Browse
(440) 299-7778
Menu
(440) 299-7778
One-year parts and labor warranty30-day money-back guaranteeEnglishEspañolContact

The text of a manual · part 3 of 4

DS6300 Quick Reference Guide, pages 21 to 30

Pages 21 to 30 of 39

The original document is the PDF (2.6 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.

For the Datalogic DS-6300

Page 21

DS6300 DEVICENET MODEL SOFTWARE FEATURES Readable Codes Interleaved 2/5 Code 39 standard Codabar Code 128 EAN 128 Code 93 (Standard & Full ASCII) EAN/UPC (including Add-on 2 and Add-on 5) Code Selection Up to 10 codes during one reading phase Headers and Up to 128-byte headers and128Terminators byte terminators Operating On Line, Automatic, Test Modes Config. Mode Genius™ utility program Param. Storage Non-volatile internal FLASH PHYSICAL FEATURES Dimensions mm (inch) Weight

Std Models 110x113x99 (4.33x4.45x3.9)

Oscill. Mirror 113x180x104.5 (4.45x7.08x4.11)

1.5 kg (3.3 lb)

2.0 kg (4.4 lb)

ENVIRONMENTAL FEATURES Operating 0° to +40 °C (+32 to +104 °F) Temperature -20° to +70 °C Storage Temperature (-4° to +158 °F) Humidity 90% non condensing Ambient Light 3500 lux Immunity Vibration Resistance 14mm @ 2 to 10Hz IEC 68-2-6 test FC 1.5 mm @13 to 55 Hz 2 g @ 70 to 200 Hz 2 hours on each axis Shock Resistance 30 g; 11 ms IEC 68-2-27 test EA 3 shocks on each axis IP64 Protection Class

Accessories: Name CAB-6011 CAB-6012 CAB-6015 C-BOX 100 INT-30 GFC-60 GFC-600 GFX-60 PWR-120 BTK-6000 PG6002 PG6001 PG6000 FBK-6000 US-60 PH-1 MEP-543 OEK-2 OEK-1

Description Cable to C-BOX100 1 m Cable to C-BOX100 2 m Cable to C-BOX100 5 m Passive connection box 20 mA C.L. interface board for C-BOX 100 90° mirror 90° mirror close distance X-pattern mirror Power unit 110/230 V AC - 24 V DC Terminator kit (5 pcs) Single unit power supply – US Single unit power supply – UK Single unit power supply – EU Fast bracket kit (2 pcs) Mounting bracket kit (5 pcs) for multisided stations Photocell kit - PNP Photocell kit – NPN Optical encoder (10 m cable + spring) Optical encoder kit + 10 m cable

Part Number 93A051221 93A051222 93A051223 93ACC1510 93A151022 93A201100 93A201102 93ACC1730 93ACC1530 93ACC1710 93ACC1718 93ACC1719 93ACC1720 93ACC1721 93ACC1729 93ACC1791 93ACC1728 93ACC1770 93ACC1600

Electrical Connections: The DS6300 DeviceNet reader provides a 26-pin male D-sub connector for connection to power supply and input/output signals. A DeviceNet connector is used for connection to the remote Host, while a local Lonworks 9-pin female connector connects the DeviceNet master to the first slave reader of the system.

When using DeviceNet, the Main serial interface is disabled and must not be physically connected. NOTE

19

Page 22

DS6300 DEVICENET MODEL The details of the connector pins are indicated in the following table: 26-pin D-Sub Connector Pinout Pin

Name

Function

Chassis - internally connected to GND Cable shield connected to chassis 20 RXAUX Receive data of auxiliary RS232 (referred to GND) 21 TXAUX Transmit data of auxiliary RS232 (referred to GND) 8 OUT 1+ Configurable digital output 1 – positive pin 22 OUT 1Configurable digital output 1 – negative pin 11 OUT 2+ Configurable digital output 2 – positive pin 12 OUT 2Configurable digital output 2 – negative pin 9 1 16 OUT 3A Configurable digital output 3 – polarity insensitive 10 18 19 17 OUT 3B Configurable digital output 3 – polarity insensitive 26 18 EXT_TRIG/PS A External trigger (polarity insensitive) for PS 26-pin male D-sub Connector 19 EXT_TRIG/PS B External trigger (polarity insensitive) for PS 6 IN2/ENC A Input signal 2 (polarity insensitive) for Encoder 10 IN2/ENC B Input signal 2 (polarity insensitive) for Encoder 14 IN3A Input signal 3 (polarity insensitive) 15 IN4A Input signal 4 (polarity insensitive) 24 IN_REF Common reference of IN3 and IN4 (polarity insensitive) 9, 13 VS Supply voltage – positive pin 23, 25, 26 GND Supply voltage – negative pin 1

