The text of a manual · part 9 of 11

MS-3 CCD Reader User's Manual, pages 81 to 90

Pages 81 to 90 of 106

The original document is the PDF (1.3 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.

For the Microscan MS-3

Page 81

Appendices

Appendix C, Connectivity Standalone Setup with IC-332 and IB-131 1 2 3

Scanner

5

6

Network

4

Host

In this configuration the IC-332 transforms incoming 10 to 28 VDC to 5VDC for the MS-3 CCD reader’s use. • (1) MS-3 CCD reader. • (2) IC-332 interface converter, FIS-0001-0035G. • (3) IB-131 Interface box, P/N 99-000018-01. • (4) Host computer. • (5) Null modem configuration cable, 61-300026-03, DB-25 plug to DB-9 socket, 6 ft. • (6) Power supply, 97100004-15 (90-264 VAC, 24VDC).

7

Figure A-4 MS-3 CCD/IB-131 Typical Setup

• (7) Optional object detector P/N 99-000017-01 is shown here.

IC-332 Interface Converter 5

1

6 (plug)

15

“READER” connector to MS-3 (with fixed nuts). See Figure A-5, IC-332 READER Pinouts

3.08" (78.2 mm)

1

5

10 (plug) 15

“EXTERNAL” Connector to IB-131 (with thumbscrews). See Figure A-6, IC-332 EXTERNAL Pinouts (to IB131)

MS-3 CCD Reader User’s Manual

A-5

Page 82

Connectivity Figure A-5 IC-332 READER Pinouts Pin No. 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15

A-6

Figure A-6 IC-332 EXTERNAL Pinouts (to IB-131)

Host 232

In/Out

Power + VDC TXD RXD Power/Signal Ground NC RTS Not used by MS-3 CCD Not used by MS-3 CCD Trigger CTS Not used by MS-3 CCD Not used by MS-3 CCD Chassis Ground Not used by MS-3 CCD Not used by MS-3 CCD

Out In Out

In In Out Out Out In Out In

MS-3 CCD Reader User’s Manual

Pin No. 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15

Host 232

In/Out

Power + 10 to 28 VDC TXD RXD Power/Signal Ground Trigger (-) RTS Not used by MS-3 CCD Not used by MS-3 CCD Trigger (+) CTS Not used by MS-3 CCD Not used by MS-3 CCD Chassis Ground Not used by MS-3 CCD Not used by MS-3 CCD

In Out In

In Out Out In In In In Out Out

Page 83

Appendices

Optoisolator Trigger Inputs for IC-332 Trigger inputs can be fully isolated pulses as NPN circuits. Minimum

Maximum

VIN-HIGH/IIN-HIGH 4.5 V/ 3.0 mA

V/ 23 mA

VIN-LOW/IIN-LOW

0 V/0 mA

2 .0 V/1 mA

Pulse Widthmin

48 µS

Input (+) I load

Input (–)

Input Examples Fully Optoisolated

+V

Isolated Reader V CC

NPN Source (trigger only)

Isolated Reader

PNP Source

V CC

+V Input (+)

Signal

Input (+)

Input (–)

Input (–)

Signal

Not Optoisolated Non-isolated Reader V CC

PNP Source Power (+) Input (+)

+V

V CC Signal

NPN Source (trigger only)

Non-isolated Reader Power (+)

+V

Input (+) Input (–)

Input (–)

Power ground

Signal

Power ground

Maximum Operating Power: 2 Watts Power Input: 5 VDC ± 5%, 200 mV p-p max. ripple, 150mA @ 5 VDC (typ.) Trigger Inputs: 4.7 to 24 V rated (0 µA @ 5 VDC, –600 µA@ 0 V)

MS-3 CCD Reader User’s Manual

A-7

Page 84

Connectivity

IB-131 Interface The IB-131 interface simplifies connecting to the reader by providing separate ports for the host, power supply, trigger and network. The network port is used for multidrop or daisy chain configurations. See the following pages for configuration diagrams and a list of cables offered by Microscan for ease of connectivity when using the IB-131. Table A-2 Host 25-pin Connector Pin 1 2 3 4 5 6 7 8 9 10 11 13 14 15 16 17 19 25

Function External ground Transmit data (RS-232) Receive data (RS-232) RTS CTS Not used by MS-3 CCD Signal ground Not used by MS-3 CCD Trigger (–) Trigger (+) Not used by MS-3 CCD Not used by MS-3 CCD Not used by MS-3 CCD Not used by MS-3 CCD Not used by MS-3 CCD Not used by MS-3 CCD Not used by MS-3 CCD Not used by MS-3 CCD

Table A-3 Trigger 4-pin Connector Pin 1 2 3 4

Function Power + 10 to 28 VDC (out)a Trigger (–) (in)b Power Ground Trigger (+) (in)a a. For NPN type, connect pins 1 and 4. b. For PNP type, connect pins 2 and 3.

