Page 77

Operating Instructions

Operation

Chapter 6

CLX490

The output format of the reading result is the same as that of the Reading mode. Fig. 6-3, Page 6-16 explains the structure and function of the reading diagnosis data. 6.5.3

Background teach-in

If the Autofocus function is used in the DIFFERENTIAL BACKGROUND or DIFFERENTIAL BACKGROUND AND TRACKING mode, the CLX must first learn the ambient background in its reading field to narrow the reading field even further. When the internal distance profile is set, the reading field of the CLX must not contain any objects. In this mode, the CLX does not output any reading results. The background teach-in function can be started as a device function in the submenus under CLX490, DEVICE FUNCTIONS, BACKGROUND, TEACH-IN or press the [F2] key. Starting the background teach-in: 1.

Mount the CLX at the reading station and align it with an object containing a bar code.

2.

Remove the object so that the CLX has an unobstructed view of the background.

3.

Choose CLX490, DEVICE FUNCTIONS, BACKGROUND, TEACH-IN – or – press the [F2] key. The EXECUTE DEVICE FUNCTIONS window opens. Choose BACKGROUND, TEACH-IN in the EXECUTE menue. The Device Ready LED extinguishes. The CLX then scans its background and learns the ambient background so that it can create a suitable distance profile. The CLX outputs a message once the teach-in procedure has been successfully completed (see Fig. 6-6). The CLX has then learned the ambient background.

Fig. 6-6:

4.

8009237/RB47/2007-08-03

CLV-Setup: Dialog window for running the background teach-in

Choose CLOSE to close the window. If the parameters in the CLX were modified as a result of the teach-in procedure, CLVSetup asks you whether you want to copy and display the modified parameter set by uploading it from the CLX.

© SICK AG · Division Auto Ident · Germany · All rights reserved

6-19

Page 78

Operation

Chapter 6

Operating Instructions CLX490 Compact OMNI Scanner

5.

Confirm the query with YES. The CLX resumes Reading mode and the "Device Ready" LED lights up.

Displaying the taught-in background: The new background distance profile can be uploaded and displayed in CLV-Setup via the auxiliary interface of the CLX. CLV-Setup displays the background without any parameterized limitations ("Autofocus space"). 1.

Choose the READING CONFIGURATION tab.

2.

In the AUTOFOCUS section, click the SHOW BACKGROUND button. The SHOW BACKGROUND dialog box appears. An example of a distance profile is shown in Fig. 6-7, with part A of the scan line.

Fig. 6-7:

Note

6-20

CLV-Setup: Display of the learned background

The CLX adjusts the entire profile of the taught-in background by minus 100 mm, that is, the entire background curve is moved toward the CLX by this amount (also in the display in CLVSetup) to compensate measurement inaccuracies. The parameterized limits of the applicable autofocus space are not corrected. The CLX also shifts the edges of the measured profile toward the center by 2 angular elements. If no background has been taught in, the CLX uses the default background (2,200 mm across the entire distance profile).

© SICK AG · Division Auto Ident · Germany · All rights reserved

8009237/RB47/2007-08-03

Page 79

Operating Instructions

Operation

Chapter 6

CLX490

6.5.4

Displaying and editing operating data

This function enables statistical operating data, which the CLX maintains in the form of counters during the reading procedure, to be displayed and reset. The CLX does not output a reading result in this mode. 1.

Click in the tool bar. The CLX cancels the Reading mode and the "Device Ready" LED is extinguished. The OPERATING DATA dialog box appears (Fig. 6-8).

2.

After checking/resetting the desired counters, click OK to confirm any changes made. The CLX then returns to the Reading mode and the "Device Ready" LED lights up.

Fig. 6-8:

6.5.5

CLV-Setup: "Operating Data" dialog box

Reading diagnosis

Function of the auxiliary interface. In this mode, the CLX outputs the data contents of all the bar codes that were read – including those that, according to the evaluation criteria, were not read properly and were thus incorrect – together with the associated reading diagnosis data via the auxiliary interface. The number of bar codes output can, therefore, be higher than the number of those that were sent via the host interface in the reading result. In the default setting, this mode is selected for the auxiliary interface. Fig. 6-3, Page 6-16 shows the output format of the auxiliary interface. In the default setting, the CLX doesn’t output any reading diagnosis data via the host interface. If the reading diagnosis is to remain inactive, this can be set via the AUXILIARY INTERFACE tab.

8009237/RB47/2007-08-03

1.

In the AUXILIARY INTERFACE drop-down list, choose READ DIAGNOSTICS.

2.

Perform a download to the CLX by clicking in the toolbar. The DOWNLOAD PARAMETERS dialog box is then displayed.

3.

Choose the PERMANENT storage option in the dialog box.

