Read Rate You can check the percentage read rate by accessing Read Rate from the Utilities pull down menu or clicking the Utilities button and Read Rate tab to see the following:

Figure 4-2 Read Rate

With a readable symbol in place and Percent enabled, click Start to see the read rate percentage. <C> Decodes/Second Read Rate Test Click on Decodes/Sec. under Single Label and click Start or send <C> to enter Decodes/Second Read Rate Test to output the decodes per second and label data (if any) for single labels. The Decodes/Second Read Rate Test processes 25 images and records the time required to do so. The decodes per second is then calculated from this value. <Cp> Percent Read Rate Test

<J> Exit Read Rate Test Click on Stop or send <J> to end Decodes/Second Read Rate Test <C> or Percent Read Rate Test <Cp>.

Quadrus 2D Code Reader User’s Manual

4-5

4–Utilities

Click on Percent and click Start or send <Cp> to output the percentage of scans decoded and label data (if any). The Percent Read Rate Test processes 25 images and calculates the percentage of good decodes.

Page 92

Chapter 4

Counters You can request counter numbers or reset the counters by accessing Counters from the Utilities pull-down menu or by clicking the Utilities button and Counters tab to see the following:

Figure 4-3 Counter

The Xs in all counter commands denote a numeric value from 00000 to 65,535. After reaching the maximum numeric limit of 65,535, an error message will be displayed and the counter will automatically rollover and start counting again at 00000. To obtain the cumulative total of counts after the rollover has occurred, add 65,536 per each rollover (the reader does not keep track of the number of rollovers) to the current count.

4–Utilities

Note: All counter values will be lost if you cycle power to the reader or send an <A> command. Clicking Reset for any counter will reset that counter to 0. Clicking Request for any counter will return the current number for that counter. <N> Noread Counter Request Clicking Request or sending serial command <N> displays the total number of noreads that have occurred since power-on or the last Noread Counter Reset command. <O> Noread Counter Reset Clicking Reset or sending serial command <O> sets Noread Counter to 00000. <T> Trigger Counter Request Clicking Request or sending serial command <T> displays the total number of triggers since power-on or the last Trigger Counter Reset command.

4-6

Quadrus 2D Code Reader User’s Manual

Page 93

<U> Trigger Counter Reset Clicking Reset or serial command <U> sets the trigger counter to 00000. <V> Match Code/Good Read Counter Request Clicking Request or sending serial command <V> displays the total number of good reads matching the master label or, if Match Code is not enabled, the number of good reads since power-on or the last Match Counter Reset command. This counter is always enabled, but will only work as a match count when Match Code is enabled. If Match Code is not enabled, this counter records the number of good reads. This count can be requested at any time. <W> Match Code/Good Read Counter Reset Clicking Reset or sending serial command <W> sets the Match Counter to 00000. <X> Mismatch Counter Request Clicking Request or sending serial command <X> displays the message the number of labels successfully read that do not match the master label since power-on or the last Mismatch Counter command. <Y> Mismatch Counter Reset Clicking Reset or sending serial command <Y> sets the Mismatch Counter to zero.

4–Utilities

Quadrus 2D Code Reader User’s Manual

4-7

Page 94

Chapter 4

Trigger A serial trigger initiates a read cycle in the reader. It is useful testing the reader in a triggered mode. You can select, set up, and send serial trigger characters by accessing Trigger from the Utilities pull-down menu or by clicking the Utilities button and the Trigger tab to see the following:

4–Utilities

Figure 4-4 Serial Trigger Character

Setup Clicking SETUP loads the current serial trigger character.

