Page 92
DS6300
5
5 MAINTENANCE 5.1 CLEANING Clean the laser beam output window (Figure A, 7) periodically for correct operation of the scanner. Dust, dirt, etc. on the window may alter the reading performance. Repeat the operation frequently in particularly dirty environments. Use soft material and alcohol to clean the window and avoid any abrasive substances.
Clean the window of the DS6300 when the scanner is turned off or at least when the laser beam is not active. WARNING
5.2 AUTOMATIC SCANNER REPLACEMENT (ASR) The Datalogic Automatic Scanner Replacement (ASR) procedure allows restoring system functioning automatically after one or more scanners are replaced in a Master/Slave Lonworks network. The Master must be prepared at the time of installation in order for this procedure to work correctly.
5.2.1
ASR Network Configuration
1. On the Master scanner, check the Modify & Backup Lon Slave Scanner Configuration parameter in Genius™ and configure the Lonworks Slave Scanner Common Parameters (Code and Reconstruction Parameters). 2. Enable the ASR procedure through the Enable A.S.R. parameter in the Master configuration. 3. Send the configuration to the Master EEPROM to force the Slave Operating Mode, Code Reading Symbologies and Reconstruction parameters. Now the Slave scanners are configured through the Master and the ASR procedure is implemented. 4. Save this configuration to file (.ddc).
78
Page 93
MAINTENANCE
5.2.2
5
Scanner Replacement Procedure
The ASR procedure requires replacing one scanner at a time. NOTE
Slave 1. Power down the entire system. 2. Replace the Slave scanner with a new one (default settings). 3. Power up the system and wait for initialization.
Master 1. Load the saved configuration from file (.ddc) to the new Master. 2. Power down the entire system. 3. Replace the Master scanner with the new one. 4. Power up the system and wait for initialization.
The ASR works only if both the Master and Slave devices have software 6.40 or later. NOTE
79
Page 94
DS6300
6
6 TROUBLESHOOTING
NOTE
Before contacting your local Datalogic office or Datalogic Partner or ARC, it is suggested to save the device configuration to a *.ddc file by means of the Genius™ software configuration program and check the device exact model and serial number.
TROUBLESHOOTING GUIDE Problem Power On: the “Power On” LED is not lit.
Suggestion • Is power connected? • If using a power adapter (like PG6000), is it connected to AC source? • If using rail power, does rail have power? • If using C-BOX 100, does it have power (check switch and LED)? Check if you are referring to the 25/26-pin connector or to the C-BOX 100 spring clamp connectors. • Measure voltage either at pin 13 and pin 25 (for 25/26-pin connector) or at spring clamp 1 and (for C-BOX 100). • Check carefully if you are referring to the On Line Mode: 25/26-pin connector or to the C-BOX 100 the Master’s “Phase On” LED is not lit spring clamp connectors. (when external trigger activates). • Is sensor connected to EXT TRIG/PS input? • Is power supplied to photo sensor? • Is power supplied to one out of the two EXT TRIG/PS (NPN output)? • Is one out of the two EXT TRIG/PS grounded (PNP output)? • Are the photo sensor LEDs (if any) working correctly? • Is the sensor/reflector system aligned (if present)? • Is the software configuration consistent with On Line Mode: the application condition (operating mode, the Master’s “Phase On” LED is etc..)? correctly lit but nothing happens (no In the Genius™ software configuration reading results). program select the OPERATING MODES folder and check for related parameters. • In the Genius™ program select the Serial On Line Mode: OPERATING MODE folder and check if the reader is not triggered (no reading serial on line is enabled as “On Line options” results). parameter value. • Are the Start-Stop string correctly assigned? • Is the serial trigger source correctly connected and configured) 80
Page 95
TROUBLESHOOTING
6 TROUBLESHOOTING GUIDE
Problem Suggestion On Line Mode and Serial On Line • In the Genius™ software configuration program select the OPERATING MODES Mode: folder and check the “Reading Phase the reader does not respond correctly to Timeout” parameterization. the expected external signal end. • Check synchronization of reading pulse with Reading: object to read. it is not possible to read the target • Is the scan line correctly positioned? barcode (always returns No Read) • Place barcode in the center of scan line and run TEST MODE (selectable by Genius™ as Operating Modes). If you still have troubles, check the following: • Is the reading distance within that allowed (see reading diagrams)? • Is the Tilt angle too big? • Is the Skew angle less than 10° (direct reflection)? • Choose the CODE tab and enable different code types (except Pharmacode). LENGTH = Variable. • Is the barcode quality sufficient? • If you had no success, try to perform the test using the BARCODE TEST CHART included with the product. • Is serial cable connected? Communication: the device is not transmitting anything to • Is correct wiring respected? the host. • If using MAIN RS232 or RS485 interface, is the reference ground connected to proper SGND Main Isolated (also referred to as GND_ISO)? Be careful that it is not completely different from GND power ground. • If using C-BOX 100, be sure the RS485 termination switch is positioned to OFF. • Are serial host settings equivalent to serial device setting? • In the Genius™ program enable the DATA Communication: COMMUNICATION SETTINGS/MAINdata do not appear on the terminal. AUXILIARY PORT\DATA TX parameter. • In the Genius™ program select the DATA Communication: COMMUNICATION SETTINGS/DATA data transferred to the host are FORMAT folder and check for HEADER, incorrect, corrupted or incomplete. TERMINATOR, SEPARATOR and FILL CHAR values. • Check the CODE FIELD LENGTH value, too. • Are the COM port parameters correctly assigned?
81
Page 96
DS6300
6 TROUBLESHOOTING GUIDE
Problem Suggestion How do I obtain my units’ serial • The device serial number is printed on the device identification label that is affixed to the numbers? reader (Figure A, 2). • The serial number is also displayed when connecting the device through the Genius™ program. • Serial numbers consist of 9 characters: one letter, 2 numbers, another letter followed by 5 numbers.
82