9.13 Changing password To be able to transmit configuration data and monitoring areas to the LSI you must be logged on as an “Authorised Client” or as “SICK Service”. This requires a password (default: “SICK_PLS”). To protect your LSI system against manipulation, you should change the default password and keep it in a safe location accessible only to authorised persons. To change the password:
• • •
From the menu choose LSI – User Category, or click on the “Logon/Logoff User Category” button on the toolbar. Log on as an “Authorised Client”, using the old password (e.g. “SICK_PLS”). Choose LSI – Password – Change for Authorised Client.
This dialogue box appears.
• •
Type in the new password twice – once in each box. Onscreen it is shown only as a series of asterisks. Confirm with “OK”.
The new password is stored in the LSI.
• •
Log off. Make a note of the new password at a location accessible only to authorised persons.
Note: The new password takes immediate effect. Please be sure to always log off before leaving your workstation! Only if you do so can the password protect your LSI system against manipulation.
8 008 310/M848/06-05-02 Technical Description • LSI © SICK AG • Industrial Safety Systems • Germany • All rights reserved
51
Page 52
9.14 Changing screen view You can enlarge, reduce or move the fields on-screen to make them look as you want. You can also choose between a circular and a rectangular grid pattern in the background. Enlarge or reduce view
•
Click on the “Zoom In” button on the toolbar.
The zoomed segment is zoomed one stage further with every click of the mouse.
•
Or click on the “Zoom Out” button on the toolbar.
The zoomed segment is unzoomed one stage further with every click of the mouse.
Move view
• •
Click on the “Move” button on the toolbar. The cursor changes shape into a four-pointed arrow. Keep the mouse button pressed down and move the screen segment to where you want it.
… or:
•
Move the view using the scroll bars at the right and bottom of the screen.
Centre view
•
52
Click on the “Centre” button on the toolbar. The zero point is now back in the centre of the screen.
8 008 310/M848/06-05-02 Technical Description • LSI © SICK AG • Industrial Safety Systems • Germany • All rights reserved
Page 53
Change grid pattern You can switch between rectangular and circular grid patterns.
• • • • • • • •
From the menu choose View – Options . In the dialogue box select the “View” tab. Check the checkbox for a rectangular or circular grid pattern. Set your desired grid width. Select the marker size for editing the fields. Set your desired maximum visible area. Check the checkbox for metric or imperial units. Check the checkboxes to display the grid pattern, labels and visible area on-screen.
Restore defaults: Click on the “Default” button. All values are reset to their defaults.
•
To confirm settings: Click on “OK”.
•
Set display resolution You can set how many measured values of the space contour are to be displayed in monitoring and editing of the fields. Note: The more measured values you display, the more accurate the display will be, but the slower it will be, too.
• • • •
From the menu choose View – Options . Select the “Space Contour” tab. Under “Display Resolution”, select how many measured values of the space contour you want to evaluate. Enter the time intervals at which new values are to be requested during editing.
8 008 310/M848/06-05-02 Technical Description • LSI © SICK AG • Industrial Safety Systems • Germany • All rights reserved
53
Page 54
9.15 Interrogating the fault memory (system diagnosis) If your LSI system is not functioning as you would like it, you can interrogate the fault memory of the LSI in a number of different steps to localise possible faults. The fault table in section 11.2 will tell you what to do to rectify the fault. Initial fault diagnosis You can carry out a simple diagnosis to localise faults.
•
From the menu choose File – Page View.
This dialogue box appears.
•
Click on “Execute”.
The diagnosis is carried out and the window displays information on the current status of your LSI system. Move the scroll bar at the right of the screen to scroll through all the entries in the fault memory. The fault codes are given at the bottom of the list. To find out what the fault codes mean, and whether you can rectify the fault yourself, refer to the fault table in section 11.2. If necessary you can also carry out a second diagnosis to get more detailed information. The more detailed diagnosis is described below. To interrogate the fault memory and the sensor status of the configured PLS units, check the “Request PLS states via LSI” box.
Interrogating the LSI fault memory
•
From the menu choose LSI – SICK Diagnosis.
This dialogue box appears.