Send Clicking Send sends a trigger character to the reader. Note: The trigger character can also be defined by serial configuration command as <Ki...> (see “Serial Trigger Character ” on page 3-11). <t...> Multiple Trigger Delays Format: <ttrigger delay # [1 to 10], trigger delay value [0 - 65,535 in 46 µs incr.]> You can enter the trigger delays for the multiple triggers from a single trigger event mode of operation. You enter the trigger delay number, followed by a trigger delay value. The multiple triggers from a single trigger mode functions as follows:

4-8

Quadrus 2D Code Reader User’s Manual

Page 95

The multiple triggers from a single trigger mode functions as follows: 1st Trigger Delay = Normal Trigger delay set up using the <KG> command. 2nd Trigger Delay = Multiple trigger delay set up using the <t1,XXXXX> command. 3rd Trigger Delay = Multiple trigger delay set up using the <t2,XXXXX> command. 4th Trigger Delay = Multiple trigger delay set up using the <t3,XXXXX> command. 5th Trigger Delay = Multiple trigger delay set up using the <t4,XXXXX> command. 6th Trigger Delay = Multiple trigger delay set up using the <t5,XXXXX> command. 7th Trigger Delay = Multiple trigger delay set up using the <t6,XXXXX> command. 8th Trigger Delay = Multiple trigger delay set up using the <t7,XXXXX> command. 9th Trigger Delay = Multiple trigger delay set up using the <t8,XXXXX> command. 10th Trigger Delay = Multiple trigger delay set up using the <t9,XXXXX> command. 11th Trigger Delay = Multiple trigger delay set up using the <t10,XXXXX> command. <t#?> Request Trigger Delay Value You can request the trigger delay value for any trigger by entering a question mark (?) immediately following the trigger delay number. For example, if you wanted the trigger delay value for the 5th trigger delay, enter <t5?>. If a value of 3500 had been entered, the Quadrus™ would return <t5/X3500>. If no value had been assigned, it would return <t5/0>. Note: Trigger Delay Value is returned in brackets with a forward slash regardless of the status of Operation Command Output Format.

4–Utilities

Quadrus 2D Code Reader User’s Manual

4-9

Page 96

Chapter 4

Master Label Note: Match code operations are intended for use when the reader is in a triggered mode, that is with an external or serial trigger enabled. If Match Code is enabled while the reader is in Continuous Read, the reader defaults to Continuous Read 1 Output and label data must change before the reader will output data again,unless a timeout, if enabled, occurs. When a master label has been defined, scanned labels are compared with the master label and under certain conditions relay driver signals may be output and internal counters advanced.

4–Utilities

You can set up master label functions by accessing Master Labels from the Utilities pull down menu or clicking the Utilities button and Master Label tab to see the following:

Figure 4-5 Master Label Commands

<E> Enable Match Code You can enable Match Code by clicking Setup in MENU\Operations or by sending <E>. This allows you to subsequently enter a master label and begin comparing labels against the master label.1 You can also enable Match Code (if not enabled) anytime you click the Program button in any of the 10 Master Label text blocks. You will see the following message:

1. If no master label is entered following the Enable Master Label, every subsequent decoded label will be a “mismatch” and will increment the mismatch counter by one.

4-10

Quadrus 2D Code Reader User’s Manual

Page 97

<F> Disable Match Code You can disable match code by clicking Disable in MENU\Operations\Match Code Type or by sending an <E> command. After enabling Match Code, a master label can be sent to the reader by entering the master label data by keystroke (SEND) or by instructing the reader to load the next label scanned as the master label (NEXT LABEL/MASTER). <)XXXX)> Send Master Label Data Typing in data into the first master label text block of the Master Label dialogue (figure 4-5) and clicking Send Text or sending a serial command <)XXXX)> (with data in place of the Xs) sends master label information as Master Label #1 in the master label data base from the host or a terminal. Master label information can be sent at any time and saved in nonvolatile memory with a <Z> command. A stored master label will not affect standard operations unless Match Code is enabled. The Xs denote alphanumeric data. <)> Request Master Label Data Clicking Setup in the Master Label dialogue (figure 4-5) or sending serial command <)> immediately sends master label #1 data to the host. Note: If Operation Command Output Format is set for Bracketed, the data from the reader will be returned with brackets and forward slash, such as <)/XXXX>. Note: If the master label information has previously been stored in nonvolatile memory (by a <Z> command), cycling the power will restore that information.

