Page 41
1.
UEC (Unused Error Correction): This is the percentage of error correction capability that is available for further incorrect modules. The assignment of grade is according to the following table: UEC % > 62 > 50 (but less than 62) > 37 (but less than 50) > 25 (but less than 37) < 25
Grade A B C D F
2. SC (Symbol Contrast): This is the difference in reflectivity between the brightest module and the darkest module. The assignment of grade is according to the following table: SC % > 70 > 55 > 40 > 20 < 20
(but less than 70) (but less than 55) (but less than 40)
Grade A B C D F
3. MOD and RM (Modulation): This is a grade based on the amount of variability in reflectivity of the modules. A multi-step process is used to get the modulation grade. First the reflectivity of each module is compared to the global threshold and the overall symbol contrast according to the following formula: MOD = 2 * (abs(R - GT)) / SC The Global Threshold GT is the midpoint between the reflectance of the brightest module and the reflectance of the darkest module. Next, the grade level for each module is determined from the MOD value according to the following table: MOD % > 50 > 40 (but less than 50) > 30 (but less than 40) > 20 (but less than 30) < 20
Grade A B C D F
Finally, the value of the grade for the MOD parameter will be the highest modulation level for which the modules meeting that level will result in an Unused Error Correction grade of that level or higher. The module with the lowest MOD is reported as Contrast Uniformity (CU) in General Characteristics to facilitate conformance testing to the requirements of ISO/IEC 15426-2. 4. ANU (Axial Non-uniformity): This is the amount of “out of square” a symbol is, or in other words, a measure of the overall aspect ratio of the symbol.
Page 42
ANU % ≤6 <8 (but more than 6) < 10 (but less than 8) < 12 (but less than 10) > 12
Grade A B C D F
5. GNU (Grid Non-uniformity): This is the worst-case distance between the calculated center of a module and the ideal location for the center of the module based on perfectly evenly spaced modules. The calculated center of the module is determined using clock tracks. The value is reported as a percentage of a module size. GNU % ≤ 38 < 50 (but less than 38) < 63 (but less than 50) < 75 (but less than 63) > 75
Grade A B C D F
6. FPD (Fixed Pattern Damage): This is the overall grade for all the fixed pattern components. This grade is equal to the lowest grade of all the components listed below. The following is a list of components of the finder pattern. 7. LLS (Left ‘L’ Side): This is a grade based on imperfections in the left ‘L’ side of the finder pattern. There are two checks required to pass. The first requires gaps to be three modules or less and that gaps are separated by stretches of at least four correct modules. The second assigns a grade based on the overall percentage of correct modules according to the following table: % of incorrect modules 0 <9 (but more than 0) < 13 (but more than 9) < 17 (but more than 13) > 17
Grade A B C D F
The grade is the highest modulation level in which the first (gap test) passes and the correct module percentage results in a grade of that level or higher. 8. BLS (Bottom ‘L’ Side): This is a grade based on imperfections in the bottom ‘L’ side of the finder pattern (see Left ‘L’ Side). 9. LQZ (Left Quiet Zone): This is a grade based on imperfections in the quiet zone, which is a one-module area to the left of the left ‘L’ side. The grade is based on the percentage of modules, which are correct using the same grading table as for the ‘L’ sides. 10. BQZ (Bottom Quiet Zone): This is a grade based on imperfections in the quiet zone, which is a one-module area below the bottom ‘L’ side.
Page 43
11. TQZ (Top Quiet Zone): This is a grade based on imperfections in the quiet zone, which is a one-module area above the top clock track.
a. ULQZ (Upper Left Quiet Zone): This is the top quiet zone above the upper left quadrant (Used only for 2 and 4 quadrant symbols, this is the grade based on the segment of the quiet zone above the top clock track of the left quadrant).
b. URQZ (Upper Right Quiet Zone): This is the top quiet zone above the upper right quadrant (Used only for 2 and 4 quadrant symbols, this is the grade based on the segment of the quiet zone above the top clock track of the right quadrant). 12. RQZ (Right Quiet Zone): This is a grade based on imperfections in the quiet zone, which is a one-module area to the right of the Right Clock Track.
a.
