Page 71
MS-3 Laser E.book Page 11 Monday, August 30, 2004 11:09 AM
Chapter 4 Read Cycle
Trigger Filter Duration Usage:
Trigger Filter Duration is useful where trigger bounce could cause false triggers.
Definition:
Trigger Filter Duration can prevent trigger bounce from falsely triggering the scanner by limiting the time in which trigger pulses can be received.
Serial Cmd:
<K200,trigger mode,trigger filter duration>
Default:
244 (0.0099796 seconds)
Options:
0.0000818 to 2.6803815 seconds
External Trigger State Note: External Level, External Edge, or Serial Data or Edge trigger mode must be enabled for External Trigger to take effect. Usage:
Users can select the trigger polarity that will operate with their systems.
Definition:
Determines whether a positive or negative transition will initiate the read cycle.
Serial Cmd:
<K202,external trigger state>
Default:
Positive
Options:
0 = Negative
1 = Positive
MS-3 Laser Scanner User’s Manual
4-11
Page 72
MS-3 Laser E.book Page 12 Monday, August 30, 2004 11:09 AM
Serial Trigger
Serial Trigger Usage:
Allows the user to define the trigger character that initiates the read cycle.
Definition:
An on-line host command composed of a single ASCII host serial trigger character that initiates or ends the read cycle. Trigger characters are of two types: delimited and non-delimited. A delimited character is entered within angle brackets delimiters < >. A non-delimited trigger character is one that either starts or ends the read cycle and is NOT enclosed by delimiters such as < >.
Serial Trigger Character Usage:
Useful in applications where a single serial character enclosed in angle brackets initiates the read cycle.
Definition:
A trigger character is one that starts the read cycle, is enclosed by delimiters < > and can only be sent when the trigger mode is set to Serial Data or Serial Data & Edge.
Serial Cmd:
<K201,serial trigger character>
Default:
GS (corresponds to ^] displayed in the embedded menu)
Options:
Any single ASCII character, including control characters, except NUL (00H), an existing host command character, or an on-line protocol character.
Note: Serial Data or Serial Data & Edge triggering mode must be enabled for Serial Trigger Character to take effect.
4-12
MS-3 Laser Scanner User’s Manual
Page 73
MS-3 Laser E.book Page 13 Monday, August 30, 2004 11:09 AM
Chapter 4 Read Cycle
Non-delimited Start and Stop Characters Note: If the serial trigger is set to an existing command, it will override the command. The command will no longer be valid. For example, if the serial trigger is <Z>, parameters will no longer be saved, but when the save command is sent (<Z>), the reader will trigger. Note: In multidrop and user-defined protocols, the start and stop characters must be delimited. Both Start and Stop non-delimited characters can be defined and will function according to the trigger event, as follows: When defining Start and Stop trigger characters, the following rules apply: • In External Edge the reader looks only for the Start Trigger Character and ignores any Stop Trigger Character that may be defined. • In External Level, the Start Trigger Character begins the read cycle and Stop Trigger Character ends it. Note that even after a symbol has been decoded and the symbol data transmitted, the reader remains in External Level trigger read cycle until a Stop character is received. • In Serial Data & Edge trigger mode, either a Start Trigger Character or a hardware trigger can start an Edge trigger read cycle.
Start Character (non-delimited) Usage:
It is useful in applications where different characters are required to start and end a read cycle.
Definition:
A single ASCII host serial trigger character that starts the read cycle and is not enclosed by delimiters such as < and >.
Serial Cmd:
<K229,start trigger character>
Default:
Null (disabled)
Options:
Two hex digits representing an ASCII character except <, >, XON and XOFF.
Stop Character (non-delimited) Usage:
It is useful in applications where different characters are required to start and end a read cycle.
Definition:
A single ASCII host serial trigger character that ends the read cycle and is not enclosed by delimiters such as < and >.
Serial Cmd:
<K230,stop trigger character>
Default:
Null (disabled)
Options:
Two hex digits representing an ASCII character except <, >, XON and XOFF.
See Appendix F, “ASCII Table” for ASCII character information.
MS-3 Laser Scanner User’s Manual
4-13
Page 74
MS-3 Laser E.book Page 14 Monday, August 30, 2004 11:09 AM
End of Read Cycle
End of Read Cycle Note: When to Output must be set to End of Read Cycle before this command can be active. Definition:
The read cycle is the time during which the scanner will attempt to read and decode a symbol. At the end of a read cycle the scanner stops reading symbols and sends the symbol data or noread message to the host.
End of Read Cycle Mode Definition:
A read cycle can be terminated by a new trigger, a timeout, or a combination of both.
Serial Cmd:
<K220, end of read cycle,read cycle timeout>
Options:
0 = Timeout 1 = New Trigger 2 = Timeout & New Trigger
Note: When operating in Continuous Read or Continuous Read 1 Output, the scanner is always in the read cycle.
