Page 31
INSTALLATION
2.3.1
2
Main/Aux. Serial Interface and I/O Connector
The DS6300 master/slave model is equipped with a 25-pin male D-sub connector for connection to the host computer, power supply and input/output signals. The DS6300 fieldbus models (Ethernet, DeviceNet, Profibus) adopt a 26-pin male connector instead of the 25-pin one. The details of the connector pins are indicated in the following table: 13
1 9
1 10 19
18
26
Figure 17 - 26-pin Connector
25
14
Figure 18 - 25-pin Connector
DS6300 25-pin/26-pin D-sub connector pinout Pin
Name
Function Chassis - internally connected to GND 1 CHASSIS Cable shield connected to chassis 20 RXAUX Receive data of auxiliary RS232 (referred to GND) 21 TXAUX Transmit data of auxiliary RS232 (referred to GND) 8 OUT 1+ Configurable digital output 1 - positive pin 22 OUT 1Configurable digital output 1 - negative pin 11 OUT 2+ Configurable digital output 2 - positive pin 12 OUT 2Configurable digital output 2 - negative pin 16 OUT 3A Configurable digital output 3 - polarity insensitive 17 OUT 3B Configurable digital output 3 - polarity insensitive 18 EXT_TRIG/PS A External trigger (polarity insensitive) for PS 19 EXT_TRIG/PS B External trigger (polarity insensitive) for PS 6 IN 2/ENC A Input signal 2 (polarity insensitive) for Encoder 10 IN 2/ENC B Input signal 2 (polarity insensitive) for Encoder 14 IN 3A Input signal 3 (polarity insensitive) 15 IN 4A Input signal 4 (polarity insensitive) 24 IN_REF Common reference of IN3 and IN4 (polarity insensitive) 9,13 VS Supply voltage - positive pin 23,25,26 GND Supply voltage - negative pin Main Interface Connector Pinout RS485 RS485 Pin 20 mA C.L. RS232 (INT-30 with C-BOX 100 only) Full Duplex Half Duplex 2 TX TX485 + RTX485 + see INT-30 instructions 3 RX * RX485 + 4 RTS TX485 RTX485 5 CTS * RX485 7 GND_ISO GND_ISO GND_ISO Pin 26 is only available for fieldbus models (Ethernet, DeviceNet, Profibus). * Do not leave floating, see par. "RS485 Full-Duplex Interface" for connection details.
17
Page 32
DS6300
2
Main Interface The main serial interface is compatible with the following electrical standards: RS232 RS485 full-duplex RS485 half-duplex (20 mA current loop) The 20 mA Current Loop interface is available by using the C-BOX 100 with the optional INT-30 accessory installed in it. The scanner communicates to the C-BOX 100 through the RS232 interface and the INT-30 converts the signals. The main serial interface type and its relative parameters (baud rate, data bits, etc.) are selected via software using the Genius™ utility program. For more details refer to the section "Main Serial Port" in the Genius™ Help On Line. Details regarding the connections and use of the main interface selection are given in the next paragraphs.
RS232 Interface The main serial interface is used for communication with the Host computer and allows both transmission of code data and configuring the reader. The overall maximum cable length should not exceed 15 m (50 ft). The following pins of the 25-pin and 26-pin connector are used for RS232 interface connection depending on the reader model: Pin 2 3 4 5 7
Name TX RX RTS CTS GND_ISO
Function Transmit Receive Request to send Clear to send Main signal ground
The RTS and CTS signals control data transmission and synchronize the connected devices. If the RTS/CTS hardware protocol is enabled, the DS6300 activates the RTS output to indicate a message can be transmitted. The receiving unit must activate the CTS input to enable the transmission. DS6300
USER INTERFACE 2
TX
3
RX
4
RTS
5
CTS
7
GND_ISO
1
CHASSIS
RXD TXD
SGND Main Isolated Earth Ground
Figure 19 - RS232 Connections
18
Page 33
INSTALLATION
2
RS485 Full-Duplex Interface The RS485 full-duplex interface is used for non-polled communication protocols in point-to-point connections over longer distances than those acceptable for RS232 communications or in electrically noisy environments. The overall maximum cable length should not exceed 1200 m (3937 ft). The following pins of the 25-pin and 26-pin connector are used for RS485 full-duplex interface connection: Pin 2 3 4 5 7
Name TX485 + RX485 + TX485 RX485 GND_ISO
Function RS485 output (+) RS485 input (+) RS485 output (-) RS485 input (-) Main signal ground
DS6300
USER INTERFACE 2
TX485+
3
RX485+
4
TX485-
5
RX485-
7
GND_ISO
1
CHASSIS
RX485+ TX485+ RX485TX485-
