4 REFERENCES 4.1 RS232 PARAMETERS 4.1.1 Handshaking Hardware handshaking (RTS/CTS) The RTS line is activated by the decoder before transmitting a character. Transmission is possible only if the CTS line (controlled by the Host) is active.
Signals at EIA levels RTS TX Transmitted data CTS
Transmitted data
Host busy
RTS/CTS handshaking
Software handshaking (XON/XOFF) During transmission, if the Host sends the XOFF character (13 Hex), the decoder interrupts the transmission with a maximum delay of one character and only resumes when the XON character (11 Hex) is received.
Transmitted data
Transmitted data
TX
RX XOFF Host busy
XON Host ready
XON/XOFF handshaking
4.1.2 Radio RX Lock This parameter is used to control radio reception from the guns to the cradle when the Host uses either RTS/CTS or XON/XOFF Handshaking. If enabled, this command blocks radio reception to the cradle when the Host has interrupted RS232 data communication (CTS low or XOFF). The effect is that guns cannot send data to the cradle until the Host resumes data communication (CTS high or XON). This is similar to FIFO buffering References - 4.1
Page 83
DLL6000-R/OM6010-R SYSTEM
DATALOGIC disabled, where data can be collected only when communication is possible. In any case, prior to interruption, data is buffered in the cradle. This command is only effective if handshaking is enabled. If used in a Multidrop Network, it only works for the Master cradle. If disabled, guns continue to send data to the cradle which buffers them even if data communication has been interrupted by the Host (CTS low or XOFF). If the buffer becomes full, the gun signals an error and any further data is discarded until communication is restored. This is similar to FIFO enabled where data collection continues even though communication is interrupted.
4.1.3 RX Timeout The timeout in data reception can be used to automatically end data reception after the specified period of time. If no character is received from the Host, after the timeout expires, any incomplete string (any string not terminated by <CR>) is flushed from the Cradle receive buffer and if being configured, the Cradle exits configuration mode.
4.2 WEDGE PARAMETERS 4.2.1 IBM AT - Alt Mode Interface The IBM AT - Alt Mode interface allows barcodes sent to the PC to be interpreted correctly independently from the Keyboard nationality used, therefore no keyboard nationality selection is required. When selecting the IBM AT - Alt Mode interface, make sure the Num lock parameter selection matches the Num lock key status on your keyboard.
4.2.2 Inter-character Delay When IBM SURE1 interface is selected, the default value for the Intercharacter Delay is forced to 5 msec. With this interface no lower value can be programmed for this parameter.
4.2 - References
Page 84
DATALOGIC
DLL6000-R/OM6010-R SYSTEM
4.3 PEN EMULATION PARAMETERS 4.3.1 Minimum Output Pulse This parameter sets the duration of the output pulse corresponding to the narrowest element in the barcode. In this way the code resolution is controlled by the signal sent to the decoder, independently of the physical resolution of the code read. The shortest pulse (200 µs) corresponds to a high resolution code emulation and therefore a shorter transfer speed to the decoder (for decoders able to work on high resolution codes). Likewise, longer pulses correspond to low resolution code emulation and therefore a longer transfer time to the decoder.
4.3.2 Overflow This parameter generates a white space before the first bar and after the last bar of the code. The selections are as follows: narrow = space 10 times the minimum output pulse. medium = space 20 times the minimum output pulse. wide = space 30 times the minimum output pulse.
References - 4.3
Page 85
DLL6000-R/OM6010-R SYSTEM
4.3.3 Output and Idle Levels
The following state diagrams describe the different output and idle level combinations for Pen emulati idle OUTPUT: Normal IDLE: Normal OUTPUT: Normal IDLE: Inverted idle OUTPUT: Inverted IDLE: Normal idle OUTPUT: Inverted IDLE: Inverted idle on:
space
DATALOGIC bar space space space
+
ULI LU barcode output, - » 7 I] ] barcode output, - »> bar barcode output, - > bar barcode output, - »>
Output and Idle Levels
4.4 CRADLE AND GUN OPERATING PARAMETERS
4.4.1 Output Data Format
The output data format towards the Host is:
[Header] [Gun_Addr] [Gun_Addr_delimiter] ] [Cradle_Addr] [Cradle_Addr_delimiter]
[Time stamp] [Ts_delimiter] [Code ID] CODE [Terminator]
[Items in square brackets are
4.4 - References optional.]
