Operating Instructions: CLV490 Bar Code Scanner, Advanced Line: 5 Electrical installation, pages 61 to 70

Pages 61 to 70 of 266, from Operating Instructions: CLV490 Bar Code Scanner, Advanced Line

The original document is the PDF (5.1 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.

For the Sick CLV490

Page 61

Electrical installation

Operating Instructions

Chapter 5

CLV 490 Bar Code Scanner

CLV with heater on CDM 490

5.3.5

Cable no. 2 030 065/2 031 034 (connector cover with parameter memory)

Pin assignment of 15-pin D Sub HD cable plug ("Host/Term" connection): 6 1

11

5 10

15

Signal

15-pin plug

Connector cover 15-pin socket

VS

1

–

RxD (RS 232), Aux

2

2

TxD (RS 232), Aux

3

3

Term (RS 422/485)

4

4

GND

5

–

RD+ (RS 422/485), Host

6

6

RD– (RS 422/485), Host RxD (RS 232), Host

7

7

TD+ (RS 422/485), Host

8

8

TD– (RS 422/485), Host TxD (RS 232), Host

9

9

CAN H

10

10

Reserved

11

11

CAN2 H

12

12

CAN2 L

13

13

Reserved

14

14

CAN L

15

15

Shield

Housing

Housing

Table 5-10: Pin assignment: 15-pin D Sub HD plug of the cable no. 2 030 065/no. 2 031 034

Pin assignment of 15-pin D Sub HD cable socket ("I/O" connection): 10 5

1

15

11

6

Signal

15-pin socket

Connector cover 15-pin plug

VS

1

–

IN 1

2

2

Sensor

3

3

Result 1

4

4

GND

5

5

IN 0

6

6

IN 2

7

7

Result 2

8

8

INGND

9

9

Result 3

10

10

IN 3

11

11

IN 4

12

12

I2C SDA

13

–

I2C SCL

14

–

Result 4

15

15

Shield

Housing

Housing

Table 5-11: Pin assignment: 15-pin D Sub HD socket of the cable no. 2 030 065/no. 2 031 034

8 009 993/O824/20-12-2004

© SICK AG · Division Auto Ident · Germany · All rights reserved

5-11

Page 62

Chapter 5

Electrical installation

Operating Instructions CLV 490 Bar Code Scanner

Open end (power supply): Signal

Wire

Ferrules

Connector cover 15-pin plug

15-pin socket

VS

1 red

1

1

GND

2 black

5

5

Shield

3 grey

Housing

Table 5-12: Pin assignment: open end of the cable no. 2 030 065/no. 2 031 034

5-12

© SICK AG · Division Auto Ident · Germany · All rights reserved

8 009 993/O824/20-12-2004

Page 63

Electrical installation

Operating Instructions

Chapter 5

CLV 490 Bar Code Scanner

CLV with heater on CDM 490

5.3.6

Cable no. 2 027 543 (connector cover with parameter memory)

Pin assignment of 15-pin D Sub HD plug in connector cover "AMV/BMV": 6 1

11

5 10

15

Signal

15-pin plug

Connector cover 15-pin socket

VS

1

1

RxD (RS 232), Aux

2

2

TxD (RS 232), Aux

3

3

Term (RS 422/485)

4

4

GND

5

5

RD+ (RS 422/485), Host

6

6

RD– (RS 422/485), Host RxD (RS 232), Host

7

7

TD+ (RS 422/485), Host

8

8

TD– (RS 422/485), Host TxD (RS 232), Host

9

9

CAN H

10

10

Reserved

11

–

Reserved

12

–

Reserved

13

–

Reserved

14

–

CAN L

15

15

Shield

Housing

Housing

Table 5-13: Pin assignment: 15-pin D Sub HD plug of the cable no. 2 027 543

Pin assignment of 15-pin D Sub HD socket in connector cover "AMV/BMV": 10 5

1

15

11

6

Signal

15-pin socket

Connector cover 15-pin plug

VS

1

–

IN 1

2

2

Sensor

3

3

Result 1

4

4

GND

5

5

IN 0

6

6

IN 2

7

7

Result 2

8

8

INGND

9

9

Result 3

10

10

IN 3

11

11

IN 4

12

12

Reserved

13

–

Reserved

14

–

Result 4

15

15

Shield

Housing

Housing

Table 5-14: Pin assignment: 15-pin D Sub HD socket of the cable no. 2 027 543

8 009 993/O824/20-12-2004

© SICK AG · Division Auto Ident · Germany · All rights reserved

5-13

Page 64

Electrical installation

Chapter 5

Operating Instructions CLV 490 Bar Code Scanner

CLV without heater on power pack 10 5

1

15

11

6

5.3.7

Cable no. 2 020 303

Pin assignment of 15-pin D Sub HD socket/open end ("Host/Term" connection): Signal