© SICK AG · Division Auto Ident · Germany · All rights reserved

6-21

Page 80

Operation

Chapter 6

Operating Instructions CLX490 Compact OMNI Scanner

The auxiliary interface is now set to the "Read Diagnostics" mode. 6.5.6

Monitor Host Interface

Function of the auxiliary interface. In this mode, the CLX outputs the data traffic of its host interface via the auxiliary interface. Protocol driver handshakes and protocol-specific data, such as start and stop characters, are suppressed here. Each data string is displayed on a separate line on the screen. Two monitoring modes are available that differ with regard to the data output format (Table 6-9). Mode

Direction of data

Output format

Representation on screen

"Monitoring host interface"

CLX receives from host CLX sends to host

<STX> I Data <CR> <LF> <ETX> <STX> O Data <CR> <LF> <ETX>

I Data O Data

"Monitoring host interface 2"

CLX receives from host CLX sends to host

<STX> Data <ETX> <STX> Data <ETX>

Data Data

Table 6-9: "Monitor Host Interface" function

The MONITOR HOST INTERFACE or MONITOR HOST INTERFACE 2 function is activated via the AUXILIARY INTERFACE tab.

Note

1.

In the AUXILIARY INTERFACE drop-down list, choose MONITOR HOST INTERFACE or MONITOR HOST INTERFACE 2.

2.

Perform a download to the CLX. To do so, click the desired option in the list again with the right mouse button and choose DOWNLOAD PARAMETER in the context menu. CLV-Setup copies the parameter to the CLX temporarily. The auxiliary interface then operates in the choosed "Monitor Host Interface" mode until the CLX is switched off again.

On good read, the CLX only outputs the number of read bar codes being defined for maximum on the CODE CONFIGURATION tab, in the NUMBER OF CODES section. Displaying the data traffic of the host interface in the Terminal Emulator: The data traffic of the host interface can be displayed in the Terminal Emulator of CLV-Setup. Fig. 6-9 shows an example of how the reading result can be output.

6-22

1.

Choose the serial interface as the external trigger source for the reading pulse. See Chapter 6.5.1 Reading mode (standard operating mode), Page 6-14.

2.

Click in the toolbar. The Terminal Emulator window is then displayed. The CLX is in the Reading mode.

3.

Click on the SW-TRIGGER ON button or press [F7]. The red scan line then appears as a cross.

4.

Present the bar code pattern from Fig. 6-1, Page 6-3.

5.

Click the SW-TRIGGER OFF button or press [F8]. CLV-Setup outputs the reading result in the Terminal Emulator. Example: "O 0123412345".

© SICK AG · Division Auto Ident · Germany · All rights reserved

8009237/RB47/2007-08-03

Page 81

Operation

Operating Instructions

Chapter 6

CLX490

Fig. 6-9:

Note

CLV-Setup: Displaying the reading result of the host interface in the Terminal Emulator with direction identifier at the beginning (in this case: O = Output)

If a large amount of data is being read in at high speed and with a high transmission rate, the host interface’s data trafficing may no longer be completely displayed on the auxiliary interface (display "..."). This is due to the slower transmission rate of the auxiliary interface (9,600 bd). In the default setting, the CLX sends the following in the data output string of the host interface: •

On Good Read: Data content of one bar code

•

On No Read: Character string "NOREAD"

The "Header", "Code-Info/Separator", "Splitter" and "Terminator" blocks are blank. Tip

8009237/RB47/2007-08-03

The structure of the data output string of the host interface can be configured via the DATA STRINGS tab. •

Up to 10 elements, consisting of constants (letters, digits, control characters) and/or reading diagnosis data, can be entered in the "Header", "Code-Info/Separator", "Splitter" und "Terminator" blocks.

•

Up to 5 elements, consisting of constants (letters, digits, control characters) can be entered in the "Splitter" blocks.

•

The position of the "Code-Info/Separator" block can be choosen before or after the code.

1.

Choose the DATA STRINGS tab.

2.

Click the HEADER field. The EDIT PARAMETER: TFH dialog box is then displayed.

3.

Click the desired constants or placeholders for the reading diagnosis data. The selected elements appear in the text box at the top of the dialog box in the sequence in which they were selected.

4.

Confirm the entries made by clicking OK.

5.

Repeat this procedure for the "Code-Info/Separator", "Splitter" und "Terminator" blocks.

© SICK AG · Division Auto Ident · Germany · All rights reserved

6-23

Page 82

Chapter 6

Operation

Operating Instructions CLX490 Compact OMNI Scanner

6.

Perform a download to the CLX by clicking in the toolbar. The DOWNLOAD PARAMETERS dialog box is then displayed.

7.

Choose the PERMANENT storage option in the dialog box. The CLX outputs the selected elements in the data output string of the host interface with the next reading result.

