Operating Instructions: CLV490 Bar Code Scanner, Advanced Line, pages 141 to 148

Pages 141 to 148 of 266: 7 Maintenance; 8 Troubleshooting

The original document is the PDF (5.1 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.

For the Sick CLV490

Page 141

Maintenance

Operating Instructions

Chapter 7

CLV 490 Bar Code Scanner

7

Maintenance 7.1

Cleaning the CLV during operation

To ensure optimum reading performance, the reading window of the CLV must be clean at all times. For this reason, we recommend that you check the reading window at regular intervals, particularly if the device is operated under extreme conditions (dust, abrasion, humidity, finger prints). Laser beam can cause blindness. The CLV uses a class 2 red-light laser. Looking directly at the laser beam can seriously damage your eyesight. ¾

Never look directly into the path of the beam (similar to sunlight).

¾

Switch off the device while it is being cleaned.

¾

Never look directly into the reading window while the device is in operation.

Damage on the reading window The reading window is made of glass. Scratches and streaks on the window will reduce the reading performance. ¾

Use a mild, non-scouring detergent.

¾

Never use scratching or scouring motions to clean the reading window.

Cleaning the reading window: ¾

Clean the reading window at regular intervals with a mild, non-scouring detergent, e. g. an antistatic glass cleaning fluid. Fig. 7-1 shows the areas that require cleaning. Use soft, lint-free cloth.

¾

If necessary, clean the LEDs on the rear of the device.

Clean here

Front reading window Fig. 7-1:

Note

8 009 993/O824/20-12-2004

Side reading window

Cleaning the reading window

Don’t open the device. The producer warranty will be forfeited if the device is opened.

© SICK AG · Division Auto Ident · Germany · All rights reserved

7-1

Page 142

Chapter 7

Maintenance

Operating Instructions CLV 490 Bar Code Scanner

Cleaning other optical surfaces: ¾

If an external reading pulse generator and/or object-height detection sensors are used (e. g. photoelectric reflex switches), make sure to clean the relevant optical surfaces (Fig. 7-2). Contamination on these surfaces can cause switching errors.

Clean here

Trigger start

Trigger stop (optional)

Fig. 7-2:

Cleaning the external optical sensors (reading pulse generator, object-height detector)

Incremental encoder If an incremental encoder is used, check the correct position on the conveyor. ¾

7.2

Ensure that the incremental encoder always contact the conveyor so that the friction wheel turns without slipping.

Maintenance

The CLV does not require any maintenance except the described cleaning measures. Its self-monitoring functions ensure fault-free operation over long periods of time. The CLV indicates any faults or malfunctions by outputting coded system and error messages (e. g. "CLV SYS-Failure: 081") via its terminal interface. The messages are displayed in the "CLV-Setup" program window in the fourth field from the left on the bottom, in the status line (see also Chapter 8.3 Error messages, Page 8-2). CLV-Setup logs the messages in a file.

7.3

Disposal

Irreparable devices or devices that are no longer required must be disposed of in an environmentally-friendly manner: 1.

Always observe the applicable national waste disposal regulations.

2.

Remove the CLV housing.

3.

Remove the electronic modules.

4.

Remove the glass panel from the reading window and send it to be recycled.

5.

Send the chassis and cover (aluminium) to be recycled.

6.

Send the electronic modules for disposal as problem waste.

At present, SICK AG does not accept any unusable or irreparable devices.

7-2

© SICK AG · Division Auto Ident · Germany · All rights reserved

8 009 993/O824/20-12-2004

Page 143

Operating Instructions

Troubleshooting

Chapter 8

CLV 490 Bar Code Scanner

8

Troubleshooting 8.1

Overview of possible errors and malfunctions

8.1.1

Mounting errors

•

CLV aligned incorrectly with the object carrying the bar code (e. g. blanking)

•

Reading pulse sensor positioned incorrectly (e. g. internal reading interval starts too soon or too late)

•

With event-controlled focus control: sensors for object height detection positioned incorrectly

8.1.2 •

Electrical installation errors

Interfaces on the CLV connected incorrectly (wiring error in CDB 420 or CDM 490)

8.1.3

Parameter errors

•

Functions not adjusted to local conditions, e. g. communication parameters on the host interface are set incorrectly

