The text of a manual · part 6 of 18

Operating Instructions CLX490 Compact OMNI Scanner for Bar Codes: 5 Electrical installation, pages 39 to 48

Pages 39 to 48 of 158, from Operating Instructions CLX490 Compact OMNI Scanner for Bar Codes

The original document is the PDF (3.4 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.

For the Sick CLX490

Page 39

Electrical installation

Operating Instructions

Chapter 5

CLX490

5

Electrical installation 5.1

Installation sequence

•

Connect the CLX to a SICK Connection Module from the series AMV/S, BMV/BMH10 or BMS20, or using a customer-specific wiring configuration

•

Connect the data and function interfaces of the CLX in the module

•

Connect the PC to the Connection Module (at the auxiliary interface on the CLX)

•

Connect the power supply to the Connection Module

5.1.1

SICK Connection Modules (overview)

CLX Type

Temperature range

Connection module

Purpose see chapter

CLX without heater

0 to +40 °C

AMV60

– Connecting one CLX – Operating voltage 18 to 30 V DC – Enclosure rating IP 30, max. IP 54

5.2.3

AMS60

– Connecting one CLX – Operating voltage 230 V AC (115 V)/24 V DC – Enclosure rating IP 30, max. IP 54

5.2.3

AMV100

– Connecting one CLX – Operating voltage 18 to 30 V DC – Enclosure rating IP 52, with connector cover max. IP 65

5.2.3

BMV/BMH10

– Connecting one CLX to the Profibus-DP – Operating voltage 20 to 30 V DC – Enclosure rating IP 52, with connector cover max. IP 65

5.2.4

– Connecting one CLX to the Interbus-S – Operating voltage 24 V DC – Enclosure rating max. IP 54

5.2.4

AMV30-071

– Connecting two CLX – Operating voltage 18 to 30 V DC – Enclosure rating IP 30, max. IP 54

5.2.3

AMV200

– Connecting two CLX – Operating voltage 18 to 30 V DC – Enclosure rating IP 52, with connector cover max. IP 65

5.2.3

AMV100

– Connecting one CLX – Operating voltage 24 V DC +20%/–10% – Enclosure rating IP 52, with connector cover max. IP 65

5.2.5

AMV200

– Connecting two CLX – Operating voltage 24 V DC +20%/–10% – Enclosure rating IP 52, with connector cover max. IP 65

5.2.3

Profibus-DP

BMS20 Interbus-S

CLX with heater

–30 to +35 °C

Table 5-1: Connection Modules for the CLX

8009237/RB47/2007-08-03

© SICK AG · Division Auto Ident · Germany · All rights reserved

5-1

Page 40

Electrical installation

Chapter 5

Operating Instructions CLX490 Compact OMNI Scanner

5.2

Electrical connections and cables

The electrical connections on the CLX consist of two 15-pin D Sub HD connections on the housing, one plug, and one socket. These connections are used to route the following interfaces: •

Three serial data interfaces (host interface, CAN interface 1 and 2 as well as auxiliary interface)

•

Six switching inputs (external reading pulse and multifunctional inputs)

•

Four switching outputs (for result status function, for connecting to a PLC for example)

•

Power supply

5.2.1

Wire cross-sections

CLX without heater: ¾

All connections must be wired with copper cables with a minimum wire diameter of 0.15 mm2 !

CLX with heater: ¾

Connect the power supply terminals (Pin 1/Pin 5) using copper wires with a minimum cross-section of 0.75 mm2 at a maximum length of 10 m!

