Page 161

Symbol Quality

Decode Time Definition: Serial Cmd: Default: Options:

The time in milliseconds required to decode a symbol. If enabled, it is appended to the symbol data. <K710,percent cell damage,total read time,capture time,locate time,decode time,pixels per element,EC level,matrix size,quiet zone> Disabled 0 = Disabled 1 = Enabled

Pixels Per Element (Data Matrix Only) Definition: Serial Cmd: Default: Options:

The number of pixels for each element, either dark or light for both x and y directions. If enabled, it is appended to the symbol data. <K710,percent cell damage,total read time,capture time,locate time,decode time,pixels per element,EC level,matrix size,quiet zone> Disabled 0 = Disabled 1 = Enabled

Error Correction Level (Data Matrix Only) Definition: Serial Cmd: Default: Options:

Outputs the ECC type symbology. If enabled, it is appended to the symbol data. <K710,percent cell damage,total read time,capture time,locate time,decode time,pixels per element,EC level,matrix size,quiet zone> Disabled 0 = Disabled 1 = Enabled

Matrix Size (Data Matrix Only) Definition: Serial Cmd: Default: Options:

Defines the symbol matrix size, in number of pixels in both the x and y axis. If enabled, it is appended to the symbol data. <K710,percent cell damage,total read time,capture time,locate time,decode time,pixels per element,EC level,matrix size,quiet zone> Disabled 0 = Disabled 1 = Enabled

FIS-6300 ESD Imager User’s Manual

8-11

Page 162

wenglor Symbol Quality Output

Quiet Zone (Data Matrix Only) Definition: Serial Cmd: Default: Options:

8-12

If enabled, the size of the quiet zone is evaluated and a PASS or FAIL message is appended to the symbol data. <K710,percent cell damage,total read time,capture time,locate time,decode time,pixels per element,EC level,matrix size,quiet zone> Disabled 0 = Disabled 1 = Enabled

FIS-6300 ESD Imager User’s Manual

Page 163

Symbol Quality wenglor Symbol Quality Output by ESP ESP’s Symbol Quality interface allows you to evaluate Data Matrix symbols for compliance with a rigorous set of standards, such as Locate Time, Capture Time, and Decode Time. First, determine which parameters you want to evaluate, using the Symbol Quality tree control.

Once you’ve set these evaluation parameters, go to the Preferences tab and set your Data Matrix Grading Report output preferences.

After your output preferences are set, click the Data Matrix Grading button.

There will be a wait of a few seconds, and then the evaluation results will appear at the lower right of the Symbol Quality view, in a format similar to the one shown below.

If you want to view the results in a report format, click the Generate Report button. The report will be output in the format you chose in the Preferences dialog (Default Report Type).

FIS-6300 ESD Imager User’s Manual

8-13

Page 164

8-14

FIS-6300 ESD Imager User’s Manual

Page 165

7 I/O Parameters Contents I/O Parameters by ESP ................................................................................................................ 7-2 I/O Parameters Serial Commands ............................................................................................... 7-3 Symbol Data Output ..................................................................................................................... 7-4 Noread Message .......................................................................................................................... 7-6 Bad Symbol Message .................................................................................................................. 7-8 No Symbol Message .................................................................................................................... 7-8 1D / Stacked Symbology Qualification ......................................................................................... 7-9 2D Symbology Qualification ....................................................................................................... 7-10 Read Duration Output ................................................................................................................ 7-16 Output Indicators ........................................................................................................................ 7-17 Beeper........................................................................................................................................ 7-21 LED Configuration ...................................................................................................................... 7-22 Serial Verification ....................................................................................................................... 7-23 EZ Button ................................................................................................................................... 7-25 EZ Button Modes........................................................................................................................ 7-27 Configurable Output 1 ................................................................................................................ 7-29 Trend Analysis Output 1............................................................................................................. 7-32 ISO/IEC 16022 Symbol Quality Output 1 ................................................................................... 7-35 Diagnostic Output 1.................................................................................................................... 7-38 Configurable Output 2 ................................................................................................................ 7-39 Trend Analysis Output 2............................................................................................................. 7-39 ISO/IEC 16022 Symbol Quality Output 2 ................................................................................... 7-39 Diagnostic Output 2.................................................................................................................... 7-39 Configurable Output 3 ................................................................................................................ 7-40 Trend Analysis Output 3............................................................................................................. 7-40 ISO/IEC 16022 Symbol Quality Output 3 ................................................................................... 7-40 Diagnostic Output 3.................................................................................................................... 7-40 Power-On/Reset Counts ............................................................................................................ 7-41 Time Since Reset ....................................................................................................................... 7-42 Service Message........................................................................................................................ 7-43 Frame Information ...................................................................................................................... 7-44 Image Output.............................................................................................................................. 7-45 Database Identifier Output ......................................................................................................... 7-48

