Page 61

MS-3 Laser E.book Page 1 Monday, August 30, 2004 11:09 AM

Chapter

Read Cycle

4 Chapter Contents Read Cycle by ESP .....................................................................................4-2 Read Cycle by Serial Command ................................................................4-2 Multisymbol .................................................................................................4-3 Number of Symbols .....................................................................................4-4 Serial Trigger .............................................................................................4-12 End of Read Cycle ....................................................................................4-14 Decodes Before Output .............................................................................4-16 Scanner Setup...........................................................................................4-17 Laser Setup ...............................................................................................4-21

After you’ve established communications and completed basic read rate testing, you will need to address the spatial and timing parameters associated with your application. In a typical operation a bar coded item moves along a line past a reader. A trigger or timer activates a read cycle during which the reader actively searches for symbols. You will need to decide how to initiate the read cycle and how and when to end it. Note: When assigning characters in user-defined fields, the characters NULL <> and , can only be entered through embedded menus, not through ESP or serial commands. Note: You can learn the current setting of any parameter by inserting a question mark after the number, as in <K100?>. To see all “K” commands, send <K?>.

MS-3 Laser Scanner User’s Manual

4-1

Page 62

MS-3 Laser E.book Page 2 Monday, August 30, 2004 11:09 AM

Read Cycle by ESP

Read Cycle by ESP

Click this Button to bring up the Read Cycle/ Triggering menu

To change a setting, double-click the setting and use your cursor to scroll through the options.

To open nested options, single-click the +.

Read Cycle by Serial Command Command Title Multisymbol Trigger Mode External Trigger State Serial Trigger Character Start Serial Trigger Character End Serial Trigger Character End of Read Cycle Decodes Before Output Automatic Gain Control Scan Speed Symbol Detect/Transition Maximum Element Scan Width Enhance Laser Controls

4-2

Format <K222,number of symbols,multisymbol separator> <K200,trigger mode,trigger filter duration> <K202,external trigger state> <K201,serial trigger character> <K229,start trigger character> <K230,end trigger character> <K220,end of read cycle,read cycle timeout> <K221,number before output,decodes before output mode> <K504, gain level,AGC sampling mode,AGC min,AGC max> <K500,scan speed> <K505,symbol detect status,transition counter> <K502,maximum element> <K511,scan width enhance> <K700, laser on/off status,laser framing status,laser on position,laser off position,laser power>

MS-3 Laser Scanner User’s Manual

Page 63

MS-3 Laser E.book Page 3 Monday, August 30, 2004 11:09 AM

Chapter 4 Read Cycle

Multisymbol Usage:

Commonly used in shipping applications where a shipping symbol contains individual bar codes for part number, quantity, etc. This feature allows one trigger to pick up all the symbols. AIAG and UCC/EAN-128 are two application standards that address this need.

Definition:

Allows the user to set up conditions that will allow more than one symbol to be decoded and output.

Conditions

The following conditions apply: 1. All noread messages are posted at the end of the data string. 2. If more than one symbol is within the scan beam at the same time, symbol data may not be displayed in the order of appearance. 3. If Matchcode Type is set to Sequential or if Trigger is set to Continuous Read 1 Output, Number of Symbols will default to 1 (if set to any number greater than 1). 4. The maximum number of characters in any one bar code (other than PDF417) is 64. 5. The maximum number of characters in a single scan line is 102 (Code 39). 6. The maximum number of characters for all symbols is 392, including preamble, separators, and LRC.

MS-3 Laser Scanner User’s Manual

4-3

Page 64

MS-3 Laser E.book Page 4 Monday, August 30, 2004 11:09 AM

Number of Symbols

Number of Symbols Definition:

Number of Symbols allows the user to define up to 12 bar code symbols that can be read in a single read cycle.

Serial Cmd:

<K222,number of symbols,multisymbol separator>

Default:

1

Options

1 to 6

4-4

MS-3 Laser Scanner User’s Manual

Page 65

MS-3 Laser E.book Page 5 Monday, August 30, 2004 11:09 AM

Chapter 4 Read Cycle

Multisymbol Separator Usage:

Used to delimit or separate data fields with a user defined character.

Definition:

The character that’s inserted between each symbol scanned when Multisymbol is set to any number greater than 1.

ESP:

To select a new multisymbol separator, double-click on Separator and select a character in the popup window.

Serial Cmd:

<K222,number of symbols,multisymbol separator> Note: If Multisymbol Separator has been changed to any character other than the default comma and you wish to re-define the separator as a comma, use ESP or the embedded menu.

Default:

, (comma)

Options:

Any available ASCII character, except < > NUL.

MS-3 Laser Scanner User’s Manual

4-5

Page 66

MS-3 Laser E.book Page 6 Monday, August 30, 2004 11:09 AM

Number of Symbols

Trigger Trigger Mode Definition:

The type of trigger event that will initiate the read cycle.

Serial Cmd:

<K200,trigger mode,trigger filter duration>

Default:

Continuous Read

Options:

0 = Continuous Read 1 = Continuous Read 1 Output

3 = External Edge 4 = Serial Data

2 = External Level

5 = Serial Data & External Edge

Continuous Read Usage:

Continuous Read is useful in testing symbol readability or reader functions. It is not recommended for normal operations.