*

CHASSIS

Pin

RS232

RS485 Full-Duplex

RS485 Half-Duplex

2 3 4 5 7

TX RX RTS CTS GND_ISO

TX485+ * RX485+ TX485* RX485GND_ISO

RTX485+

20 mA C.L (INT-30 with C-BOX 100 only)

see INT-30 instructions

RTX485GND_ISO

Do not leave floating, see Reference Manual for connection details. 9-pin Lonworks Connector Pinout Pin

Name

1

CHASSIS

9 2 6 3 4 5 7 8

VS GND VS_I/O Ref_I/O SYS_ENC_I/O SYS_I/O LON A LON B

Function Cable shield internally connected by capacitor to chassis Supply voltage – positive pin Supply voltage – negative pin Supply voltage of I/O circuit Reference voltage of I/O circuit System signal System signal Lonworks line (polarity insensitive) Lonworks line (polarity insensitive)

1

5

9

6

9-pin female Local Lonworks Connector

5-pin DeviceNet Connector Pinout Pin 2 5 1 4 3

NOTE

20

Name V+ CAN_L SHIELD CAN_H V-

Function Supply voltage – positive pin CAN bus data line – L Shield CAN bus data line – H Supply voltage – negative pin

4

3

1

2

5

5-pin male DeviceNet Connector

The power supplied on pin V+ and V- is used only to propagate power to the section of the DeviceNet board directly connected to the Bus. It is completely isolated from the DS6300 power which must be supplied on pin 9, 13 and pin 23, 25 of the 26-pin Main/Aux connector.

Page 23

DS6300 DEVICENET MODEL

Connectivity: Point-to-Point Layout Remote Host

Fieldbus Network DS6300 C-BOX 100 CAB-601X

*P.S PG6000

*

P.S. (Presence Sensor) connected to External Trigger/PS input.

Pass Through Layout Remote Host

*P.S.

*P.S. Gryphon

DS4600A

<

<

EN T

CAB-601X

C-BOX 100 AUX

DS4600A

EN T

<

DS6300

<

Fieldbus Network

C-BOX 100

C-BOX 100

*P.S.

2

1

2

1

PWR-120

1 Main Serial Interface

*

2 Auxiliary Serial Interface

P.S. (Presence Sensor) connected to External Trigger/PS input.

21

Page 24

DS6300 DEVICENET MODEL Local Lonworks Network Remote Host

Fieldbus Network

Slave 2***

Slave 1***

Master

C-BOX 100** CAB-610X

CAB-610X

CAB-60XX

P.S.*

CAB-610X

Encoder**** CAB-610X

CAB-610X

CAB-63XX

BTK-6000

Slave 5***

Slave 3***

Slave 4***

PWR-240

CAB-63XX

Fieldbus Small Synchronized Network

*

P.S. (Presence Sensor) connected to External Trigger/PS input.

**

C-BOX 100 modified to accept scanner power.

***

The Slave scanners are Master/Slave models, which allow Lonworks network propagation.

****

Encoder connected to IN2/ENC input.

22

Page 25

DS6300 OSCILLATING MIRROR MODEL

DS6300-105-0XX OSCILLATING MIRROR MODEL

1

Figure A 1 Laser Beam Output Window

Oscillating mirror models are used when coverage of a large reading area is required, mainly in picket fence applications. The DS6300 scanner mounts a dedicated optic head with integrated oscillating mirror driven by a linear motor. The speed, precision, repeatability, and reliability of this driving technology assure high level performance. The new oscillating mirror is completely software controlled and software programmable. The Genius™ software tool allows adjusting the linear motor speed (oscillating frequency) and the upper and lower limits of the oscillation by defining the top and bottom line limit angles. When the oscillating mirror is programmed to read barcode labels at very small angles, position the reader to assure at least 10° for the Skew angle (see DS6300 Reference Manual). This angle refers to the most inclined or external laser line, so that all other laser lines assure more than 10° Skew. This avoids the direct reflection of the laser light emitted by the reader.

10° Oscillating Mirror Skew Angle

Otherwise, the scanner can be mounted at an angle of inclination of 17.5° in order to attain symmetrical deflection ranges.

10 7. 5°

17.5°

Oscillating Mirror Reading Position

In the above case, the zone where the scan line is perpendicular to the reflecting surface corresponds to a neutral zone at the center of the reading field.

