Page 21
Configuring the hardware You can edit the received configuration or create a new one. To create a new configuration:
• •
From the menu choose File – New and click on “LSI Configuration”. Click on “OK”.
The following dialogue box appears: This is where you set the sensor type you want to work with.
• •
Select your sensor type or have the sensor automatically detected by clicking on “Detect”. Click on “Continue”.
The remaining steps to follow and dialogue boxes which appear are the same as the following steps under “Edit configuration”.
To edit the received configuration:
• •
From the menu choose LSI – Configuration – Edit . Or choose the “Edit Configuration” button from the toolbar.
The following dialogue box appears: This is where you set the address under which the LSI and the sensors are to be addressed.
•
•
Select whether you want to use a universal or single address. Universal address (zero): This is the recommended setting. If you set “Universal address”, the stored configuration can be transferred at a later time to any other LSI. Single address (between 5 and 126): If you set “Single address” you assign the LSI the address specified here for transferring the configuration. It is then only possible to reload a configuration at a later time if the address specified here and the one stored in the LSI match. This makes sense when you want to ensure that a configuration saved as a file can only be transmitted to a specific LSI unit. Click on “Continue”.
The following dialogue box appears: This is where you define the area of application of your LSI system.
• •
•
Select whether you want to use the LSI for area protection or for protection on a vehicle (e.g. an automated guided vehicle – AGV). Select which evaluation you want to use. Standard: This is the standard evaluation. Pixel suppression: In this setting, objects detected only by a single pixel per scan are ignored. This can be useful in preventing unintended shutdowns. Click on “Continue”.
8 008 310/M848/06-05-02 Technical Description • LSI © SICK AG • Industrial Safety Systems • Germany • All rights reserved
21
Page 22
The following dialogue box appears: This is where you set which inputs and outputs you are using.
• • •
•
Check the checkbox to indicate if you have connected incremental encoders. You can use incremental encoders if you are operating the LSI on a vehicle (see section 9.7). Select which inputs and outputs you are using by placing a check against them. Check the checkbox to indicate if external device monitoring (EDM) is to be active for the outputs you are using. How to connect the external device monitor is described in section 8.2. Click on “Continue”.
The following dialogue box appears: This is where you set the restart response of the outputs and the start-up testing.
•
•
•
For the configured outputs, select how the LSI is to restart after a protective field infringement. With restart inhibit: The system only restarts after a protective field infringement or a reset when the protective field is free and the restart button is pressed. Without delay (without restart inhibit): The system restarts as soon as the protective field is free after the preset response time (multiple scans). Restart delayed by n seconds: The system only restarts when the time set here has elapsed after the protective field has become free again. You can also select whether you want to work with start-up testing. If you select this option, you must intentionally interrupt the protective field once after switching on the system. In this way you can stipulate that, after switching on the system, the machine operator must first verify the detection of the PLS by intruding into the protective field fefore beginning work. Click on “Continue”.
The following dialogue box appears: This is where you log on the sensors you are using. When the system is shipped one sensor is already logged on. To log on other sensors:
• • •
22
Click on “Add”. Enter a name for the sensor and click on “OK”. The sensor is logged on. Click on “Continue”.
8 008 310/M848/06-05-02 Technical Description • LSI © SICK AG • Industrial Safety Systems • Germany • All rights reserved
Page 23
Defining monitoring areas The following dialogue box appears: This is where you define the monitoring areas to be used. One monitoring area is already defined when the system is shipped (see also section 9.4, “Expanding the configuration”). To define other monitoring areas:
• •
Click on “Add”. Enter a name for the monitoring area and click on “Continue”.
The following dialogue box appears:
•
Select the shape of the protective field and warning field. You can define the sizes of the fields now or later. If you want to define them now, click on “Dimensions” and enter the dimensions you want.
Note: You will find more details on editing the protective fields and warning fields in section 9.5.
• •
Click on “Continue”. Confirm your settings with “OK”.
The monitoring area is set.
•
Click on “Continue”.
8 008 310/M848/06-05-02 Technical Description • LSI © SICK AG • Industrial Safety Systems • Germany • All rights reserved
23
Page 24
Defining monitoring cases The following dialogue box appears: This is where you define the monitoring cases to be used. One monitoring case is already defined when the system is shipped. You can change its settings as you wish. To change a monitoring case:
•
Click on “Edit” and change the settings as you need.
To define other monitoring cases:
• •
Click on “Add”. Enter a name for the monitoring case and click on “Continue”.
