CBL 2 Technical Reference: Technical Specifications for Sensors
Pages 10 to 16 of 63, from CBL 2 Technical Reference
The original document is the PDF (424 KB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.
For the Texas Instruments CBL2 System
Page 10
Technical Specifications for Sensors TI Light Sensor The TI light sensor uses a phototransistor to measure relative irradiance. The units of irradiance are milliwatts per square centimeter. The light sensor’s output is a voltage that is linearly proportional to the amount of irradiance it senses. The range of light over which the sensor is sensitive is 10µW/cm2 to 1mW/cm2. The auto-ID resistor in the sensor causes the CBL 2 software to automatically convert the measured voltage to relative units. The sensor is direction dependent and achieves the highest output when the end of the sensor is pointed directly at the light source. The light sensor is sensitive in the visible and near-infrared (IR) light range. This means you can use it with IR emitting diodes as well as all visible light sources. The light sensor is designed to work in air only - it is not waterproof. The light sensor returns relative readings, not absolute irradiance readings. Values may vary from light sensor to light sensor. The light sensor readings are also sensitive to temperature.
TI Light Sensor Specifications Channels
Connects to CH1, CH2, CH3 (analog channels)
Current drain
5 mA max.
Voltage range
0–5 Volts
Irradiance range
10µW/cm2 to 1mW/cm2 (approximately)
Spectral response range
300nm to 1100nm (nanometers) (non-flat response)
Chemical tolerance
None (air only)
Pins used
2 4 5 6 ground auto-ID resistor +5 Volts DC Signal
Stainless Steel Temperature Sensor The Stainless Steel Temperature Sensor is an auto-ID general-purpose laboratory temperature sensor that comes with your CBL 2. The sensor is rugged and durable, and is designed to be used as you would use a thermometer for experiments in chemistry, physics, biology, earth science, and environmental science.
10
CBL 2 Technical Reference
Page 11
This probe uses the 20 kΩ NTC Thermistor. The thermistor is a variable resistor whose resistance decreases nonlinearly with increasing temperature. The best-fit approximation to this nonlinear characteristic is the Steinhart-Hart equation. The CBL 2 or CBL interface measures the resistance value, R, at a particular temperature, and converts the resistance using the Steinhart-Hart equation: T = [K0 + K1(ln 1000R) + K2(ln 1000R)3]-1 – 273.15 where T is temperature (°C), R is the measured resistance in kΩ, K0 = 1.02119 X 10-3, K1 = 2.22468 X 10-4, and K2 = 1.33342 X 10-7. Fortunately, CBL 2 and CBL take care of this conversion for you, and provide readings in °C (or other units, if you load a different calibration).
Stainless Steel Temperature Sensor Specifications Channels
Connects to CH1, CH2, CH3 (analog channels)
Current drain
0.5 mA max.
Temperature range
-25 to 125°C (-13 to 257°F)
Maximum temperature sensor can tolerate without damage
150°C
10-bit resolution
0.32°C (-25 to 0°C) 0.12°C (0 to 40°C) 0.4°C (40 to 100°C) 1.0°C (100 to 125°C)
Temperature sensor
20 kΩ NTC Thermistor
Accuracy
±0.2°C at 0°C, ±0.5°C at 100°C
Response time
95% of full reading: 11 seconds 98% of full reading: 18 seconds 100% of full reading: 30 seconds
Probe dimensions
Probe length (handle plus body): 16 cm Stainless steel body: length 11 cm, diameter 4.0 mm Probe handle: length 5.0 cm, diameter 1.25 cm
Pins used
2 3 4 6
CBL 2 Technical Reference
Ground Vres auto-ID resistor Signal
11
Page 12
Temperature Accuracy This probe provides very accurate temperature readings. Near 0°C, readings are accurate to ±0.2°C; near 100°C, readings are accurate to 0.5°C. Important: Because of the non-linear nature of the Stainless Steel Temperature Probe, you cannot re-calibrate this sensor. Probe-specific calibrations should not be necessary when using this sensor.
