Page 143

Conducting the Test 1.

Select and run the test.

2.

Select a barcode that the scanner is configured to read. Orient the label so the beam sweeps across the barcode from quiet zone to quiet zone. Every 100 scans, you should see a result similar to the following: LLLLLLL x=nnnnn y=mmmm t=pppp dsp=q height=r where: LLLLLLL - The first seven characters of the barcode read and nnnnn, mmmmm, pppp, q and r are numbers. x & y - Represent the barcode position actually, the first edge of the barcode detected by the scanning beam in a coordinate system centered under the scanner, in mils (0.001" units). t - The time from the start of the scan when the barcode was first detected. r - The height zone above the conveyor in which the barcode is being read. If your installation includes height-sensing photoeyes, you must block all of the photoeyes that would be blocked by a box corresponding to the height of the label in order for the label position determination to be accurate. q - The decoding (DSP) processor that detected the barcode by its associated laser beam. Single line scanners use only one DSP.

Tachometer Status Test If your installation includes a tachometer, this test is useful for troubleshooting and for verifying the accuracy of parameter 702 MILS/TIC.

Conducting the Test Select and run the test. When this test is run, you will see a result similar to the following: nnnn COUNTS mmmm MILS pppp FT/MIN where: nnnn - The number of tachometer pulses received in one second. mmmm - The distance in mils that this corresponds to, as determined by parameter 702 MILS/TIC. The number of digits reported may vary among scanner types. pppp - The calculated conveyor speed in feet per minute. If parameter 702 MILS/TIC is set to zero, gap tracking will be disabled. If the value is not accurate, the distance and speed calculations will be incorrect.

Tachometer Calibration Procedure If your installation includes a tachometer, this procedure may be used to set parameter 702 MILS/TIC. This will be useful when the parameter cannot be derived from other known values,

Rev F, February 2002

CiMAX 75xxA/76xx Scanner Technical Reference Manual

8-9

8 Diagnostics

Introduction

Page 144

Chapter

8

Diagnostics such as the number of pulses/revolution and the gearing ratio between the driven shaft and actual conveyor movement.

To Run Calibration 1.

Select a box or article with a barcode that the scanner is configured to read.

2.

Make sure that the conveyor is running at normal speed and that the tachometer is functioning and connected correctly.

3.

Make sure that the following parameters are set to appropriate values. Parameter 702 MILS/TIC must have a preliminary value entered. A best guess or a value of 50 is appropriate. Parameters 510-517 INPUT1-8 TIME and parameter 518 INPUT DBNC TYPES must be set to the appropriate presence and tachometer input debouncing times and types. Parameters 706 INZONE-CENTER and 707 OUTZONE-CENTER, the distances between the inzone and the outzone photoeyes and the centerline of the scan zone, must have accurate values entered.

4.

Select and run the test.

5.

When the test is started, the scan zone between the inzone and outzone photoeyes should be empty. The message Waiting for inzone... will be displayed on the terminal. If there are boxes in the zone when the test is started, you may see the message Waiting for zone to clear... until the boxes exit from the scan zone.

6.

Pass several boxes through the scan zone on the conveyor, one box at a time. As each box blocks the inzone photoeye, the message Inzone will be displayed. When the leading edge of the box blocks the outzone photoeye, a message similar to the following will be displayed: Box x: nnn counts mmm mils/tic ppp ft/min ht zone: q where: x - The number of the box passed through the zone nnn - The number of encoder pulses recorded mmm - The calculated distance traveled per pulse ppp - The resulting conveyor speed q - The height zone indicated by the height-sensing photoeye tree, if your application requires this. As subsequent boxes are passed through the zone, the mils/tic values will be averaged. Ending the test will cause this average to be entered into parameter 702 MILS/TIC.

8-10

CiMAX 75xxA/76xx Scanner Technical Reference Manual

Rev F, February 2002

Page 145

Introduction

8

The scanner has 11 digital pots. The pots are 100-position digital potentiometers with non-volatile memory. Refer to parameters 444-459 in the Scanner Parameters Reference Guide for more information.

Diagnostics

Set Digital Potentiometers

The pots cannot be set to an absolute position using the Diagnostics menu. They must be incremented or decremented to their top or bottom end by incrementing or decrementing 100 times, if you don't know where they are positioned initially, increment or decrement them to the desired position. The pots can also be set by entering numerical positions using the Setup menu as described in “Non-Intelligent ASCII Terminal” on page 5-7. You can observe pot settings and also adjust them, with the exceptions noted below, through the Diagnostics menu. This will allow you to increment or decrement pot settings while watching the results on a scope, or while running a diagnostics test such as Read Efficiency. •

To protect the laser diode, the scanner has a jumper that must be re-positioned before their LSR PWR LIMIT pots can be adjusted. The jumper should be returned to their Safe positions after adjustment.