This section includes instructions on setting up conditions for changing input/output electrical transitions for control of the Imager’s internal and external devices. A discrete I/O (in/out) signal is an electrical transition from one voltage level to another so that digital switching can occur.

FIS-6300 ESD Imager User’s Manual

7-1

Page 166

I/O Parameters by ESP

I/O Parameters by ESP

Click this button to bring up the I/O menu.

To change a setting, double-click the setting and use your cursor to scroll through the options.

To open nested options, single-click the +.

7-2

FIS-6300 ESD Imager User’s Manual

Page 167

I/O Parameters

I/O Parameters Serial Commands Power On/Reset Counts

<K406,power-on,resets,power-on saves,power-on flash saves>

Time Since Reset

<K407,hours,minutes>

Service Message

<K409,status,service message,threshold,resolution>

Serial Verification

<K701,serial command echo status,serial command beep status,control/hex output>

Beeper

<K702,status>

Quality Output

<K704,quality output separator,decodes/trigger status>

Symbol Data Output

<K705,symbol data output status,when to output>

Read Duration Output

<K706,status,separator>

Noread Message

<K714,status,message>

Bad Symbol Message

<K715,[unused],message>

No Symbol Message

<K716,[unused],message>

1D/Stacked Symbology Qualification

<K717,minimum number of bars,minimum number of qualified scan lines,start/stop status>

2D Symbology Qualification

<K718,finder pattern status,symbol size status,symbol size 1,symbol size 2,symbol size tolerance,dimension status,dimension 1,dimension 2,dimension tolerance,orientation status,orientation value>

Frame Information

<K734,output frame number,output coordinates>

LED Configuration

<K737,LED mode,ISO/IEC 16022 grade>

Image Output

<K739,image output mode,communication port,file format,jpeg quality>

Target/Flood LEDs

<K750,green flash LED status,target pattern LED status,green flash duration>

Database Identifier Output

<K759,status,separator>

EZ Button

<K770,global status,default on power-on,load configuration database,save for power-on>

EZ Button Modes

<K771,single beep,two beeps,three beeps,four beeps> <K780,trend analysis mode,trigger evaluation period,number to output on>

Trend Analysis Output 1 Trend Analysis Output 2

<K781,trend analysis mode,trigger evaluation period,number to output on>

Trend Analysis Output 3

<K782,trend analysis mode,trigger evaluation period,number to output on>

Diagnostic Output 1

<K790,unused 1,service unit>

Diagnostic Output 2

<K791,unused 1,service unit>

Diagnostic Output 3

<K792,unused 1,service unit>

ISO/IEC 16022 Symbol Quality Output 1

<K800,output on symbol contrast,symbol contrast threshold,output on print growth,print growth threshold,output on axial non-uniformity,axial non-uniformity threshold,output on UEC,UEC threshold>

ISO/IEC 16022 Symbol Quality Output 2

<K801,output on symbol contrast,symbol contrast threshold,output on print growth,print growth threshold,output on axial non-uniformity,axial non-uniformity threshold,output on UEC,UEC threshold>

ISO/IEC 16022 Symbol Quality Output 3

<K802,output on symbol contrast,symbol contrast threshold,output on print growth,print growth threshold,output on axial non-uniformity,axial non-uniformity threshold,output on UEC,UEC threshold>

Configurable Output 1

<K810,output on,output state,pulse width,output mode>

Configurable Output 2

<K811,output on,output state,pulse width,output mode>

Configurable Output 3

<K812,output on,output state,pulse width,output mode>

FIS-6300 ESD Imager User’s Manual

7-3

Page 168

Symbol Data Output

Symbol Data Output Symbol Data Output Status Note: Symbol Data Output relates to data and should not be confused with Outputs 1, 2, and 3 listed in the Output Parameters which describe output states and functions. Usage: Definition: Serial Cmd: Default: Options:

Useful when the host needs symbol data only under certain conditions. Defines the conditions under which decoded symbol data is transmitted to the host. <K705,symbol data output status,when to output> Good Read 0 = Disabled 1 = Match 2 = Mismatch 3 = Good Read

Note: Symbol Data Output Status, if set to Match or Mismatch, will not take effect unless Matchcode Type is enabled and a master symbol is loaded into memory.

