For the Microscan MS-3

Page 91

Appendices

Appendix G, Defaulting/Saving/Resetting Understanding and controlling your reader’s active, saved, and default settings is critical to the operation of your reader. Table A-9 Software Reset/Save/Recall Commands <A> <Z> <Zrd>

Reset (does not save for power-on) Save current settings for power-on Recall Microscan default parameters and save for power-on

Power-on Parameters Power-on parameters (saved by the <Z> command) are saved to NOVRAM and recalled and loaded into current parameters when power is cycled or the <Arp> command is issued.

Default/Reset Definitions Defaults are original or saved customer settings that can be recalled, either by software reset <Zrd> or hardware reset (“Hardware Default” on page A-16). Defaulting the reader by <Zrd> resets the configuration parameters to Microscan default values. Defaulting might be necessary if you have make temporary changes, communications between the reader and another device are lost or interrupted, or you are using incompatible equipment (for example, a terminal that is set at 38.4 K baud communicating with a reader that is set at 115.2 K baud).1 Resets (“A” commands) affect only the current settings (active memory) and are not saved for power on or recall. Active Memory is where the reader’s active settings are stored during use. These are not available on power-on unless saved by a “Z” command.

1. There are no menu options or host commands for resetting the configuration program. MS-3 CCD Reader User’s Manual

A-15

Page 92

Defaulting/Saving/Resetting

Hardware Default Caution: Be certain that you short the correct pins. Shorting the wrong pins can cause serious damage to the unit.

Procedure: Important: You must complete the default within one minute after power-on. 1. Locate and mark the default pins or wires connecting to default pins. If using an IB-131 Interface, use pins 11 and 7 (figure A-2) on the HOST 25-pin connector. Results will be more consistent if you attach a push button switch. Pin 11

Pin 7

Figure A-2 Host Connector, 25-pin (on side of IB-131 box) 2. Turn power to the reader ON. 3. Connect pins 11 and 7 on the 25-pin Host connector of the IB-131. You should see the amber status light turn ON. 4. Within one second after the amber status light goes OUT, connect the pins briefly again. You should see the LED illuminators flash OFF and turn back ON, indicating a successful reset. Note: If not successful, re-cycle power to the reader before attempting to default again. 5. Momentarily short the default pins and release. Watch the LED. After it flashes OFF, short the default pins again.

A-16

MS-3 CCD Reader User’s Manual

Page 93

Appendices

Appendix H, Object Detector In a typical operation, a reader will wait for symbol data only during a triggered read cycle. A read cycle is initiated by a “trigger” and can be in the form of a serial command from the host (internal trigger) or a signal from an object detector (external trigger). When an object detector (also called a sensor, package detector, etc.) is used, it is set up so that its beam will bounce off the approaching object and the resulting pulse will be sent to the reader to begin the read cycle. Typically, a detector is positioned so that it will detect the presence of an object before its symbol can be read. An object detector is mounted in almost any position relative to the object as long as (1) the object passes within range of the detector and (2) direct or reflected light from the detector does not interfere with the reader’s reception. As the item continues to move down the line, its symbol moves into the reader beam and is read by the reader.

Bar coded item

Object detector

Bar coded item

MS-3

Figure A-1 Object Detector

MS-3 CCD Reader User’s Manual

A-17

Page 94

Formulas for Number of Decodes

Appendix I, Formulas for Number of Decodes To ensure reliable scanning, apply a minimum of five decodes to each symbol. Use the formulas below to calculate the number of decodes that your symbol will receive. If the number of decodes you derive from one of these calculations is less than the minimum for your application, plug in the minimum number of decodes (5) and solve for another parameter that might be changed, such as symbol speed or scans per second. Note: While the formulas given here solve for the predicted number of decodes, you may also use the formulas to solve for other parameters that might be changed, such as symbol speed, symbol length, etc.

Single Line Ladder Calculation For single scan line ladder scanning, use the following formula:

Direction of label travel

Scan Line

1 SH

------ SS × DR – 3 = ND (number of decodes)

SH (Symbol Height) (ladder formula only) is a measurement of the height of individual bars. SS (Symbol Speed) is the distance per second that a symbol Figure A-2 Ladder moves through the scan. DR (Decode Rate) is the number of decodes per second that a given symbol receives (as derived from a stationary scan test in “Test Decode Rate” on page A-43).

