Page 111

Illuminate The Illuminate dialog allows you view the symbol on screen, adjust parameters and immediately review the changes to the image. 1.

Illuminate dialog by clicking the ILLUMINATE tab.

2.

Click Send to display the image. (figure 5-2).1

Note: A blurry image may indicate that the symbol is not at the correct focal length from the reader; however, since this is a compressed version of the image that Quadrus™ has captured, it may give an erroneous indication of an out-of-focus condition. See "Image Flow” on page 6-4 for instructions on receiving a full resolution image from Quadrus™.

Histogram To generate a histogram for the first time, place the tip of your mouse pointer on the image and double click. Subsequently you just click the Histogram button or the Image button to toggle back and forth. You can also click the Statistics button to see associated histogram data. See sample symbol

Figure 5-2 Typical Overlay Image

5–Calibration

1. You only need click the Send button the first time to generate an image. After this you make adjustments to either Shutter Time, Offset, Threshold, or Contrast and an updated image is automatically generated. Quadrus 2D Code Reader User’s Manual 5-3

Page 112

Chapter 5

Auto Calibration If - when in the Illuminate dialog (under Calibration) - you click Auto Cal, a dialog like the one in figure 5-3 will appear which lets you set the shutter time or allow the Auto Calibrate routine to calculate it for you by clicking Calculate. In general, the faster the application (or line speed), the shorter the shutter time. However it is recommended that you click Calculate. If you click Calculate, you will see a dialog like the one in figure 5-4. Enter the Density Type of your model.

Figure 5-3 Auto Calibrate

Note: The FIS number on your Quadrus’ label will identify your model type by its ending: -UHD is “Ultra High” density, -HD is “High” density, -LD is “Low” density, etc. Click OK to set label speed (figure 5-5). Enter your label speed in inches per second (2.54 cm/inch for conversion). Figure 5-4 Scanner Density

Click OK and OK. The Quadrus will go through several iterations before picking the optimum settings for your application.

Data Base

Figure 5-5 Label Speed

5–Calibration

If you click Data Base, an IP (image processing) data base dialog is opened, such as seen in figure 5-4.

5-4

Quadrus 2D Code Reader User’s Manual

Page 113

Figure 5-6 IP Data Base

The IP Data Base allows you setup and store a data base with up to 20 different IP (Image Processing) settings for various combinations of symbols and conditions. Clicking the chevrons >> or << at the bottom of the dialog will advance or return you to the next group of 5 settings. Notice that your current settings are displayed in the block at the top of the dialog. If you click Load Current IP, you will see the current settings being applied to the first data base row 1. Each time you click Load Current IP, your current settings are applied to data base row 1 and all data base row settings shift down one so that row 2 now has row 1’s settings, and so forth on down to the last row whose earlier settings are discarded. In operation, the Quadrus reads the current settings first. If successfully decoded, it will output the data; if not, it continues on, using the settings from data base row 1, then 2, and so on until the number of settings enabled is reached. When a settings row successfully decodes the symbol, it is moved to Current Settings and the current settings shift down to #1. The settings from #1 drop down into #2, and so forth until all the rows enabled by # of Settings are updated. In this manner, the last successful settings are always applied first.

When you are in Calibration mode - doing Locator, Illuminate, or Dynamic - you will need to click Active in Calibration after you set # Of Settings in order to have Quadrus cycle through the data base. Quadrus 2D Code Reader User’s Manual

5-5

5–Calibration

For each row of settings you will need to click Send IP [#] to save these settings to Quadrus. Or you can click Send All to send all settings to the reader at once.

Page 114

Chapter 5

Manual Settings You can also make adjustments manually while in the Illuminated dialog to Shutter TIme, Contrast, and Background (explained in Symbol menu) and, by clicking More Settings, you can also set Background Offset and Threshold Type. At the core of image decoding, each pixel on the Quadrus™ CCD receives a quantity of light that must be designated light or dark. The distribution of these pixels can be understood graphically by histograms. See Figure 5-8, “Illumination Adjustment Flow,” on page 5-8 for sequence of adjustment for static labels.

Shutter Time Default:

1/500

Options:

1/60, 1/125, 1/ 250, 1/500, 1/1000, 1/2000, 1/4000, 1/10K, 1/20K, 1/50K, and 1/100K.

The duration that light is received by the reader, similar to the shutter speed of a conventional camera. When shutter time is shortened, less light is absorbed and more pixel values appear on the dark side of the histogram. (For shutter times 1/ 20K or less, exposure time is controlled by the LED or external illumination duration.) In general, you increase shutter time to lighten the image. However, for moving labels, too long of a shutter time can cause blurring (see "Adjust Shutter Speed” on page 5-14).

Contrast Default:

200

Options:

0 to 255

Controls the photo-sensitivity of Quadrus™. Similar to Offset in that higher values lighten the image and vice versa, but all changes are amplified. (In terms of histograms, increasing contrast expands pixel values to the right by a multiplier affect.) Increase contrast when contrast is low (pixels values tightly clustered to the left). Find the upper and lower range of Contrast and set at the midpoint.

5–Calibration

Background Default:

200

Options:

0 to 255

If the finder pattern is White (or light, relative the background), the background color should be set to Black. The terms Black and White are relative and you should think in terms of lighter or darker.

5-6

Quadrus 2D Code Reader User’s Manual

Page 115

Background Offset Default:

Medium

Options:

Medium (100), Light (150), Dark (50), and Custom (0-255)

Controls overall darkness/lightness. Increasing Offset lightens the overall image and vice versa. (In terms of histograms, pixel values are changed as an additive affect.) The Find the upper and lower range of Offset and set at the midpoint.

