The technical reference for the Texas Instruments CBL 2 (Calculator-Based Laboratory), written for users who program the CBL 2 and TI graphing calculators themselves. It explains the realtime, non-realtime and FastMode data collection modes, beep sequences and the FLASH archive, gives specifications for the light, stainless steel temperature and voltage sensors and custom sensor pinouts, documents every CBL 2 command with programming examples, and lists error messages and DataMate sensor defaults.
Its parts
- Introduction
- Miscellaneous Reference Information
- Technical Specifications for Sensors
- Programming the CBL 2
From the manual
Key facts from this manual
Each line names its page in the PDF.
Texas Instruments technical reference (2000) for the CBL 2 calculator-based data collection interface, written for people who program the CBL 2 themselves from TI graphing calculators. It gives the data collection modes, beep codes, sensor specifications and connector pinouts, the command set with sample programs, error codes, and DataMate default sensor settings.
- What it is
- Data collection interface for TI graphing calculators; its software sits in FLASH and can be upgraded p. 5
- Connectors
- Analog CH1, CH2, CH3, SONIC (right-hand) and DIG IN/OUT (left-hand), all 6-pin British Telecom-style p. 16
- Analog inputs
- 10-bit A/D; Vin ±10 Volts at 19.6 mV resolution, Vin-low 0 to 5 Volts at 5.6 mV p. 16
- Collection modes
- Realtime, non-realtime or FastMode (one analog channel at a very fast rate) p. 5
- Data memory
- 24K dedicated to data, up to 12,000 samples per collection p. 6, 58
- Sample rate
- Non-realtime: as fast as 1E-4 second on one channel; realtime: no more than a few points each second p. 6
- Analog current limit
- 160 mA at most from all three analog channels p. 15
- Digital output
- 4 bits wide, CMOS (0-5V) compatible, fed from a 32-element circular buffer p. 17
- Temperature sensor
- Stainless Steel Temperature Sensor, included: 20 kΩ NTC thermistor, -25 to 125°C, ±0.2°C at 0°C, ±0.5°C at 100°C p. 10, 11
- Light sensor
- TI light sensor: 10µW/cm2 to 1mW/cm2, 300nm to 1100nm, relative readings, air only p. 10
- Voltage sensor
- TI voltage sensor reads ±10 Volts on CH1, CH2 or CH3 p. 13
- Motion detector
- Channel 11 (SONIC): ½ meter to 6 meters, or 1½ feet to 18 feet p. 14
- Custom sensor parts
- Analog Probe Kit CBL/CA/D; for digital sensors, a cut CBR-to-CBL cable CBR/CA/C p. 15
- Power
- Batteries or the AC9920 adapter p. 58
What this manual walks you through
- Choosing realtime, non-realtime or FastMode data collection p. 5, 6, 7
- What the beep sequences mean p. 8, 9
- Archiving data sets and programs in FLASH memory p. 9
- Specifications of the light, temperature and voltage sensors p. 10, 11, 12, 13
- Auto-ID resistor values and ranges of supported sensors p. 14
- Building a custom analog or digital sensor p. 15
- Sensor connector pinouts p. 16
- Programming the digital output buffer p. 17
- Summary of the CBL 2 commands p. 22
- Error numbers and their causes p. 56, 57, 58, 59
Good to know
- Never connect the analog input grounds to different potentials; they are common and the CBL 2 may be damaged. p. 13
- The digital probe kit of the original CBL does not work with the CBL 2. p. 15
- The Stainless Steel Temperature Probe cannot be recalibrated. p. 12
- In FastMode, a GET sent before sampling ends aborts the sampling with an error. p. 7
- The light sensor works in air only and is not waterproof. p. 10
Questions this manual answers
How do I find out which error the CBL 2 has?
Send the request for status (Command 7) and read the error parameter of the returned list. An error usually sounds the low tone and flashes the red LED three times. p. 22, 56
What does CBL 2 error 80 mean?
The battery voltage is too low to safely write to FLASH memory; replace the batteries. p. 58
What does CBL 2 error 99 mean?
The current load on the analog or digital ports is more than the unit can supply, so it turned the power off. Correct the problem before sampling again. p. 59
How many samples can the CBL 2 store?
Up to 12,000 samples in non-realtime collection, from 24K of data memory. p. 58
Does the original CBL temperature sensor work with the CBL 2?
Yes. It auto-identifies as the Stainless Steel Temperature sensor and uses the same built-in calibration. p. 13
Contents
Chapters and sections
Each line opens the PDF at its page.
- Title Page1
- Introduction5
- Miscellaneous Reference Information5
- Technical Specifications for Sensors10
- Programming the CBL 217
- Digital Output Buffer17
- Triggering and Thresholds18
- Measuring Period and Frequency19
- Asynchronous/ Synchronous Triggering versus Record Time21
- CBL 2 Command Summary22
- Command 0 Reset CBL 2 RAM23
- Command 1 Channel Setup23
- Command 2 Data Type26
- Command 3 Trigger Setup26
- Command 4 Conversion Equation Setup (Analog)28
- Command 4 Sonic Temperature Compensation (Sonic)30
- Command 5 Data Control30
- Command 6 System Setup31
- Command 7 Request System Status31
- Command 8 Request Channel Status33
- Command 9 Request Channel Data33
- Command 10 Advanced Data Reduction33
- Command 12 Digital Data Capture34
- Command 102 Power Control Command41
- Command 115 Check Set-up Information42
- Command 116 Check Long Sensor Name42
- Command 117 Check Short Sensor Name43
- Command 1998 Set LED Command43
- Command 1999 Sound Command43
- Command 2001 Direct Output to Digital-Out Port44
The whole text · 9 parts
Read the manual here as text
Its text as extracted from the PDF, unedited, in parts; each page opens the PDF at that page.
- Title Page1-4
- Introduction5-9
- Technical Specifications for Sensors10-16
- Programming the CBL 2, pages 17 to 2617-26
- Programming the CBL 2, pages 27 to 3627-36
- Programming the CBL 2, pages 37 to 4637-46
- Programming the CBL 2, pages 47 to 5447-54
- Pages 55 to 5955-59
- Appendix C: DataMate Sensor Setup Default Settings60-63