Single Scan Line Picket Fence Calculation For single scan line picket fence scanning use the following for-

Direction of label travel mula: Label placement area

SW – SL ) (-------------------------× DR – 3 = number of complete decodes

SS

Scan line

Figure A-3 Picket Fence

1. The -3 component in the formula is included to allow for AGC acquisition, an incomplete first scan, and an incomplete last scan. This applies only if the number inside the parentheses equals 4 or more. If the number equals 3, then only subtract 2 to derive 1 good scan.

A-18

MS-3 CCD Reader User’s Manual

Page 95

Appendices

Angled Picket Fence Calculation

Direction of symbol travel

The number of complete scans for angled picket fence is calculated the same as that for picket fence, with the exception that the scan width is shortened in proportion to scan tilt.

Scan Line

Improving the Number of Decodes

Figure A-4 Angled Picket Fence

After changing any of the parameters described in this section, recalculate the number of decodes.

Range Adjusting the symbol’s range, if possible, is one of the quickest and most effective ways to improve decode rates. However, in some applications you may need to select a less than optimum range, or one that is beyond the fringes of the ranges.

Scan Width Increasing scan width will increase the number of scans in a picket fence oriented application. Scan width is linked with scan range and changing one will usually require a change in the other.

Symbol Speed Applies to both picket fence and ladder oriented symbols. If your application allows it, slowing symbol speed (the time in seconds that a symbol is fully within the scan width of the reader) is an effective way to increase the number of decodes.

Symbol Dimensions, Symbol Density, and Symbol Ratio Not usually an option in most applications, but changes to symbol parameters can affect number of decodes calculations and possibly decode rates. If your application allows it, shortening the length of a picket fence symbol means the symbol will be in the scan range longer and hence receive a greater number of scans. Increasing the height of a ladder symbol means it will receive more scans. Changing symbol density and/or symbol ratio is another way ranges, decode rates, etc. can be altered.

MS-3 CCD Reader User’s Manual

A-19

Page 96

Operational Tips

Appendix J, Operational Tips Do: • Check inputs (symbol speed, length, height, etc.) to ensure the desired number of decodes per symbol. • Connect reader to host before connecting power to the reader; disconnect power before disconnecting the host from the reader. • For optimum decodes, mount reader so that your symbols pass through the center of the depth-of-field (minimum/maximum range). You find the center by moving your symbol in and out during a read rate test. • Avoid excessive tilt, pitch, and skew of the symbol. • Check the symbol for readability by doing a decode rate test. If there is any question about the symbol's readability, contact your Microscan representative at helpdesk@microscan.com. • After changing any parameter that might affect decode rate, repeat decode rate test. • Clean the reader window with a clean, dry Q-tip or cotton cloth on a regular basis.

Do Not: • Aim the reader into direct light or sunlight. • Aim the reader into an external object detector or other light-emitting device. • Obstruct the reader window with mounting hardware or other objects. • Connect chassis of reader and host to different ground potentials. • Operate the reader in excessive temperature environments.