The following dialogue box appears:
• • •
Select the sensor you want to be active in this monitoring case. Select the monitoring area you want to monitor on the selected sensor. Click on “Continue”.
The following dialogue box appears:
•
For the existing inputs, define the input information on which you want activation of the monitoring case to be based: X = low or high 0 = low 1 = high
Note: Each input consists of two signals of which the states must always be inverted, e.g. for input A signals A1 and A2: – Input A low: A1 = 1 and A2 = 0 – Input A high: A1 = 0 and A2 = 1 The value set in the dialogue box always corresponds to the signal level of the assigned input A2, or B2, C2, D2. (For the wiring of the inputs see section 8.2.)
•
24
Click on “Continue”.
8 008 310/M848/06-05-02 Technical Description • LSI © SICK AG • Industrial Safety Systems • Germany • All rights reserved
Page 25
The following dialogue box appears:
• •
Select the output to be switched in this monitoring case if there is an object in the protective field. Click on “Continue”.
The following dialogue box appears: This is where you set the order in which the monitoring cases are to be activated. Note: You can alter this setting later. For more details on this and on switching between several monitoring cases refer to “Additional monitoring cases” in section 9.4.
• •
Select “Any” as the order, for example. Click on “Continue”.
The following dialogue box appears: This is where you set how often (that is, in how many consecutive scans) the sensor must detect an object in the protective field before the configured outputs are shut off (between 2 and 16 scans are possible). (If you are using the LSI together with PLS type 101-316 for vehicle protection, the response time of the LSI is fixed at 270 ms.)
•
•
Enter the desired number of scans. Note: For safety reasons, always select the lowest possible setting! If you increase the setting the system will become more stable, but will also respond more slowly. The current response time is shown in the box. Click on “Continue” and confirm your settings with “OK”.
The monitoring case is defined.
• •
Click on “Continue”. Click on “OK” to accept the configuration.
The configuration is defined. You can now transmit the settings to the LSI, as described in the following section.
8 008 310/M848/06-05-02 Technical Description • LSI © SICK AG • Industrial Safety Systems • Germany • All rights reserved
25
Page 26
Sending a configuration to the LSI
• •
Make sure you are logged on as an “Authorised Client” (see status bar at bottom of screen). If you are not logged on as an “Authorised Client”, choose LSI – User Category from the menu, or click on the “Logon/ Logoff User Category” button on the toolbar.
The following dialogue box appears:
• •
Choose “Authorised Client” from the category list. Enter the password “SICK_PLS” and click on “Logon”.
You are now logged on and can send data to the LSI.
•
From the menu choose LSI – Configuration – Send to LSI, or click on the “Send Configuration” button on the toolbar.
The screen displays a multi-page overview of the configuration settings for you to check through once again. To correct settings:
•
Click on “Cancel” to close the overview and change the settings.
To confirm settings:
•
Click on “Confirm”.
The configuration data are sent to the LSI and stored there.
The following dialogue box appears:
• •
26
If you want to transmit the monitoring areas unchanged, click on “Yes”. You can then skip the next section, “Editing a monitoring area”. If you also want to alter the shape and size of the monitoring areas, click on “No”. You can then edit the protective and warning fields, as described in the following section.
8 008 310/M848/06-05-02 Technical Description • LSI © SICK AG • Industrial Safety Systems • Germany • All rights reserved
Page 27
Editing a monitoring area The screen displays the protective field you have defined. You can edit the sizes of all protective fields and warning fields after defining them.
• • •
•
From the menu choose Monitoring area – Edit, or click on the “Edit Monitoring area” button on the toolbar. From the “Active Field” list select the protective field or warning field you want to edit. From the “Active Background Field” list select the protective field or warning field that you want to see in the background for comparison purposes. From the “Active Sensor” list select the sensor of which you want to see the current defined space contour in the background for comparison purposes.
The selected fields and the space contour of the selected sensor are shown on-screen. The status bar at the bottom of the screen contains a colour key for on-screen display of the protective field and warning field. Note: The space contour of the active sensor serves only as an aid to defining the protective and warning fields. The sensor you select here has no influence on the linking of monitoring areas and sensors in the monitoring cases.
Convert field shape:
•
To convert a field into a different shape, choose Edit – Convert Into from the menu.