When you access the Set Digital Potentiometers test, you will see the following (your parameter values may be different): Current settings: (select # to change) 1 MOTOR SPEED = 50 2 LSR PWR (FINE) = 0 3 LSR PWR (COARSE) = 0 4 LSR PWR LIMIT = 0 5 PHASE = 99 6 RASTER FREQUENCY = 0 7 RASTER AMPLITUDE = 0 8 OPT RCVR LIMIT = 99 9 OPT RCVR THRESH = 72 10 OPT RCVR SHIFT = 20 11 OPT RCVR GAIN = 0 Note: Potentiometers #6 and #7 are only used in a CiMAX 7555A/7655 with the programmable raster option. In older scanners, these pots were used for laser modulation. Using the default values without programmable raster option may cause loss of laser power. To prevent this, set pots #6 and #7 to a value of 99. Once you have selected a pot by entering its number followed by Enter, the following will then be displayed: U to increment, D to decrement, T for max, B for min E to run Read Efficiency, X to stop Read Efficiency To change the pot settings, press the keys listed below. U - Moves the pot up one increment. D - Moves the pot down one increment.

Rev F, February 2002

CiMAX 75xxA/76xx Scanner Technical Reference Manual

8-11

Page 146

Chapter

8

Diagnostics

T - Moves the pot to its maximum position B - Moves the pot to its minimum position Pressing E while a pot is selected will run the Read Efficiency test. Refer to “Control Commands from the Host” on page 6-1 for more information. Pressing X will stop the test. It can be useful to monitor read efficiency while adjusting certain pots by pressing U and D to increment or decrement them.

Decoder Processor Test (CIDSP DIAGS) Decoder Processor Test displays the current software version and program being used by the decoding processor, also referred to as the DSP. The processor decodes the bars and spaces of barcodes into ASCII data that is presented to the host.

Conducting the Test 1.

Select and run the test. If CIX (stitching) is not enabled, the two lines below will be sent to the terminal when the test is run: CIDSP-Non-Stitching Auto-distinguish (5 symbologies) for Scanner: Ver X.X CONFIRM If CIX (stitching) is enabled, the two lines below will be sent to the terminal when the test is run: CIDSP- Stitching Auto-distinguish (C128, I25, C39, UPC, C93) for Scanner: Ver X.X CONFIRM The first line reports the program type (Stitching or Non-Stitching), the version of the scanner software in use and, if CIX is enabled, the five symbologies that can be read, but have not necessarily been selected to be read. The second line asks if you want to run an extensive decoding processor (DSP) communication and RAM test.

2.

When you enter Y, the test will cause the scanner to reboot. This test is no longer available with version 2.0 or later (76xx), or version 3.0 or later (75xx). Earlier versions will display the @ character at regular intervals on the terminal. Each @ indicates that a group of tests have been successfully run. If you see the message DSP FAIL while running the test, there may be a problem with the scanner. Contact SICK Auto Ident, Inc. Customer Service for assistance. Press any key to end the test.

Can Diagnostics (75xxA Only) CAN DIAGS allows some limited diagnostics for the DeviceNet/SDS board. When this diagnostic is run, you may see: Can Device is Offline This indicates that the network cable is not connected, or the DeviceNet or SDS master is not running, and the diagnostic will return to the menu. 8-12

CiMAX 75xxA/76xx Scanner Technical Reference Manual

Rev F, February 2002

Page 147

Otherwise, you will see the following prompt: Network Type: DeviceNet, Software Version: X.X, Device State: Online Enter 8 digits (hex format) to drive outputs Input = 00000000 00000000 00000000 00000000 You will see 'SDS' in place of DeviceNet if an SDS board is installed. For 'Device State', you may see other messages besides 'Online'. Initialize No Can Power Autobaud Can Bus Off Online Offline The last line will be repeated every one second, reflecting the current state of the expansion inputs according to any DeviceNet or SDS messages that are received. You will see either a '1' or a '0' in each of the 32 bits. You can force outputs from the diagnostic by typing in an 8-digit hex format number. For example, typing in '80000001' will cause outputs 1 and 32 to be forced on. In this example, the message: Output= 10000000 00000000 00000000 00000001 will then be displayed. Reading the inputs and setting the outputs from within this diagnostic can be very helpful in testing PLC ladder logic or PC flowchart logic in the field.

Exhaustive RAM Test The Exhaustive RAM Test completely tests the microprocessor and non-volatile memory. When this test is performed, all downloaded C programs and parameter values stored in memory will be permanently erased. Once started, the test cannot be aborted except by turning the scanner off, which causes it to reset to its default (factory-set) parameter values.

Conducting the Test 1.

Select EXHAUSTIVE RAM test by typing R - Enter while EXHAUSTIVE RAM is shown. The following confirmation message will be displayed: WARNING! ALL PARAMETER SETTINGS WILL BE ERASED. WARNING! DO YOU WISH TO PROCEED? (Y/N)

2.