A-20

MS-3 CCD Reader User’s Manual

Page 97

Appendices

Appendix K, Glossary of Terms Analog Gain Adjustment (AGC). Adjustment to signal strength that seeks to maintain a constant level regardless of the range of the symbol. Application Record. A variation of UCC/EAN-128 which adds an application identifier to symbol data, including user-definable separators, brackets, and padding. Autodiscriminate. The ability to decode several different symbologies without changing configuration. Auxiliary Port. Provides RS-232 connections to an auxiliary terminal or device for remote viewing, for the transfer of data to and from the host, and under certain conditions as a configuration port. Bar Code. The medium or label on which a symbol is printed or etched. Bar Code Data. The information that is transmitted from a decoded bar code symbol. Baud Rate. The number of discrete signal events per second. Bits per second. Calibration. A routine that cycles through various optical settings and selects the combination that produce the best read rate. Capture. The act of grabbing or recording a frame by a sensor. A frame or succession of frames that are captured. Check Digit. A Modulus 43 or Modulus 10 digit that is added to the symbol message for additional data integrity. Configuration. A setup or process of changing a reader’s settings to conform to a specific application. A physical arrangement of components. Concentrator. Intermediary device that relays data from readers to a host and commands from the host to the reader or other devices. Connector. Physical device (plug or socket) on a device or cable to provide in/out connectivity for various circuits and pins. Counter. Memory space provided to keep track of reader events. Daisy Chain. Linkage of master and slave readers to allow data to be relayed up to the host via auxiliary port connections. Decode. A good read. The successful scanning and decoding of the information encoded in symbol. Default. (verb) Restore Microscan or customer default settings, initialize serial commands and reset all counters. Defaults. (noun) The settings that are restored whenever a default takes place. Delimited. A command or field that is bracketed by pre-defined characters. Decode Rate (DR) The number of decodes per second that a given reader is capable of receiving. Depth of Field. The distance between the minimum and maximum range in which symbols have been read. Discrete I/O. Inputs and outputs characterized by discrete signal transitions from one voltage level to another so that digital switching can occur. MS-3 CCD Reader User’s Manual

A-21

Page 98

Glossary of Terms Dynamic Setup. Testing and configuration done with symbols in motion. EPROM. Erasable, programmable, read only memory. Embedded Memory. Onboard memory device such as EPROM or flash. End of Read Cycle. The time or condition at which the reader stops expecting symbol information to decode. Edge. Allows a read cycle to be initiated by a trigger signal from an object detector when it detects the appearance of an object (rising edge). The read cycle ends with a good read, a timeout, or a new trigger. ESP. Easy Setup Program. A portable proprietary Windows/NT based program developed by Microscan Systems, Inc. Falling Edge. A change of state (to inactive) associated with a level trigger in which the reader stops searching for symbols. (See Rising Edge.) Fixed Code Length. Increases data integrity by ensuring that only one symbol length will be accepted. Flash Memory. Memory that can be changed by downloading new code and recalled on power-on. Focal Length. The distance measured from the reader to the center of the depth of field, or focal point. Focus. The point at which the tip of the scan beam is at its narrowest. Full Duplex. Auxiliary port data is sent directly to the host but not displayed on the auxiliary port screen. Gain. Optimal signal strength. Good Match. The event occurring when a scanned symbol matches the master symbol information that is stored in the memory of the device. Good Read. A decode. The successful scanning and decoding of the information encoded in a symbol. Half Duplex. Auxiliary port data is sent directly to the host and displayed on the auxiliary port screen. Host. A computer, PLC, or other device that is used to execute commands and process data and discrete signals. Host Port. The pins or connections on a reader or other device that physically connect with a host and - using the RS-232, RS-422, or RS-485 standards - pass data and serial commands from one device to another. Initialize. Implement serial configuration commands into the reader’s active memory. Input. A channel or communications line. Decoded data or a discrete signal that is received by a device. See Output. Intercharacter Delay. The time interval in milliseconds between individual characters transmitted from the reader to the host. Intercharacter Gap. The extra space between the last element of one character and the first element of the adjacent character of a specific bar code symbol. Ladder Symbol Orientation. A bar code symbol in which the bars are parallel to the symbol’s direction of travel. Large Intercharacter Gap. Allows the reader to read symbols with gaps between

