Page 80
LS 3070 Product Reference Guide: Programming
OCIA Transmit Timeout When connected to an OCIA terminal, the RL 470 base/interface will attempt to transmit the data to the host. If the host is not prepared to accept data at that time (i.e., the host data buffer may be full), the RL 470 will try to retransmit the data for up to 3 seconds. If transmission is not successful, the data will be discarded, and the RL 470 will issue a transmission error. This parameter allows you to increase the time the RL 470 will attempt transmission before discarding the data. You can select up to a 30 second delay in one second increments. During the time the RL 470 is attempting to transmit data, no new data can be scanned.
NCR 2152 Fast Transmit Selects the data transmission speed to the NCR 2152 POS terminal. Depending on the version of NCR 2152, selecting this option may increase the possibility of lost or mis-transmitted data.
IBM 4683/93 Magstripe Communications The RL 470 base/interface communicates with the IBM 4683/84/4693/94 as a scanner device, and also as a magstripe reader device. The RL 470/IBM 4683/ 93 magstripe communications can be enabled and disabled via this parameter. If the RL 470 magstripe capability is needed, this parameter should be enabled. If not, it should be disabled to avoid contention with any other magstripe reader which may be connected to the host.
International Keypad Emulation This parameter is supported by IBM AT/XT/PS2 terminals only. The layout of some international keyboards differ from that of American keyboards. To send a given character, regardless of its physical location on the keyboard, enable this parameter.
International Keypad Emulation Fast Transmit This parameter is used when International Keypad Emulation is enabled. Some IBM AT/XT/PS2s allow for faster transmission. With this parameter enabled, the RL 470 transmits at a faster rate. Not all IBM AT/XT/PS2s support a faster transmission rate. 6-17
Page 81
LS 3070 Product Reference Guide: Programming
National Keyboard Types Use this parameter to set the national character type for keyboard characters. Selections include U.S. English, French, German, French International, Spanish, Italian, Swedish, and U.K. English. The following terminals do not support Italian or French International but do support the other six options: IBM 3680, 316X, 319X; HP 700/92, 2392; Telex 122; Wyse 50, 60, 160. These terminals support all options but French International: VT 220, 320, 420. These terminals support all eight options: IBM PC XT, PC AT, PS/2. The table below indicates hosts supported by industrial versions of the scanner. Table 6-1. Hosts Supported by Industrial Scanners Host Interface
U.S.
U.K.
Fr.
Ger.
Spn.
Swe.
Ital.
Fr. In.
PC AT
Y
Y
Y
Y
Y
Y
Y
Y
PC XT
Y
Y
Y
Y
Y
Y
Y
Y
PS/2
Y
Y
Y
Y
Y
Y
Y
Y
Single RS-232
Y
N
N
N
N
N
N
N
Dual RS-232
Y
N
N
N
N
N
N
N
Dec VT 220
Y
Y
Y
Y
Y
Y
Y
N
Dec VT 320
Y
Y
Y
Y
Y
Y
Y
N
Dec VT 420
Y
Y
Y
Y
Y
Y
Y
N
HP 700/9XX
Y
Y
Y
Y
Y
Y
Y
N
HP 2392N
Y
Y
Y
Y
Y
Y
Y
N
IBM 3151
Y
N
N
N
N
N
N
N
IBM 347X
Y
N
N
N
N
N
N
N
IBM 3179
Y
N
N
N
N
N
N
N
IBM 3180
Y
N
N
N
N
N
N
N
IBM 3179
Y
N
N
N
N
N
N
N
IBM 3180
Y
N
N
N
N
N
N
N
6-18
Page 82
LS 3070 Product Reference Guide: Programming
Table 6-1. Hosts Supported by Industrial Scanners Host Interface
U.S.
U.K.
Fr.
Ger.
Spn.
Swe.
Ital.
Fr. In.
PC AT
Y
Y
Y
Y
Y
Y
Y
Y
PC XT
Y
Y
Y
Y
Y
Y
Y
Y
PS/2
Y
Y
Y
Y
Y
Y
Y
Y
Single RS-232
Y
N
N
N
N
N
N
N
Dual RS-232
Y
N
N
N
N
N
N
N
Dec VT 220
Y
Y
Y
Y
Y
Y
Y
N
Dec VT 320
Y
Y
Y
Y
Y
Y
Y
N
Dec VT 420
Y
Y
Y
Y
Y
Y
Y
N
HP 700/9XX
Y
Y
Y
Y
Y
Y
Y
N
IBM 316X
Y
N
N
N
N
N
N
N
IBM 319X
Y
N
N
N
N
N
N
N
IBM 3178
Y
N
N
N
N
N
N
N
IBM 3278
Y
N
N
N
N
N
N
N
Telex - 88
Y
N
N
N
N
N
N
N
Telex - 122
Y
N
N
N
N
N
N
N
Wyse 60/PC
Y
Y
Y
Y
Y
Y
N
N
Wyse 60/ASCII
Y
Y
Y
Y
Y
Y
N
N
Wyse 60/ANSI
Y
Y
Y
Y
Y
Y
N
N
Wyse 85/ANSI
Y
Y
Y
Y
Y
N
Y
N
Wyse 150/ANSI
Y
N
N
N
N
N
N
N
Wyse 150+/PC
Y
N
N
N
N
N
N
N
Wyse 150+/ASCII
Y
N
N
Y
Y
N
N
N
Wyse 150+/ANSI
Y
N
N
Y
Y
N
N
N
Wyse 185/ANSI
Y
N
N
N
N
N
N
N
6-19
Page 83
LS 3070 Product Reference Guide: Programming
Set Transmission Frequency Use this parameter to set an initial transmission frequency to avoid interference on the default channel (channel 50). During operation, the transmission channel changes automatically whenever interference is encountered 80% of the time or more over a continuous five-minute period. The selected frequency channel must be between 01 and 82.