Table A-4 Power 3-pin Connector Pin Function 1 Power Ground 3 Power + 10 to 28 VDC (in)

A-8

MS-3 CCD Reader User’s Manual

Table A-5 Reader 15-pin Connector Pin 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15

Function Power + 10 to 28 VDC (out) Transmit Receive Power/Signal Ground Trigger (–) (out) RTS Not used by MS-3 CCD Not used by MS-3 CCD Trigger (+) (out) CTS Not used by MS-3 CCD Not used by MS-3 CCD External ground Not used by MS-3 CCD Not used by MS-3 CCD

Table A-6 Network 25-pin Connector Pin 1 2 3 6 7 8 13 14 15 16 17 19

Function Ground Not used by MS-3 CCD Not used by MS-3 CCD Not used by MS-3 CCD Signal ground Not used by MS-3 CCD Not used by MS-3 CCD Not used by MS-3 CCD Not used by MS-3 CCD Not used by MS-3 CCD Not used by MS-3 CCD Not used by MS-3 CCD

Page 85

Appendices Electrical: Voltage Input: Regulated +10 to 28 VDC Mounting holes 3 places 1.25" (3.18 cm)

Cabling:

Trigger Power Supply

Length: 3.2 in. (8.13 cm) Width: 3.15 in. (8 cm) Height: 0.75 in. (1.9 cm)

3.2" (8.13 cm) Scanner

Mechanical:

Network

RS-232 Cable: 61-300026-03 Multidrop Cable: 61-100030-03 Daisy Chain Cable: 61-100029-03

Host

0.17: (0.43 cm) dia. 3. places

3.15" (8 cm)

.75" (1.9 cm)

Figure A-7 IB-131 Mechanical

MS-3 CCD Reader User’s Manual

A-9

Page 86

Serial Configuration Commands

Appendix D, Serial Configuration Commands The following table is a list of all the available serial configuration commands (also called “K” commands) in alphabetical order. These commands are also listed at the beginning of each chapter, as applicable. For utility (operational) commands see “Summary of Utilities Commands” on page 5-2. Table A-7 MS-3 CCD Serial Configuration Commands Command Title

Format Communications

Host Port Parameters Host Protocol Preamble Postamble

<K100,baud,parity,stop bits,data bits> <K140,protocol> <K141,preamble status,preamble> <K142,postamble status,postamble>

Number of Symbols Time Between Identical Decodes Trigger Mode Serial Trigger Character Start Character Stop Character No Decode Timeout Decodes Before Timeout Noread Message Automatic Gain Control Lighting Intensity Filter Mode

<K222,number of symbols>

Read Cycle <K503,TBID> <K200,trigger mode> <K201,serial trigger character> <K229,start trigger character> <K230,stop trigger character> <K220,no decode timeout> <K221,number before output> <K714,status,noread message> <K522,shutter speed,minimum speed,maximum speed> <K523,percent lighting> <K524,filter mode,filter list>

Symbology Code 39 Code 128/EAN Interleaved 2 of 5 Codabar UPC/EAN Code 93 Industrial 2 of 5 MSI Code Plessey

A-10

<K470, status,check digit status,check digit output status,unused,unused,minimum symbol length,full ASCII set> <K474, status,unused,minimum symbol length,1,1,unused,application record separator character> <K472, status,check digit status,check digit output status,symbol length #1,symbol length #2> <K471,status,unused,start & stop output status,unused,symbol length status,symbol length,check digit type,check digit status,check digit output status> <K489,UPC-A status,UPC-E status,EAN-8 status,EAN-13 status,supplemental status,supplemental 2 status,supplemental 5 status,UPC-A as EAN-13, UPC-E as A,EAN-8 as EAN-13> <K475,status,unused,minimum length> <K488, status,check digit status,check digit output status,format,symbol length #1,symbol length #2> <K486,status,check digit mode,check digit output status,unused,minimum symbol length> <K487, status,check digit output status,unused,minimum symbol length>

MS-3 CCD Reader User’s Manual

Page 87

Appendices Command Title RSS Expanded RSS Limited RSS-14 Symbology ID

Format

<K484, status> <K483, status> <K482, status> <K450,0,symbology identifier status>

Serial Communication Programs You can send serial commands from your terminal window in ESP. You can also use a terminal program such as Hyperterminal to send serial commands from your PC to the reader. From your PC host computer, you can initiate communications with the reader by with a terminal communications program such as Hyperterminal™ provided in the Windows™ operating system. In order to communicate with the reader, you will need to use the following communications settings: Baud = 9600 Parity = Even Stop Bits = One Data Bits = Seven Flow Control = None

Figure A-8 Hyperterminal Dialog

MS-3 CCD Reader User’s Manual

A-11

Page 88

Serial Command Format

Appendix E, Serial Command Format Serial commands are of two types: utility and configuration.

Rules that apply to both utility and configuration commands • A less than < and greater than > characters enclose the commands unless nondelimited commands are used. (See “Trigger Mode” on page 3-6.) • Commands and data are “case sensitive.” That is, characters must be entered as upper or lower case, as specified.