Various field shapes are available to choose from: – Segmented field: You can select various resolutions. The more segments a field has, the higher is its resolution. You can define the co-ordinates for each segment individually. – Rectangle: This is the default. Newly defined fields are always this shape, unless you select a different setting. You can adjust the height of the rectangle and the width at the right and left. – Semi-circle: Here you define the radius. – Segmented rectangle / Segmented semi-circle: You define a rectangle or a semi-circle and select the desired resolution (number of segments). The field is automatically converted into a segmented field with the appropriate dimensions. Notes: Warning fields are always segmented fields. If you have defined a warning field as a rectangle or semi-circle, it is automatically converted into a segmented field with the appropriate dimensions. When you convert a field or change the resolution of a segmented field, the shape of the field may change minimally. The following sections set out only the basic means of defining the sizes of the various field shapes. For more information on editing fields, refer to section 9.5.
8 008 310/M848/06-05-02 Technical Description • LSI © SICK AG • Industrial Safety Systems • Germany • All rights reserved
27
Page 28
Define rectangular field:
•
To set or move a corner point: Double-click on the desired position with the mouse.
Note: In the example a rectangular grid pattern was chosen for the screen view. You can switch between circular and rectangular grid patterns. For more details refer to section 9.14.
… or:
•
From the menu choose Edit – Field Co-ordinates.
This dialogue box appears, showing the dimensions of the rectangle.
• •
Enter the dimensions you want. Confirm with “OK”.
The dimensions of the rectangle are changed accordingly.
Define semi-circle:
•
Double-click with the mouse on the desired position to define the radius of the field.
… or:
•
From the menu choose Edit – Field Co-ordinates.
This dialogue box appears, showing the radius of the semi-circle.
• •
Enter the dimension you want. Confirm with “OK”.
The radius of the semi-circle is changed accordingly.
28
8 008 310/M848/06-05-02 Technical Description • LSI © SICK AG • Industrial Safety Systems • Germany • All rights reserved
Page 29
Define segmented field:
• • •
To set a point with the mouse: Double-click on the desired position. To move a point with the mouse: Select the point and drag it to the desired position. To delete a point: Select the point and click on the “Delete” button on the toolbar.
… or:
•
From the menu choose Edit – Field Co-ordinates.
This dialogue box appears, showing the co-ordinates of all defined points. You can add any points to the list, or you can select a point and edit it or remove it.
• • •
To set a point: Click on the “Add” button and enter your desired coordinates in the dialogue box. To move a point: Select the point in the list and click on the “Edit” button. Enter your desired co-ordinates in the dialogue box. To delete a point: Select the point in the list and click on the “Remove” button.
Note: You can also select a point with the mouse before choosing Edit – Field Co-ordinates. The co-ordinates of the point in question are then already highlighted in the list.
•
When you have defined the protective and warning fields as you want, deactivate the Monitoring area – Edit menu function, or deactivate the “Edit Monitoring area” button on the toolbar.
Note: Intrusion into needle-shaped protective fields or parts of protective fields (segments comprising just one pixel) are ignored by the system for the sake of greater availability. If protective field shapes of such kinds are unavoidable, always programme at least one additional adjacent pixel.
8 008 310/M848/06-05-02 Technical Description • LSI © SICK AG • Industrial Safety Systems • Germany • All rights reserved
29
Page 30
Sending a monitoring area to the LSI
•
From the menu choose Monitoring area – Send to LSI, or click on the “Send Monitoring area” button on the toolbar.
The following dialogue box appears:
•
Click on “Add All”, or – if you only want to transmit individual fields – select the fields you want from the list at the top and click on “Add”.
The fields are entered in the list at the bottom.
•
Click on “Send” and confirm for each individual field.
The following dialogue box appears: The transmitted fields are now marked with asterisks in the list.
• •
Check that all protective fields and warning fields are marked with an asterisk, and so have been transmitted correctly. If you already sent the configuration before, you can click on “No”. If you have not yet sent the configuration, click on “Yes” and transmit it as previously described under “Send configuration to LSI”.
When you have transmitted the configuration and all monitoring areas, the system is ready for operation.
Note: After programming, check on the plant or vehicle that all monitoring areas are the correct size and shape! You can do this by intruding intentionally into the monitoring areas. Only start up the plant or vehicle when you are sure all monitoring areas are operating effectively! When leaving your workstation log off by way of the “LSI – User Category” menu function! Also change the “Authorised Client” logon password. Make a note of the new password at a location accessible only to authorised persons. This will prevent unauthorised persons from manipulating the LSI system. (How to change the password is described in section 9.13).