Wait for Host Interface Response Time The wait for host interface response timeout is the amount of time it takes for the base to communicate with the host interface and the base to send an acknowledgment back to the scanner. The amount of time required varies with the host device. Based on the installed interface type, the scanning system automatically calculates a wait or timeout duration for host response. Under normal operating conditions, there should be no need to program a wait for host response timeout value. However, use this parameter when the interface’s data stream carries an additional ADF pause duration. The more pause duration added to the data stream, the longer the wait for host response timeout needs to be. Typically, most interfaces timeout in an average of three seconds. If a pause duration is added, this may cause RF communications errors (6 beeps); increase the host wait response timeout to eliminate RF communication errors. Just start by programming at least a 5-second timeout. If errors continue, increment the wait for host response timeout until the RF communications errors stop occurring. Wait for host response timeout values vary from 1 second to 99 seconds. After scanning the Wait for Host Interface Response Time bar code, scan two number keypad bar codes for the two-digit response time (in seconds). If you program a value of zero, there will be an automatic calculation of the host wait response timeout value, based on the selected host interface type.
6-20
Page 84
LS 3070 Product Reference Guide: Programming
Parameter Selections Supported features for each host type. Table 6-2. Host Supported Parameters Decode Parameters
RS-232C
IBM 4683*
OCIA
OCR
Keyboard Wedge
Add Codes to be Decoded
X
X
X
X
X
ASCII Format
X
Baud Rate
X
Beep After Good Decode
X
X
X
X
X
Beeper Volume
X
X
X
X
X
Check Parity
X
Codabar Any Length
X
X
X
X
Codabar Length W/I Range
X
X
X
X
Codabar One Length
X
X
X
X
Codabar Two Lengths
X
X
X
X
Code 39 Buffering
X
Code 39 Check Digit
X
Code 39 Any Length
X
X
X
X
X
X
X
X
X
X
X
X
Code 39 Length W/I Range
X
X
X
X
X
Code 39 One Length
X
X
X
X
X
Code 39 Two Lengths
X
X
X
X
X
Code 128 Any Length
X
X
X
X
X
Convert UPC-E to A
X
X
X
X
Data Transmission Format
X
X
X
X
X
D 2 of 5 Any Length
X
X
X
X
D 2 of 5 Length W/I Range
X
X
X
X
D 2 of 5 One Length
X
X
X
X
D 2 of 5 Two Lengths
X
X
X
X
6-21
Page 85
LS 3070 Product Reference Guide: Programming
Table 6-2. Host Supported Parameters Decode Parameters
RS-232C
IBM 4683*
OCIA
OCR
Keyboard Wedge
Decode Redundancy
X
X
X
X
X
EAN Zero Extend
X
X
X
X
X
Hardware Handshaking
X
Host Interface Code
X
X
X
X
X
X
X
X
X
IBM 468X/9X Mgstrpe Comm. Ignore Unknown Chars.
X
X
Intercharacter Delay
X
I 2 of 5 Any Length
X
X
X
X
I 2 of 5 Length Within Range
X
X
X
X
I 2 of 5 One Length
X
X
X
X
I 2 of 5 Two Lengths
X
X
X
X
ITF-14/EAN-13 Conversion
X
X
X
X
X
X
Int’l Keyboard Emulation
X
Int’l Keyboard Emul Fast Xmit
X
Laser Off Time-out
X
X
X
X
X
Laser On Time-out
X
X
X
X
X
Magstripe Prefix
X
X
X
X
X
Magstripe Suffix
X
X
X
X
X
X
X
NCR 2152 Fast Transmit Number of Stop Bits
X
OCIA Clock Polarity
X
OCIA Transmit Time-out
X
Parity
X
Pause Duration
X
RTS Line State
X
Scan Prefix
X
X
X
6-22
X
X
X
Page 86
LS 3070 Product Reference Guide: Programming
Table 6-2. Host Supported Parameters Decode Parameters
RS-232C
IBM 4683*
OCIA
OCR
Keyboard Wedge
Scan Suffix
X
X
X
X
X
Serial Response Time-out
X
Software Handshaking
X
Transmit AIM Code ID
X
X
X
X
X
Transmit Code ID Chars.
X
X
X
X
X
Transmit MSI Check Digit
X
X
X
X
X
Transmit No Decode Message
X
X
X
X
X
Transmit UPC-A Check Digit
X
X
X
X
X
Transmit UPC-E Check Digit
X
X
X
X
X
UPC/EAN Security Level
X
X
X
X
X
UPC/EAN Supplemental
X
X
X
X
X
UPC-A Preamble
X
X
X
X
X
UPC-E Preamble
X
X
X
X
X
* Some parameters are programmed through host computer.
6-23
Page 87
LS 3070 Product Reference Guide: Programming
Beeper Definitions Standard Use Beeper Sequence
Indication
1 Beep - short high tone
A bar code symbol, or magstripe data was decoded (if decode beeper is enabled).
2 Beeps - long high tone
Mis-match between the selected host and the interface cable.
4 Beeps - long high tone
This signifies either a host interface error or a format or transmission error in the magnetic stripe card or in a scanned symbol. In that case, the data is ignored. This will occur if a unit is not properly configured. Check option settings.
5 Beeps - long high tone
Data conversion error. When Ignore Unknown Characters is disabled, any attempt to transmit data not supported by the selected host will produce these error beeps.
6 Beeps - short low tone
RF communication error.
6-24