15-pin socket

Wire color

VS

1 red

RxD (RS 232), Aux

2 white

TxD (RS 232), Aux

3 brown

Term (RS 422/485)

4 violet

GND

5 blue

RD+ (RS 422/485), Host

6 green

RD– (RS 422/485), Host RxD (RS 232), Host

7 yellow

TD+ (RS 422/485), Host

8 grey

TD– (RS 422/485), Host TxD (RS 232), Host

9 black

CAN H

10 grey-pink

Reserved

11 red-blue

CAN2 H

12 white-green

CAN2 L

13 brown-green

Reserved

14 white-yellow

CAN L

15 yellow-brown

Schirm

Housing orange

Table 5-15: Pin assignment: 15-pin D Sub HD socket/wire colors of cable no. 2 020 303

CLV without heater on power pack 6 1

11

5 10

15

5.3.8

Cable no. 2 020 264

Pin assignment of 15-pin D Sub HD plug/open end ("I/O" connection): Signal

15-pin plug

Wire color

VS

1 red

IN 1

2 white

Sensor

3 brown

Result 1

4 violet

GND

5 blue

IN 0

6 green

IN 2

7 yellow

Result 2

8 grey

INGND

9 black

Result 3

10 grey-pink

IN 3

11 red-blue

IN 4

12 white-green

I2C SDA

13 brown-green

I2C SCL

14 white-yellow

Result 4

15 yellow-brown

Shield

Housing orange

Table 5-16: Pin assignment: 15-pin D Sub HD plug/wire colors of cable no. 2 020 264

5-14

© SICK AG · Division Auto Ident · Germany · All rights reserved

8 009 993/O824/20-12-2004

Page 65

Operating Instructions

Electrical installation

Chapter 5

CLV 490 Bar Code Scanner

CLV without heater on power pack

5.3.9

Cable no. 2 020 981 (connector cover with parameter memory)

Cable 1, open end ("Host/Term" connection):

Cable 2

Cable 1

Signal

Wire color

Connector cover 15-pin socket

VS red and pink

1

RxD (RS 232), Aux white

2

TxD (RS 232), Aux brown

3

Term (RS 422/485)

violet

4

GND blue and grey-brown

5

RD+ (RS 422/485), Host green

6

RD– (RS 422/485), Host RxD (RS 232), Host yellow

7

TD+ (RS 422/485), Host grey

8

TD– (RS 422/485), Host TxD (RS 232), Host black

9

CAN H grey-pink

10

Reserved red-blue

11

CAN2 H white-green

12

CAN2 L brown-green

13

Reserved white-yellow

14

CAN L yellow-brown

15

Shield orange

Housing

Table 5-17: Wire colors of cable no. 2 020 981, cable 1 ("Host/Term" connection)

Cable 2, open end ("I/O" connection): Signal

Wire color

Connector cover 15-pin plug

VS red and pink

1

IN 1 white

2

Sensor brown

3

Result 1 violet

4

GND blue and grey-brown

5

IN 0 green

6

IN 2 yellow

7

Result 2 grey

8

INGND black

9

Result 3 grey-pink

10

IN 3 red-blue

11

IN 4 white-green

12

Reserved

–

13

Reserved

–

14

Result 4 yellow-brown

15

Shield orange

Housing

Table 5-18: Wire colors of cable no. 2 020 981, cable 2 ("I/O" connection)

8 009 993/O824/20-12-2004

© SICK AG · Division Auto Ident · Germany · All rights reserved

5-15

Page 66

Electrical installation

Chapter 5

Operating Instructions CLV 490 Bar Code Scanner

CLV with/without heater on power pack

Cable 1

Cable 2

5.3.10

Cable no. 2 021 267 (connector cover)

Cable 1, open end (connection for data and function interfaces): Signal

Wire color

Connector cover 15-pin plug

15-pin socket

IN 1 white

2

–

Sensor brown

3

–

Result 1 green

4

–

IN 0 yellow

6

–

IN 2 grey

7

–

Result 2 pink

8

–

INGND blue

9

–

Result 3 red

10

–

IN 3 black

11

–

IN 4 violet

12

–

Result 4 grey-pink

15

–

RxD (RS 232), Aux red-blue

–

2

TxD (RS 232), Aux white-green

–

3

Term (RS 422/485)

brown-green

–

4

RD+ (RS 422/485), Host white-yellow

–

6

RD– (RS 422/485), Host RxD (RS 232), Host yellow-brown

–

7

TD+ (RS 422/485), Host white-grey

–

8

TD– (RS 422/485), Host TxD (RS 232), Host grey-brown

–

9

Shield orange

Housing

Table 5-19: Wire colors of cable no. 2 021 267, cable 1 (connection for data and function interfaces)

Cable 2, open end (connection for power supply): Signal

Wire color

Connector cover 15-pin plug

15-pin socket

VS red

–

1

GND black

–

5

Table 5-20: Wire colors of cable no. 2 021 267, cable 2 (connection for power supply)

The signals of the CAN interface (CAN H, CAN L, CAN2 H, CAN2 L) are not available.

