Page 161
501 - 560 Inputs/Outputs Parameters
503 504
INZONE TRIGGER OUTZONE TRIGGER INZONE TRIGGER, OUTZONE TRIGGER Valid Values
0 or 1
Default Value
1
Description
The default inputs for the inzone and outzone Presence functions are input 1 and input 2, respectively. You can change the polarity of the signal’s inzone and outzone using these parameters. Table E–15 lists the INZONE/OUTZONE TRIGGER valid values and their descriptions. TABLE E–15. INZONE/OUTZONE TRIGGER Valid Values and Descriptions
Description
0
Input must be active, or true.
1
Input must be inactive, or false.
The device digital inputs are active, or true, when they are pulled low. They are usually connected to normally open relays or open collector photoeyes, which effectively close a switch to ground when the photoeye is blocked. Entering a zero for that input may use the inverse polarity.
Rev A, January 2002
MHI-2000 Technical Reference Manual
E-41
E Parameters
Value
Page 162
Appendix
505 506
E
Parameters
ZONE ERROR CHAR 1 ZONE ERROR CHAR 2 ZONE ERROR CHARACTER 1 and 2 Valid Values
Decimal 000 - 255
Default Value
ZONE ERROR CHAR 1, Decimal 042 ( * ) ZONE ERROR CHAR 2, Decimal 063 ( ? )
Description
The device allows the designation of two ASCII characters as zone error characters. Both characters are appended to the end-of-barcode or NO READ data to alert the host or controller that zone synchronization errors are occurring. There are two zone errors that may occur: •
The inzone Presence detection device may signal a second time before the outzone device has signaled once. In this case, the error characters will be appended in the order of the two parameters (defaults: *?).
•
The outzone device may signal prior to an inzone signal. In this case, the characters will be appended in reverse order (?*).
This scheme indicates which error condition was detected.
Example Assuming that the ZONE ERROR CHARACTER ONE and TWO parameters are left at their default values (decimal 042 and 063, respectively), an inzone error will be reported as: ...<DATA>...*? and an outzone error will be reported as: ...<DATA>. .?* Note: Zone errors have a different meaning when you are using gap tracking with a tachometer, since there may be multiple packages in the scan zone at the same time. The zone error characters described above will not be sent in this situation. There are some zone error situations that can be recognized when using gap tracking. A message of ****** (six asterisks) is sent as an indicator that a box was tracked through the zone, but did not arrive at the outzone within the parameter 703 BOX TOLERANCE value of distance traveled as determined by the tachometer count. In other words, the box was early or late at the outzone photoeye. This could happen if an inzone signal occurred but an outzone signal did not because, for example, a box did not travel all the way through the zone. This error message does not get sent by default. To enable it, set parameter 505 to 255, with gap tracking enabled.
E-42
MHI-2000 Technical Reference Manual
Rev A, January 2002
Page 163
501 - 560 Inputs/Outputs Parameters
510 511 512
INZONE DEBOUNCE (Input 1) OUTZONE DEBOUNCE (Input 2) INPUTS DEBOUNCE (Inputs 3 - 8) INZONE DEBOUNCE, OUTZONE DEBOUNCE, INPUTS DEBOUNCE Valid Values
0 - 255
Default Value
020 (milliseconds) (parameters 510, 512) 000 (milliseconds) (parameter 511) These parameters are the debounce times for their respective inputs. Debouncing allows you to eliminate multiple edges from a noisy input source that may take some time to settle to its final value. Separate times are allowed to cover situations where some inputs might need to have the minimum response time and others need longer debouncing (push buttons and lever-type divert verify switches, for instance).
E
000 = no debounce time, respond immediately to any input change Debounce times help the device deal with the problem of detecting multiple transitions of contact closure inputs, or photoeyes that do not give a logic low level of less than 0.7 V. A larger debounce time helps filter out this noise, but will not always solve the problem. We recommend using only the photoeyes that SICK Auto Ident, Inc. provides, or using solid state input modules instead of relays. This problem can also occur at the tachometer input. Use the tachometer/encoder provided by SICK Auto Ident, Inc., or make sure that the signal swings below 0.7 V and above 2.5 V.
Rev A, January 2002
MHI-2000 Technical Reference Manual
E-43
Parameters
Description
Page 164
Appendix
E
520-524
Parameters
INPUT (4-8) TRIGGER
INPUT X TRIGGER Valid Values
0 or 1
Default Value
1
Description
You can change the polarity of the signals input 4 through input 8 using these parameters. Table E–16 lists the valid values and their descriptions. TABLE E–16. INPUT 4-8 TRIGGER Valid Values and Descriptions
Value
Description
0
Input must be active, or true.
