, each with channel indication and unit of measure. A typical display of measurement data is:

B2.11µS

All 200pH/CR meters are calibrated from the factory and normally require no further calibration. If QA/QC practice requires it, the instrument be calibrated after installation. See Chapter 8 for more information on meter calibration.

B2.10µS

This display indicates that channel A is measuring 7.76pH and channel B is measuring 2.10µS/cm. The display of the other measurements can be achieved by pressing the UP or DOWN keys. In the menus, an underline cursor and flashing (bold) characters will indicate a field that can be changed. A typical menu appears as follows:

SP1=17.00 M High This menu indicates that setpoint #1 is programmed at a value of 17.00 M (million) and is set as a high limit. The cursor is under the digit “7” indicating that the UP and DOWN keys can be used to change it. The RIGHT and LEFT keys will move the cursor to the next or previous field.

Display Contrast Adjustment The contrast quality of the display can change with ambient temperature. The display contrast is adjusted from the factory for operation at standard room temperature (25°C). If the meter is operated at an ambient temperature that is much different then it may be necessary to make an adjustment. A potentiometer is accessible from the back side of the instrument to change the contrast. Use a small slotted screwdriver to gently turn the potentiometer. A counter-clockwise turn will increase the contrast and a clock-wise turn will decrease the contrast. The rear panel is shown in Figure 3.1.

Chapter 3

Using the 200pH/CR

7

Page 14

Figure 3.1: 200pH/CR Rear Panel

THE KEYPAD The 200pH/CR is equipped with an 11-key keypad as shown in Figure 3.2.

Figure 3.2: 200pH/CR Front Panel

8

Using the 200pH/CR

Chapter 3

Page 15

The keypad has 6 keys that provide direct access to specific menus as follows: 1.

MEASURE MODE measurement modes. - menus menus for to change

2.

SETPOINTS setpoints.

programming

3.

RELAYS - menus for programming relays.

4.

OUTPUTS - menus for programming outputs.

5.

CALIBRATE - menus to perform calibration.

6.

MENUS - all other menus (cell constants, security, compensation, averaging, etc.).

The other keys are referred to as control keys and are used to make changes within a menu. 1.

OK/NEXT Key - used to accept a selection and proceed to the next menu level.

2.

UP Key - up arrow is used to scroll up through a list of options.

3.

DOWN Key - down arrow is used to scroll down through a list of options.

4.

LEFT Key - left arrow is used to move the cursor to the left within a menu.

5.

RIGHT Key - right arrow is used to move the cursor to the right within a menu.

USING THE MENUS There are six menu keys across the bottom of the 200pH/CR front panel. The first five of these keys (MEASURE MODE, SETPOINTS, RELAYS, OUTPUTS, AND CALIBRATION) are used to enter specific menus. These menus allow the modification of parameters most frequently used by the operator. The sixth key labeled MENUS allows access to all other menus for various functions such as setting compensation methods, security levels, etc. The UP and DOWN arrow keys scroll vertically through the menus. Part or all of the display changes to the next option whenever an UP or DOWN arrow key is pressed. A field is defined as a section of the display that can be changed. The characters of the field will also blink. The LEFT and RIGHT arrow keys move the underline cursor across the display from one field to the next.

Chapter 3

Pressing the OK/NEXT arrow key causes the instrument to accept the options that are displayed and move to the next menu. Numbers are set one digit at a time using the arrow keys. The LEFT and RIGHT arrow keys are used to position the underline cursor below the digit to be changed. The UP and DOWN arrow keys are then used to change the value of the digit. Each digit can be scrolled through the values: .(decimal point), 0, 1, 2, 3, 4, 5, 6, 7, 8, and 9 The first digit of any number can also be set to a negative sign (-). To exit the menus either scroll completely through a set of menus with the OK/NEXT key or press any of the six menu keys at any time. The meter will display a prompt asking if the recent changes should be saved.

Save Changes Yes To save the changes press the OK/NEXT key with “Yes” on the display. To discard the changes use the UP or DOWN arrow keys to change “Yes” to “No”, then press the OK/NEXT arrow key. If the instrument is displaying a menu and a key is not pressed for two minutes, the instrument will automatically exit the menus without saving any changes. When performing a calibration the operator may need to wait for a measurement to stabilize so the menu time-out feature is automatically disabled.

Menu Example Press the MEASURE MODE key and the display will show:

A=S/cm (AUTO) This menu indicates that the Channel A primary measurement is set for conductivity (S/cm) with auto ranging. The section “S/cm” is the field to be changed and will be flashed as long as the cursor is under it. Pressing the UP arrow key will change the “S/cm” to “Ω -cm”. The RIGHT arrow key will move the cursor to the “(AUTO)” field.

Using the 200pH/CR

A=Ω/cm (AUTO) 9

Page 16

The OK/NEXT key is used to accept the entry and move to the next menu. When the last menu level is reached the following message is displayed:

a = channel A secondary measurement B = channel B primary measurement b = channel B secondary measurement

Save Changes Yes Pressing the OK/NEXT key will save the changes and exit the menus. The UP and DOWN arrow keys can be used to change the “Yes” to “No”. Pressing the OK/NEXT key with “No” will discard the changes and exit the menus.