•
Make sure zero is entered as the device address and “LSI” as the device type, and confirm with “OK”. (This assumes the device address has not yet been changed. If you have assigned the LSI a different address, you must enter the new address here accordingly.)
The “Execute Diagnosis” dialogue box appears. In it you will find various telegrams with which you can execute additional commands and functions.
54
8 008 310/M848/06-05-02 Technical Description • LSI © SICK AG • Industrial Safety Systems • Germany • All rights reserved
Page 55
• •
In the list of available telegrams select the “Read fault memory” telegram and click on “Add”. Select the “Read sensor status” telegram and click on “Add”.
The two telegrams are now entered one under the other in the list of telegrams for diagnosis.
•
Click on “Run”.
The fault memory of the LSI and the sensor status are read, and the result log is displayed in the box at the bottom of the screen. You can find out what the listed fault codes mean from the fault table in section 11.2.
You can add supplementary information to the result log, print it, or save it as a file. Enter supplementary information: Click on “Input” and enter the text you want. Your text is then appended to the result log.
•
Print result log: Click on “Print”.
•
Save result log as file: Click on “Save” and enter a file name and destination directory/folder.
•
Note: When you have rectified the fault reset the LSI: Choose LSI – Initialise. Or execute the “Reset and initialisation” telegram in the SICK diagnosis. The LSI system is then reset.
Interrogating the fault memory of a sensor If a fault is signalled by rapid flashing of the yellow LED (»4Hz) of a sensor, the sensor fault memory can be interrogated as follows:
•
From the menu choose LSI – SICK Diagnosis and enter zero as the device address and “LSI” as the device type, as described above. Click on “OK”.
The “Run Diagnosis” dialogue box appears.
•
Insert the following telegrams in the execution list one after the other: – “Switch operating mode: Diagnosis mode” – “Set device and address” – “Read fault memory” – “Read sensor status” – “Set device and address”
•
Click on “Run“.
8 008 310/M848/06-05-02 Technical Description • LSI © SICK AG • Industrial Safety Systems • Germany • All rights reserved
55
Page 56
While the telegrams are being executed, a dialogue box appears in which you need to enter the device address of the sensor:
• •
Enter the device address of the desired sensor (1 to 4) and enter “PLS” as the device type. Click on “OK”.
A second dialogue box appears in which you need to enter the device address of the LSI:
• •
Enter zero as the device address and “LSI” as the device type. Click on “OK”.
The fault memory and the sensor status of the PLS are read, and the result is displayed in the box at the bottom of the screen. For the sensor fault codes refer to the Technical Description of the PLS.
•
Eliminate the fault and reset the LSI system as described below:
In case of PLS faults:
•
Switch the power supply to the PLS off and back on again. Then choose LSI – Initialise from the menu to reset the LSI system.
In case of LSI faults:
•
From the menu choose LSI – Initialise to reset the LSI system, or execute the “Reset and initialisation” telegram in the SICK diagnosis. The LSI system is then reset.
Reading the permanent fault memory Up to 20 initial faults from the fault memory are summarised in short form and stored in the non-volatile permanent fault memory. As a result, fault codes can be displayed even after the system has been reset a number of times. If more than 20 entries are registered, the oldest is deleted.
•
From the menu choose LSI – Sick Diagnosis, as described above.
•
Insert the following telegrams in the execution list one after the other: – “Change mode: Diagnosis mode” – “Read permanent error memory”
•
Click on “Run”.
Note: The time in “days, hrs, min & sec” indicates the elapsed operating time in monitoring mode.
56
8 008 310/M848/06-05-02 Technical Description • LSI © SICK AG • Industrial Safety Systems • Germany • All rights reserved
Page 57
9.16 Deconfiguring a sensor A sensor previously connected to an LSI system is programmed for operation on an LSI. Before it can be disconnected from the LSI and re-used as a stand-alone unit, you will need to deconfigure it using the SICK diagnosis function.
•
From the menu choose LSI – SICK Diagnosis and enter zero as the device address and “LSI” as the device type, as described above. Click on “OK”.
The “Run Diagnosis” dialogue box appears.