Stainless Steel Temperature Sensor Chemical Tolerance The body of this sensor is constructed from grade 316 stainless steel (0.08% carbon, 2.0% manganese, 0.75% silicon, 0.04% phosphorus, 0.03% sulfur, 16-18% chromium, 10-14% nickel, 2-3% molybdenum, and 0.1% nitrogen). This high-grade stainless steel provides a high level of corrosion resistance for use in the science classroom. Here are some general guidelines for using this probe:
1. The probe handle is constructed of molded plasticized Santoprene®. While this material is very chemical resistant, we recommend that you avoid submerging the probe beyond the stainless steel portion.
2. Always wash the probe thoroughly after use. 3. The probe can be left continuously in water at temperatures within the range of –25° to 125°C. Continuous usage in saltwater will cause only minor discoloration of the probe, with no negative effect on performance.
4. You can leave the probe continuously in most organic compounds, such as methanol, ethanol, 1-propanol, 2-propanol, 1-butanol, n-hexane, lauric acid, paradichlorobenzene, phenyl salicylate, and benzoic acid. The probe should not be left in n-pentane for more than 1 hour.
5. The probe can be left in strong basic solutions, such as NaOH, for up to 48 hours with only minor discoloration. We do not recommend usage in basic solutions that are greater than 3 M in concentration.
6. The following chart provides the maximum length of time we recommend for probe exposure to some common acids. Probes left in an acid longer than these times may bubble and/or discolor, but will still be functional. We do not recommend probes be left in any acid longer than 48 hours.
Acid
12
Maximum Exposure Time
1 M HCI
20 minutes
2 M HCI
10 minutes
3 M HCI
5 minutes
1 M H 2SO 4
48 hours
2 M H 2SO 4
20 minutes
CBL 2 Technical Reference
Page 13
Acid
Maximum Exposure Time
3 M H 2SO 4
10 minutes
1 M HNO 3
48 hours
2 M HNO 3
48 hours
3 M HNO 3
48 hours
1 M CH 3CO OH
48 hours
2 M CH 3CO OH
48 hours
3 M CH 3CO OH
48 hours
1 M H 3PO 4
48 hours
2 M H 3PO 4
48 hours
3 M H 3PO 4
48 hours
7. Cole Parmer has an extensive listing of chemical compatibility of grade 316 stainless steel on their web site (www.coleparmer.com/techinfo). This listing can be used for general guidelines not covered in this summary.
TI Temperature Sensor Note If a TI Temperature sensor (the flexible temperature sensor that came with the original CBL) is used with CBL 2, it will auto-ID as the Stainless Steel Temperature sensor. Both sensors use the same calibration that is built into the CBL 2.
TI Voltage Sensor The TI voltage sensor is a generic sensor that you can use to read any voltage between ±10 Volts. The auto-ID resistor contained in the sensor causes the CBL 2 software to automatically measure voltage. No conversion equation is loaded. The black hook should be connected to ground and the red hook to the signal voltage. Channels
Connects to CH1, CH2, CH3 (analog channels)
Voltage range
±10 Volts
Chemical tolerance
None (air only)
Pins used
1 2 4
Signal Ground auto-ID resistor
Note: It is very important that the ground connections of the analog inputs are never connected to different potentials. These ground connections are all in common. Connecting the grounds to different potentials may damage the CBL 2.
CBL 2 Technical Reference
13
Page 14
Auto-ID Sensors The CBL 2 contains provisions for the auto-ID sensor resistor values listed below. If needed, a conversion equation is loaded automatically for some of the auto-ID values.