•

Technical limits of the device exceeded, e. g. relationship between the scanning frequency and the reading distance

8.1.4

Malfunctions

•

Start/stop mode: no external reading trigger, more than one object in the reading field

•

Tracking mode: gap between the objects in conveying direction fallen below minimum

•

Laser timeout for incorrectly terminated reading pulse exceeded

•

Device error (hardware/software)

8.2

Monitoring error and malfunctions

The CLV is self-monitoring:

8 009 993/O824/20-12-2004

•

After the power supply has been switched on, the CLV automatically carries out a selftest, in which it checks important hardware components, before it is initialized (parameter set loaded and device functions initialized). The self-test can be repeated at any time by cancelling Reading mode. See Chapter 6.5.13 Self-test, Page 6-58.

•

If the CLV detects an error during the self-test or at any other time, it outputs the error status ST= 3 on the host interface. Prerequisite: the error status is enabled for transmission. This is disabled by default. The CLV displays the diagnosed error as a numeric combination via the terminal interface (see Table 8-1, Page 8-2).

•

While it is in operation, the CLV constantly monitors the laser diode and speed of the polygon mirror wheel. In addition, a watchdog circuit responds to error statuses. The CLV carries out distance measurement referencing at regular cycles. To do so, the laser diode is switched on for max. 10 s.

•

A timer (laser timeout) automatically deactivates the laser diode in Reading mode ("Sensor input" and "Serial interface trigger" mode) if the reading interval has not ended after 10 minutes (default setting). However, it does not end the reading interval. In this case, the CLV outputs the message: "Laser safety timeout" on the terminal interface. The reading interval must be terminated by resetting the trigger signal. The laser diode is activated again by the next reading trigger.

© SICK AG · Division Auto Ident · Germany · All rights reserved

8-1

Page 144

Chapter 8

Troubleshooting

Operating Instructions CLV 490 Bar Code Scanner

8.3

Error messages

The CLV outputs coded error message (number combinations) via the terminal interface only. Table 8-1 contains a list of the messages. Displaying messages: To display the messages on the PC, proceed as follows in CLV-Setup: ¾

Click in the toolbar. The Terminal Emulator is launched.

CLV-Setup logs the error messages automatically in a file (sys_msg.log). 8.3.1 Message

CLV without external parameter memory

Meaning

"CLV SYS-Failure: xxx" The CLV has diagnosed a system error. The "Device Ready" LED does not light up or extinguishes. The system is restarted.

Possible cause Error key: 011 012 013 051

1)

061 071 091 to 095 101 to 118 130 131 132 133 201 to 215

Data error (RAM) Address error (RAM) RAM to small Speed out of tolerance (mirror wheel) Mirror interval out of tolerance Error: laser shutter Error: focus control Error: external parameter memory Amplifier error Background profile invalid Center position CP 50 implausible Brightness overflow DSP error

Remedy If the error occurs again after the CLV is restarted, contact the SICK Service department.

1) Deactivate the CAN interface to suppress error numbers above 300 (not relevant in stand-alone mode). To do so, upload the parameter set from the CLV using CLV-Setup, choose the CAN INTERFACE tab and deactivate the function in the CAN INTERFACE FUNCTION section. Download the parameter set to the CLV!

Table 8-1: Error messages output on the terminal interface

8-2

© SICK AG · Division Auto Ident · Germany · All rights reserved

8 009 993/O824/20-12-2004

Page 145

Troubleshooting

Operating Instructions

Chapter 8

CLV 490 Bar Code Scanner

8.3.2

LED error messages for the external parameter memory (connector cover)

The CLV also reports errors accessing the external parameter memory by means of the "Device Ready" and "Read Result" LEDs. LED behavior

Meaning

Possible cause

1. When the device is switched on, the "Device Ready" LED blinks for approx. 10 s and then lights up constantly.

– Start option: START WITH EXTERNAL PARAMETERS. The CLV has successfully loaded the external parameter set to its RAM.

2. When the device is switched on, the "Device Ready" and "Read Result" LEDs blink for approx. 10 s. The "Device Ready" LED then lights up constantly.