Serial Utility Commands These are sent during operations and are not followed by a <A> or <Z>. See Table A1, “Summary of Utility Serial Commands,” on page A-3.

Serial Configuration “K” Commands These begin with a single K character followed by a three digit numeric, a comma, data fields separated by commas, and typically, an initializing command, as follows: <K three digit numeric,data,data,...etc.><initializing command>1 An initializing command of type “A” (not saved to non-volatile memory) or type “Z” (saved to non-volatile memory) may follow the command. For example, to enable UPC and save the change for power-on, send: <K473, 1><Z>.

Serial Configuration Command Conventions: • All data fields (except the last) must be followed by a comma (without a space). • The following characters cannot be used: , < > NUL. • All fields preceding a modified field must be included. • If there is no change in preceding fields, then commas alone can be entered in these fields. For example, if only the last field in the following command is changing, <K100,4,1,0,0> can be entered as <K100,,,0>. • All fields following a modified field can be omitted. For example, to change Baud Rate, send <K100,3>.

1. In many commands a K command will automatically be saved for power-on with the MS-3 CCD.

A-12

MS-3 CCD Reader User’s Manual

Page 89

Appendices

Concatenating Configuration Commands Commands can be concatenated (added together) in a single string or data block.

Serial Command Status Request The status of a specific serial command can be requested by entering the command followed by a question mark, for example <K101?>. To see all K commands, send a <K?> command. With the MS-3 CCD, each K command is framed with a preamble and postamble and may be mixed with bar code symbol data. To prevent this, send an <I> command to disable scanning before sending the K request. Follow the K request with a <H> command to restore scanning.

Entering Special Characters in Serial Commands To enter control characters within a serial command, hold down the control key while typing the desired character.

Entering Special Characters in Embedded Menus Control Characters Control characters entered on the command line are displayed in the menu as mnemonic characters, such as: <CR><LF><NUL><NUL>. Press SP (the space bar) once, then enter the control character by holding down the control key and simultaneously pressing the desired character. For example to define a line feed, press SP, then Control and J simultaneously. It is displayed as ^J on the command line and as <LF> in the menu when the screen is refreshed.

To Define a Carriage Return as a Character Press SP, then CR. It is displayed as ^M on the command line and as <CR> in the menu when the screen is refreshed.

To Define a Space as a Character Press SP twice. It is displayed as a blank space in the menu when the screen is refreshed. While it appears that nothing has been assigned, the hex value 20 will be sent during data transmission.

To Select NUL as the Character Press SP, then a 0 (zero). It is displayed as <NUL> in the menu when the screen is refreshed.

MS-3 CCD Reader User’s Manual

A-13

Page 90

ASCII Table

Appendix F, ASCII Table Table A-8 ASCII Table with Control Characters Dec 00 01 02 03 04 05 06 07 08 09 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31

A-14

Hex 00 01 02 03 04 05 06 07 08 09 0A 0B 0C 0D 0E 0F 10 11 12 13 14 15 16 17 18 19 1A 1B 1C 1D 1E 1F

Mne NUL SOH STX ETX EOT ENQ ACK BEL BS HT LF VT FF CR SO SI DLE DC1 DC2 DC3 DC4 NAK SYN ETB CAN EM SUB ESC FS GS RS US

Ctrl ^@ ^A ^B ^C ^D ^E ^F ^G ^H ^I ^J ^K ^L ^M ^N ^O ^P ^Q ^R ^S ^T ^U ^V ^W ^X ^Y ^Z ^[ ^\ ^] ^^ ^_

Dec 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63

Hex 20 21 22 23 24 25 26 27 28 29 2A 2B 2C 2D 2E 2F 30 31 32 33 34 35 36 37 38 39 3A 3B 3C 3D 3E 3F

MS-3 CCD Reader User’s Manual

Ch SP ! “ # $ % & ' ( ) * + , . / 0 1 2 3 4 5 6 7 8 9 : ; < = > ?

Dec 64 65 66 67 68 69 70 71 72 73 74 75 76 77 78 79 80 81 82 83 84 85 86 87 88 89 90 91 92 93 94 95

Hex 40 41 42 43 44 45 46 47 48 49 4A 4B 4C 4D 4E 4F 50 51 52 53 54 55 56 57 58 59 5A 5B 5C 5D 5E 5F

Ch @ A B C D E F G H I J K L M N O P Q R S T U V W X Y Z [ \ ] ^ _

Dec 96 97 98 99 100 101 102 103 104 105 106 107 108 109 110 111 112 113 114 115 116 117 118 119 120 121 122 123 124 125 126 127

Hex 60 61 62 63 64 65 66 67 68 69 6A 6B 6C 6D 6E 6F 70 71 72 73 74 75 76 77 78 79 7A 7B 7C 7D 7E 7F

Ch ` a b c d e f g h i j k l m n o p q r s t u v w x y z { | } ~ D