The text of a manual · part 7 of 7
200pH/CR pH, ORP, Conductivity & Resistivity Instrument Instruction Manual, pages 61 to 64
Pages 61 to 64 of 64
The original document is the PDF (443 KB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.
Filed in the catalog under the Thornton 200pH
Page 61
200PH/CR SPECIFICATIONS Functional Cond./Resist. Ranges:
pH & ORP Ranges: Temperature Range: Temperature Compensation:
Outputs Setpoints/Alarms: Relays:
Analog Output Signals:
Serial Output:
Performance Accuracy: Repeatability: Update Rate: Ratings/Approvals: Environmental Storage Temperature: Operating Temperature: Humidity: Enclosure Display/keypad: Material: Rating: Panel Cutout: Weight: Sensor Patch Cord Length:
0.1 Constant 2-E Cell:
0.02 µS/cm to 1200 µS/cm 833 Ω-cm to 50 MΩ-cm 50 Constant 2-E Cell: 100 µS/cm to 1.0 S/cm 1.0 Ω-cm to 0.01 MΩ-cm 4-E Cell: 5 µS/cm to 500,000 µS/cm 2 Ω-cm to 0.2 MΩ-cm TDS: covers equivalent conductivity ranges Concentrations: HCl: 0-20%, NaOH: 0-15%, H2SO4: 0-20%, -1 to 15 pH, -1400 to +1500 mV -40° to 200°C, -40 to 392°F Automatic, referenced to 25°C for Resistivity, Conductivity, Percent Rejection and TDS. Field selectable for standard high purity (Thornton/Light), cation/ammonia/ETA (power industry), isopropyl alcohol, or Light 84 (special microelectronics applications). pH temperature compensation for Nernst electrode output effects plus adjustable solution temperature compensation for high purity water ionization effects, referenced to 25°C. Four, may be set as high, low or USP limits on any measurement. Any relay can be programmed to operate from multiple setpoints 2 SPDT relays, rated 5 amp max. resistive load up to 30 VDC or 250 VAC, standard; optional additional: 2 AC-only Solid State, rated 1.5 amp max, 250 VAC resistive load, 10 mA minimum current Two optional powered 4-20 mA outputs (may be recalibrated to 0-20 mA), 500 ohm load maximum, freely scalable to any parameter, isolated from input and from ground; accuracy ±0.05 mA RS232 maximum distance 50 feet. RS422 maximum distance 4,000 feet. Field selectable up to 19.2 k baud. Requires external isolation if using sensors other than 0.1/cm conductivity. Conductivity/Resistivity: ±0.5% of reading up to 18 Mohm-cm, ±0.25°C pH/ORP: ±0.03 pH, ±2 mV, ±0.3°C ±0.5% of reading for cond./resist, ±0.075°C, ±0.02 pH, ±1 mV All parameters are updated once per second UL listed, CE compliant. -40° to 70°C (-40° to °158F) -10° to 55°C (14 to 131°F) 0 to 95% RH (non-condensing)
16 character, backlit LCD (4.8 mm x 9.6 mm);11 tactile feedback keys ABS-PC alloy IP65 front panel seal, and rear with accessory back cover. 3.78” x 7.56” (96 x 192 mm) DIN 1.9 lbs. (0.9kg) 200 feet (61 m) maximum, 2-E, pH and ORP sensors 50 feet (15 m) maximum, 4-E sensors
55
Page 62
Power:
90-130 VAC or 180-250 VAC 12 Watts maximum, 50-60Hz or 12-30 VDC, 300 mA steady state, 600 mA inrush at 24 VDC. With sensors other than 0.1/cm conductivity, the DC power supply must be isolated from earth ground. On power loss all stored values are retained in non-volatile memory without batteries.
200pH/CR Models Part Number
Relays
Analog Outputs
Line Power
6520-1
2 SPDT
None
110 VAC (24 VDC)
6520-2
2 SPDT
None
220 VAC (24 VDC)
6522-1
2 SPDT
2
110 VAC (24 VDC)
6522-2
2 SPDT
2
220 VAC (24 VDC)
6542-1
2 SPDT and 2 Solid State, AC only
2
110 VAC (24 VDC)
6542-2
2 SPDT and 2 Solid State , AC only
2
220 VAC (24 VDC)
Pending CE-Declaration of Conformity Thornton Inc., 1432 Main-Street, Waltham, MA 02451, USA declares that the 200pH/CR pH, ORP and Conductivity/Resistivity Instruments, Models 6520-1, 6520-2, 6522-1, 6522-2, 6542-1, 6542-2, 6520K1, 6522K1 and 6542K1 meet the intent of Directive 89/336/EEC for Electromagnetic Compatibility. They have been designed to meet Directive 73/23/EEC for Low Voltage. Testing for compliance has been done to the following specifications. EN 50081-2 Emissions: EN 55011
EN 50082-2 Immunity: IEC 1000-4-2 IEC 1000-4-3 IEC 1000-4-4
Radiated and Conducted, Class A (Class B with accessory ferrite bead kit installed)
Electrostatic Discharge RF-Radiated Electrical Fast Transients
Additional testing for compliance will be done to the following specification. Until that testing is complete, these instruments will not bear the CE mark. IEC 61010-1 Safety
56
Page 63
UL Listing Thornton Inc., 1432 Main Street, Waltham, MA 02451, USA has obtained Underwriters Laboratories’ Listing for 200pH/CR pH, ORP and Conductivity/Resistivity Instruments. They bear the UL Listed mark signifying that the products have been evaluated to the applicable UL standards for electrical process control equipment, UL3121-1.
WARRANTY Thornton Inc. warrants products it manufactures against defects in materials or workmanship for one year from start-up or 18 months from the date of shipment from Thornton, whichever is sooner. Warranties do not apply to limited life components such as batteries. Some non-Thornton manufactured resale items may have shorter warranties than one year. Thornton honors only the warranty period of the original manufacturer. Catalog descriptions, although accurate, should not be taken as a guarantee or warranty. Thornton’s obligation under the warranty shall be to repair at its facility or replace any products which Thornton finds to be defective. Items returned for warranty must be properly packaged, shipped prepaid and insured, and must be accompanied by the Return Authorization Number described below. Returned Goods: Please contact us for a Return Authorization Number before any item is returned. Items returned for credit or exchange must be in salable condition. Items returned after 30 days from date of invoice, will be subject to a 10% rehandling charge. Note: Substitution or modification of cables voids all warranties. THE ABOVE WARRANTY IS THE ONLY WARRANTY MADE BY THORNTON INC. AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING, WITHOUT LIMITATION, IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
57
Page 64
Thornton Inc. 1432 Main Street Waltham, MA 02451-1623 (781) 839-6400 Smartline: 800-642-4418 Faxline: 781-890-5507 Email: info@thorntoninc.com
Part No. 84383 Rev. B 12/00