Definition:

In Continuous Read, trigger input options are disabled, the reader is always in the read cycle, and it will attempt to decode and output every scan crossing a symbol.When To Output and Noread options have no affect on Continuous Read.

Serial Cmd:

<K200,0>

4-6

MS-3 Laser Scanner User’s Manual

Page 67

MS-3 Laser E.book Page 7 Monday, August 30, 2004 11:09 AM

Chapter 4 Read Cycle

Continuous Read 1 Output Usage:

Continuous Read 1 Output can be useful in applications where it is not feasible to use a trigger and all succeeding symbols contain different information. It is also effective in applications where the objects are hand-presented.

Definition:

In Continuous Read 1 Output the scanner self-triggers whenever it decodes a new symbol or a timeout occurs. If End Of Read Cycle is set to Timeout and the symbol doesn't change, the scanner will repeat the output at the end of each timeout period. For example, if Timeout is set to one second, the scanner sends the symbol data immediately and repeats the output at intervals of one second for as long as the symbol continues to be scanned. If End Of Read Cycle is set to New Trigger, the scanner will send the current symbol data immediately, but send it only once. A new symbol appearing in the scanner’s range will be read and sent immediately provided it is not identical to the previous symbol read.

Serial Cmd:

<K200,1>

Caution: In automated environments, Continuous Read 1 Output is not recommended because there is no one to verify that a symbol was missed. Note: If Trigger is set to Continuous Read 1 Output, Number of Symbols will default back to 1 (if set to any number greater than 1).

MS-3 Laser Scanner User’s Manual

4-7

Page 68

MS-3 Laser E.book Page 8 Monday, August 30, 2004 11:09 AM

Number of Symbols

External Level Usage:

This mode is effective in an application where the speeds of the conveying apparatus are variable and the time the reader spends scanning each object is not predictable. It also allows the user to determine if a noread has occurred.

Initiate Read Cycle: Object # 1, moving in front of the detector beam, causes a change in the trigger state, initiating the read cycle.

1

Object detector 1

Object detector

End Read Cycle: The same object, moving out of the detector beam, causes another change in the trigger state, ending the read cycle.

Positive 0 Associated waveforms assume External Trigger State is set to Positive.

Positive 0

Figure 4-2 External Level Trigger Definition:

External Level allows the read cycle to begin when a trigger (change of state) from an external sensing device is received. The read cycle endures until the object moves out of the sensor range and the active trigger state changes again.

Serial Cmd:

<K200,2>

Important: Level and Edge apply to the active logic state (Positive or Negative) that exists while the object is in a read cycle, between the rising edge and falling edge. Rising edge is the trigger signal associated with the appearance of an object. Falling edge is the trigger signal associated with the subsequent disappearance of the object.

4-8

MS-3 Laser Scanner User’s Manual

Page 69

MS-3 Laser E.book Page 9 Monday, August 30, 2004 11:09 AM

Chapter 4 Read Cycle

External Trigger Edge Usage:

This mode is highly recommended in any application where conveying speed is constant or if spacing, object size, or timeouts are consistent.

Definition:

External Trigger Edge, as with Level, allows the read cycle to begin when a trigger (change of state) from an external sensing device is received. However, the passing of an object out of sensor range does not end the read cycle. The read cycle ends with a good read output or, depending on End of Read Cycle setting, a timeout or new trigger occurs. Initiate Read Cycle: Object # 1, moving in front of the detector beam, causes a change in the trigger state, initiating the read cycle.

1

Object detector 2

Object detector

Active 0

Initiate Second Read Cycle: Associated waveforms Object # 2, moving in front of the assume External State detector beam, causes another is set to Positive. change in the trigger state. This Active signal initiates a new read cycle and ends the previous read cycle unless Timeout is enabled and a good read or timeout has not 0 occurred.

Figure 4-3 Trigger Edge Serial Cmd:

<K200,3>

Important: Level and Edge apply to the active logic state (Positive or Negative) that exists while the object is in a read cycle, between the rising edge and falling edge. Rising edge is the trigger signal associated with the appearance of an object. Falling edge is the trigger signal associated with the subsequent disappearance of the object. Important: Level and Edge apply to the active logic state (Positive or Negative) that exists while the object is in a read cycle, between the rising edge and falling edge. Rising edge is the trigger signal associated with the appearance of an object. Falling edge is the trigger signal associated with the subsequent disappearance of the object.

MS-3 Laser Scanner User’s Manual

4-9

Page 70

MS-3 Laser E.book Page 10 Monday, August 30, 2004 11:09 AM

Number of Symbols

Serial Data Usage:

Highly recommended in any application where conveying speed is constant or if spacing, object size, or timeouts are consistent.

Definition:

In Serial Data, the reader accepts an ASCII character from the host or controlling device as a trigger to start a read cycle. A Serial Data trigger behaves the same as an External Edge trigger.

Serial Cmd:

<K200,4>

Serial Data or External Edge Usage:

Serial Data or External Edge is seldom used but can be useful in an application that primarily uses an external sensing device but occasionally needs to be manually triggered. An auxiliary terminal can be connected to the aux port so the user can send the serial trigger character through the scanner to the host.

Definition:

In this mode the reader accepts either a serial ASCII character or an external trigger pulse to start the read cycle.

Serial Cmd:

<K200,5>

4-10