Page 86
DATALOGIC
DLL6000-R/OM6010-R SYSTEM
Header/Terminator Selection The header/terminator selection is not effected by the reading of the restore default code. In fact, header and terminator default values depend on the interface selection: RS232: no header, terminator CR-LF WEDGE: no header, terminator CR These default values are always restored through the reading of RS232 or WEDGE interface selection code, step 10 of the System Start Up procedure.
HEX 83 84 85 86 87 88 89 8A 8B 8C 8D 8E 8F 90 91 92 93 94 95 96 97 98 99 9A
EXTENDED KEYBOARD TO HEX CONVERSION TABLE IBM AT IBM 3153 IBM XT IBM 31xx, 32xx, 34xx, 37xx KEY KEY KEY ENTER ENTER FIELD EXIT TAB TAB TAB F1 F1 F1 F2 F2 F2 F3 F3 F3 F4 F4 F4 F5 F5 F5 F6 F6 F6 F7 F7 F7 F8 F8 F8 F9 F9 F9 F10 F10 F10 F11 ESC F11 F12 BACKSPACE F12 HOME HOME ENTER END END RESET PG UP PG UP INSERT PG DOWN PG DOWN DELETE FIELD
FIELD +
ENTER (Paddle)
ESC ESC CTRL (Right) CTRL (Right)
References - 4.5
Page 87
DLL6000-R/OM6010-R SYSTEM
DATALOGIC
Gun/Cradle Address Stamping It is possible to include the gun and cradle addresses in the message sent to the host. The Gun Address Stamping and the Cradle Address Stamping parameters consist of a 3-digit number in the range 000 to 126. For message output format, refer to the example on page 4.4.
Gun/Cradle Address Delimiter The Address Delimiters allow a character to be included to separate the Gun and Cradle Address stamping fields from the next fields in the message. Any character can be included in the hexadecimal range from 00 to FE. For message output format, refer to the example on page 4.4.
Time Stamping Format The Time Stamping parameter sets the format for hour and date information. It consists of 1 or 2 groups of numbers, each one made up of 6 decimal digits. For example, setting the Hour/Minutes/Seconds/Month/Day/Year format, the information 17:03:16 on June 12, 2000 will be formatted as 170316061200.
Time Stamping Delimiter The Time Stamping Delimiter allows a character to be included to separate the Time Stamping field from the next field in the message. Any character can be included in the hexadecimal range from 00 to FE.
4.6 - References
Page 88
DATALOGIC
DLL6000-R/OM6010-R SYSTEM
Code Identifier CODE IDENTIFIER TABLE DATALOGIC AIM STANDARD STANDARD 2/5 interleaved N ]Iy 2/5 industrial P ]Xy 2/5 normal 5 bars O ]Sy 2/5 matrix 3 bars Q ]Xy EAN 8 A ]E4 EAN 13 B ]E0 UPC A C ]Xy UPC E D ]Xy EAN 8 with 2 ADD ON J ]E5 EAN 8 with 5 ADD ON K ]E6 EAN 13 with 2 ADD ON L ]E1 EAN 13 with 5 ADD ON M ]E2 UPC A with 2 ADD ON F ]Xy UPC A with 5 ADD ON G ]Xy UPC E with 2 ADD ON H ]Xy UPC E with 5 ADD ON I ]Xy Code 39 V ]Ay Code 39 Full ASCII W ]Ay CODABAR R ]Fy ABC CODABAR S ]Xy Code 128 T ]C0 EAN 128 k ]C1 Code 93 U ]Gy CIP/39 Y ]Xy CIP/HR e ]Xy Code 32 X ]Xy CODE
NOTE AIM standard identifiers are not defined for all codes: the X identifier is assigned to the code for which the standard is not defined, y value depends on the selected options (check digit tested or not, check digit tx or not, etc.).
