For the Microscan MS-3

Page 31

Chapter 3 Read Cycle

External Level Usage:

This mode is effective in an application where the speeds of the conveying apparatus are variable and the time the reader spends scanning each object is not predictable. It also allows the user to determine if a noread has occurred.

Initiate Read Cycle: Object # 1, moving in front of the detector beam, causes a change in the trigger state, initiating the read cycle.

1

Object detector 1

Object detector

End Read Cycle: The same object, moving out of the detector beam, causes another change in the trigger state, ending the read cycle.

Active 0

Active 0

Figure 3-1 External Level Trigger Definition:

External Level allows the read cycle to begin when a trigger (change of state) from an external sensing device is received. The read cycle endures until the object moves out of the sensor range and the active trigger state changes again.

Serial Cmd:

<K200,2>

Important: Level and Edge apply to the active state of the trigger that exists while the object is in a read cycle. Negative is the electrical state (0 V/gnd) of the trigger signal associated with the appearance of an object. Positive is the electrical state of the trigger signal received by the reader and associated with the subsequent disappearance of the object.

MS-3 CCD Reader User’s Manual

3-7

Page 32

Time Between Identical Decodes

External Trigger Edge Usage:

This mode is highly recommended in any application where conveying speed is constant or if spacing, object size, or timeouts are consistent.

Definition:

External Trigger Edge, as with Level, allows the read cycle to begin when a trigger (change of state) from an external sensing device is received. However, the passing of an object out of sensor range does not end the read cycle. The read cycle ends with a good read or with a noread when a No Decode Timeout occurs.

Initiate Read Cycle: Object # 1, moving in front of the detector beam, causes a change in the trigger state, initiating the read

1

Active 0 V/gnd

Object detector 2

Object detector

Initiate Second Read Cycle: Object # 2, moving in front of the detector beam, causes another change in the trigger state. This signal initiates a new read cycle and ends the previous read cycle unless Timeout is enabled

Active 0 V/gnd

Figure 3-2 Trigger Edge Serial Cmd:

<K200,3>

Important: Level and Edge apply to the active state of the trigger that exists while the object is in a read cycle. Negative is the electrical state (0 V/gnd) of the trigger signal associated with the appearance of an object. Positive is the electrical state (+5 V) of the trigger signal associated with the subsequent disappearance of the object.

3-8

MS-3 CCD Reader User’s Manual

Page 33

Chapter 3 Read Cycle

Serial Data Usage:

Highly recommended in any application where conveying speed is constant or if spacing, object size, or timeouts are consistent.

Definition:

In Serial Data, the reader accepts an ASCII character from the host or controlling device as a trigger to start a read cycle. A Serial Data trigger behaves the same as an External Edge trigger.

Serial Cmd:

<K200,4>

Serial Data or External Edge Usage:

Serial Data or External Edge is seldom used but can be useful in an application that primarily uses an external sensing device but occasionally needs to be manually triggered.

Definition:

In this mode the reader accepts either a serial ASCII character or an external trigger pulse to start the read cycle.

Serial Cmd:

<K200,5>

Serial Trigger Character Usage:

Useful in applications where a single serial character enclosed in angle brackets initiates the read cycle.

Definition:

A trigger character is one that starts the read cycle, is enclosed by delimiters < > and can only be sent when the trigger mode is set to Serial Data or Serial Data & Edge.

Serial Cmd:

<K201, serial trigger character>

Default:

^] (corresponds to <GS>

Options:

Any single ASCII character, including control characters, except NUL (00H), an existing host command character, or an on-line protocol character.

Note: Serial Data or Serial Data & Edge triggering mode must be enabled for Serial Trigger Character to take effect.