5.2.2 CLX Type

Temperature range

Connection module

Optional cables

Length

Type

AMV/S60

2 x No. 2020302 or 1 x No. 2020307

3m 3m

Connecting cable (plug/socket) Ext. parameter memory with 2 cables (plug/socket)

BMV/BMH10 Profibus-DP

1 x No. 2020307 or 1 x No. 2021298

3m 3m

Ext. parameter memory with 2 cables (plug/socket) Connector cover to connector cover

AMV30-071

2 x No. 2020302 or 1 x No. 2020307

3m 3m

Connecting cable (plug/socket) Ext. parameter memory with 2 cables (plug/socket)

BMS20 Interbus-S

1 x No. 2020264 + 1 x No. 2020265 or 1 x No. 2020308

3m 3m 3m

Connecting cable with plug/open end Connecting cable with plug/socket Ext. parameter memory with 2 cables (plug/open end)

Non-SICK Power pack

1 x No. 2020303 + 1 x No. 2020264 or 1 x No. 2020981 or 1 x No. 2021267

3m 3m 3m 3m

Connecting cable with socket/open end Connecting cable with plug/open end Ext. parameter memory with 2 cables (open ends) Connector cover with 2 cables (open ends)

– 30 to +35 °C AMV100

1 x No. 2021298 or 1 x No. 2021689

3m 3m

Connector cover to connector cover Ext. parameter memory to connector cover

AMV200

2 x No. 2021298 or 2 x No. 2021689

3m 3m

Connector cover to connector cover Ext. parameter memory to connector cover

Non-SICK Power pack

1 x No. 2021267

3m

Connector cover with 2 cables (open ends)

CLX without 0 to +40 °C heater

CLX with heater

Prefabricated cables (overview)

Table 5-2: Cables for connecting the CLX

For technical data on the cables, see Chapter 10.12.5 Cables, external parameter memories and plug cover, Page 10-37.

5-2

© SICK AG · Division Auto Ident · Germany · All rights reserved

8009237/RB47/2007-08-03

Page 41

Operating Instructions

Electrical installation

Chapter 5

CLX490

5.2.3

Connections/cables for the AMV/S Connection Module

The AMV/S Connection Module is suitable for connecting the CLX to peripherals (distribution function) and the power supply. The module can be used to establish a connection to the host (point-to-point) or integrate the device in a SICK network or daisy-chain configuration (pass-through or master/slave configuration). The module is available in several variants (see Table 5-1, Page 5-1). Fig. 5-1 shows the connection principle of the AMV/S60 for one CLX. CLX490

Scanner

AMV/AMS60 connection module

Decoder Photoelectric switch Reading pulse

PC "Terminal"

Interface "Terminal" "Host"

HOST/PLC

"Result 1 ... 4"

Photoelectric switch Focal position Conveyor increment Reading pulse end

"Host"

"Sensor" "In 0 ... 4" DC 18 ... 30 V

"Result 1 ... 4"

PLC

"Sensor" "In 0 ... 4"

– – Cable if required Fig. 5-1:

18 to 30 V DC (AMV60-011) 230 V AC 50 Hz (AMS60-013) 115 V AC 50/60 Hz (AMS60-012)

Block diagram: Connection of the CLX to the AMV/S60 connection module

Connecting the CLX to the AMV/S Two cables no. 2020302 are required to connect the CLX to the AMV/S. Alternatively, the device can be connected via the external parameter memory no. 2020307. See Chapter 5.2.5 Connections/cables for the external parameter memory (connection to AMV/S or BMV10/BMS20), Page 5-4. a) CLX without heater: Connection modules: AMV/S60 and AMV30-071. The cables no. 2020302 can be extended to 10 m (auxiliary interface: RS 232!). b) CLX with heater: Connection modules: AMV100 and AMV200. The cables no. 2020302 must not be extended! If longer cables are required, the entire power supply system (Pin 1/Pin 5) must be connected with a wire cross-section of at least 0.75 mm2 with a maximum length of 10 m. Recommendation

Use the cable no. 2021298 (with two connector covers, length 3 m). The procedures for connecting and configuring the AMV/S Connection Modules are described in the following documentation:

8009237/RB47/2007-08-03

•

Operating Instructions for the "AMV/S60 Connection Module" (no. 8008296)

•

Operating Instructions for the "AMV30-071 Connection Module" (no. 8008648)

•

Operating Instructions for the "AMV100/200 Connection Module" (no. 8008879)