Start option: START WITH EXTERNAL PARAMETERS. The CLV has successfully loaded the external parameter set with tolerated errors1) to its RAM.

Remedy Not applicable.

Transfer ok, but different parame- Upload parameter set to CLVter versions in the external mem- Setup and check parameters1). ory and CLV.

3. The "Device Ready" LED • Start option: START WITH EX• External parameter memory not • OK if the device is to be lights up constantly when TERNAL PARAMETERS. connected. operated without an external the device is switched on. The CLV has not loaded exparameter memory. ternal parameter set. Instead, • Check whether the external it has loaded the internal paparameter memory is connectrameter set to its RAM. ed correctly (free wiring: check wire assignment). • Start option: START WITH INTER- • Start ok. • None. NAL PARAMETERS. The CLV has successfully loaded the internal parameter set to its RAM. • Start option: START WITH EX4. When the device is • TERNAL PARAMETERS. switched on, the "Device The CLV has not loaded the • Ready" and "Read Reexternal parameter set. Insult" LEDs blink contistead, it has loaded the internuously 2). nal parameter set to its RAM. • The CLV, however, does not output data on the host interface.

External parameter memory not • See 3 Remedy. connected. The parameter set in the exter- • Use the correct external panal memory is not from a rameter memory or overwrite CLV 490. the current one. External parameter memory is • Repeat the startup procedure. defective. If the error is still present after following 4 Remedy, 1st and 2nd paragraph, contact the SICK Service department.

1) The software of the parameter set in the CLV is older than that of the parameter set in the external memory: print out the parameter set using CLV-Setup. If necessary, display the ID of the unknown/errored command group(s) using the <STX> 8?ERR <ETX> command in the Terminal Emulator window of CLV-Setup. Decode the ID using the reference manual entitled "Command Language of the CLV Bar Code Scanners" and check the relevant parameter values in the tabs. The software of the parameter set in the CLV is newer than that of the parameter set in the external memory: all of the new parameters in the CLV are set to the default values. If these parameters are necessary, edit the values in the relevant tabs in CLV-Setup and download them to the CLV and the external memory. 2) Flashing stops when the device switches from Reading mode to Parameter mode (e. g. during download from CLV-Setup).

Table 8-2: LED error messages for access to the external parameter memory (connector cover)

8 009 993/O824/20-12-2004

© SICK AG · Division Auto Ident · Germany · All rights reserved

8-3

Page 146

Chapter 8

Troubleshooting

Operating Instructions CLV 490 Bar Code Scanner

LED behavior

Meaning

Possible cause

4. When the device is swiched on, the "Device Ready" LED and "Read Result" LEDs blink continuously 2). – continued –

• Start option: START WITH COPY • External parameter INTERN –> EXTERN. memory is not connected. The CLV was not able to copy the internal parameter set to the external memory. • Data transmission error. It has loaded the internal parameter set to its RAM.

5. When the device is switched on, the "Device Ready" and "Read Result" LEDs blink alternately for approx. 10 s. The "Device Ready" LED then lights up constantly.

Start option: START WITH COPY INTERN –> EXTERN. The CLV has successfully copied the internal parameter set to the external memory3) and set the option for the next start to START WITH EXTERNAL PARAMETERS. It has loaded the internal parameter set to its RAM.

Remedy

• Ensure that the external parameter memory is connected correctly (free wiring: check wire assignment). • Repeat the startup procedure. If the error is still present after following 4 Remedy, 1st and 2nd paragraph, contact the SICK Service department. • Contact the SICK Service • Internal parameter set too large department. to be copied to the external memory (memory is full). • See 4 Remedy, 3rd paragraph • The external parameter memory is defective. Transfer ok.

Not applicable.

2) Flashing stops when the device switches from Reading mode to Parameter mode (e. g. during download from CLV-Setup). 3) The parameter set stored in the external memory is not checked automatically to determine whether it can be read. To check the parameter set, switch off the power supply to the CLV briefly and monitor the "Device Ready" LED as described under 1.

Table 8-2: LED error messages for access to the external parameter memory (connector cover) (contd.)