4.4.2 Radio Timeout After a code has been read and transmitted, the radio remains active for the amount of time set by the Radio Timeout parameter. This is particularly useful when the Host is expected to send a reply message to the gun. This parameter depends heavily on the application and is therefore provided here for system optimization according to your application. The default value for this parameter is 008 (about 1/2 sec.). This is a reasonable value for most applications, but it could need to be changed according to the following considerations:
References - 4.7
Page 89
DLL6000-R/OM6010-R SYSTEM
DATALOGIC
If your application does not require the host to send messages to the gun(s), (which is true for all Wedge applications), the best setting for the radio timeout is the minimum value 001, which is about 1/16 of a second. This allows maximum battery autonomy.
When your application requires bi-directional communications, the radio timeout must be set according to the number of guns and the amount of traffic so that no messages are lost. NOTE Setting the Radio Timeout to 000 causes the radio to always be ON. If you do this, the gun will accept messages from the host at any time, but the batteries will discharge quickly.
4.5 READING PARAMETERS 4.5.1 Trigger Signal Trigger signal is useful to determine the modality of the reader ON state: trigger level: the reader goes ON when the trigger is pressed and goes OFF when it is released; trigger pulse: the reader goes ON at the first trigger press and goes OFF only at a second press.
4.5.2 Trigger Timeout When this timeout is selected, the reader which is triggered ON but not decoding turns the laser OFF automatically after the desired period of time.
4.5.3 Reads per Cycle A reading cycle depends on the trigger signal selection (see par. 4.5.1) and on the trigger timeout selection (see par. 4.5.2). When one read per cycle is selected, the scanner turns off as soon as a valid code is decoded. To turn the reader on again, release and again press the trig-
4.8 - References
Page 90
DATALOGIC
DLL6000-R/OM6010-R SYSTEM ger in case the scanner is operating in ‘trigger level mode’, pull the trigger if the reader is operating in ‘trigger pulse mode’. When multiple reads per cycle is selected, the scanner turns off after a good decoding for the time necessary to transmit the code and activate the beeper, then it turns on again. The scanner turns off after a trigger press according to the ‘trigger signal’ selection or when the timeout expires. The Safety Time parameter can be used in this case to avoid unwanted multiple reading of the same code, see safety time below.
4.5.4 Safety Time Safety time prevents the device from immediately decoding the same code more than once. Same code consecutive reading can be disabled requiring the reader to be removed from the code (no decoding) for at least 400 ms, or a timeout can be set up to 9.9 seconds before the decoder will accept the same code. Reading is immediate if the code changes.
4.5.5 Single-Store When single-store mode is enabled, if the DLL6000-R fails to transmit a code to the cradle, it enters a special operating mode that prevents the user from reading barcodes. When such operating mode is entered, the trigger no longer enables barcode reading but is used to retry the transmission itself. Once the transmission is successful the gun returns to the standard mode. Single-store may be useful if you often read codes at the limit of the coverage area and there is a chance that code transmission can fail. In such case single-store allows you to move to a more favorable position or location (i.e. closer to the cradle) and retry transmission without the necessity of re-reading the code since it is already stored in the gun. Conversely, if single-store is disabled, and the user wants to retry transmission, the code must be read again, and therefore the attempt must be made from basically the same location. If the user gives up, he doesn't know if the transaction was successful. (Actually the transmission could have been successful but the cradle may have been unable to acknowledge the message). There are applications in which there is no risk of transmission failure. In such cases it may be
References - 4.9