RUQZ (Right Quiet Zone to the right of the upper right quadrant): Only for 2 and 4 quadrant symbols, this is the grade based on the segment of the quiet zone to the right of the upper right quadrant.
b. RLQZ (Right Quiet Zone to the right of the lower right quadrant): Only for 4 quadrant symbols, this is the grade based on the segment of the quiet zone to the right of the lower left quadrant. 13. TTR (Top Transition Ratio): This is a grade based on imperfections in the top clock track, with relation to its adjoining quiet zone. The ratio is the number of transitions from light to dark or dark to light in the quiet zone divided by the number of transitions in the clock track. Since the number of transitions in the quiet zone should be zero, the ideal value for this parameter is zero. However, a small number of transitions can be tolerated as long as the ratio remains relatively low. As the number of teeth in the clock track increases (larger symbols) more transitions in the quiet zone can be tolerated. Also, more transitions in the clock track (which are really imperfections) will tend to improve this measurement. The grading scheme for this transition ratio is: Transition Ratio % <6 <8 (but more than 6) < 10 (but more than 8) < 12 (but more than 10) > 12
Grade A B C D F
The value of the grade will be the highest modulation level for which the ratio gives a grade from the above table of that modulation level or higher. a. ULQTTR (Transition ratio for Upper Left Quadrant Top Clock Track): Only for 4 quadrant symbols, this is the grade based on the clock track segment at the top of the upper left quadrant. For a 2 quadrant symbol, this will be labeled LQTTR. b. URQTTR (Transition ratio for Upper Right Quadrant Top Clock Track): Only for 4 quadrant symbols, this is the grade based on the clock track segment at the top of the upper right quadrant. For a 2 quadrant symbol this will be labeled RQTTR. c. LLQTTR (Transition ratio for Lower Left Quadrant Top Clock Track): Only for 4 quadrant symbols, this is the grade based on the clock track segment at the top of the lower left quadrant.
Page 44
d. LRQTTR (Transition ratio for Lower Right Quadrant Top Clock Track): Only for 2 and 4 quadrant symbols, this is the grade based on the clock track segment at the top of the lower right quadrant. 14. RTR (Right Transition Ratio): Transition ratio (see Top Transition Ratio) for the right clock track in relation to the right quiet zone.
a. ULQRTR (Transition ratio for Upper Left Quadrant Right Clock Track): Only for 2 and 4 quadrant symbols, this is the grade based on the clock track segment to the right of the upper left quadrant. For a 2 quadrant symbol, this will be labeled LQRTR.
b. URQRTR (Transition ratio for Upper Right Quadrant Right Clock Track): Only for 2 and 4 quadrant symbols, this is the grade based on the clock track segment to the right of the upper right quadrant. For 2 quadrant symbols, this will be labeled RQRTR.
c. LLQRTR (Transition ratio for Lower Left Quadrant Right Clock Track): Only for 4 quadrant symbols, this is the grade based on the clock track segment to the right of the lower left quadrant.
d. LRQRTR (Transition ratio for Lower Right Quadrant Right Clock Track): Only for 4 quadrant symbols, this is the grade based on the clock track segment to the right of the lower right quadrant. 15. TCT (Top Clock Track): This is a grade based on imperfections in the top clock track. Some imperfections in the clock track can be tolerated. However, the rule that must be maintained for a passing grade is that three out of every five modules (on a consecutively rolling window of five modules) must be correct. The value of the grade will be the highest modulation level for which this test passes.
a. ULQTCT (Top Clock Track for Upper Left Quadrant): Only for 2 and 4 quadrant symbols, this is the grade based on the clock track segment at the top of the upper left quadrant. For 2 quadrant symbols this will be labeled LQTCT.
b. URQTCT (Top Clock Track for Upper Right Quadrant): Only for 2 and 4 quadrant symbols, this is the grade based on the clock track segment at the top of the upper right quadrant. For 2 quadrant symbols this will be labeled RQTCT.
c. LLQTCT (Top Clock Track for Lower Left Quadrant): Only for 4 quadrant symbols, this is the grade based on the clock track segment at the top of the lower left quadrant.