<G> Store Next Label Scanned as Master Label With Serial Data or External Edge enabled, clicking Setup followed by Program under the #1 master label text block in the Master Label dialogue, or sending serial command <G> causes the reader to read the next bar code label scanned as the master label #1and all subsequently decoded labels are compared against the new master label data. (See “Multilabel” on page 3-12.) Master label # can be any number from 1 to 10. For example if <G5> is entered the next label read will be saved as master Label #5.

Quadrus 2D Code Reader User’s Manual

4-11

4–Utilities

<))> Delete Master Label Data Selecting and deleting the data on data line 1 in the Master Label dialogue (figure 4-5) or sending serial command <))> deletes master label #1.

Page 98

Chapter 4 Master Label Data Base All match code operations, except sequential, utilize up to 10 labels that can be saved to non-volatile memory for access on power on.

Table 4-2 Maximum Characters

Master Label Number

Maximum Characters

#1

2048

#2

1024

#3

682

#4

512

#5

409

You can send up to 10 master labels to the master label data base by using this command or by using the Master Label dialogue (figure 4-5) by clicking Serial Data or External Edge and clicking Program for each master label that you want to add.

#6

341

#7

292

#8

256

#9

227

<G...> Store Next Label Scanned as Master Label #____

#10

204

It is important to note that changing the number of master labels in the data base changes the maximum length for each master label due to a fixed data base size. The maximum characters per each master label as the numbers change, as can be seen in table 4-2 or in figure 4-5. Max. # of Chars. per Label: XXX is updated as you change the total number of master labels.

Allows you to assign a master label number (1 through 10) for the next label you are scanning. <M...,...> Send Master Label Information Clicking OK or entering <Mmaster label #,master label data> 1

4–Utilities

<M...?> Request Master Label Data You can request master label information about a specific master label by following the master label # with a question mark. For example, to request data for master label #5, enter <M5?>. Quadrus™ will display master label #5 data in brackets. If no master label data is present then the label number <M5> (or <M/5> if Operation Command Output Format is enabled) would be displayed without label data. <M...,> Delete Master Label Data You can delete specific master labels. For example, to delete master label # 5, enter <M5,>. The blank master label data field following the comma tells the reader to delete the indicated master label from the data base.

1. You can set up your number of master labels by configuration command “Master Label Database Size ” on page 3-13; it will only allow the user Master Labels up to the number of master labels available. So, if the number of master labels has been set to 5, then this command will not allow you to enter a master label # (index) above 5.

4-12

Quadrus 2D Code Reader User’s Manual

Page 99

Checksum/Part Number You can request the software part # or check sum figures by accessing Checksum from the Utilities pull down menu or clicking theUtilities button and Checksum tab to see the following:

Figure 4-6 Checksum/Part Number

<#> Part Number Request Command Clicking Part # or sending <#> displays the software part number as 35-XXXXXX-XX, or as <#/35-XXXXXX-XX> if Operation Command Output Format is enabled. <!> Check Sum Request Command Clicking Checksum or sending <!> returns a four-digit hex number (specific to a given software version) used to verify a reader’s software.

4–Utilities

Quadrus 2D Code Reader User’s Manual

4-13

Page 100

Chapter 4

Interface You can control pulses and perform resets by accessing Interface from the Utilities pulldown menu or by clicking the Utilities button and the Interface tab to see the following:

Figure 4-7 Interface

Software Reset

4–Utilities

Caution: Record all data that you wish to save prior to sending these commands. A reset will cause the numeric counters in use to lose their count. For your convenience certain factory settings (Shutter Time, Offset, Contrast, and AGC Max) are saved apart from all other operating parameters. You can define and save your own set of factory settings as “Saved Factory Settings” or recall “Factory Defaults,” those set in the factory and not changeable, at any time. In the following diagrams the oval RAM box represents volatile memory and the rectangular boxes, non-volatile memory. The two rectangles on the right hold the special factory settings, those on the left apply to all other parameters. The following are accessed in Utilities/Interface menu under Software Reset: <A> Reset Clicking Reset in the Interface menu or sending serial command <A> resets all configuration commands in RAM and resets all counters and operating parameters.

<A>

Saved Factory Settings

Saved Parameters RAM Default Parameters

4-14

Initialize

Quadrus 2D Code Reader User’s Manual

Factory Defaults