Timeout Usage:
Typically used with Serial or Edge and Continuous One. It is effective in highly controlled applications when the maximum length of time between objects can be predicted. It assures that a read cycle ends before the next bar coded object appears, giving the system extra time to decode and transmit the data to the host.
Definition:
Timeout ends the read cycle when the time set in Read Cycle Timeout elapses or “times out.” If in Continuous Read 1 Output, a timeout initiates a new read cycle and allows the same symbol to be read again. With either External Edge, Serial Data, or Serial Data & Edge enabled, a timeout ends the read cycle and symbol data or a noread message is sent to the host. With External Level enabled, the read cycle does not end until the falling edge trigger or a timeout occurs. The next read cycle does not begin until the next rising edge trigger.
New Trigger Usage:
4-14
New Trigger is an effective way to end a read cycle when objects move past the scanner at irregular intervals (not timing dependent).
MS-3 Laser Scanner User’s Manual
Page 75
MS-3 Laser E.book Page 15 Monday, August 30, 2004 11:09 AM
Chapter 4 Read Cycle Definition:
New Trigger ends the current read cycle and initiates a new one when a new trigger occurs. New Trigger refers only to a “rising edge” trigger. With either External Edge, Serial, or Serial or Edge enabled, an edge or serial trigger ends a read cycle and initiates the next read cycle. In the case of External Level, a falling edge trigger ends the read cycle but the next read cycle does not begin until the occurrence of the next rising edge trigger.
Note: When New Trigger is enabled and if not in External Level mode, Laser On/ Off will have no effect. When noreads occur, the laser will remain on.
Timeout or New Trigger Usage:
Useful in applications that require an alternative way to end the read cycle. For example, if an assembly line should stop completely or the intervals between objects are highly irregular.
Definition:
Timeout or New Trigger is identical to Timeout, except that a timeout or a new trigger (whichever occurs first) ends the read cycle.
Read Cycle Timeout Usage:
It is useful in many tightly controlled applications which require a read cycle to end before the next object appears and therefore need the flexibility of a timeout adjustment.
Definition:
Read Cycle Timeout is the time span of the read cycle and is represented in 10 mS increments. It is used in conjunction with External Edge or Serial Trigger.
Serial Cmd:
<K220, end of read cycle,read cycle timeout>
Default:
100 (Corresponds to one second or 1000 mS.)
Options:
0 to 65535. (Divide any positive number entered by 100 to determine the time in seconds.)
Timeout does not function when in Continuous or Level read cycle modes. Continuous 1: The scanner will only read a given symbol once until a timeout occurs or a different symbol is presented. Note: A minimum setting of 2 is recommended. Note: Timeout or Timeout or New Trigger under End of Read Cycle must be enabled for Read Cycle Timeout to take effect.
MS-3 Laser Scanner User’s Manual
4-15
Page 76
MS-3 Laser E.book Page 16 Monday, August 30, 2004 11:09 AM
Decodes Before Output
Decodes Before Output Note: When setting up, determine if the reader’s scan rate is capable of scanning your longest symbol the required number of times.
Decodes Before Output Mode Usage:
Useful when the highest level of reliability is required.
Definition:
When consecutive is enabled, the number set in Decodes Before Output must be achieved consecutively before decoded data is sent. For example, if set to Number Before Output is set to 5, then the data will be output whenever five decodes occur in succession. When the default non-consecutive is set, whenever a cumulative number of decodes is achieved, decoded data is sent. For example, if non-consecutive is enabled and Decodes Before Output is set to 10, then data will be sent when the total number of decodes reaches 10 regardless of the number of bad reads between decodes.
Serial Cmd:
<K221,number before output,decodes before output mode>
Default:
Non-consecutive
Options:
0 = Non-consecutive
1 = Consecutive
Number Before Output Usage:
This is a very useful feature to increase reliability of reads for symbologies that do not have internal error checking such as Pharmacode.
Definition:
The number of decodes required per symbol before a symbol’s decoded data is sent. It requires the scanner to successfully decode a symbol a designated number of times before it is sent. If it doesn’t achieve the number of good reads during the read cycle, then a noread will be sent. Note: Higher settings will decrease throughput speed.
Serial Cmd:
<K221,number before output,decodes before output mode>
Default:
1
Options:
1 to 255
4-16
MS-3 Laser Scanner User’s Manual
Page 77
MS-3 Laser E.book Page 17 Monday, August 30, 2004 11:09 AM
Chapter 4 Read Cycle
Scanner Setup Scanner Setup includes Gain, Scan Speed, Laser Power, and AGC.
Gain Level Usage:
Can be useful for fine-tuning gain or when portions of analog signals spike, as in the case of specular reflection or extraneous noise.
Definition:
Sets a voltage amplitude value for the analog circuitry.
Serial Cmd:
<K504, gain level,AGC sampling,AGC min,AGC max>
Default:
110 (high density scanner) or 90 (low density scanner). Note: Gain Level is optimized at the factory before shipment.