© SICK AG · Division Auto Ident · Germany · All rights reserved

5-3

Page 42

Electrical installation

Chapter 5

Operating Instructions CLX490 Compact OMNI Scanner

5.2.4

Connections/cables for the Bus Connection Modules BMV10 and BMS20

The BMV10 Bus Connection Module (depending on the model) is used to connect a CLX to the Profibus-DP or to the DeviceNet for use in industrial applications. The BMS20 Bus Connection Module enables the CLX to be integrated in the Interbus-S. Connecting the CLX to the BMV10/BMS20 BMV10: two cables no. 2020302 are required. BMS20: the cables no. 2020265 and no. 2020264 are required. Alternatively, via external parameter memory no. 2020307 (see Chapter 5.2.5 Connections/cables for the external parameter memory (connection to AMV/S or BMV10/ BMS20)). a) CLX without heater: Connection modules: BMV 10and BMS20. The cables no. 2020302 can be extended to 10 m for CLX without a heater (host and auxiliary interface: RS-232!). b) CLX with heater: Connection module: BMV10. The BMV10 may only be used if it is mounted outside the refrigeration area (temperature range 0 to +40 °C). The cables no. 2020302 must not be extended. If longer cables are required, the entire power supply system (Pin 1/Pin 5) must be connected with a wire cross-section of at least 0.75 mm2 at a maximum of 10 m. Recommendation

Use the cable no. 2021298 (with two connector covers, length 3 m). The BMS 20 is not suitable for powering CLX with heater! The procedures for connecting and configuring the Bus Connection Modules are described in the following documentation: •

Operating Instructions for the "BMV/BMH10 Bus Connection Module for Profibus DP" (no. 8008825)

•

Operating Instructions for the "BMV/BMH10 Bus Connection Module for DeviceNet" (no. 8008972)

•

Technical Information for the "BMS20 Bus Connection Module for Interbus-S" (no. 8007546)

5.2.5

Connections/cables for the external parameter memory (connection to AMV/S or BMV10/BMS20)

The external parameter memory (external accessory) is located in a connector cover with 15-pin D Sub HD plug connections. The cover covers the terminals on the CLX so that the enclosure rating IP 65 is achieved. The connector cover with the external parameter memory is prefabricated with two cables, each 3 m in length, and is available with four different cable ends:

5-4

• with two 15-pin D Sub HD connections (pin assignment identical to that of the CLX terminals), no. 2020307

• with a connector cover, no. 2021689

• with two open cable ends, no. 2020981

• with one 9-pin D Sub cable connector and one open end, no. 2020308

© SICK AG · Division Auto Ident · Germany · All rights reserved

8009237/RB47/2007-08-03

Page 43

Operating Instructions

Electrical installation

Chapter 5

CLX490

Connecting the CLX with external parameter memory a) CLX without heater: Connection Modules: AMV/S60, AMV30-071, BMV10 ¾

Cover the connections on the CLX with the connector cover (parameter memory), no. 2020307. Connect the plug/socket on the cables ("Host/Term" and " I/O" connections) with the appropriate connections on the AMV/S or BMV10.

Connection Module: BMS20 ¾

Recommendation

Cover the connections on the CLX with the connector cover (parameter memory) no. 2020308. Connect the 9-pin plug ("Host/Term" connection) to the socket on the BMS20. Connect the wires of the free cable end ( "I/O" terminal) to the terminal strips in the BMS20.

Use the external parameter memory no. 2020981 (open cable ends) for connecting the device to non-SICK Power packs/wiring configurations. b) CLX with heater: Connection Module: AMV100/200 ¾

AMV100: Cover the connections on the CLX with the connector cover (parameter memory), no. 2021689, labeled ”SCANNER”. Connect the other connector covers, labeled "HOST", to the corresponding connections on the AMV100.

¾

AMV200: see AMV100, proceed in the same way with the second CLX.