5-16

© SICK AG · Division Auto Ident · Germany · All rights reserved

8 009 993/O824/20-12-2004

Page 67

Operating Instructions

Electrical installation

Chapter 5

CLV 490 Bar Code Scanner

5.4

Preparations for electrical installation

5.4.1

Requirements for the host interface

The host interface of the CLV can be operated as an RS 422/485 or an RS 232 interface. Table 5-21 shows the recommended maximum cable lengths as a function of the selected data transfer rate. Interface type

Transfer rate

Distance from host

RS 232

Up to 19,200 bd 38,400 to 57,600 bd

Max. 10 m (32.8 ft) Max. 3 m (9.84 ft)

RS 422/4851)

Max. 38,400 bd Max. 57,600 bd

Max. 1,200 m (3,936 ft) Max. 500 m (1,640 ft)

1) with suitable line termination according to specifications

Table 5-21: Maximum cable lengths between the CLV and host

¾

To prevent interference, do not lay the cables parallel with power supply and motor cables over long distances, e. g. in cable ducts.

5.4.2

Supply voltage

The CLV requires a supply voltage of 18 to 30 V DC for operation without a heater and 24 V DC +20 %/–10 % for operation with a heater (functional extra-low voltage in accordance with IEC 364-4-41). The functional extra-low voltage can be generated by using a safety transformer pursuant to IEC 742. The power consumption of the individual types is as follows: Type

Voltage

Scanning method

Heater

Power consumption1)

Connection module

CLV 490-0010/-2010/-6010 CLV 490-1010/-3010/-7010

18 to 30 V DC 18 to 30 V DC

Line scanner Line scanner with oscillating mirror

No No

11 W (typ.)/max. 16 W 13 W (typ.)/max. 18 W

CDB 420 or CDM 490

CLV 490-0011/-2011/-6011 CLV 490-1011/-3011/-7011

24 V DC +20 %/–10 %

Line scanner Line scanner with oscillating mirror

Yes Yes

75 W (typ.)/max. 90 W 75 W (typ.)/max. 100 W

CDB 420 or CDM 490

1) switching outputs not connected

Table 5-22: Power consumption of the CLV

Via the CDB 420 or CDM 490 Connection Module the CLV without heater is supplied with 18 to 30 V DC, the CLV with heater is supplied with 24 V DC +20 %/−10 %. If the CMP 490 Power Supply Module in the CDM 490 is used for the CLV without heater, the input voltage is 100 to 250 V AC/50 to 60 Hz. Power-up delay The selected device number (default: 1) affects the power-up delay of the device. This is useful if a large number of CLVs (e. g. in the SICK CAN Scanner Network) are to be supplied from one power source. Table 5-23 contains a list of the available intervals. Device number GN

Power-up delay

Device number GN

Power-up delay

1; 11; 21; 31

0 ms

6; 16; 26

2,000 ms

2; 12; 22

400 ms

7; 17; 27

2,400 ms

3; 13; 23

800 ms

8; 18; 28

2,800 ms

4; 14; 24

1,200 ms

9; 19; 29

3,200 ms

5; 15; 25

1,600 ms

10; 20; 30

3,600 ms

Table 5-23: Power-up delay as a function of the device number GN

8 009 993/O824/20-12-2004

© SICK AG · Division Auto Ident · Germany · All rights reserved

5-17

Page 68

Electrical installation

Chapter 5

Operating Instructions CLV 490 Bar Code Scanner

Tip

The device number can be selected on the DEVICE CONFIGURATION tab in the "CLV-Setup" program. 5.4.3

Non-SICK Power supply unit/connections without the Connection Module

Power output If an non-Sick Power supply unit is used, it must be capable of providing the following voltage and power values: •

For CLV without heater: 18 to 30 V DC, min. 20 W continuous power output

•

For CLV with heater: 24 V DC +20 %/–10 %, min. 100 W

The non-Sick Power supply unit must provide the functional extra-low voltage in accordance with IEC 364-4-41. The output circuit must be reliably electrically isolated from the input circuit. To do so, use a safety isolating transformer pursuant to IEC 742. Wire cross-section The wire cross-section for the power supply (Pin 1/Pin 5) should be at least 0.15 mm2 (approx. 26 AWG) or 0.75 mm2 (approx. 20 AWG) for CLVs with heater, with a max. length of 10 m (32.8 ft).