SGND Main Isolated Earth Ground
Figure 20 - RS485 Full-Duplex Interface Connections
For applications that do not use RX485 signals, do not leave these lines floating but connect them to GND_ISO as shown below. NOTE
DS6300
USER INTERFACE 2
TX485+
4
TX485-
3
RX485+
5
RX485-
7
GND_ISO
1
CHASSIS
RX485+ RX485-
SGND Main Isolated Earth Ground
Figure 21 - RS485 Full-Duplex Interface Connections Using Only TX Signals
19
Page 34
DS6300
2
RS485 Half-Duplex Interface The RS485 half-duplex interface can be used for multidrop connections with a Datalogic multiplexer or it can also be used for a master/slave layout. The overall maximum cable length should not exceed 1200 m (3937 ft). The following pins of the 25-pin and 26-pin connector are used for RS485 half-duplex interface connection: Pin 2 4 7
Name RTX485 + RTX485 GND_ISO
Function RS485 input/output (+) RS485 input/output (-) Main signal ground
DS6300
MULTIPLEXER 2
RTX485+
4
RTX485-
7
GND_ISO
1
CHASSIS
RTX485+ RTX485RS485REF Earth Ground
Figure 22 – RS485 Half-Duplex Interface Connections
20
Page 35
INSTALLATION
2
Auxiliary Interface The auxiliary serial interface is equipped with RS232 full-duplex interface connections. The interface type is exclusive and is selectable through the Genius™ configuration program. The overall maximum cable length should not exceed 15 m (50 ft). The following pins of the 25-pin and 26-pin connector are used for RS232 full-duplex interface connection: Pin 20 21 23
Name RXAUX TXAUX SGND AUX
Function Receive data Transmit data Auxiliary signal ground
DS6300
USER INTERFACE 20
RXAUX
21
TXAUX
23
SGNDAUX
1
CHASSIS
TXD RXD GND Earth Ground
Figure 23 - RS232 Auxiliary Interface Connections
21
Page 36
DS6300
2
Inputs The inputs of the reader are on the 25-pin and 26-pin connector (Figure D, 1 and Figure E, 1) of the DS6300. These inputs are called EXT_TRIG/PS, IN2/ENC, IN3 and IN4. Pin 18 19 6 10 14 15 24
Name EXT_TRIG/PS A EXT_TRIG/PS B IN2/ENC A IN2/ENC B IN3A IN4A IN_REF
Function External trigger (polarity insensitive) for PS External trigger (polarity insensitive) for PS Input signal 2 (polarity insensitive) for Encoder Input signal 2 (polarity insensitive) for Encoder Input signal 3 (polarity insensitive) Input signal 4 (polarity insensitive) Common reference of IN3 and IN4 (polarity insensitive)
IN2/ENC is normally used for the Encoder input. The maximum Encoder frequency is 2 KHz. EXT_TRIG/PS is the main presence sensor. When active, this input tells the scanner to scan for a code and that decoding can take place. The yellow LED (Figure C, 3) indicates the EXT_TRIG/PS is active. IN3 and IN4 can be used as the stop signal for the reading phase. All inputs are optocoupled, polarity insensitive, and driven by a constant current generator; the command signal is filtered through an anti-disturbance circuit which generates a delay which can be set to 5 ms or 500 µs. In particular, EXT_TRIG/PS, IN3 and IN4 share the same value which usually corresponds to 5 ms when using a photoelectric sensor, while IN2/ENC has a different value which is set to 500 µs when this input is used for the Encoder. Vext
DS6300
EXTERNAL TRIGGER/ENCODER V
A/B + 5V
~ +
-
~
B/A
Ground
Figure 24 – PNP Command Input Connection using External Power
DS6300
EXTERNAL TRIGGER/ENCODER VS
V
A/B + 5V
~ + ~
-
B/A
Ground
GND
Figure 25 - PNP Command Input Connection using Scanner Power
22
Page 37
INSTALLATION
2 DS6300
EXTERNAL TRIGGER/ENCODER Vext A/B
+ 5V
~
V
-
+ ~
B/A
Ground
Figure 26 - NPN Command Input Connection using External Power
EXTERNAL TRIGGER/ENCODER
DS6300 VS A/B + 5V
~ +
V
-
~
B/A GND
Ground
Figure 27 - NPN Command Input Connection using Scanner Power
Vext
DS6300
EXTERNAL DEVICE V
IN3A + 5V
~ +
-
~
Ground
Vext V IN4A + 5V
~ + ~
-
INREF
Ground
Figure 28 - IN3/IN4 PNP Input Command using External Power
23
Page 38
DS6300
2 DS6300
EXTERNAL DEVICE VS
INREF
+ 5V
~ +
V
-
~
IN3A Ground
EXTERNAL DEVICE + 5V
~ +
V
-
~
IN4A GND
Ground
Figure 29 - IN3/IN4 NPN Input Command using Scanner Power