5.2.6

Connections/cables for the IP 65 connector cover (connection to AMV100/200 or BMV10)

The connector cover (optional accessory) with 15-pin D Sub HD plug connections is used to cover the terminals on the CLX so that the enclosure rating IP 65 is provided. The connector cover is prefabricated with two cables, each of 3 m in length, and is available with two different cable ends: • with one additional connector cover at the other end (pins assignment identical to that of the CLX terminals), no. 2021298

• with two open cable ends, no. 2021267

CLX with heater: Both connector cover variants can be used with temperatures up to max. –50 °C if the CLX is stationary and the cables are not moved. If the cables are moved due to changes in the position of the CLX, the temperature must not drop below max. –40 °C. Recommendation

8009237/RB47/2007-08-03

Use the external parameter memory no. 2020267 (open cable ends) for connecting the device to non-SICK Power packs/wiring configurations.

© SICK AG · Division Auto Ident · Germany · All rights reserved

5-5

Page 44

Electrical installation

Chapter 5

Operating Instructions CLX490 Compact OMNI Scanner

5.3

Connector pin assignment

5.3.1

Terminals on the CLX

Device plug: "Host/Term" connection 6 1

Pin

5 10

1)

11

1 2 3 4 5 6 7

15

8 9 10 11 12 13 14 15 Housing

Signal VS

Function Supply voltage

RxD (RS 232), Auxiliary TxD (RS 232), Auxiliary Term (RS 422/485) GND RD+ (RS 422/485), Host RD– (RS 422/485), Host RxD (RS 232), Host TD+ (RS 422/485), Host TD– (RS 422/485), Host TxD (RS 232), Host CAN H Reserved CAN2 H CAN2 L Reserved CAN L –

Auxiliary interface (receiver) Auxiliary interface (transmitter) Termination host interface Ground Host interface (receiver+) Host interface (receiver–) Host interface (sender+) Host interface (sender–) CAN Bus interface 1 (IN/OUT) – CAN Bus interface 2 (IN/OUT) CAN Bus interface 2 (IN/OUT) – CAN Bus interface 1 (IN/OUT) Shield

1) Pin 1 is jumpered with Pin 1 of the "I/O" connection in the CLX

Table 5-3: Pin assignment of the 15-pin D Sub HD "Host/Term" plug

Device socket: "I/O" connection L

10 5

1

6

Pin

Signal

Function

VS

Supply voltage

N1

Switching input (trigger for focus control)

3

Sensor

Switching input (external reading pulse)

4

Result 1

Switching output, variable function

5

GND

Ground

6

IN 0

Switching input (trigger for focus control)

7

IN 2

Switching input (trigger for focus control)

8

Result 2

Switching output, variable function

9

INGND

Common ground for all inputs

10

Result 3

Switching output, variable function

11

IN 3

Switching input, variable function

12

IN 4

Switching input, variable function

13

I2C SDA

I2C Bus (for external parameter memory)

14

I2C SCL

I2C Bus (for external parameter memory)

15

Result 4

Switching output, variable function

Housing

–

Shield

1 15

11

1)

2

1) Pin 1 is jumpered with Pin 1 of the "Host/Term" connection in the CLX

Table 5-4: Pin assignment of the 15-pin D Sub HD "I/O" socket

5-6

© SICK AG · Division Auto Ident · Germany · All rights reserved

8009237/RB47/2007-08-03

Page 45

Electrical installation

Operating Instructions

Chapter 5

CLX490

5.3.2

External parameter memory no. 2020307/2021689 (optional accessory) connector cover no. 2021298 (optional accessory)

Cable plug: "Host/Term" connection 6 1

5 10

Pin 1)

11

1 2 3 4 5 6 7

15

8 9 10 11 12 13 14 15 Housing

Signal VS

Function Supply voltage

RxD (RS 232), Auxiliary TxD (RS 232), Auxiliary Term (RS 422/485) GND RD+ (RS 422/485), Host RD– (RS 422/485), Host RxD (RS 232), Host TD+ (RS 422/485), Host TD– (RS 422/485), Host TxD (RS 232), Host