5-18

© SICK AG · Division Auto Ident · Germany · All rights reserved

8 009 993/O824/20-12-2004

Page 69

Operating Instructions

Electrical installation

Chapter 5

CLV 490 Bar Code Scanner

5.5

Electrical installation procedure

5.5.1

Individual steps

•

Connect the power supply

•

Connect the host interface

•

Connect the PC (connect the terminal interface)

•

Connect the "Sensor" and "IN 0" to "IN 4" switching inputs

•

Connect the "Result 1" to "Result 4" switching outputs

5.5.2

Tools

•

Tool

•

Digital measuring device (current/voltage measurement)

5.5.3

Connecting the supply voltage a) SICK CDB 420 or CDM 490 Connection Module If the CLV is powered via the SICK Connection Modules using the cables no. 2 027 046, no. 2 030 023, no. 2 020 307, or no. 2 027 543, the supply voltage does not have to be wired separately. Using the cables with open ends no. 2 033 126, no. 2 033 127, no. 2 030 065, or no. 2 031 034, the supply voltage must be wired separately at the screw terminal strips in the connection module. Connecting the CLV without connector cover: 1.

Make sure that the power supply to the connection module is switched off.

2.

Connect the "Host/Term" and "I/O" connections on the CLV to the corresponding connections on the CDM 490 Connection Module using two cables no. 2 020 302 or no. 2 021 815 and secure them tightly on both sides.

Connecting the CLV with connector cover/connector cover with parameter memory: 1.

Connect the connector cover of the cable no. 2 027 046, no. 2 030 023, no. 2 033 126, or no. 2 033 127 to the "Host/Term" and "I/O" connections on the CLV and secure it tightly.

2.

Connect the free 15-pin cable plug to the corresponding socket on the CDB 420 Connection Module and secure them tightly. For the cable no. 2 033 126 or no. 2 033 127, lead additionally the cable with open end through one of the cable glands and connect it to the screw terminals 3, 4, and 6. – or –

1.

Connect the connector cover of the cable no. 2 020 307, no. 2 030 065, no. 2 031 034, or no. 2 027 543 to the "Host/Term" and "I/O" connections on the CLV and secure it tightly.

2.

Connect the free 15-pin cable connections to the corresponding connections on the CDM 490 Connection Module and secure them tightly. For the cable no. 2 030 065 or no. 2 031 034, lead additionally the cable with open end through one of the cable glands and connect it to the screw terminals 3, 4, and 6.

The data and function interfaces of the CLV are connected to the Connection Module.

8 009 993/O824/20-12-2004

© SICK AG · Division Auto Ident · Germany · All rights reserved

5-19

Page 70

Chapter 5

Electrical installation

Operating Instructions CLV 490 Bar Code Scanner

See also the Fitting Instructions for the "Cold-resistent cable no. 2 033 126/no. 2 033 127" (order no. 8 010 971, German/English) used for CDB 420 respectively the Fitting Instructions for the "Cold-resistent cable no. 2 030 065/no. 2 031 034" (order no 8 010 972, German/English) used for CDM 490. b) Power supply via non-SICK Power pack Connecting the CLV without connector cover: 1.

Connect the cable socket on cable no. 2 020 303 on the "Host/Term" connection and screw it tightly into position.

2.

Connect the cable plug on the cable no. 2 020 264 to the "I/O" connection and screw it tightly into position.

3.

Connect the power supply to the red wire (Pin 1, VS) and blue wire (Pin 5, GND) on cable no. 2 020 303. See also Table 5-15, Page 5-14

Connecting the CLV with connector cover/connector cover with parameter memory: 1.

Connect the connector cover of the cable no. 2 020 981 or no. 2 021 267 to the "Host/Term" and "I/O" connections on the CLV and secure it tightly.

2.

For the cable no. 2 020 981 connect the power supply to the red/pink wire (Pin 1, VS) and the blue/grey-brown wire (Pin 5, GND) of cable 1. See also Table 5-17, Page 5-15. For the cable no. 2 021 267 connect the power supply to the red wire (Pin 1, VS) and the black wire (Pin 5, GND) of cable 2. See also Table 5-20, Page 5-16.

The CLV is connected to the supply voltage.

5-20

© SICK AG · Division Auto Ident · Germany · All rights reserved

8 009 993/O824/20-12-2004