INSTALLING A SENSOR Each probe is equipped with a cell and temperature sensor. Each of these elements have calibration constants (or cell constants) that must be programmed into the meter for proper operation. These factors are printed on the sensor as well as a Certificate of Calibration supplied with each conductivity sensor. The label may look like this:

Upper case letters are used to indicate the primary measurements and lower case letters are used to indicate the secondary measurements. Each of the four calculated measurements can be one of the following: 1.

pH

2.

ORP

3.

Resistivity

4.

Conductivity

5.

Total Dissolved Solids (TDS)

6.

Degrees C

7.

Degrees F

8.

% Rejection

9.

Difference (A-B or B-A)

RES M=1.0034 TEMP M=1.0015

10. Ratio (A/B or B/A)

RES M is the conductivity cell constant, and TEMP M is the temperature sensor constant. pH sensors will have pH M (multiplier), pH A (adder) and TEMP M (multiplier) constants. See ENTERING/EDITING CELL CONSTANTS in Chapter 9 for information on entering cell constants.

11. %HCl

The 200pH/CR instrument will measure four fundamental signals during each measurement cycle. These measurements are the conductivity or pH and temperature of the probe on channel A and the conductivity or pH and temperature of the probe on channel B. The 200pH/CR can process and display four calculated measurements. They are referred to as A primary, A secondary, B primary and B secondary. These measurements are designated by a single letter as follows:

10

13. %H2S04

DISPLAYING MEASUREMENTS Changing the Display of Measurements

MEASUREMENT DESIGNATIONS

A = channel A primary measurement

12. %NaOH

The 200pH/CR display can show either one or two measurements at a time. The display of measurements can be changed by using the UP or DOWN arrow keys. Pressing one of these keys will cause the meter to change the display mode (show an alternative set of data). The display modes for two measurements per line are: Mode #1: A primary and B primary (three significant digits displayed for each parameter):

A1.76µS

B2.11µS

Mode #2: A secondary and B secondary (three significant digits displayed for each parameter).

Using the 200pH/CR

Chapter 3

Page 17

a25.2°C b25.1°C

Menus use arrows

Mode # 3: A Primary and A Secondary (four significant digits displayed for each parameter):

Press the UP arrow key until “Display Menus” is displayed.

A1.764µS

Display Menus

25.10°C

Mode #4: B primary and B Secondary (four significant digits displayed for each parameter):

B2.109µS

25.12°C

Note that when two measurements from the same channel are displayed, the secondary measurement indicator ( a or b) is not displayed. This allows for greater precision in the display of the primary measurement.

Press the OK/NEXT key to access this menu. Use the Up or DOWN keys to toggle the field until “Disp Format” appears. Press OK/NEXT to access this menu.

Set: Disp Format Use the UP and DOWN keys to toggle the field between “1” and “2”.

Measure per Line: 1

The default display setting (after a system reset) is mode #1 (A Primary & B Primary). The display modes for one measurement per line are:

Press OK/NEXT when done. The meter will ask if changes should be saved.

Save Changes Yes

Mode #1: A Primary:

A 1.765µS/cm C

Press OK/NEXT key to save the changes and return to the display of measurement data.

Mode #2: A Secondary:

Measurement Display Scrolling a 25.25 deg C The 200pH/CR has an automatic display scrolling feature for measurement data. With this feature enabled, the display will show channel A data for 5 seconds and then show channel B data for 5 seconds. This process is repeated indefinitely.

Mode #3: B Primary:

B 2.109µS/cm C

To enable or disable this feature:

Mode # 4: B Secondary:

Press the MENUS key and the following menu will appear:

b 25.12 deg C Setting the Number of Measurements per Display Line

Menus use arrows Press the UP arrow key until “Display Menus” is displayed.

The 200pH/CR can be set to display either one or two measurements per line.

Display Menus

To change this feature: Press the MENUS key and the following menu will appear:

Chapter 3

Press the OK/NEXT key to access this menu

Using the 200pH/CR

11

Page 18

Set: Auto Scroll Use the Up or Down arrow keys to toggle the field until “Auto Scroll” appears. Press OK/NEXT to access this menu.

Auto Scroll=off Use the UP or DOWN arrow keys to toggle the field from “Off” to “On”. Press the OK/NEXT key when done. The meter will ask if changes should be saved.

Save Changes Yes Press the OK/NEXT key to save the changes and return to the display of measurement data.

ALARM INDICATIONS A setpoint can be programmed as a high limit, a low limit or a USP limit. When a measurement is higher than a high (or USP) point or lower than a low point then the setpoint is in alarm state. This condition is indicated by flashing the corresponding measurement value in the normal operating display. See Chapter 5: Using Setpoints.