Input devices can be supplied by either scanner power (VS and GND) or external power supplies (Vext). Electrical isolation between the input command logic and the scanner is maintained when powering the input devices from an external supply voltage (Vext). The driving logic of the input signals may be powered, for convenience, with the voltage supply between pins 9 (VS) and 23 (GND) of the 26-pin I/O connector. In this case, however, the device is no longer electrically isolated. The voltage available on the 26-pin I/O connector, is physically the same as used to power the scanner. The electrical features of these inputs are: Maximum voltage 30 V Maximum current 10 mA
Outputs Three general purpose outputs are available. Pin 8 22 11 12 16 17
Name OUT 1+ OUT 1OUT 2+ OUT 2OUT 3A OUT 3B
Function Configurable digital output 1 – positive pin Configurable digital output 1 – negative pin Configurable digital output 2 – positive pin Configurable digital output 2 – negative pin Configurable digital output 3 – polarity insensitive Configurable digital output 3 – polarity insensitive
The function of the three outputs OUT1, OUT2 and OUT3 can be defined by the user. Refer to Genius™ Help On-Line for further details.
24
Page 39
INSTALLATION
2
By default, OUT1 is associated with COMPLETE READ event, which activates when the code has been read correctly. In case the reader has been programmed to read several codes within the same reading phase, the event activates when all codes have been read. OUT2 is associated with NO READ event, which activates when no code has been read. OUT3 is associated with NONE, which means that the output is always in Line State. The OUT1 and OUT2 electrical features are given below: Collector-emitter voltage Collector current (pulse) Collector current (continuous) Saturation voltage (VCE) Maximum power dissipation
30 V Max. 130 mA Max. 40 mA Max. 1 V at 10 mA Max. 90 mW at 50°C (Ambient temperature).
The limit requested by the maximum power dissipation is more important than that of the maximum collector current: if one of these outputs is continuously driven, the maximum current must not be more than 40 mA although 130 mA may be reached in pulse conditions. DS6300
USER INTERFACE Vext 30 Vdc max
+
-
Figure 30 – Output 1 and Output 2 Interface
When the load is powered by an external power supply, the voltage must be less than 30 V. OUT3 has different electrical features, since it is a bi-directional solid state relay with built-in current limit protection. If this output is continuously driven, the maximum current must be not more than 200 mA although more than 300 mA may be reached in pulse conditions for an ambient temperature of 25°C. At the maximum ambient temperature of 50°C the maximum respective current is 150 mA continuous and 240 mA pulse. The OUT3 electrical features are given below: Maximum voltage Collector current (pulse) Collector current (continuous) R on R off Off-state leakage current Maximum power dissipation
± 100 V 240 mA Max. 150 mA Max. 6 – 15 Ω > 500 Ω < 1 µA 550 mW at 50°C (Ambient temperature).
25
Page 40
DS6300
2 DS6300
USER INTERFACE Vext 100 Vdc max
A
B
Figure 31 – Output 3 Interface
The command signal is filtered and generates a delay of about 50 µs for OUT1 and OUT2 and 1 ms for OUT3.
2.3.2
Lonworks Connectors
Do not connect an RS232 port to the 9-pin Lonworks Connector. This may damage your Laptop PC. CAUTION The local network used by DS6300 exploits a Lonworks standard communication system requiring only two wires (polarity insensitive) to enable a connection. The connector also provides a positive and a negative supplying wire. In this way, all the slave readers can be powered by the master through the Datalogic standard cables. When working in applications requiring enhanced synchronization capabilities, the DS6300 master reader (output) transmits two system signals named Sys_I/O and Sys_Enc_I/O to the slave readers (input). For example, when working with applications requiring an encoder the signal is received by the master and directly transmitted to the slaves through the cable. The internal circuits generating the system signals are externally supplied by means of the VS_I/O and REF_I/O pins and are isolated from the reader supply voltage. The use of these system circuits is not required in all the operating modes (see par. 2.7 for details). Anyway, for a correct system functioning it is suggested to use Datalogic cables and accessories and follow the description of the typical layout (see par. 2.7 for details). 1
5
9
6
6
Female (all models)
9
Male (Master/Slave model)
Figure 32 – 9-pin Local Lonworks Connectors
26
5
1