Options:
40 to 255
Scan Speed Usage:
To ensure a minimum number of scans, faster scan speeds are typically used for fast moving symbols and/or longer symbols with larger bar widths. See “Formulas for Number of Decodes” on page A-27. Note: Scan beams move faster across symbols further out in the scan range since the moving beam is being projected from a spinning mirror.
Definition:
Allows the user to set the number of scans per second by controlling the spinning mirror motor speed.
Serial Cmd:
<K500,scan speed>
Default:
50 (x 10) Low density scanner 40 (x 10) High density scanner
Options:
300 to 1000
MS-3 Laser Scanner User’s Manual
4-17
Page 78
MS-3 Laser E.book Page 18 Monday, August 30, 2004 11:09 AM
Scanner Setup
Laser Power Usage:
Used to optimize read performance at different distances. Typically higher power is used at greater distances.
Definition:
Allows the user or calibration routine to select laser power setting.
Serial Cmd:
<K700,laser on/off status,laser framing status,laser on position,laser off position,laser power>
Default:
High Power (Low density scanner) Medium Power (High density scanner)
Options:
0 = Low
1 = Medium
2 = High
Automatic Gain Control (AGC) AGC Sampling Mode Usage:
AGC Sampling allows you to specify how automatic gain control sampling will be done.
Definition:
AGC is the acronym for Automatic Gain Control. AGC increases the depth of field by attempting to maintain signal strength at a constant level regardless of the range of the bar code symbol.
Serial Cmd:
<K504,gain level,AGC sampling mode,AGC min,AGC max>
Default:
Continuous
Options:
0 = Disabled
1 = Leading Edge
2 = Continuous
Disabled Usage:
Disabled is not recommended, but can be useful in certain applications in which symbols do not require a large depth of field.
Definition:
When AGC Sampling is set to Disabled, gain levels can only be manually controlled by the Gain command.
Leading Edge Usage:
Leading Edge is rarely used, but in some cases can improve scan rates when symbols are skewed and the leading edge of a symbol is difficult to locate.
Definition:
Finds the leading edge of a symbol by looking for a 40 µS quiet zone followed by the number of transitions set in “Transition Counter” on page 4-20, stores the highest value of the samples, and adjusts the AGC accordingly at the end of the scan.
4-18
MS-3 Laser Scanner User’s Manual
Page 79
MS-3 Laser E.book Page 19 Monday, August 30, 2004 11:09 AM
Chapter 4 Read Cycle Continuous Usage:
Continuous is the default and the preferred mode for sampling analog signal amplitude.
Definition:
Samples AGC throughout the scan at the rate set in “Transition Counter” on page 4-20, averages the sample values, and adjusts the AGC value at the end of the scan.
AGC Minimum (Read Only) Usage: Definition:
Limits the minimum range of the available gain.
Serial Cmd:
<K504,gain level,AGC sampling mode,AGC min,AGC max> Send <K504?> for a readout.
Default:
70
Options
40 to 250
AGC Maximum (Read Only) Usage: Definition:
Limits the maximum range of the available gain.
Serial Cmd:
<K504,gain level,AGC sampling mode,AGC min,AGC max> Send <K504?> for a readout.
Default:
245
Options
60 to 255
Symbol Detect Status Definition:
During the read cycle, the scanner counts the number of bar and space transitions defined in the Transition Counter. If the count matches or exceeds the threshold, the scanner will perform AGC routines. When enabled, a bad symbol or no symbol message can be output rather than a noread message.
Serial Cmd:
<K505,symbol detect status,transition counter>
Default:
Disabled
Options:
0 = Disabled
1 = Enabled
MS-3 Laser Scanner User’s Manual
4-19
Page 80
MS-3 Laser E.book Page 20 Monday, August 30, 2004 11:09 AM
Scanner Setup
Transition Counter Usage:
Can be useful in certain application where extraneous objects in the field of view might require a higher threshold.
Definition:
Determines the number of bar/space transitions that are sampled before AGC can be resolved.
Definition:
Determines the number of bar/space transitions that must be read before a symbol is considered present and AGC can be resolved.
Serial Cmd:
<K505,symbol detect status,transition counter>
Default:
14
Options:
0 to 255
Note: Transition Counter is used for both AGC and symbol detection routines (see “No Symbol Message” on page 6-10.
Maximum Element Usage:
Caution: Do not change this parameter unless instructed by a your technical representative.
Definition:
Maximum Element represents the maximum bar element width, as measured in micro seconds, that’s allowed before a video reset occurs (default to white).
Serial Cmd:
<K502,maximum element>
Default:
1500 (45.75 µS)
Options:
1 to 65535 (.035 µS to 2294 µS increments)
Scan Width Enhance Usage:
This option should be enabled if large portions of scan width are needed. For example, if a large character count symbol is used to the edge of the scan or is using 70-80% of total scan width.
Definition:
Scan width Enhance tends to trade depth for width. Disable if outside edge/long range performance is needed.
Serial Cmd:
<K511,scan width enhance>
Default:
Enabled
Options:
0 = Disabled
4-20
1 = Enabled
MS-3 Laser Scanner User’s Manual