12

Using the 200pH/CR

Chapter 3

Page 19

CHAPTER 4: MAKING MEASUREMENTS MEASUREMENT PROCESS The 200pH/CR will process two measurements from each of the two channels. The measurements of each channel are referred to as the primary and the secondary measurement. The instrument will process a total of four different measurements per cycle. Measurements are designated as follows:

pH is displayed with fixed range. ORP is displayed in fixed range millivolts (mV).

Resistivity Resistivity is expressed in ohms-centimeter (Ωcm). This measurement can be displayed with a prefix in front of the units. The prefixes are k (kilo or 1,000) and M (Mega or 1,000,000). The display can be set for a fixed range such as Ω-cm, KΩ-cm (1,000Ω-cm), and M (Mega or 1,000,000). The 200pH/CR can also be set for auto ranging where the range will be automatically adjusted for the most appropriate display. The range is set via the measure mode menus.

A = channel A primary measurement a = channel A secondary measurement B = channel B primary measurement b = channel B secondary measurement Note that upper case letters are used to indicate the primary measurements and lower case letters are used to indicate the secondary measurements.

1,000,000 Ω-cm = 1,000KΩ-cm = 1MΩ-cm

MEASUREMENT TYPES

Conductivity is expressed in siemens per centimeter (S/cm) and is the reciprocal of resistivity. This measurement can be displayed with a prefix in front of the units. The prefixes are m (milli or 1/1,000) and µ (micro or 1/1,000,000). The 200pH/CR can also be set for auto ranging where the range will be automatically adjusted for the most appropriate display. The range is set via the measure mode menus.

Each of the four measurements (channel A primary, etc.) can be programmed as one of the following: 1.

pH

2.

ORP

3.

Resistivity

4.

Conductivity

5.

Total Dissolved Solids (TDS)

6.

Degrees C

7.

Degrees F

8.

% Rejection

9.

Difference (A-B or B-A)

10. Ratio (A/B or B/A) 11. %HCl - Hydrochloric Acid 12. %NaOH - Sodium Chloride 13. %H2S04 - Sulphuric Acid pH and ORP Chapter 4

Conductivity

1 S/cm = 1,000 mS/cm = 1,000,000 µS/cm

Total Dissolved Solids Total Dissolved Solids (TDS) is another way to measure and display conductivity/resistivity data. TDS is the equivalent of Sodium Chloride (NaCl) required to produce the measured conductivity-approximately 0.46 ppm TDS per µS/cm. If some other conversion is desired, it is necessary to adjust the cell constant to give direct readout. For example, if a conversion of 0.6 ppm TDS per µS/cm is desired, the cell multiplier to be entered into the 200pH/CR is 0.6/0.46 x Multiplier on sensor label. See Chapter 9, Entering/Editing Sensor Constants. TDS is measured in parts per billion (ppb), parts per million (ppm), or parts per thousand (ppk). A

Making Measurements

13

Page 20

TDS reading of 10 ppm is equivalent to 10 milligrams per liter. Because of space limitations, the following abbreviations are used to display TDS units: PB = parts per billion

CONDUCTIVIY SENSOR

B

CONDUCTIVITY SENSOR

REVERSE OSMOSIS MEMBRANE

A

PM = parts per million PK = parts per thousand FEED

PRODUCT

The 200pH/CR automatically converts to TDS based on REJECT

Temperature Temperature can be measured in degrees Celsius (°C) or degrees Fahrenheit (°F). The 200pH/CR normally works with a 1000 ohm DIN platinum RTD sensor which is built into most Thornton conductivity sensors. Alternatively, the 200pH/CR can automatically recognize and measure with a 500 ohm Ni-Fe RTD temperature sensor. When configured for a 50/cm constant cell only, the 200pH/CR automatically changes its characteristic to measure from the 262 ohm @ 25°C thermistor supplied in those sensors.

Figure 4.1: % Rejection IMPORTANT: When preparing the system to perform a percent rejection measurement, the product monitoring sensor must be installed in the channel that will measure percent rejection. If the product conductivity sensor is installed in channel A, then percent rejection must be measured in channel A. Likewise if the product sensor is installed in channel B, then the percent rejection measurement must also be programmed in channel B.

Difference (A-B or B-A)

% Rejection For reverse osmosis (RO) applications, percent rejection is measured in conductivity to determine the ratio of impurities removed from product water to the total impurities in the incoming feed water. The formula for obtaining Percent Rejection is: [1 - (Product/Feed)] X 100 = % Rejection Where Product is the conductivity measurement of the first sensor and Feed is the conductivity of the second sensor. Figure 4.1 shows a diagram of an RO installation with sensors installed for Percent Rejection.

The difference measurement is computed as: Difference on channel A = A-B. or Difference on channel B = B-A. When the difference is assigned to one channel, the meter will measure the same type of measurement mode of the other channel as a basis. For example, if channel A is set to measure the difference and channel B is measuring conductivity, then the 200pH/CR will measure conductivity on both channels before computing the difference.

Ratio (A/B or B/A) This measurement is similar to the difference measurement. Ratio on channel A = A/B. Ratio on channel B = B/A.

14

Making Measurements

Chapter 4