•
Insert the following telegrams in the execution list one after the other: – “Change mode: Setup mode” – “LSI mode on/off: LSI mode OFF”
•
Click on “Run”.
All connected sensors are deconfigured. You can then disconnect them from the LSI and use them as stand-alone units. Note: If you want to use sensors without the LSI, refer to the Technical Description of the PLS.
8 008 310/M848/06-05-02 Technical Description • LSI © SICK AG • Industrial Safety Systems • Germany • All rights reserved
57
Page 58
10 Tests 10.1 Checking the LSI These tests are required in order to check that the safety devices are functioning correctly and are properly integrated into the machine/plant control, as well as to reveal any alterations or manipulations in the system. The following instructions must be observed to ensure compliance with the intended use: The units must be installed and their electrical connections made only by qualified personnel. Qualified in this context means that personnel have undergone specialist training and have gained experience of power-driven machinery/equipment. They are also required to be sufficiently familiar with the relevant national health and safety regulations, accident prevention regulations, directives and codes of practice to be able to assess the safe condition of any power-driven machinery/ equipment. Such persons are normally qualified personnel of the manufacturers of the electro-sensitive protective equipment (ESPE) or personnel who have been trained accordingly by the device manufacturers and are engaged primarily for the testing of electro-sensitive protective equipment and have been commissioned by the device operators. 1. Testing of the protective device by an expert, before initial start-up of the machine: – Test before initial start-up to confirm that the safety requirements demanded by national/international regulations, in particular, Machinery Directive and Provision and Use of Work Equipment Directive are fulfilled (EC Declaration of Conformity). – Test the effectiveness of the protective device on the machine in all operating modes programmable on the machine, based on the supplied checklist. – The personnel operating the machine protected by the safety device must be instructed by qualified personnel prior to starting work. The instruction is the responsibility of the machine operating company. Test your LSI system using the checklist reproduced in the Technical Description. 2. Regular testing of the protective device by experts:
3. Daily testing of the protective device by authorised and instructed personnel: To test your LSI system correctly: 1. The test must be carried out for the relevant preset monitoring case. 2. Check the mechanical installation to ensure that all mounting screws are secure and that the PLS is properly aligned. 3. Check each PLS unit for visible changes, such as damage, manipulation etc. 4. Switch on the machine/plant. 5. Watch the LEDs on each PLS (red, green, yellow). 6. If at least one LED is not permanently lit when the machine/plant is switched on, it is to be assumed that there is a fault in the machine or plant. In this case the machine must be shut down immediately and checked by a specialist. 7. Deliberately obstruct the protective field while the machine is running in order to test the effectiveness of the entire system. The LEDs of each tested PLS unit must change from green to red and the hazardous movement must stop immediately. Repeat this test at different points in the danger area and on all PLS units. If you discover any non-conformance of this function, the machine/plant must be shut down immediately and checked by a specialist. 8. For stationary applications, check that the danger area marked out on the floor matches the shape of the protective field stored in the PLS and that any gaps are protected by additional safety measures. In the case of mobile applications, check that the moving vehicle actually stops at the field limits which are set in the LSI and listed on the data plate in the vehicle or in the configuration log. If you discover any non-conformance of this function, the machine/plant/vehicle must be stopped immediately and checked by a specialist. 9. This test replaces the test specified in the PLS Technical Description/Operating Instructions.
– Testing must be carried out in accordance with the valid national and international regulations and standards at the intervals specified therein. The purpose of these tests is to reveal any modifications or manipulations of the protective device since the initial commissioning. – The tests must be carried out in the event of any major modifications to the machine or protective device, as well as after re-fitting or repair in the event of damage to the housing, front screen, connecting cable, etc. Test your LSI system using the checklist reproduced in the Technical Description.