Type Y - Enter. Entering any other response will abort the test. A second confirmation message will be displayed: YOU ANSWERED "Y". TO PROCEED WITH RAM TEST/CLEAR, NOW TYPE 'CLEAR'

Rev F, February 2002

CiMAX 75xxA/76xx Scanner Technical Reference Manual

8 Diagnostics

Introduction

8-13

Page 148

Chapter

8

Diagnostics

3.

To perform the test, type CLEAR or clear as a security check. Any other response will abort the test. Observe the display. If there is a memory failure, the terminal will display the message: MEM FAILED AAAA where: AAAA - The failed memory address.

4.

Allow the test to run through several displays of @ before terminating it.

Exiting the Test To end the test, turn the scanner off. Failures in the scanner's memory mean that it must be serviced. Contact your SICK Auto Ident, Inc. representative immediately.

Flash Operating System Download This procedure should only be used by SICK Auto Ident, Inc. Service Personnel. Note: All parameters will be reset to defaults after this procedure. Parameters should be uploaded to a file for future reference. All setup parameters will be set to default following new OS download. Verify that you have all group setup strings to a file allowing you to properly restore setup parameters. The scanner Operating System (O.S.) software is stored in FLASH read only memory (ROM). This technology allows the scanner to be reprogrammed in the field. The scanner may be upgraded with new releases of the operating system using this procedure.

Performing the Download 1.

Select FLASH O.S. Download by typing R - Enter. The following message will be displayed: WARNING! OPERATING SYSTEM WILL BE ERASED. NEW OPERATING SYSTEM CODE MUST THEN BE DOWNLOADED, USING TERMINAL PORT SET 9600,N,8,1, XON/XOFF WARNING! DO YOU WISH TO PROCEED? (Y/N)

Note: Since the operating system will be erased, you must be absolutely sure you have the necessary new release from SICK Auto Ident, Inc. You also must be sure that you are connected to a computer with a terminal emulation program that can perform either a straight ASCII download without any outgoing character translations, or an Xmodem file transfer. 2.

Press Y or y - Enter to proceed. Press any other key(s) to cancel the procedure. The following will be displayed: YOU ANSWERED 'Y' TO PROCEED WITH ERASING OPERATING SYSTEM, NOW TYPE 'FLASH'

3.

8-14

Type FLASH or flash - Enter to start the procedure. The following message will be displayed: CiMAX 75xxA/76xx Scanner Technical Reference Manual

Rev F, February 2002

Page 149

Please Cycle power (turn OFF, then ON) To Start FLASH Download Procedure 4.

Cycle power to the scanner. The following will be displayed. RVSI CiMatrix FLASH programming monitor BOOT MONITOR V X.X Operating System must be reprogrammed. Code checksum: XXXXXXXX Actual checksum: XXXXXXXX Erasing main block 1... erase OK Erasing main block 2... erase OK Erasing main block 3 erase OK Erasing main block 4 erase OK Waiting for download of new system code.... or start of Xmodem/Xmodem1K transfer.......

5.

At the PC, load and run a terminal emulation program such as ProComm for Windows or DOS, or Windows Terminal.

Note: After the old O.S. is erased, the Setup/Terminal/Slave port will revert to default serial port parameter values; 9600 baud, 7 data bits, space parity (equivalent to 8 data bits, no parity). Because O.S. downloads using 38.4K baud are 4x faster than at 9600 baud, you may want to speed up the download process by changing the default values. 6.

(Optional) Change the default serial port parameters. To change the baud rate, enter the following command: ~?x00n~ where: x - 1 or 2, for Host or Terminal ports, respectively n - The values of n are shown in Table 8–2.

TABLE 8–2. n Values n

Values

0

14.4k

1

28.8k

7

9600

8

19.2k

9

38.4k

Rev F, February 2002

CiMAX 75xxA/76xx Scanner Technical Reference Manual

8 Diagnostics

Introduction

8-15

Page 150

Chapter

8

Diagnostics

To change the parity and number of stop bits, enter: ~?x02nn~ where: x - 1 or 2 for Host or Terminal ports respectively nn - the values for nn are shown in Table 8–3. TABLE 8–3. nn Values nn

Parity

Stop Bit(s)

0 space

1

1 mark

1

2 even

1

3 odd

1

4 none

1

8 space

2

9 mark

2

10 even

2

11 odd

2

12 none

2

The new OS can be downloaded over the Host or Terminal ports using ASCII, Xmodem or Xmodem1K protocols. Under ideal conditions of no noise and a relatively short RS-232 cable, you can use the straight ASCII download. This method is almost twice as fast as either of the Xmodem protocols, but it does not allow error correction. If you use a modem or if your on-site communications are noisy, you should use either the Xmodem or Xmodem1K protocol. Every data packet sent using these protocols is tested for errors by a checksum comparison. Xmodem1K is slightly faster than Xmodem, but it is not universally supported. 7.