23

Page 26

DS6300 OSCILLATING MIRROR MODEL The mirror can be deflected up to 40°. Oscillation with respect to the output window median axis is asymmetrical (see figure below).

° 37.5

40°

-2.5°

0°

Oscillating Mirror Maximum Aperture and Asymmetry

By configuring the oscillating speed up to the maximum value of 19 Hz, raster emulation can be performed for reading fast moving objects. Hz

Max. Aperture

0-5 6-10 11-15 16-19

40° 30° 20° 10°

By limiting the raster width to the minimum necessary, the number of scans on the reading surface is increased. NOTE

Oscillating angles are selected in software where the minimum and maximum angles correspond to –2.5° and +37.5°.

+37.5° +17.5°

The scanner can be tilted in order for the 17.5° software setting to correspond with the 0° horizontal plane.

-2.5°

Oscillating Mirror Extreme Angle Positions

These models provide higher scanning speed (1200 scans/sec) compared to standard models and the reading performance is not adversely effected by the oscillating mirror. The example represents the selection of an angle of +10° for the bottom line and an angle of +20° for the top line (see figure beside).

+37.5° +27.5° +17.5°

Oscillating Mode

24

Page 27

COMMON FEATURES

COMMON FEATURES C-BOX 100 Pinout for DS6300: The table below gives the pinout of the C-BOX 100 terminal block connectors. Use this pinout when the DS6300 reader is connected in a network by means of the C-BOX 100: C-BOX 100 Terminal Block Connectors Power 1, 3, 5 2, 4, 6 7, 8 20, 40 27 28 29 30 31, 33 32, 34 36 21 22 23 24 25 26 35 37 38, 39

11, 15 12, 16 17 18 10, 14, 19 9, 13 *

VS GND EARTH GROUND Reserved Inputs EXT TRIG/PS A (polarity insensitive) for PS EXT TRIG/PS B (polarity insensitive) for PS IN 2/ENC A (polarity insensitive) for Encoder IN 2/ENC B (polarity insensitive) for Encoder IN 3A (polarity insensitive) IN 4A (polarity insensitive) IN 3B/IN 4B Reference (polarity insensitive) Outputs OUT 1+ OUT 1OUT 2+ OUT 2OUT 3A (polarity insensitive) OUT 3B (polarity insensitive) Auxiliary Interface TX AUX RX AUX GND Main Interface RS232

RS485 Full-Duplex

RS485 Half-Duplex

TX 232 RTS 232 RX 232 CTS 232 SGND Main Isolated

TX 485+ TX 485* RX 485+ * RX 485SGND Main Isolated RS485 Cable Shield

RTX 485+ RTX 485-

20 mA C.L. (with INT-30 only)

see INT-30 instructions SGND Main Isolated RS485 Cable Shield

Do not leave floating, see Reference Manual for connection details.

25

Page 28

COMMON FEATURES

Mechanical Installation: TM

The DS6300 reader can be positioned and installed in the best way possible as a result of the Step-A-Head feature. Thanks to the separation between Head and Base, you can modify the orientation of the decoder base, and therefore display-keypad and connector panels, while keeping the optic head in the correct reading position. The reading head and the decoder base can be rotated independently from each other allowing the installation even in the most critical locations. To rotate the head follow the given procedure: 1.

detach the head from the base by unscrewing the four fixing screws;

2.

rotate the head in the desired position;

3.

loosen but don't remove the two screws on top of the head;

4.

affix the head onto the base carefully aligning the four fixing screws and progressively tightening them about half-way;

5.

completely tighten the two screws on top of the head;

6.

completely tighten the four fixing screws.

Head Screws

Fixing Screw (4)

Step-A-Head™ Feature

The following diagrams give the overall dimensions of the reader standard model, oscillating mirror model and mounting bracket. They may be used for their installation: 42 1.65 10 0.4