58
8 008 310/M848/06-05-02 Technical Description • LSI © SICK AG • Industrial Safety Systems • Germany • All rights reserved
Page 59
10.2 Checklist Checklist for machine manufacturer/installer for the installation of Electro Sensitive Protective Equipment (ESPE) Dependent upon the application, the below listed checks are a minimum when placing an ESPE in operation for the first time. For reference purposes the checklist should be retained or stored with the machine documents. 1. Are the relevant safety standards incorporated into the machine build? Will they satisfy the Regulations? Yes o
No o
2. Are the standards listed in the Declaration of Conformity? Yes o
No o
3. Is the ESPE the correct Type and interfaced to the correct Category? No o Yes o
4. Is access to the danger zone/point of danger only possible through the ESPE? No o Yes o
5. Are measures in place to prevent standing between the ESPE and the danger zone (mechanical point-of-operation guarding)? If so, are these measures secured against removal? No o Yes o
6. Are all monitoring areas selectable by way of the LSI configured such that the measures in place as per item 5 remain operational? No o Yes o
7. Are additional mechanical means positioned to avoid reaching over, under or around, and are they secure against manipulation? No o Yes o
8. Has the overall machine stopping time been checked and documented (on the machine and/or in the machine documents? No o Yes o
10. Is the ESPE correctly fixed and secured against movement after setting in its fixed position? Yes o
No o
11. Are the required protection measures against electric shock in place (protection class)? Yes o
No o
12. Are the monitoring case switching signals for the LSI inputs A, B, C, D of two-channel design (For more details see LSI Technical Description, section 8.2)? No o Yes o
13. Is the re-set / re-start switch for the ESPE installed and fitted to the correct standard? No o Yes o
14. Are the OSSDs of the ESPE connected in accordance with the machine circuit diagram? No o Yes o
15. Have the protective functions been inspected in accordance with the inspection instructions of this document? Yes o
No o
16. Are the switching elements controlled by the ESPE, e. g. contactors, valves, monitored? No o Yes o
17. Is the ESPE effective during the entire hazardous state? Yes o
No o
18. Is the Daily Check Requirement sign positioned in a place visible to the operator? No o Yes o
This checklist does not replace the initial commissioning or regular inspections by qualified personnel.
9. Is the resultant safety distance observed between the danger point and the ESPE in all operating modes (monitoring cases)? No o Yes o 8 008 310/M848/06-05-02 Technical Description • LSI © SICK AG • Industrial Safety Systems • Germany • All rights reserved
59
Page 60
11 Care and maintenance Every time the front screen of a PLS is replaced (see PLS Technical Description, section 11, “Care and maintenance”), the contamination measurement must be calibrated.
• •
From the menu choose LSI – Tools – Calibrate Contamination Measurement. Select the PLS on which you want to calibrate the contamination measurement via the LSI.
11.1 LEDs on the LSI On the LSI and the sensors there are light-emitting diodes (LEDs) which deliver important information indicating whether your system is working correctly. You can check here to see what the functions of the LEDs are.
Note: The contamination measurement may only be calibrated directly after replacement of the front screen.
Red Red Green Yellow Green
LSI LEDs: Status
OSSD WEAK/ ERROR (green) (yellow)
RES (yellow)
OSSD (red)
Yellow
Yellow
Protective field free Object in protective field 1 Hz
Contamination warning * Contamination *
»4 Hz
Fatal error **
RES
Start-up testing 1 Hz
Waiting for reset/restart PLS LEDs: Status
Green
Yellow
A
OSSD WEAK/ OSSD B ERROR
RES
LEDs on the LSI
Red
Protective field free Object in protective field 1 Hz
Contamination warning * Contamination *
»4 Hz
Fatal error **
Red Yellow Green
Start-up testing 1 Hz
Waiting for reset/restart Output level on LSI: Status
Output OSSD
Output Warning field
Output ERROR
Protective field free Warning field free Object in protective field Object in warning field ***
LEDs on the PLS sensor
Contamination warning * Contamination * »4 Hz
Fatal error ** Start-up testing Waiting for reset/restart
= LED lit 1 Hz = LED flashing slowly »4 Hz = LED flashing rapidly = Output switches to high = Output switches to low = Output is constantly low »4 Hz = Output alternates between high and low * Clean front screen with plastic cleaner and a soft cloth. ** System error: Carry out system diagnosis (see section 9.15). *** Object in warning zone: not indicated by the yellow LEDs of the PLS and LSI.