Page 101

<Ad> Restore Default Operating Parameters Sending serial command <Ad> performs an <A> reset and copies default parameters (except factory settings) to RAM.

<Ad>

Initialize Saved Factory Settings

Saved Parameters RAM

Factory Defaults

Default Parameters

<An> Restore Saved Settings Clicking on Restore Saved Settings in the Interface menu or Sending serial command <An> performs an <A> reset and copies saved parameters to RAM.

<An>

Initialize Saved Factory Settings

Saved Parameters RAM

Factory Defaults

Default Parameters

<Af> Read Saved Factory Settings Sending serial command <Af> performs an <A> reset and copies saved factory settings to RAM.

<Af>

Initialize Saved Factory Settings

Saved Parameters RAM

Factory Defaults

Default Parameters

<Z> Reset/Save

<Z>

Initialize Saved Factory Settings

Saved Parameters RAM Default Parameters

Quadrus 2D Code Reader User’s Manual

Factory Defaults

4-15

4–Utilities

Clicking Reset/Save in the Interface menu or sending serial command <Z>, performs an <A> reset and saves parameters (except factory settings) to non-volatile memory for power on.

Page 102

Chapter 4 <Zd> Restore Default Operating Parameters/ Save for Power-on Clicking Default in the Interface menu or sending serial command <Zd> performs an <A> reset, restores default parameters, and saves them (except factory settings) to non-volatile memory for power on. <Zp> Save Factory Settings Sending serial command <Zp> performs an <A> reset and saves ONLY factory settings to non-volatile memory for power on.

<Zd>

Initialize Saved Factory Settings

Saved Parameters RAM

Factory Defaults

Default Parameters

<Zp>

Initialize Saved Factory Settings

Saved Parameters RAM

Factory Defaults

Default Parameters

<Zf> Restore Factory Defaults & Save for Power-on Sending serial command <Zf>, performs an <A> reset, copies special factory defaults to RAM, and saves ONLY the factory settings for power-on.

Outputs

<Zf>

Initialize Saved Factory Settings

Saved Parameters RAM Default Parameters

Factory Defaults

4–Utilities

<L1> Output # 1 Pulse Clicking Output # 1 Pulse or sending serial command <L1> sends a pulse to pin 6 (at any time regardless of Match Code or Output # 1 Driver status). <L2> Output # 2 Pulse Clicking Output # 2 Pulse or sending serial command <L2> sends a pulse to pin 8 (at any time regardless of Match Code or Output # 2 Driver status). <L3> Output # 3 Pulse Clicking Output # 3 Pulse or sending serial command <L3> sends a pulse to pin 15 (at any time regardless of Match Code or Output # 3 Driver status).

4-16

Quadrus 2D Code Reader User’s Manual

Page 103

Download You can select and send image files or application code by serial command or by accessing Send from the Utilities pull-down menu or by clicking the Utilities button and the Send tab.

Send Application Code Caution: Do not update the ap code unless you are certain the .hex file is the correct file for your application and compatible with your ESP™. If you need to update your application code, use the <d1> serial command, or 1. Click Send Application Code button and use the dialog box to locate the “ap” code file (ending in .hex). 2. Click OK. You should see a solid bar moving across the menu and a message below, as shown in figure 4-8. 3. Be sure to wait for the beep.

Immediately after receiving the Send command, a switch to 8 bit data length (if not already present) must be made for serial communication. The answer from the reader will be ACK (in 8 bit data length). The following format will then be sent to your reader: <length indicator(4 bytes), data (binary), CRC (4 bytes)>

Length Indicator A 4 Byte binary value indicating the length of the binary data file to be sent, most significant byte first. A maximum length of 65,535 can be sent (Length = 0xFFFF). Note that the length does not include the check sum.

Quadrus 2D Code Reader User’s Manual

4-17

4–Utilities

Figure 4-8 Download

Page 104

Chapter 4 Data 8 bit binary data, either an image file or code.

CRC Cyclical Redundancy Check. A 32 bit string of the length value and binary data file received. If the Receive was successful the reader will respond with an ACK, otherwise a NAK will be transmitted. Afterwards the communications will return to the previous data bit length. Current status and response messages are displayed in the Message Box.