Auxiliary interface (receiver) Auxiliary interface (transmitter) Termination host interface Ground Host interface (receiver+) Host interface (receiver–)

CAN H2) Reserved CAN2 H2 CAN2 L2 Reserved CAN L2) –

Host interface (sender+) Host interface (sender–) CAN Bus interface 1 (IN/OUT) – CAN Bus interface 2 (IN/OUT) CAN Bus interface 2 (IN/OUT) – CAN Bus interface 1 (IN/OUT) Shield

1) Pin 1 is jumpered with Pin 1 of the "I/O" connection in the CLX 2) not when using cable no. 2021689, no. 2021298 and no. 2021267

Table 5-5: Pin assignment of the 15-pin D Sub HD "Host/Term" cable plug

Cable socket: "I/O" connection 10 5

1

6

Pin

Signal

Function

1

VS

Supply voltage

2

IN 1

Switching input (trigger for focus control)

1)

15

11

3

Sensor

Switching input (external reading pulse)

4

Result 1

Switching output, variable function

5

GND

Ground

6

IN 0

Switching input (trigger for focus control)

7

IN 2

Switching input (trigger for focus control)

8

Result 2

Switching output, variable function

9

INGND

Common ground for all inputs

10

Result 3

Switching output, variable function

11

IN 3

Switching input, variable function

12

IN 4

13

Switching input, variable function 2

I2C Bus (for external parameter memory)

2

I2C SDA )

14

I2C SCL )

I2C Bus (for external parameter memory)

15

Result 4

Switching output, variable function

Housing

–

Shield

1) Pin 1 is jumpered with Pin 1 of the "Host/Term" connection in the CLX 2) not connected

Table 5-6: Pin assignment of the 15-pin D Sub HD "I/O" cable socket

8009237/RB47/2007-08-03

© SICK AG · Division Auto Ident · Germany · All rights reserved

5-7

Page 46

Electrical installation

Chapter 5

Operating Instructions CLX490 Compact OMNI Scanner

5.4

Preparations for electrical installation

5.4.1

Requirements for the host interface

The host interface of the CLX can be operated as an RS 422/485 or an RS 232 interface. Table 5-7 shows the recommended maximum cable lengths as a function of the selected data transfer rate. Interface type

Transfer rate

Distance from host

RS 232 up to 19,200 bd 38,400 to57,600 bd max. 10 m max. 3 m

RS 422/4851)

max. 38,400 bd max. 57,600 bd max. 1,200 m max. 500 m

1) with suitable line termination according to specifications

Table 5-7: Maximum cable lengths between the CLX and host

¾

To prevent interference, do not lay the cables parallel with power supply and motor cables over long distances, e. g. in cable ducts.

5.4.2

Supply voltage

The CLX requires a supply voltage of 18 to 30 V DC for operation without a heater and 24 V DC +20 %/–10 % for operation with a heater (functional extra-low voltage in accordance with IEC 364-4-41). The functional extra-low voltage can be generated by using a safety transformer pursuant to IEC 742. The power consumption of the individual types is as follows: Type

Voltage

Heater

Power consumption1)

Connection Modules

CLX490-0010

18 to 30 V DC no

11 W (typ.)/ max. 16 W

AMV/S60 or AMV30-071 or BMV/BMH10 or BMS20

CLX490-0011

24 V DC +20 %/–10 % yes

75 W (typ.)/ max. 90 W

AMV100/200

1) Switching outputs not connected

Table 5-8: Power consumption of the CLX

Power-up delay The selected device number (default: 1) affects the power-up delay of the device. This is useful if a large number of CLX (e. g. in the SICK network) are to be supplied from one power source. Table 5-9 contains a list of the available intervals. Device number GN

Power-up delay

Device number GN

Power-up delay

1; 11; 21; 31

0 ms

6; 16; 26

2,000 ms

2; 12; 22

400 ms

7; 17; 27

2,400 ms

3; 13; 23

800 ms

8; 18; 28

2,800 ms

4; 14; 24

1,200 ms

9; 19; 29

3,200 ms

5; 15; 25

1,600 ms

10; 20; 30

3,600 ms

Table 5-9: Power-up delay as a function of the device number GN

Tip

5-8

The device number can be selected on the DEVICE CONFIGURATION tab in the "CLV-Setup" program.