8-4

© SICK AG · Division Auto Ident · Germany · All rights reserved

8 009 993/O824/20-12-2004

Page 147

Troubleshooting

Operating Instructions

Chapter 8

CLV 490 Bar Code Scanner

8.3.3

Messages for errors accessing the external parameter memory (connector cover)

The CLV outputs further error messages, in addition to the general error messages in Chapter 8.3.1 CLV without external parameter memory, Page 8-2, if problems arise when the external parameter memory is accessed. Message

Meaning

Possible cause

Remedy

1. "error: access to external • Start option: START WITH EXThe external parameter memory is Repeat the startup procedure. If TERNAL PARAMETERS. parameters not successful, defective. the error is still present, contact The CLV has not loaded the internal parameters the SICK Service department. external parameter set. Inloaded" stead, it has loaded the inter(Error type: 104, 105, nal parameter set to its RAM. 106, 111, 112,115) The CLV, however, does not output data on the host interface. • Start option: START WITH COPY INTERN –> EXTERN. The CLV was not able to copy the internal parameter memory. It has loaded the internal parameter set to its RAM. 2. "error: external parameter memory full, internal parameters loaded" (Error type: 108)

The internal parameter set is too Start option: START WITH COPY INTERN –> EXTERN. large for the external memory The CLV was not able to copy (memory is full). the internal parameter memory. It has loaded the internal parameter set to its RAM.

Contact the SICK Service department.

3. "error: external parameters not valid for device, internal parameters loaded" (Error type: 116)

Start option: START WITH EXTERThe parameter set in the external NAL PARAMETERS. memory is not from a CLV 490. The CLV has not loaded the external parameter set. Instead, it has loaded the internal parameter set to its RAM. The CLV, however, does not output data on the host interface.

Use the correct external parameter memory or overwrite the current one.

• 4. "error: no external para• Start option: START WITH EXTERNAL PARAMETERS. meter memory connected, The CLV has not loaded the internal parameters external parameter set. Inloaded" stead, it has loaded the inter(Error type: 102) nal parameter set to its RAM. • Start option: START WITH COPY • INTERN –> EXTERN. The CLV was not able to copy the internal parameter set to the external memory. It has loaded the internal parameter set to its RAM. 5. "error: writing parameters CLV -> EXT, internal parameters loaded" (Error type: 110, 111, 112)

Start option: START WITH COPY INTERN –> EXTERN. The CLV was not able to copy the internal parameter set to the external memory. It has loaded the internal parameter set to its RAM.

External parameter memory not • OK if the device is to be operconnected. ated without an external parameter memory. • Check whether the memory is connected correctly (free wiring: check wire assignment). External parameter memory not • See 4 Remedy, 2nd paragraph connected.

Data transmission error.

Check whether the external parameter memory is connected correctly (free wiring: check wire assignment). Repeat the startup procedure. If the error is still present, contact the SICK Service department.

Table 8-3: Messages for problems accessing the external parameter memory (connector cover)

8 009 993/O824/20-12-2004

© SICK AG · Division Auto Ident · Germany · All rights reserved

8-5

Page 148

Chapter 8

Troubleshooting

Operating Instructions CLV 490 Bar Code Scanner

Message

Meaning

Possible cause

Remedy

6. "warning: additional unknown parameter(s), external parameters loaded" (Error type: 101, 114)

Start option: START WITH EXTERNAL PARAMETERS. The CLV has successfully loaded the external parameter set with tolerated errors1) to its RAM.

Transfer ok but different parameter versions in the external memory and the CLV.

Upload the parameter set to CLVSetup and check the values1).

1) The software of the parameter set in the CLV is older than that of the parameter set in the external memory: print out the parameter set using CLV-Setup. If necessary, display the ID of the unknown/errored command group(s) using the <STX> 8?ERR <ETX> command in the Terminal Emulator window of CLV-Setup. Decode the ID using the reference manual entitled "Command Language of the CLV Bar Code Scanners" and check the relevant parameter values in the tabs. The software of the parameter set in the CLV is newer than that of the parameter set in the external memory: all of the new parameters in the CLV are set to the default values. If these parameters are necessary, edit the values in the relevant tabs in CLV-Setup and download them to the CLV and the external memory.

Table 8-3: Messages for problems accessing the external parameter memory (connector cover) (contd.)

8-6