MHI-2000 Technical Reference Manual: APPENDIX E Parameters, pages 181 to 190

Pages 181 to 190 of 207, from MHI-2000 Technical Reference Manual

The original document is the PDF (2.6 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.

For the Sick MHI2000

Page 181

640 - 697 Ethernet Network Parameters

665

TELNET LOCAL OPTS TELNET LOCAL OPTS Valid Values

000 - 007

Default Value

000

Description

There are three bit settable Telnet Local Options. TABLE E–19. Values and Descriptions

Value

Description

001

Disable inactivity timeout. By default, the Telnet connection will be closed if there is no activity for 10 minutes. This feature can be disabled by setting this option.

002

Change CR to CR LF on input.

004

Change CR to CR LF on output.

Parameters

E

Rev A, January 2002

MHI-2000 Technical Reference Manual

E-61

Page 182

Appendix

666

E

Parameters

NAME BROADCAST PORT NAME BROADCAST PORT Valid Values

0 - 65535

Default Value

08509

Description

This parameter defines the broadcast port address for a UDP name message. In general, the factory default setting of 08509 should be unchanged. This feature can be disabled by entering an address of 0. This feature provides a method for our service technicians to make an Ethernet connection when the IP address of the device is not known (without having to make a serial connection).

E-62

MHI-2000 Technical Reference Manual

Rev A, January 2002

Page 183

640 - 697 Ethernet Network Parameters

667

TCP/IP KEEP ALIVE TCP/IP KEEP ALIVE Valid Values

5 - 20000

Default Value

00060

Description

This parameter defines the number of seconds the TCP/IP stack uses to determine if a connection is lost.

Parameters

E

Rev A, January 2002

MHI-2000 Technical Reference Manual

E-63

Page 184

Appendix

668

E

Parameters

TCP/IP CONNECT TIME TCP/IP CONNECT TIME

E-64

Valid Values

1- 99

Default Value

05

Description

This parameter defines the number of seconds before a timeout occurs when attempting to open a new socket.

MHI-2000 Technical Reference Manual

Rev A, January 2002

Page 185

640 - 697 Ethernet Network Parameters

670 680 690

NET 1 DEST IP NET 2 DEST IP NET 3 DEST IP NETWORK x DESTINATION IP Valid Values

X.X.X.X (where X = 0 - 255)

Default Value

000.000.000.000

Description

This string provides the TCP/IP address assigned to the server. When the MHI-2000 boots, it will attempt to establish a TCP connection with the server assigned this address. Up to 3 TCP/IP sockets are provided through these parameter settings. Additional sockets can be created with a user program. With the default setting of 000.000.000.000, no socket connection is attempted.

Parameters

E

Rev A, January 2002

MHI-2000 Technical Reference Manual

E-65

Page 186

Appendix

671 681 691

E

Parameters

NET 1 LOCAL SOURCE NET 2 LOCAL SOURCE NET 3 LOCAL SOURCE NETWORK x LOCAL SOURCE Valid Values

0-2

Default Value

0

Description

This setting is used to select the Ethernet controller that will be used for establishing this connection. If you want the operating system to determine which controller to use, select the default setting of 0. TABLE E–20. Values and Descriptions

E-66

Value

Description

0

Determine controller to use in the O/S.

1

Use the on-board (built-in) Ethernet controller.

2

Use the optional add-on Ethernet controller.

MHI-2000 Technical Reference Manual

Rev A, January 2002

Page 187

640 - 697 Ethernet Network Parameters

672 682 692

NET 1 PORT NET 2 PORT NET 3 PORT NETWORK x PORT Valid Values

0 - 65535

Default Value

08510

Description

The device expects to establish a TCP (Ethernet) connection with a server using this well-known port value and the IP address of the server. The default port value of 08510 can be used, providing the server uses the same port value.

E Parameters

The port value must not conflict with port values used by existing server applications such as FTP or Telnet, or use port numbers reserved for applications having root privileges.

Rev A, January 2002

MHI-2000 Technical Reference Manual

E-67

Page 188

Appendix

673 683 693

E

Parameters

NET 1 PROTOCOL NET 2 PROTOCOL NET 3 PROTOCOL NETWORK x PROTOCOL Valid Values

0 or 7

Default Value

0

Description

This parameter indicates the protocol used for reporting data over Ethernet. TABLE E–21. Values and Descriptions

E-68

Value

Description

0

Default Ethernet protocol (TCP)

7

PC-based Tunnel Protocol over Ethernet (TCP)

MHI-2000 Technical Reference Manual

Rev A, January 2002

Page 189

640 - 697 Ethernet Network Parameters

674 675 684 685 694 695

NET 1 MSG PREFIX NET 1 MSG SUFFIX NET 2 MSG PREFIX NET 2 MSG SUFFIX NET 3 MSG PREFIX NET 3 MSG SUFFIX NETWORK x MESSAGE PREFIX and SUFFIX Valid Values

Decimal 000 - 255

Default Value

PREFIX Decimal 002 (STX) SUFFIX Decimal 003 (ETX) All command and data messages sent over the TCP (Ethernet) client connection must begin with the prefix character and end with the suffix character. The default ASCII value of 002 (STX) can be used, providing the server also uses this PREFIX default. The default ASCII value of 003 (ETX) can be used, providing the server also uses this SUFFIX default. Setting the message prefix and suffix to 0 disables block mode communication.

Rev A, January 2002

MHI-2000 Technical Reference Manual

E-69

E Parameters

Description

Page 190

Appendix

676 686 696

E

Parameters

NET 1 MSG CHAR NET 2 MSG CHAR NET 3 MSG CHAR NETWORK x MESSAGE CHARACTER Valid Values

Decimal 000 - 255

Default Value

Decimal 126 ( ~ )

Description

All command messages sent to the device must be framed with this character. The default ASCII character 126 can be used, providing the server application also uses 126. The message character is always required and cannot be disabled by setting it to 255. For example, "~?S~ instructs the device to enter the Setup menu. In this example, the '~' characters are command prefix and suffix indicators (the prefix and suffix are always the same character). It is important to understand that the message character should only appear in a data stream when it is intended to be part of a command.

E-70

MHI-2000 Technical Reference Manual

Rev A, January 2002