Page 11
INSTALLING THE SCANNER TO THE HOST (OPTIONAL FEATURES) Hand-held Port Option A Non-Decode Hand-Held Scanner can be connected to an IS8000 series fixed scanner. To avoid potential problems, do not power up the scanner until the communication cable is secured to the host. 1.
Turn off the host system.
2.
Connect the communications cable ® to the HoloTrak and to the host.
3.
Connect the communication cable of the hand-held scanner to the HoloTrak.
4.
Check the AC input requirements of the power supply for the scanner to make sure the voltage matches the AC outlet.
5.
Connect the power supply to the HoloTrak.
6.
Plug the power supply into the AC outlet.
7.
Power up the host system. When the hand-held scanner first receives power, it will immediately go through a self-diagnostic routine, then the green and red LEDs will flash, and the unit will beep once.
Caution: To maintain compliance with applicable standards, all circuits connected to the scanner must meet the requirements for SELV (Safety Extra Low Voltage) according to EN 60950. To maintain compliance with standard CSA C22.2 No. 950/UL 1950 and norm EN 60950, the power source should meet applicable performance requirements for a limited power source.
7
Page 12
CONFIGURATION TO THE HOST SYSTEM The scanner is shipped from the factory configured to a set of default conditions noted in the Default Settings section of this guide. In order for the scanner to communicate with the host system, it will need to be properly configured. Since each host system is unique, the scanner must be configured to match the host system requirements. Use the MetroSet®2 Scanner Configuration Program provided to configure the scanner.
Primary/Secondary Configuration Note During configuration, the secondary scanner’s communication protocol will be established automatically through the primary scanner. If the configuration is to be downloaded from the host to the scanners, it is required that both the primary and the secondary scanners are powered up and ready to scan. The appropriate communications cables should be connected between the host and the primary scanner and between the primary and the secondary scanner. Once a configuration is downloaded to the primary scanner, the primary will automatically configure the HoloTrak secondary scanner. This step is necessary in order to configure the secondary scanner properly. For all practical purposes, the scanners will act as one scanner. Once both units are configured, the scanner settings are stored in non-volatile RAM, and will not need to be configured again.
INSTALLING AND RUNNING THE METROSET® 2 PROGRAM Installation 1.
Close all open applications.
2.
® Insert the MetroSet 2 CD into your CD drive.
3.
The installation wizard will display the Welcome dialog box, select Next to proceed to with the installation. If the wizard does not automatically start, go to the windows Start menu, choose Run, designate the CD-ROM drive, type setup (d:\setup) and then click OK.
4.
To accept the default installation directory, select Next. To change the destination folder, select Browse to locate and choose the desired folder.
5.
To accept the default program folder, select Next. To change the default program folder type a new folder name or choose an existing folder.
6.
When the setup process is complete, select Yes, and choose Finish to restart your computer.
8
Page 13
INSTALLING AND RUNNING THE METROSET® PROGRAM How to Start To avoid potential problems, do not power up the scanner until the communication cable is secured to the host. 1.
Turn off the PC after installing the MetroSet 2 program.
2.
Connect the communication cable to the scanner and to the RS-232C serial port (COM1 or COM2) located on the PC.
3.
Check the AC input requirements of the transformer to make sure the voltage matches the AC outlet. An easily accessible socket-outlet should be installed near the equipment.
4.
Power up the scanner.
5.
Turn on the PC.
6.
From your Start menu choose Programs, MetroSet 2, MetroSet 2.
7.
Select Holotrak® as the unit to configure.
8.
Then choose Configure.
9.
For detailed instructions on the configuration options available choose Help from the menu bar.
Metrologic’s HoloSet® program can also be used to configure the scanner.
Caution: To maintain compliance with applicable standards, all circuits connected to the scanner must meet the requirements for SELV (Safety Extra Low Voltage) according to EN 60950. To maintain compliance with standard CSA C22.2 No. 950/UL 1950 and norm EN 60950, the power source should meet applicable performance requirements for a limited power source.