Disabled Usage:

Definition:

It is useful when an application only needs to use the discrete outputs and can allow the Imager to do the decision-making. When Disabled, the host does not need the symbol data and the communication lines are used only for setup and status checks. When set to Disabled, the Imager will not transmit any data that is generated during a read cycle (symbols, noreads, etc.)

Match Usage: Definition:

Match is used in an application that requires specific symbol information and needs to sort, route, or verify based on matching the specific symbol data. When set to Match, the Imager transmits symbol data whenever a symbol matches a master symbol. However, if Matchcode Type is Disabled, it transmits on any good read.

Note: A noread can still be transmitted if Enabled.

Mismatch Usage: Definition:

Mismatch is typically used as a flag within the host system to prevent an item from being routed in the wrong container. With Mismatch enabled, the Imager transmits symbol data whenever the symbol data information does NOT match the master symbol.

Note: A noread can still be transmitted if Enabled. 7-4

FIS-6300 ESD Imager User’s Manual

Page 169

I/O Parameters

Good Read Usage: Definition:

Good Read is used when an application requires all symbol data to be transmitted. It’s typically used in tracking applications in which each object is uniquely identified. With Good Read enabled, the Imager transmits symbol data on any good read regardless of Matchcode Type setting.

Note: A noread can still be transmitted if Enabled.

When to Output Symbol Data Definition: Serial Cmd: Default: Options:

This command allows the user to choose when symbol data can be sent to the host. <K705,symbol data output status,when to output> As Soon As Possible 0 = ASAP 1 = End of Read Cycle

As Soon As Possible Usage: Definition:

As Soon As Possible is useful in applications in which symbol data needs to be moved quickly to the host, typically when the host is making decisions based on symbol data. Enabling As Soon As Possible causes symbol data to be sent to the host immediately after a symbol has been successfully decoded.

Note: More than one decode might in fact be required to qualify as a good read, depending on how Decodes Before Output is set.

End of Read Cycle Usage: Definition:

End of Read Cycle is useful in timing-based systems in which the host is not ready to accept data at the time that it is decoded. Enabling End of Read Cycle means that symbol data does not get sent to the host until the read cycle ends with a timeout or new trigger.

End of read cycle

Start of read cycle Read cycle duration Host activates trigger

FIS-6300 ESD Imager User’s Manual

This is when host expects output

7-5

Page 170

Noread Message

Noread Message Usage: Definition:

Used in applications where the host needs serial verification that a symbol has not been read and especially useful in new print verification. When enabled, and if no symbol has been decoded before timeout or the end of the read cycle, the noread message will be transmitted to the host.

Noread Message Mode Serial Cmd: Default: Options:

<K714,noread message status,noread message> Enabled 0 = Disabled 1 = Enabled 2 = Detailed Noread Message

Disabled No message is output when a noread condition occurs.

Enabled The user defined noread message is output whenever a noread condition occurs.

Detailed Noread Message Important: In the case where the conditions set in Bad Symbol/No Symbol Qualification are met, then the Bad Symbol or No Symbol message output will override the Noread Message or the Detailed Noread Message, if either is enabled. Usage: Definition:

Use this option to cover more detailed reasons for decode failures. Outputs a Noread message followed by an additional message (an Image Processing Results Code) indicating the reason a symbol failed to decode.

For example, in the following output: “NOREAD d_1, E_2”, d and E are symbology i.d.s for Data Matrix and UPC respectively. The message means that Data Matrix and UPC are enabled (and other symbologies are not), and that both failed to decode. The reason for the failing Data Matrix is 1 and the reason for the failing UPC is 2. The first part of these numbers (preceding the dot) is the relevant number and is interpreted here:

2D Symbols: 1. Failed to locate symbol (no finder pattern). 2. Failed to locate four corners or failed to qualify user defined dimensions (in pixels). 3. Failed to locate clock patterns or failed to qualify user defined orientation (0-359 degree). 4. Failed to validate clock patterns or failed to qualify user defined symbol size (clock element count). 5. Failed to decode symbol. 7-6

FIS-6300 ESD Imager User’s Manual