Threshold Default:

Adaptive

Options:

Adaptive, Fixed (1-255)

Figure 5-7 More Settings

The dividing line between dark and light pixels is represented in Quadrus™ histograms as a vertical dashed line. Threshold should be midway between light and dark pixel clusters. Significant shifts in threshold can make decoding difficult or impossible. If using Fixed Threshold, find the upper and lower range of Threshold and set at the midpoint.

5–Calibration

Quadrus 2D Code Reader User’s Manual

5-7

Page 116

Chapter 5

Start here View image in Illumination mode

Decrease Contrast and/or Background Offset settings

Increase Contrast and /or Background Offset Too dark

Too light

Is symbol outlined in red or green?

No

Is image too dark, or too light?

Yes

Neither

Double click on image to view Histogram

Check other limits (symbol size, contrast, etc.) and start over

Th res ho ld

Does Histogram shape look ideal ? Adaptive? Yes Yes

No No

Find decode limits for Threshold setting, and set to middle value

5–Calibration

Ligh t

D ar k

Saturated light (or dark): adjust offset

Low contrast: Adjust contrast

Check Quality Outputs D ar k

Lig ht

D ark

Figure 5-8 Illumination Adjustment Flow

5-8

Quadrus 2D Code Reader User’s Manual

Lig ht

Page 117

Sample Symbol # 1, Thermal With a black on white background thermal label, and Using factory defaults, that is, loading an untitled file in ESP™, we did the following: 1.

Performed a Locate Label, positioned label, and obtained a 100% read rate.

2.

Generated a crisp image in Illumination (figure 5-9).

Figure 5-9 Thermal Label Image

3.

Generated a histogram that had well-separated light/dark distribution (figure 5-10).

4.

Generated a Quality Output reading of 0 121 186 14 15 (table 5-1).

Quadrus 2D Code Reader User’s Manual

5–Calibration

Figure 5-10 Thermal Label Histogram

5-9

Page 118

Chapter 5 With a zero cell damage rate, and a close balance between dark and light ratios of Pixels per Element (14/15), and 100% read rate, there seemed to be little reason to try to improve on these readings.

Sample # 2, Laser Etch

Table 5-1 Quality Outputs

Label data

Total Pixels % Cell Symbol Decode per Damage Contrast Time Element

0123456789abcdefg 0 121 186 14 15

5–Calibration

For this example, we used a gold on green background, laser etched label, and a new default ESP™ file.

dark pixels light pixels

1.

First we selected the Image menu and reversed the background to black (since the green background appeared darker than the etched gold).

2.

Then we ran a test image (in Illumination). The result was almost black.

3.

To compensate, we doubled the shutter time (from 1/500 to 1/250).

4.

Next we increased the contrast to lighten the image. At a value of 210 a green outline appeared around the image, indicating good reads.

5.

We continued to increase the contrast until the outlines turned red (at 234).

6.

We split the difference and pegged the contrast at 222 (see figure 5-11).

7.

Next we checked the histogram (this looked good) and the Quality Outputs. These were 0 139 208 15 9.

8.

The 15 9 indicated a disproportionate white to black pixel ratio and the 208 was slightly above our target of 200 ms to give us our five scans per second.

9.

We lowered the offset to 80. This darkened the image and the resulting Quality Outputs was an acceptable 198 for total decode time and dark to light numbers of 14 and 9, a slight improvement.

Figure 5-11 Adjusted Laser Etch Image

10. The resulting read rate was 100% and the decodes per second was five, both satisfactory for our purposes.

5-10

Quadrus 2D Code Reader User’s Manual

Page 119

Quality Output Analysis, Label # 3, ink jet This symbol was printed on the top of a milk cap, black with a light purple background. 1.

After locating the image, we clicked Illumination, then Send, we got a dark image that was only reading at 89% read rate. (figure 5-12).

Figure 5-12 Dark Image

Figure 5-13 Label #3 Histogram

2.

When a histogram was generated, the curves were very close together (figure 5-13). So we increased contrast to spread the curves, and decreased offset to lower the brightness.

3.

Because of extraneous objects and low contrast inherent in the label, the threshold was not holding from image to image. Therefore we unchecked the Auto Threshold Adjust and set threshold at 70.

4.

We succeeded in enhancing the separation by intensifying contrast while bringing down the offset.

5.

The result was a histogram that while separated, had too much “noise” and little or no distinction between the light and dark curves.

6.

The approach we took was to double shutter time to 1/125 (this brightened the image), compensate for this by decreasing contrast, and increase offset accordingly. Finally, we allowed threshold to float automatically to 134.

8.

This worked! The new image was crisp (figure 5-14) and the histogram curves, while low, were wide apart and distinct (figure 5-15). The Quality numbers went to 2 126 158 14 16, all acceptable. Most importantly, the read rate rose to 100%.

Quadrus 2D Code Reader User’s Manual

5-11

5–Calibration

7.

Page 120

Chapter 5

Summary

Figure 5-14 Adjusted Image

Figure 5-15 Adjusted Histogram

Quadrus™’ robust features make the task of adjusting Illumination parameters both challenging and rewarding. Once again, you can experiment with Illumination parameters by increasing contrast to widen those curves, decreasing offset to lower the overall brightness, holding threshold in one place while you work the other parameters, allowing it to float only when your histogram, quality outputs, and/or read rate have been established. Remember to double-click the image outline to generate a histogram. See Calibration Tips at the beginning of this chapter.

5–Calibration

You can further analyze and improve read rates by capturing full images and analyzing line scans with advanced features described in Chapter 6, “Evaluation.”

5-12