6.5.7

Auxiliary input

Function of the auxiliary interface. In this mode, the CLX accepts a bar code entered at the auxiliary interface (via the keyboard or handheld scanner with decoder). It sends the bar code to the host in a separate data string via its host interface. No reads can, therefore, be corrected by transmitting missing bar codes subsequently. Chapter 10.10.1 Auxiliary input via auxiliary interface, Page 10-29 describes this function and the associated procedure in greater detail. 6.5.8

Self-test

After the CLX has been switched on, it performs a self-test before it is initialized with the parameter set. The test can be called up explicitly at any time via CLV-Setup. During the selftest, the CLX checks that its hardware components are functioning correctly. A final message via the auxiliary interface provides information on the test result. The CLX does not output a reading result during the test routine. 1.

Click in the toolbar. The Terminal Emulator window is then displayed. The CLX is in the Reading mode.

2.

In the DEVICE MODE section, choose SELF TEST. The "Device Ready" LED is extinguished. The CLX cancels the Reading mode and starts the test routine. After a few seconds, the CLX outputs the encoded test result in the form of a code number (Fig. 6-10).

3.

To return to the Reading mode, click READING MODE or close the terminal emulator. The CLX then returns to the Reading mode and the "Device Ready" LED lights up.

Fig. 6-10:

6-24

CLV-Setup: Displaying the self-test result in the terminal emulator

© SICK AG · Division Auto Ident · Germany · All rights reserved

8009237/RB47/2007-08-03

Page 83

Operating Instructions

Operation

Chapter 6

CLX490

The code number ”15000” means that the self-test was completed successfully and that no faults were diagnosed. Chapter 8.3.1 CLX without external parameter memory, Page 8-2, lists the error keys together with the associated corrective measures. 6.5.9

Executing CLX functions interactively

CLV-Setup enables a number of CLX functions to be executed interactively. CLV-Setup prompts the user to carry out specific actions and provides information on the progress of the function being executed. The following functions are available: •

Background teach-in/reset (default setting)

•

Percentage evaluation

•

Copy parameter memory

•

Analyze scanner network

•

Activate factory defaults in device

¾

Choose CLX490, DEVICE FUNCTIONS and the desired funktion – or – press the [F2] key. The EXECUTE DEVICE FUNCTIONS dialog box is then displayed. Choose the desired function in the EXECUTE menu in the menu. CLV-Setup starts the function and, where appropriate, prompts the user to carry out the necessary actions.

6.6

CLX messages

The CLX outputs encoded system and error messages via the auxiliary interface to report user actions or events in the reading process. When the system is started, the CLX also outputs system messages in plain text (English). If configured accordingly, the CLX transfers the error status ST in the data output string of the host interface. 6.6.1

Displaying messages

In the 4th field from the left in the status bar at the bottom of the screen, CLV-Setup displays system and error messages that occur at the auxiliary interface of the CLX. These are recorded in a file, which is continuously updated. The messages have the following format: •

System message: "CLV SYS-Message: xxx"

•

Error message:

"CLV SYS-Failure: xxx"

The additional system messages in plain text can be displayed in the terminal emulator if CLV-Setup and the terminal emulator are started before the CLX is connected. Additional messages: When the system is started, these system messages inform the user whether the starting process was successful. No further measures are normally required. Chapter 10.5 System messages, Page 10-14 explains the contents of the messages. Error messages: Error messages indicate the following types of error: •

8009237/RB47/2007-08-03 a device defect

© SICK AG · Division Auto Ident · Germany · All rights reserved

6-25

Page 84

Operation

Chapter 6

Operating Instructions CLX490 Compact OMNI Scanner

• incorrect parameter settings

• errors during data transmission to the host

Chapter 8.3 Error messages, Page 8-2 lists the messages with the associated corrective measures. Note

To call up the most recent error to occur in the CLX, enter command "2?SF" in the command line of the terminal emulator.

6.7

Switching off the CLX

1.

If the parameter set was modified in CLV-Setup or was only stored temporarily in CLX via a download ("Temproray" option or via context menu (right mouse button)), the parameter set must be stored permanently in the CLX by choosing the PERMANENT storage option.

2.

Save the parameter set as a configuration file "*.scl" in CLV-Setup.

3.

Switch off the power supply of the AMV/S60 or unplug both of the CLX’s connecting cables from the connection module.

The last parameter set stored permanently in the CLX remains valid. Archiving the parameter set: We recommend that you print out the configuration file in order to archive the parameter set.

6-26

1.

Click in the toolbar. The PRINT FILE dialog box is then displayed.

2.

Enter a comment in the input field in order to assign the printout to the CLX. Confirm your entry with OK. The PRINT dialog box for configuring the printer is then displayed.

3.

Make the necessary entries in the dialog box and confirm these. CLV-Setup prints out the current configuration file in the form of a table.

© SICK AG · Division Auto Ident · Germany · All rights reserved