Channels 1, 2, and 3 IDENT Valu e 1
Sensor Type
Range
2.2K
Thermocouple °C
M200C to 1400¡C
33K
TI Voltage sensor
M10 to +10 Volts
2
M10 to +10 Amps
6.8K
Current sensor
3.3K
Resistance sensor
1K to 100KJ
22K
Extra long temperature sensor for °C
L50°C to 150°C
68K
CO 2 gas sensor (PPM)
0 to 5000 ppm
100K
Oxygen gas sensor (PCT)
0 to 27%
150K
C V voltage sensor (V)
M6 to +6 Volts
220K
C V current sensor (A)
M0.6 to +0.6 Amps 3
10K
Stainless steel or TI temperature sensor for °C
M25¡C to 125¡C
15K
Stainless steel or TI temperature sensor for °F
M13¡F to 257¡F
4.7K
TI Light sensor
0 to 1
1K
Ex heart rate sensor (BPM)
N/A
47K
Voltage sensor
0 to 5 Volts
1.5K
EKG
N/A
1
IDENT values are resistance values in ohms (tolerance ±5%). Operation 3 is a mathematical conversion of voltage to a current reading (1V=1A). There is no circuitry inside the CBL 2 unit to convert current to voltage; this must be done in the external probe. 3 Default units for the Stainless Steel and TI Temperature sensors is °C. 2
Channel 11 (SONIC)
1
IDENT Valu e 1
Sensor Type
Range
15K
Motion detector, meters
½ meter to 6 meters
22K
Motion detector, meters
½ meter to 6 meters
10K
Motion detector, feet
1½ feet to 18 feet
33K
Photogate sensor
N/A
IDENT values are resistance values in ohms (tolerance ±5%).
14
CBL 2 Technical Reference
Page 15
Custom Sensors To create custom-designed sensors or other circuits for the analog input channels, the sonic input channel, the digital input channel, or the digital output channel on the CBL 2, you can purchase sensor kits from TI (1-800-TI-CARES) or its Instructional Dealers. ♦
For a custom analog sensor, use the Analog Probe Kit (order entry no. CBL/CA/D). Each sensor kit includes a four-foot length of telephone cable with a connector attached to one end. The other end of the cable is not terminated.
♦
For a custom digital sensor, cut a CBR-to-CBL cable (order entry no. CBR/CA/C) into two pieces to get two lengths of cable with connectors. (The digital probe kit used with the original CBL will not work with CBL 2.)
♦
For a custom digital ID probe, contact Vernier Software and Technology (www.vernier.com) for more information.
Be very careful when designing a custom sensor or circuit. For more accurate operation, do not connect pins 1 and 6 together on the analog input channels. Pin 1 on the British Telecom-style connector is the pin farthest from the release lever as shown in the pictures below. If you design a resistance-type sensor, connect pin 3 (Vres) to pin 6 (Vin-low) (refer to “Connector Pinouts” below). Connect the resistance to be measured from the junction of these pins to pin 2 (Gnd). The resistance range for useful measurements is limited from approximately 1 Kohms to 100 Kohms. When the Operation parameter in Command 1 (page 23) is set to 2, 3, 5, 6, or 7, the data is measured on the Vin pin (pin 1). The data for all other operations is measured on the Vin-low pin (pin 6). Note: The most current that can be drained from all three analog channels is 160 mA. This is limited by the hardware.
CBL 2 Technical Reference
15
Page 16
Connector Pinouts The CBL 2 sensors use 6-pin British Telecom-style connectors.
Pin
Analog CH1, CH2, CH3
SONIC
(Right-hand Connector)
DIG IN/OUT
(Left-hand Connector)
1
Vin
Echo
DI0
2
Gnd
Init
DI1
3
Vres/Smart ID CLK
Auto-ID
DI2
4
Auto-ID
+5 Volts DC
+5 Volts DC
5
+5 Volts DC
Gnd
Gnd
6
Vin-low
Not Applicable
DI3
Vin
Vin-low
Channels:
CH1, CH 2, CH 3
CH1, CH2, CH3
Input signal:
Analog data
Analog data
Input range:
±10 Volts
0 to 5 Volts
Resolution (using CBL 2’s 10-bit A/D converter):
19.6 mV
5.6 mV
Input impedance:
1.046 MJ
>10 MJ
♦
Vres: Output reference voltage from the CBL 2 through a 15 Kohm resistor. When using this feature, Vres should be tied to Vin-low and the value to be measured should be connected between Vin-low and Gnd.
♦
Gnd: Ground (common for all channels).
♦
Auto-ID: Auto-ID sensor detection data input. (Auto-ID resistor connected from pin 4 to ground.)
♦
Echo: Ultrasonic motion detector input.
♦
Init: Distance initialization signal
♦
D0 In/Out to D3 In/Out: Input or output pins for digital pulses.
♦
Smart ID Clk: Clock to synchronize data transfer from smart probes.
16
CBL 2 Technical Reference