9
Page 14
PARTS OF THE SCANNER c Green and Red LEDs The red LED is on when the unit is done power up, the VLD is emitting light and the unit is ready to scan. The green LED flashes when the scanner has read a bar code successfully. The functions of the LED's can be reversed through special configuration with the MetroSet®2 configuration program. See the Visual and Audible Indicators section of this guide for more details.
d Speaker The speaker is configured to emit a beep when a bar code has been transmitted. The IS8800 has an additional 90 dB speaker location on the side of the case.
e Laser Output Window This aperture emits and receives laser light.
f D-type Connector Area This area contains four D-type male connectors, a 25-pin, 15-pin, and two 9-pin D-sub connectors. The 25-pin connector is designed to attach a communication cable to the host device. The 15-pin connector is used to attach two scanners as primary/secondary and/or to connect the MX001 industrial control interface to the scanner. The first 9-pin connector is used to attach the power supply to the scanner and the second 9-pin connector is used to attach a hand held 5V non-decode type scanner to the HoloTrak®. Refer to Appendix C for Pinout details.
g D-type Connector Area The 8-pin Male AMP Connector is used to attach the optional MX008 external speaker box. Refer to Appendix C for Pinout details 10
Page 15
MOUNTING BRACKET INSTALLATION
1.
Locate the mounting bolts on the sides of the scanner.
2.
Align the bolts on the scanner with the channels on the mounting bracket.
3.
Slide the unit into place until the scanner bolts sit securely in the bottom of the channels.
4.
Rotate the unit to the desired angle and secure into place with a locking bolt in the lower slot.
5.
Tighten both bolts to secure the scanner in place.
Mounting Bracket Hole Pattern
11
Page 16
VISUAL AND AUDIBLE INDICATORS When the scanner is on, the activity of the red and green LEDs indicate the status of the scanner. Flashing Red, No Green During power-up and diagnostic mode. Steady Red, No Green Unit has completed power-up, the VLD is emitting light, and the unit is ready to scan. Steady Red, Green Flash When the scanner transmits a successful read, the green LED will flash. Generally, if the green LED does not flash, then the bar code has not been successfully read. If handshaking is being used, the green LED goes out when proper handshaking has completed. When the green light turns off, communication to the host is complete.
Alternating Red and Green with razz tone This indicates a motor failure.
Steady Red, Green Flashing with a beep Scanner has entered program mode successfully. (See Running the MetroSet® 2 Program for more details on configuring the scanner).
Steady Red, Green Flashes 3 times with 3 beeps Scanner has exited program mode successfully. Note: If the scanner has been configured for reverse LED convention, via MetroSet 2, all mode indicators described above also reverse.
12
Page 17
LABELS
1S8300
1$8400/IS8500/IS8800
AVOID EXPOSURE Laser light is emitted from this aperture.
Danger - Laser light when open AVOID DIRECT EXPOSURE TO BEAM
HoloTrak® by Metrologic instruments, Inc. Blackwood, New Jersey, USA
Contains no user servicable components. Warranty void if case is opened. Complies with FCC Class A. See manual. Manufactured Blackwood, NJ.
Model: IS8XXX RS232/RS422
Barcode Scanner Serial #: 0000000000
SAMPLE
See User's Guide for patent coverage.
June 2000 (D) 12
This product complies with US DHHS Standard 21 CFR Chapter 1 Subchapter J.
L.T.E.
ACHTUNG LASERSTRAHL, LASERCLASSE 2, NICHT IN DEN STRAHL BLICKEN. LUMIERE LASER - NE PAS REGARDER DANS LE FAISCEAU APPEREIL A LASER DE CLASSE 2.
CAUTION - LASER LIGHT: DO NOT STARE INTO BEAM
CLASS 2 LASER PRODUCT
OUTPUT POWER = 7.7 MILLIWATT (PEAK) < 1 mWAVG POWER. VLO WAVELENGTH = 630 - 690 nm; EN60825-1: 1$94/A11 : 1996
CAUTION
LASER LIGHT DO NOT STARE INTO BEAM
Complies with 21 CFR Chapter 1 subchapter J.
Meets Part 15
7.7 MILLIWATT (PEAK) < 1 mWAVG POWER] Of FCC rules.
VLD WAVELENGTH = 630-690 nm CLASS Il LASER PRODUCT
See manual.
13
Page 18
DEPTH OF FIELD AND SCAN PATTERN CHARACTERISTICS IS8300 Depth of Field and Scan Pattern
The depth of field area defined is based conservatively on a 0.25 mm/0.010” bar code width. Actual scan areas will vary by label size and application.
14
Page 19
DEPTH OF FIELD AND SCAN PATTERN CHARACTERISTICS IS8400 Depth of Field and Scan Pattern
The depth of field area defined is based conservatively on a 0.33 mm/0.013” bar code width. Actual scan areas will vary by label size and application.
15
Page 20
DEPTH OF FIELD AND SCAN PATTERN CHARACTERISTICS IS8500 Depth of Field and Scan Pattern
The depth of field area defined is based conservatively on a 0.33 mm/0.013” bar code width. Actual scan areas will vary by label size and application.
16