MS-3 CCD Reader User’s Manual

3-9

Page 34

Time Between Identical Decodes

Non-delimited Start and Stop Characters Note: If the serial trigger character that you assign is already assigned as an operational command - for example a “Z” character that is used in the <Z> command - the command will no longer function as intended. Both Start and Stop non-delimited characters can be defined and will function according to the trigger event, as follows: When defining Start and Stop trigger characters, the following rules apply: • In External Edge the reader looks only for the Start Trigger Character and ignores any Stop Trigger Character that may be defined. • In External Level, the Start Trigger Character begins the read cycle and Stop Trigger Character ends it. Note that even after a symbol has been decoded and the symbol data transmitted, the reader remains in External Level trigger read cycle until a Stop character is received. • In Serial Data & Edge trigger mode, command, either a Start Trigger Character or a hardware trigger can start an Edge trigger read cycle.

Start Character (non-delimited) Usage:

It is useful in applications where different characters are required to start and end a read cycle.

Definition:

A single ASCII host serial trigger character that starts the read cycle and is not enclosed by delimiters such as < and >.

Serial Cmd:

<K229, start trigger character>

Default:

Null (disabled)

Options:

Two hex digits representing an ASCII character except < , and >.

Stop Character (non-delimited) Usage:

It is useful in applications where different characters are required to start and end a read cycle.

Definition:

A single ASCII host serial trigger character that ends the read cycle and is not enclosed by delimiters such as < and >.

Serial Cmd:

<K230, stop trigger character>

Default:

Null (disabled)

Options:

Two hex digits representing an ASCII character except <, >, XON and XOFF.

See Appendix F, “ASCII Table” for ASCII character information.

3-10

MS-3 CCD Reader User’s Manual

Page 35

Chapter 3 Read Cycle

No Decode Timeout Usage:

Useful in many tightly controlled applications requiring a read cycle to end before the next object appears.

Definition:

The duration of the read cycle as measured from either the beginning of the read cycle or from the last decode.

Serial Cmd:

<K220, 0,no decode timeout>a

Default:

2 (seconds)

Options:

1 to 4095 seconds a. The first parameter is always a zero(0) and does not change.

The MS-3 CCD times out if no symbol is decoded during the No Decode Timeout period. If a symbol is decoded before the timeout expires, then the timeout period starts over again. No Decode Timeout only applies to Edge, Serial Data, and Serial Data and Edge trigger modes. In Single symbol mode (<K222,1>), No Decode Timeout is the maximum amount of time the reader will attempt to decode after a trigger is activated. In Multiple symbol mode (<K222,0>), No Decode Timeout is the time in which there are no decodes, not the time since the trigger went active. That is, if the reader continues to decode symbols in front of it, it will not exit the read cycle unless the time between decodes exceeds the time set in No Decode Timeout. For Level trigger mode, the read cycle ends when the trigger falls, or a symbol is decoded and output in single-symbol mode.

MS-3 CCD Reader User’s Manual

3-11

Page 36

Decodes Before Output

Decodes Before Output Note: When setting up, determine if the reader’s scan rate is capable of scanning your longest symbol the required number of times. Usage:

This is a very useful feature to increase reliability of reads for symbologies that do not have internal error checking such as I 2/5.

Definition:

The number of decodes required per symbol before a its data is sent. It requires the reader to successfully decode a symbol a designated number of times before it is sent.

Serial Cmd:

<K221, number before output>

Default:

0

Options:

0 to 10 (0 = “auto adaptive”)

If in Single symbol mode it doesn’t achieve the number of required decodes during the read cycle, a noread will be sent. Note: Higher settings will decrease throughput speed.

3-12

MS-3 CCD Reader User’s Manual

Page 37

Chapter 3 Read Cycle

Noread Message Note: A Noread Message can only be transmitted when in Single symbol mode and no decodes occur.

Noread Status Usage:

Used in applications where the host needs serial verification that a symbol has not been read and especially useful in new print verification.

Definition:

When enabled, and if no symbol has been decoded before timeout or the end of the read cycle, the noread message will be transmitted to the host.