d. LRQTCT (Top Clock Track for Lower Right Quadrant): Only for 4 quadrant symbols, this is the grade based on the clock track segment at the top of the lower right quadrant. 16. RCT (Right Clock Track): This is a grade based on imperfection in the right clock track (see Top Clock Track).
a. ULQRCT (Right Clock Track for Upper Left Quadrant): Only for 2 and 4 quadrant symbols, this is the grade based on the clock track segment to the right of the upper left quadrant. For 2 quadrant symbols, this will be labeled LQRCT.
b. URQRCT (Right Clock Track for Upper Right Quadrant): Only for 2 and 4 quadrant symbols, this is the grade based on the clock track segment to the right of the upper right quadrant. For 2 quadrant symbols, this will be labeled RQRCT.
Page 45
c. LLQRCT (Right Clock Track for Lower Left Quadrant): Only for 4 quadrant symbols, this is the grade based on the clock track segment to the right of the lower left quadrant.
d. LRQRCT (Right Clock Track for Lower Right Quadrant): Only for 4 quadrant symbols, this is the grade based on the clock track segment to the right of the lower right quadrant. 17. AG (Average Grade of Damage across many parts of the Finder Pattern): This is a grade that considers the accumulated effect of damage to several parts of the finder pattern. Five values are averaged together. One of these is the lowest of all the grades associated with all the clock track segments, namely TCT, TTR, TQZ and RCT, RTR, RQZ. The other four are LLS, BLS, LQZ, and BQZ. The average must fall in the range of 0.0 through 4.0 and is given a grade according to the following: Ave Grade
Grade
Equals 4.0 3.5 (but less than 4.0) 3.0 (but less than 3.5) 2.5 (but less than 3.0) < 2.5
A B C D F
NOTE: The effect of the AG parameter is to lower the overall grade of symbols, which have several individual parameters at or near the same level. For instance, with enough B grades in individual parameters, the overall grade may come out as a C grade rather than a B. 18. Contrast Uniformity (CU): This is the value of MOD (modulation) for the worst case module selected from a 2D Matrix symbol. This is useful for process control, as way of measuring the drift in reflectivity consistency, and more importantly for testing conformance to ISO/IEC 15426-2 which requires the modulation of a specific module within a conformance test symbol to be reported. This parameter is not graded, nor listed in the Quality Detail tab. Rather, it is reported in General Characteristics.
The AIM-DPM method of grading data matrix symbols modifies the process of ISO-15415 and is more appropriate for direct part marking applications. This standard was developed to be more representative of the scanning performance of modern readers that in some cases are specifically designed for these demanding applications. In this method, the image brightness is adjusted to produce an image of the symbol that fills most or all of the dynamic range of the imager, resulting in an image that is easier to see. Additionally, the threshold between dark and light is calculated from the statistics of the image brightness histogram. Thus, the measurements calculated by AIM-DPM differ from those of ISO 15415 significantly. Some of the parameters reported in ISO 15415 are changed so drastically that in order to remove the possibility of confusion between these two methods, the parameters have been renamed. These parameters are:
Page 46
AIM-DPM Parameter Name
ISO 15415 Parameter Name
Summary of Change(s)
CC (Cell Contrast)
SC (Symbol Contrast)
Made relative to light background
CM (Cell Modulation)
MOD (Modulation)
DD (Distributed Damage)
AG (Average Grade)
MR (Minimum Reflectance)
Not necessary since SC is An absolute limit on SC of 5% added to measured on an absolute temper the relative nature of CC. scale
Threshold calculated from statistics rather than the maximum and minimum reflectance. Grading scale range set to Mean of distribution, rather than maximum and minimum reflectance Modulation overlay uses only A, B, and F levels instead of A, B, C, D, and F.