© SICK AG · Division Auto Ident · Germany · All rights reserved

8009237/RB47/2007-08-03

Page 47

Operating Instructions

Electrical installation

Chapter 5

CLX490

5.4.3

Non-SICK Power supply unit/connections without the Connection Module

Power output If a non-SICK Power supply unit is used instead of the AMS60, it must be capable of providing the following voltage and power values: • for CLX without heater: 18 to 30 V DC, min. 20 W continuous power output

• for CLX with heater: 24 V DC +20 %/–10 %, min. 100 W

The non-Sick Power supply unit must provide the functional extra-low voltage in accordance with IEC 364-4-41. The output circuit must be reliably electrically isolated from the input circuit. To do so, use a safety isolating transformer pursuant to IEC 742. Wire cross-section The wire cross-section for the power supply (Pin 1/Pin 5) should be at least 0.15 mm2 , or 0.75 mm2 for CLX with heater, with a max. length of 10 m. a) Connecting the CLX without a connector cover/external parameter memory The two cables no. 2020303 and no. 2020264 with open cable end on one side are required to connect the CLX. The wire color assignments are shown in Table 5-10 and Table 5-11, Page 5-10. The cables must not be extended for CLX with heater. Connection cable no. 2020203 ("Host/Term" connection) 15-pin D Sub HD socket and open cable end. Pin

Signal

Wire color

11)

VS

Red

2

RxD (RS 232), Auxiliary

White

3

TxD (RS 232), Auxiliary

Brown

4

Term (RS 422/485)

Violet

5

GND

Blue

6

RD+ (RS 422/485), Host

Green

7

RD– (RS 422/485), Host RxD (RS 232), Host

Yellow

8

TD+ (RS 422/485), Host

Grey

9

TD– (RS 422/485), Host TxD (RS 232), Host

Pink

10

CAN H

Grey-pink

11

Reserved

Red-blue

12

CAN2 H

White-green

13

CAN2 L

Brown-green

14

Reserved

White-yellow

15

CAN L

Yellow-brown

–

Shield

Orange

1) Pin 1 is jumpered with Pin 1 of the "I/O" connection in the CLX

Table 5-10: Wire color assignment of the cable no. 2020303

8009237/RB47/2007-08-03

© SICK AG · Division Auto Ident · Germany · All rights reserved

5-9

Page 48

Chapter 5

Electrical installation

Operating Instructions CLX490 Compact OMNI Scanner

Connection cable no. 2020264 ("I/0" connection) 15-pin D Sub HD plug and open cable end Pin

Signal

Wire color

11)

VS

Red

2

IN 1

White

3

Sensor

Brown

4

Result 1

Violet

5

GND

Blue

6

IN 0

Green

7

IN 2

Yellow

8

Result 2

Grey

9

INGND

Black

10

Result 3

Grey-pink

11

IN 3

Red-blue

12

IN 4

White-green

13

I2C SDA

Brown-green

14

I2C SCL

White-yellow

15

Result 4

Yellow-brown

–

Shield

Orange

1) Pin 1 is jumpered with Pin 1 of the "Host/Term" connection in the CLX

Table 5-11: Wire color assignment of the cable no. 2020264 b) Connecting the CLX with external parameter memory no. 2020981 CLX without heater: ¾

Cover the "Host/Term" and "I/O" connections on the CLX with the connector cover (parameter memory) no. 2020981. Connect the free cable ends accordingly. The wire color assignments are shown in Table 5-12 and Table 5-13.

CLX with heater: Available on request.

5-10

© SICK AG · Division Auto Ident · Germany · All rights reserved

8009237/RB47/2007-08-03