Serial Cmd:

<K714, status,noread message>

Default:

Enabled

Options:

0 = Disabled

1 = Enabled

Noread Message Definition:

Any combination of ASCII characters can be defined as the noread message.

Serial Cmd:

<K714, status,noread message>

Default:

NOREAD

Options:

1 to 20 ASCII characters.

Note: A Noread Message is not output when in scanning disabled mode (an <I> command has been sent).

MS-3 CCD Reader User’s Manual

3-13

Page 38

Reader Setup

Reader Setup Automatic Gain Control Allows shutter speeds to be controlled automatically, set to a specific value, or defined within a fixed range.

Shutter Speed Definition:

Shutter speed is the time that the sensor is exposed to light.

Usage:

Faster shutter speeds may be necessary in fast moving applications. Slower shutter speeds are useful in slower and lower contrast applications.

Serial Cmd:

<K522,shutter speed,minimum speed,maximum speed>

Default:

0 (0 = automatic)

Options:

70 to 65535

Minimum Speed Definition:

Slowest shutter speed and longest time that the sensor is exposed to the light.

Usage:

Applications with variable speeds, symbol sizes and distances.

Serial Cmd:

<K522,shutter speed,minimum speed,maximum speed>

Default:

70

Options:

70 to 65535

Maximum Speed Definition:

Fastest shutter speed and shortest time that the sensor is exposed to the light.

Usage:

Applications with variable speeds, symbol sizes and distances.

Serial Cmd:

<K522,shutter speed,minimum speed,maximum speed>

Default:

250 (1/250 second)

Options:

70 to 65535

3-14

MS-3 CCD Reader User’s Manual

Page 39

Chapter 3 Read Cycle

Lighting Intensity Definition:

Sets the intensity of the illumination LEDs as a percentage of the maximum intensity.

Usage:

Higher intensities can be useful in lower contrast or fast-moving symbols. Lower intensities might help decodes in high contrast, close-in symbols.

Serial Cmd:

<K523,percent lighting>

Default:

50

Options:

0 to 100

Filter Mode Definition:

Four sets of digital filtering parameters can be applied as a group or individually to improve readability for specific applications.

Usage:

Used to improve scan efficiency by narrowing the choice of filters to one.

Serial Cmd:

<K524,filter mode,filter list>

Default:

Automatic (all four filters are applied during the read cycle)

Options:

0 = Automatic

1 = Select from filter list

Filter List Definition:

Allow user to select any of four lighting filters.

Usage:

Used to improve scan efficiency by narrowing the choice of filters to one.

Serial Cmd:

<K524,filter mode,filter list>

Default:

NULL

Options:

0 = Medium bar width filter 1 = Thin bar width filter 2 = Large bar width filter 3 = Vesta filter (special for Code 39)

In general, symbols at greater distances will perform better with the medium or large bar width filters. However at optimum focal lengths, 4.75 inches for standard density units and 2 inches for high density units, some predictions can be made, as indicated in table 3-1. For specific steps in filter selection for specific application, see figure 3-3 on page 3-16. Table 3-1 Filters by Symbol Size Filter Medium bar width filter Thin bar width filter Large bar width filter

Standard Density Model 10 to 15 mil symbols 7.5 to 10 mil symbols 20 mil and larger

High Density Model 7.5 mil symbols 5 to 7.5 mil symbols 15 mil and larger MS-3 CCD Reader User’s Manual

3-15

Page 40

Reader Setup

Filter Selection When Filter Mode is set for Automatic, you can compare filter success rates in an actual scanning application by sending a <C> command from the Terminal window. Each filter will output its number of successes. In figure 3-3 the four filters (0 through 3) are shown in succession from left to right. Filter 1 acquires 95 reads, the others 0. The choice is obvious. When you enable Filter 1 in Filter List only, the other filters are omitted and the number of captures and decodes per second should significantly improve.

Filter 0

Filter 2 Filter 1

Filter 3

Figure 3-3 Filter Selection

3-16

MS-3 CCD Reader User’s Manual