A-22

MS-3 CCD Reader User’s Manual

Page 99

Appendices symbol characters that exceed three times (3x) the narrow element width. LED. Light emitting diode. Level. The condition in which a read cycle initiated by a trigger signal from an object detector remains active and ends when the object moves out of the detector’s range. Longitudinal Redundancy Check (LRC). An error-checking routine that verifies the accuracy of transmissions. Master Symbol. A symbol or symbol data that is stored in a reader or reader’s memory and is compared with subsequently read symbols. Master Reader. First reader in a daisy chain mode and linked directly to the host and in tandem to slave readers. Matchcode. The ability to compare symbol data being scanned against a master symbol that is stored in the memory of the reader. Embedded Menu. Configuration options embedded in the reader’s erasable memory. ESP Menu. Easy Setup Program. Configuration program that runs in Windows-based operating systems 95 and above. Mil. One thousandths of an inch or 0.0254 mm. In bar-coding, a measurement standard that identifies a symbol by the width of its narrowest element. Mismatch. An event that occurs when the scanned symbol does not match the master symbol that is stored in the memory of the reader. Multisymbol. A reader mode which allows a reader to read more than one symbol in a single read cycle. Multidrop. A communications protocol for networking two or more readers or other devices with a concentrator (or controller) and characterized by the use of individual device addresses and the RS-485 standard. Narrow-bar-width. The width of the narrowest bar of a specific bar code symbol, expressed in thousands of an inch (or mils) as defined by standard code types. Narrow Margins. Allows the reader to read symbols with quiet zones less than 8 times the width of the narrow bar element. Non-delimited. A command that is not bracketed by pre-defined characters. Noread. A non-read. A condition that occurs when the reader is set up to decode symbols but no symbol is scanned during the read cycle. Normally Closed. A discrete output state that is only active when open. Normally Open. A discrete output state that is only active when closed. NOVRAM. Non-volatile random access memory. Data that is “saved for power-on” is saved to NOVRAM. Null. Also spelled “Nul.” A non-printed ASCII character that acts as a space-holder. Number of Decodes. The number of times a symbol is scanned by the reader during one pass through the laser beam. Object Detector. A photo electric device used to sense the presence or absence of an object (also referred to as a package detector). Output. A channel or communications line. Data or discrete signals that are transmitted or displayed by a device. Output Format. The modification of data output according to a user-defined index of MS-3 CCD Reader User’s Manual

A-23

Page 100

Glossary of Terms Extraction and Insertion commands. Parity. An error detection routine in which one data bit in each character is set to 1 or 0 (zero) so that the total number of 1 bits in the data field is even or odd. Picket Fence Symbol Orientation. A bar code symbol in which the bars are perpendicular to the symbol’s direction of travel. Pitch. Symbol (or reader) rotation around the center of a line parallel to the scan beam. Point-to-Point. A protocol consisting of a single communications event, typically used to connect a bar code reader to a terminal or host computer. Port. Logical circuit for data entry and exit. (One or more ports may be included within a single connector.) Protocol. The rules for communication between devices, providing a means to control the orderly flow of information between linked devices. Quiet Zones. Specified “clear” (non printed) areas immediately before and after the bar code symbol. The area is usually white (for black and white symbol) and at least 10 times the width of the narrowest bar, as measured in thousands of an inch. The zones can be other than white as long as their densities remain consistent and they have the required contrast relative to the bars. RAM. (Random Access Memory) Memory that is lost after power is recycled to the unit. Read Cycle. A programmed period of time or condition during which the reader will accept symbol input. Reader. (1) A reader, a bar code device that can read and decode symbols. (2) A device that reads and decodes 2D matrix codes. Read Range. The distances in which a symbol can be reliably read, as measured from the front of the reader. See “Depth of Field.” Reed-Solomon Error Correction Code. A linear, error correcting block code, suited to the correction of character errors which could be, in bar or matrix codes, the obliteration or removal of part of the symbol. Read Range. The distances in which a symbol can be reliably read, as measured from the front of the reader. See “Depth of Field.” Relay. An electrical switch that allows a low power to control a higher one. Reset. Sets all counters to zero. Rising Edge. A change of state (to active) that initiates (and in some cases ends) a read cycle with a new trigger, an edge trigger, or the leading edge of a level trigger. (See Falling Edge.) ROM. (Read Only Memory) Memory that cannot be changed. Reader. A scanning device that is comprised of a scan head and a decoder integrated in one package. Scan Speed. The number of scans per second that a reader projects from the spinning mirror. Scan Width (SW) is the width across the scan beam at a given distance from the reader in which a symbol can be read. Send. Transmit data from one device to another. Secondary Reader. A reader linked to the master or preceding reader in a daisy chain

A-24

MS-3 CCD Reader User’s Manual