All of the Fixed Pattern Damage grading (other than AG shown above) are not renamed, but are functionally different since the global threshold and modulation grading scale are different. In general, symbols will obtain a significantly higher grade according to AIM-DPM than ISO 15415. Therefore, grading according to AIMDPM is appropriate only when called for in an application specification. Another significant difference is the allowance for a variety of illumination options. These include the foursided 45˚ light that is the default for ISO 15415. Additionally allowed are: 30˚ lighting from four sides, 30˚ from two sides (which can be either North/South or East/West), and 90˚diffuse on-axis lighting. The light source that is used is reported using a notation that includes the angle, and a letter (Q for 4, T for two). AIM-DPM also varies the size of the aperture until the symbol is decoded, and then the grading is repeated with two different aperture sizes (50% and 80%) and the better of the two grades is reported as the final grade. The parameters, which are new or significantly modified for AIM-DPM, are explained below: 1.
CC (Cell Contrast): This is the relative contrast value between bars and spaces, taken from the means of the light and dark element CC = (Lmean - Dmean) / Lmean CC % ≥ 30% ≥ 25% ≥ 20% ≥ 15% < 15%
Grade 4 3 2 1 0
2. CMOD (Cell Modulation): Similar to MOD in ISO 15415, this parameter measures the deviation in the reflectivity of dark and light elements. A range for each group (light and dark) is created from the
Page 47
global threshold to the mean reflectance of the elements. Each module is graded along this range, then error correction capability is considered to “discount” the effect of one or a few elements with low values, and a final grade for this parameter is computed. 3. DDG (Distributed Damage Grade): Similar to AG in ISO 15415, this parameter takes into account the effect of multiple segments of the fixed pattern having imperfections. Where multiple segments have a low grade, the effect of this “distributed damage” is reflected in a lower grade for DDG than the lowest of the individual segments. 4. MR (Minimum Reflectance): This is a requirement for at least 5% reflectance difference between light and dark elements, as a restraint on the purely relative CC parameter. 5. Decode: Decode grade A or F depending upon whether the reference decode algorithm succeeds in decoding the symbol with the required final aperture size.
1.
Distortion Angle (DA): The angle between the left side and the bottom side of the ‘L’ finder pattern should be 90 degrees. The distortion angle is the deviation from 90 degrees exhibited by the ‘L’ in this symbol. A distortion angle of 7 degrees or less will Pass, while a larger distortion angle will Fail.
2. Cell Fill (Size): The size of the cells must be printed at least 60% of the cell spacing, and no more than 105% of the cell spacing. 3. Ovality: For dot peen symbols (where the cells are allowed to be round) the cells should be circular. As a measure of ovality, the ratio of the widest diameter to the smallest diameter is checked and must be less than 20% different in order to Pass. 4. Center Offset: The center of each cell must not be offset from its correct grid location by more than 20% of the cell spacing in order to Pass. 5. Symbol Contrast: Symbol contrast should be at least 20% to Pass.
ISO 16022 Parameters: Decode (DCD), Symbol Contrast (SC), Bar Width Growth (BWG), Axial Non-Uniformity (ANU), Unused Error Correction (UEC)
Page 48
DCD: Indicates whether the symbol was decoded by the reference decode algorithm. Symbol Contrast: Represents the difference between reflectance of light and dark elements. BWG: Represents the bar width growth as defined in ISO 16022. ANU: Represents Axial Non-Uniformity, which represents non-square aspect ratio of the modules in the symbol. UEC: Represents Unused Error Correction, which indicates the amount of error correction used out of the budget for error correction available in the symbol.