If you are making an Xmodem transfer, configure the scanner as an Xmodem receiver by sending it the following command: ~X~ The following message will be sent in response: READY TO RECEIVE FILE FOR XMODEM DOWNLOAD The scanner will then poll the PC or host computer, and wait to begin the file transfer process. It will expect to receive data through the same port that received the ~X~ command.

8.

8-16

If you have changed the baud rate, you should change the associated baud rate on the terminal at this time. Start downloading the new Operating System .sef file from the PC to the scanner. If you are using straight ASCII protocol, the scanner will sends a line of periods (.) to the terminal to indicate each line being sent. The process takes about 15 minutes at 9600 baud. If you are using Xmodem or Xmodem1K protocol, the scanner will not send the periods.

CiMAX 75xxA/76xx Scanner Technical Reference Manual

Rev F, February 2002

Page 151

9.

When the transfer is complete, the checksum will be displayed and the system will automatically reboot. You should see the message: Download complete. Code checksum: XXXXXXXX Actual checksum: XXXXXXXX Checksums match, download successful. Starting up new operating system code. The scanner will reboot and display the sign-on messages for the new version, on your terminal. Note: When the new O.S. starts up, default serial port parameters will again be used. If you changed the baud rate or parity on the PC, you must set it back to 9600 baud, 7 data bits, space parity (equivalent to 8 data bits, no parity). If there is a problem during the download, the resulting checksum may be incorrect. If this is the case, the procedure will start over, and you must download the code again.

Restore Keysheet The Restore Keysheet procedure allows you to restore the customized keysheet parameters that were saved in FLASH memory at the factory. Because the user cannot overwrite the FLASH memory, these parameters provide a reliable backup of the original settings.

Running the Procedure 1.

1. Select RESTORE KEYSHEET by typing R - Enter while RESTORE KEYSHEET is shown. The comment block from the last keysheet that was saved will be displayed. If no keysheet was saved, the following message will be displayed: NO KEYSHEET SAVED WARNING: Restoring keysheet parameters will overwrite current settings. Do you wish to proceed? (Y/N)

2.

Type Y or y -Enter to restore to RAM the previously saved keysheet parameters. Entering any other response will abort the procedure. If no parameters were saved, you will see the message: NO KEYSHEET SAVED Otherwise, the following message will be displayed: Keysheet parameters restored, press any key to reboot

3.

Rev F, February 2002

You must reboot the scanner for the restored parameters to take effect.

CiMAX 75xxA/76xx Scanner Technical Reference Manual

8 Diagnostics

Introduction

8-17

Page 152

Chapter

8

Diagnostics

The power to the laser diode will be turned Off during restoration to avoid spikes that could shorten diode life. Note: The keysheet area in FLASH memory is not erased when a new OS version is downloaded. Therefore, you can restore the keysheet settings easily after updating the OS. However, if the new OS introduces any new or changed parameters, they may not be set correctly by the saved keysheet. SICK Auto Ident, Inc. make every effort to be backward compatible, and will try to warn you when new parameter values will need to be entered by hand.

Restore C Program RESTORE PROGRAM is used to either save the currently loaded C program into non-volatile FLASH memory, or restore the program from FLASH into RAM memory. This allows restoring a user program in the event of memory loss of the battery-backed SRAM, and after downloading a new version of the operating system. This gives a measure of security to installations where the user program source code and download file may not be given to a customer by the system integrator, or gets lost over time. This also allows scanners that are returned to the factory for service to be shipped back to the customer with their program intact.

Procedure for Restoring a User Program The RESTORE PROGRAM procedure works in a similar fashion to the RESTORE KEYSHEET procedure. When you run the procedure by entering an R from the Diagnostic menu, you will be prompted with messages similar to the following: Currently stored User Program information: Program checksum, start, end, & comment block: ec 130000 130810 1302fc Program XYZ for Company ABC, written 9/9/99 WARNING: Restoring user program will overwrite current program Do you wish to proceed? (Y/N) In this example, the comment block saved with the program is Program XYZ for Company ABC, written 9/9/99. You may type any information in this block when the program is saved. If there is no program stored in FLASH memory, the following will be displayed: NO USER PROGRAM STORED If you answer the prompt with a Y or y, a second prompt will be displayed to confirm: WARNING: Restoring user program will overwrite current program Do you wish to proceed? (Y/N) If you answer this prompt with a Y or y, the program will be restored into memory, and another prompt will be sent: User program restored, press any key to reboot After you press any key, the scanner will reboot and run the user C program.

8-18

CiMAX 75xxA/76xx Scanner Technical Reference Manual

Rev F, February 2002