Send Image File Click Send Image File or send <d3> to bring up a browse box from which you select an image that is automatically sent to the reader upon selection.

Send Boot Code

4–Utilities

Click Send Boot Code or send <d2> to send a boot code to program the in-circuit flash memory in the reader. A dialogue box appears that prompts you for a password and upon entry, a horizontal bar shows transfer progress.

4-18

Quadrus 2D Code Reader User’s Manual

Page 105

Self Test <$> Self Test Command Instructs reader to run self test and output results.

4–Utilities

Quadrus 2D Code Reader User’s Manual

4-19

Page 106

Chapter 4

Reader Status <?> Request Reader Error Status Displays certain operating parameters. After sending <?>, the reader immediately responds with <XX>, where “XX” is a two character ASCII hex value. The hex characters offer a “shorthand” code for displaying status.

Table 4-3 Hex Value to Binary Conversion

To interpret the reader’s code, follow this example:

Hex Value Digits

1. You send a <?> command to Quadrus™. 2. It responds with an ASCII hex value of 82. 3. Refer to table 4-3 to find the Hex value’s binary conversion. 4. For example, the binary bit digits for 8 is 1000 and the binary bit digits for 2 is 0010. Hence, the binary conversion of 82 is 10000010. 5. Next plug in these numbers into the Bin. Code column of table 4-4. Note: You do this “backwards.” That is, from the bottom up, the first binary digit with the last bit. 6. The binary zero indicated a zero status (not enabled), the 1 next to Command Received indicates that the Command Received flag is set.

Binary Bit Digits 7 6 5 4 3 2 1 0

0 1 2 3 4 5 6 7 8 9 A B C D E F

0 0 0 0 0 0 0 0 1 1 1 1 1 1 1 1

0 0 0 0 1 1 1 1 0 0 0 0 1 1 1 1

0 0 1 1 0 0 1 1 0 0 1 1 0 0 1 1

0 1 0 1 0 1 0 1 0 1 0 1 0 1 0 1

Bit

Reader Status

0

Command error detected

0

1

Command received

1

Table 4-4 shows status byte and the meaning of the individual bits (the bin. digits are only entered as part of the above example):

2

Communication error detected

0

<?1> Request Software Status

3

Flash sector unprotect failure

0

4

Host/Aux port buffer overflow

0

5

Reader is in a Read cycle

0

6

Watch dog reset

0

7

Parity (0=enabled, 1=disabled)

1

4–Utilities

Table 4-4 Reader Status from Binary Code

4-20

Bin. digit

Returns the following in this order: status byte, part-number, Flash checksum, current RAM parameter checksum, and Flash parameter checksum. The status byte is output with brackets around it and the rest of the data is comma delimited with brackets around the packet.

Quadrus 2D Code Reader User’s Manual

Page 107

Imaging You can choose to capture an image and decode, or decode the current image generated from image processing. <D1> Capture Image and Decode You first enable the coordinate data output by sending <KP1>. After decoding, the reader outputs the following: <number of objects,coordinate data,decode status,length of decoded data, decode data>. The following example represents two objects with one successful decode: <002,040,050,100,100,025,020,010,010,000,040,023,030,100, 125,120,110,030,001,014,ABCD1234000127> The bold values represent, in order: Number of Objects, Noread, and Good Read. Number of objects.

Three digit output indicating the number of objects in the field of view. For example, if there are 2 objects in the field of view, as in the above example, it will output as “002”. Coordinate Data

This field outputs the x, y pixel coordinates from top left, top right, bottom left, and bottom right corners of each object, in that order. Each x and y coordinate is represented by three digits and there should be eight fields for each object, separated by commas. Decode Status

Decode status. A “000” represents no decode and a “001” represents a successful decode.