1
Input must be inactive, or false.
The device digital inputs are active, or true, when they are pulled low. They are usually connected to normally open relays or open collector photoeyes, which effectively close a switch to ground when the photoeye is blocked. Entering a zero for that input may use the inverse polarity.
E-44
MHI-2000 Technical Reference Manual
Rev A, January 2002
Page 165
501 - 560 Inputs/Outputs Parameters
553-560
RELAY X HOLD TIME
RELAY TIMES Valid Values
0 - 9999 milliseconds
Default Value
0000 (Relay Disabled)
Description
This parameter specifies the length of time (in milliseconds) for which the seven device solid-state circuits, and the relays driven by them, will remain energized after they have been energized. For mechanical relays to operate, a hold time of at least 12 milliseconds is necessary. A RELAY HOLD TIME value of not less than 15 milliseconds is recommended. Setting the parameter to 0000 will disable an output. Even if an output activation signal is generated, the corresponding output will not activate because it is being told to remain unenergized.
A user C program can use the operating system’s relay hold feature by turning on an output that has a non-zero relay hold time. The OS will automatically turn it off at the end of the hold time. Setting the hold time to either 0 or 9999 will stop the OS from automatically turning it on or off. If a relay hold time is set to 9999, the output will stay in the last driven state until the next decision point changes it.
Rev A, January 2002
MHI-2000 Technical Reference Manual
E-45
E Parameters
Conversely, if a RELAY HOLD TIME value is set to 9999, the corresponding output (and relay) will remain energized until the next output activation decision is made.
Page 166
Appendix
E
Parameters
600 - 639 User Program Parameters 600
USER STACK SIZE STACK SIZE IN KILOBYTES
E-46
Valid Values
2 - 32
Default Value
02 (2 KB)
Description
This parameter sets the stack size available to the user program. Refer to the SH3 C Programmer’s Platform User’s Manual for more detailed information.
MHI-2000 Technical Reference Manual
Rev A, January 2002
Page 167
600 - 639 User Program Parameters
601
FILE SYSTEM SIZE FILE SYSTEM SIZE IN KILOBYTES Valid Values
0 - 2560
Default Value
0032 Kb (set to 0000 to disable file system)
Description
This parameter sets the RAM size available to the user program for C file system operations. The largest size is determined by the amount of RAM in the device. Refer to SICK Auto Ident, Inc. SH3 C Programmer’s Platform documentation for complete information.
Parameters
E
Rev A, January 2002
MHI-2000 Technical Reference Manual
E-47
Page 168
Appendix
602
E
Parameters
PROGRAM CONTROL PROGRAM CONTROL Valid Values
0-4
Default Value
0
Description
This parameter allows skipping the waiting for download… time, and automatically restores the program from Flash on reboot. Table E–17 lists the PROGRAM CONTROL valid values and their descriptions. TABLE E–17. PROGRAM CONTROL Valid Values and Descriptions
E-48
Value
Description
0
Normal operation.
1
Causes the device to wait for a download from the host. If the user program exists in FLASH after a ten second delay, run the user program.
2
Causes the device to load and run the user program from flash.
3
TBD (do not use)
4
TBD (do not use)
MHI-2000 Technical Reference Manual
Rev A, January 2002
Page 169
600 - 639 User Program Parameters
610-625
USER VARIABLE x
USER VARIABLE x Valid Values
0 - 65535
Default Value
00000
Description
User definable parameters can be read and written by the user C program and accessed through the OS Setup menu. These parameters save you from writing a setup menu, and provide locations that are secure from power-down to power-up. These parameters are accessed from the User Variable library C program routine. These parameters are used to save up to 16 integer values in non-volatile RAM that can be used as a simple method to save values from power-up to power-up. Since these locations are not cleared when a program is downloaded, they may be used to communicate between different programs as well.
Refer to SICK Auto Ident, Inc. SH3 C Programmer’s Platform for additional information.
Rev A, January 2002
MHI-2000 Technical Reference Manual
E-49
E Parameters
Note: Using files is a logical way to save values from power-up to power-up or between programs. Another method is to declare a block of uninitialized static variables and use the fact that these variables are non-volatile and not zeroed by the loader when the program starts up. These variables cannot be shared by different programs, however, since their locations are not (in general) known. In this case, these 16 locations are accessed using USER VARIABLE x.
Page 170
Appendix
E
630-639
Parameters
USER STRING x
USER STRING x
E-50
Valid Values
ASCII string up to 20 characters
Default Value
None
Description
User definable parameters can be read and written by the user C program and accessed through the OS Setup menu. These parameters are saved and restored from power-down to power-up.
MHI-2000 Technical Reference Manual
Rev A, January 2002