2 N° °2 .1 N Ø40.16 Ø

106°

= mm inch

113 4.45

=

50 1.96 72 2.83 100 3.93

=

=

73.2 2.88 36 1.41

74 2.85

99 3.90

30 1.18

130 5.12

TS LO S 4 S OT N° 4 SL 4 .5 8 N ° 0 .1

126 4.96

76 2.99

4 0.15

35 1.37

30 1.18

60 2.36

82 3.22 25 50 1.96 0.98 20 18 0.78 0.71 N°2

22 0.86 50 82 1.96 3.22

16.5 0.65 85 3.34

OTS °2 SL S 8.5 N °2 SLOT N .5 0.33 Ø8 .33 Ø0 mm inch

110 4.33

DS6300 Overall Dimensions

ST-237 Mounting Bracket Overall Dimensions 42 1.65 4 0.15 50 1.96 35 1.37

130

5.11

82 3.22 50 1.96 20 0.78

30 1.18 60 2.36 85 3.35

22 0.86

25 0.98

18 0.71

10 0.4

Ø4.1 0.16

16.5 0.65

11 0.43 14 0.55

99 3.90

50 1.96 72 2.83 100 3.93

69 2.72 56 2.20

114 4.48

R36 mm inch

75 2.95

102 4.01

113 4.45

R22

110.3 4.34

R5

11 0.43 14 0.55

36 1.41

63.5 2.50

104.5 4.11

R1

Ø8 mm inch

.5

180 7.08

DS6300 Oscillating Mirror Model Overall Dimensions

26

ST-210 Mounting Bracket Overall Dimensions

Page 29

COMMON FEATURES

Typical Installations: Standard Installation The DS6300 scanner is mounted on the ST-237 106° mounting bracket which guarantees a built-in Skew angle (S in the figure below) of 16° with respect to the frame plane (typically the Skew angle should be between 10° - 20°). This avoids the direct reflection of the laser light emitted by the scanner. Furthermore, the bracket guides allow adjusting the Tilt angle (T in the figure below, which is typically 0°) for the best scanner orientation:

T

S

Conveyor Direction

“45° Skew” Installation The DS6300 scanner is mounted on the ST-210 90° mounting bracket. By adjusting the mounting bracket guides, reach 45° for the Skew angle (S in the figure below) to avoid the direct reflection of the laser light emitted by the scanner:

45°

ATTENTION

S

If using the “45° Skew” installation, the scanner reading performance is not guaranteed to match that measured for the standard installation with Skew angle between 10° - 20° (see reading diagram section).

The ST-210 mounting bracket is an accessory of the DS6300 standard model available in the US-60 kit (order no. 890001020). NOTE

WARNING

When installing several scanners, take care to position them correctly so that no laser beam enters the reading window perpendicularly and at the same level of the output beam of the other scanners. This condition could occur more frequently for side mounted applications. If these precautions are not followed, it may occur that the laser of the blinded scanner starts blinking due to an internal circuit which temporarily turns the laser off when detecting a power anomaly. To resolve this problem, it is sufficient to slightly change the inclination and position of one of the two scanners involved.

27

Page 30

COMMON FEATURES

Focus Adjustment: The DS6300 provides a manual adjustment of the optics to optimize the reading performance by choosing the best focus between two extreme positions. The focus adjustment is continuous and not by step; thus, allowing an optimum adjustment around the selected position. The relative focus positions range from 0 to 100. The adjustment can be simply made through an external screw placed on the back of the optic HEAD and protected by a cap. The screw may be rotated either clockwise or counterclockwise in order to move the scanner internal lenses. In particular, a clockwise rotation causes a farther focus from the scanner, while a counterclockwise rotation causes a nearer focus to the scanner. An internal sensor tracks the exact laser beam focusing position allowing it to be shown on the reader display or through the Genius™ software program.

Do not stare at the laser beam output window during this operation to avoid hazardous visible laser light. WARNING

Refer to the following instructions when adjusting the focus: 1)

Remove the regulation screw protecting cap (see following Figure) positioned on the back of the optic Head;

2)

Press and hold both the ▲ (up arrow) and ▼ (down arrow) key for about 2 seconds to enter the Main menu;

3)

Use the ▲ (up arrow) or ▼ (down arrow) key to select “Test Mode” item, then press the ENT (enter) key to confirm. The reader enters Test Mode;

4)

Press the ENT (enter) key to toggle between the graphical (default) and numerical visualization of the focus position; Display Visualization The first line of the display shows the read code and Good Read percentage. Possible suspending commas at the end of the code mean that the code is too long to be displayed. The second line of the display indicates the value of the focus position according to the table below. The indications “Too Near” or “Too Far” are represented for values outside the focus range. Graphical Visualization A |---^----------| Æ where ^ indicates the focus position B N-------------| Æ where N indicates that the focus position is “Too Near” C |--------------F Æ where F indicates that the focus position is “Too Far”

Numerical Visualization Fxxx Æ where xxx ranges from 000 to 100 TooNear

Fxxx* Æ where xxx is greater than 100

CodeRead… 100% |------^------------------|

CodeRead… 094% F034

Graphical Visualization

Normal Visualization

5)

Rotate the focus adjustment screw to reach the desired focus position. The display is refreshed with the new values;1

6)

Press the ▲ (up arrow) key to exit the Test Mode;

1 To avoid breakage, do not use excessive force when rotating the focus adjustment screw.

28