Length of each decoded label fields (maximum of 6 fields). In the above example the decoded data has a field length of 14 which is represented as “014”. Decode Data

Decode data for each object that is successfully decoded. In the above example the decoded data is “ABCD1234000127”. <D2> Decode Current Image Similar to Decode command in Evaluation dialog. Decodes the current image in memory and output either the label data or a NOREAD message. Note, if coordinate data output is enabled, with <KP1>, then the data will be output in the above format; otherwise just the label data or Noread message will Quadrus 2D Code Reader User’s Manual

4-21

4–Utilities

Length of Decoded Data

Page 108

Chapter 4 be output. <c> Image Capture Command Similar to Capture command in Evaluation dialog. Captures the image in the field of view.

Histogram <h1> <h3> Histogram Request Command Similar to Histogram command in Calibration dialog. Allows you to define specific areas of an image for evaluation by providing coordinates in ASCII format. Format is returned as follows: <h1point1.x,point1.y,point2.x,point2.y,point3.x,point3.y,point4.x, point4.y> or <h3point1.x,point1.y,point2.x,point2.y,point3.x,point3.y,point4.x, point4.y> The coordinates, in ASCII format, are top left to bottom right corner. The Quadrus™ then calculates the histogram from the compressed data that resides inside of these coordinates. Use <h3...> format for ASCII return. Use <h1...> format for 8 bit data communications in binary format <h2> Histogram Statistics Request Command The Quadrus™ returns the statistics in this format: <h2Mean,Median,Variance,Min,Max,Low Peak,High Peak>

4–Utilities

The statistics data is returned in ASCII decimal format.

Matrix <e> Number of Pixels per element Command Outputs the number of pixels per element for current bar code as XX or if Operation Command Output Format is enabled, as <e/XX>.

4-22

Quadrus 2D Code Reader User’s Manual

Page 109

Chapter

5

Calibration

Chapter Contents Locate Label ....................................................................................... 5-2 Illuminate ............................................................................................. 5-3 Auto Calibration .................................................................................. 5-4 Dynamic Setup .................................................................................. 5-13 For many 2D labels, particularly where symbols have low contrast between light and dark elements, you will not achieve the best decode rate without some manual adjustments of the Illumination parameters: Shutter Time, Offset, Threshold, and Contrast. You can adjust these experimentally and by analyzing Quality Outputs, histograms, and line scans. Calibration is done first statically with a stationary label, and then dynamically, with labels in motion.

Calibration Tips

Quadrus 2D Code Reader User’s Manual

5–Calibration

• Always check the distance from the reader to the label. It should be 2” for UHD model and 4” for high density, and 5” for low density Quadrus™ readers. • Make certain your reader is angled at least 15° relative to a perpendicular from the label. • Shorten shutter time for faster moving labels; increase it for slower moving labels. • Ensure that the threshold is close to midway between histogram curves or use Adaptive threshold. • Adjust so that histogram curves are widely spaced. • Make experimental adjustments in small increments, generally no more than 5 at a time. • Try to balance dark and light pixel per element values. • Keep a running log of Illumination and Quality Output settings, read rates, and decodes per second and always save settings you may want to revisit.

5-1

Page 110

Chapter 5

Locate Label Refer to the start-up steps in Chapter 1, “Quick Start.” 1.

Place a test label in front of the Quadrus™ at the correct distance with at least 15° pitch.

2.

From the main menu click the CALIBRATION tab to open the Calibration menu.

3.

Click the LOCATOR tab. Quadrus™ begins capturing images and searching for symbols. When it finds an object that satisfies Quadrus™’ criteria for a symbol, Quadrus™ displays it in the Locator window as an polygon outline, red if an extraneous object or an undecoded label, green if a successfully decoded label. This is measured by the purple bar at the bottom of the image, which in this case is at the 100% read rate mark.

4.

At this point you may need to reposition your label or reader Figure 5-1 Calibration to center the image while maintaining the correct distance, and achieving the maximum read rate.

5.

Check Overlay to see a current screen image of the label similar to the image you see in Illumination.

5–Calibration

Note: Check Mirror to reverse the image to make the screen object movements congruent with your label movements relative to the reader. Note: Under Serial Trigger, you should use the default Continuous to repeat captures without timing out. Check One Shot to check daisy chain input or data from an auxiliary port. If you select Continuous you can see a graphical representation of the percentage read rate in the Read Rate bar at the bottom of the menu.

5-2

Quadrus 2D Code Reader User’s Manual