PCS, MRD, SC, MinEC and MOD – Ways of Quantifying Contrast: PCS, Print Contrast Signal, is an older and largely no longer used measure of contrast. Contrast is intended to quantify the difference between the bars and the spaces in reflectance. PCS is defined mathematically as: PCS = (Rmax-Rmin) / Rmax In other words, the percentage of the light background accounted for by the difference between the bars and spaces. This measure was defined as a measure of perceived contrast by the human eye, long before and outside the context of measuring barcode contrast. Notice that the measurement is made relative to the brightness of the background. The fact that it is relative to the background means that the darker (worse) the background color is, the higher (supposedly better) the value of PCS. This may correspond with the way people view things, but not really, how scanners work. Rather, scanners are sensitive to the absolute difference between the reflectance of bars and spaces. Scanners are especially sensitive to variations in contrast within the same scan. Another measurement, called MRD (minimum reflectance difference) quantifies the minimum difference anywhere across the barcode. MRD considers the brightest (worst) bar and the darkest (worst) space anywhere in the barcode. These worst case bar and space need not be adjacent to one another. NOTE: ANSI grading system defines yet a different way of measuring contrast: SC (symbol contrast) and MinEC (Minimum Edge Contrast.) SC is the difference between the brightest (best) space and darkest (best) bar. Minimum Edge Contrast is the smallest difference occurring between adjacent bars and spaces (across a single edge). MOD (Modulation) is simply the ratio between MinEC and SC, in other words, the percentage of available contrast that is manifest in the worst-case edge. Bar Width Growth: Bar Width Growth is a measure of ink spread. The measurement of bar width growth is different on each symbology, and requires at least one successful decode of the symbol for measurement. Generally, bar width growth is taken on groups of bars and spaces, sometimes large groups. This results in stable results when bar width growth is uniform, but does not catch isolated cases of a single bar widening or shrinking. In fact, it is possible to have a very low (good) reading of bar width growth when in fact a large number of even all the bars and spaces are incorrectly grown or reduced, as these may tend to cancel each other out. Such bar width inaccuracies will normally be detected by low grades of Decodability.
Page 49
• The TruCheck™ Omni can be used in two modes: Single-Mode and Multi-Mode. • Single-Mode will decode a single symbol at a time, and is useful for quick, simple verification of isolated symbols.
• When the application is started for the first time, it will be in “Single-Mode”.
• To switch to “Multi-Mode”, select “File->Switch to Multi-Mode”. This will restart the software in “MultiMode”. • In “Multi-Mode”, any number of symbols may be verified per verification cycle. • Before any verification has taken place, the software will show an image that was taken during start up, and the camera will be in an inactive state. To proceed to verify symbols, press the large button near the top center that reads “Go Live”. Presenting the Sample to be verified • In “Live Mode”, the camera will capture images and update the display several times every second. At this point, the symbols to be verified should be placed on the TruCheck™ Omni unit, face-down on the glass viewing window. • Images of the sample to be verified will then be visible on the screen. Adjust the sample on the window so that all symbols to be verified are within the field of view, and that the sample sheet is flat and flush against the window. Failure to ensure that the symbols are flush against the window can result in poor verification results.
Page 50
• Select the regions to be verified. To do this, click and drag starting beyond the upper left of a symbol to be verified, and releasing beyond the lower right of the symbol. Repeat this for each symbol to be verified. • If the region selected is approximately square, the selected region will be given the default settings for a 2D symbol region. If the region is more rectangular, the region will be given the default settings of a 1D region. • Selected regions can be deleted by right clicking on a region, and selecting “Delete”. To delete all regions, right click in an area outside of any selected region, and select “Delete All Selections”. • The layout and group affiliation of regions on the screen can be saved and recalled at a later time. This is useful if similar samples are frequently verified. To do this, layout the regions and assign groups as desired, and then select “File->Save Setup”. This will save a setup file that can be restored at a later time. • To restore a saved setup file, select “File->Open Setup”.
• With the regions selected, decode settings must be selected to match the intended application. Settings can be adjusted by right clicking inside a selected region, and selecting “Settings”. • Settings are common among all regions that are part of the same group. By default, all regions selected will belong to the same group, so if a large number of symbols are to be verified using the same settings, it is only necessary to change the settings on one of the regions. • Regions are always set as either “1D” or “2D”. This determines if the region will be searched for 1D or 2D symbols upon verification. This is a setting like any other, so changing this setting for one region will apply the change to all regions in that region’s group. • If two regions that are in the same group need to be verified with different settings, then it is necessary to move one of them to a different group. To do this, right click on the region and select “Switch to Group>New Group”. • This will create a new group with this region as its only member. More regions can be added to this group by instead selecting “Switch to Group->” and selecting the appropriate group. • Groups are color-coded; regions that are of the same color are members of the same group. The settings for each group apply only to the regions that are members of that group.