For the Sick CLV490

Page 207

Operating Instructions

Appendix

Chapter 10

CLV 490 Bar Code Scanner

10.4.4

Electrical installation

Cable cross-section The connection cables for the power supply must have a cross-section of at least 0.75 mm2 (approx. 20 AWG). See also Chapter 5.4.2 Supply voltage, Page 5-17. Power consumption The power consumption depends on the variant: •

Line scanner with heater: typically 75 W and max. 90 W

•

Line scanner with oscillating mirror and heater: typically 75 W and max. 100 W

These values apply to operation without the switching outputs connected. CDB 420 or CDM 490 Connection Modules The CDB 420 or CDM 490 Connection Modules are suitable to power the CLV with heater. An external power supply pack with a higher power output must be used to provide the required voltage of 24 V DC +20%/–10%. The incomming/outgoing cables must be protected by corresponding fuses. The optional CMP 490 Power Supply Module is not suitable for CLVs with heater. See Chapter 3.1.3 System requirements for stand-alone configuration, Page 3-2. For connecting the CLV with heater to the CDB 420 or CDM 490 see also the "CDB 420 Connection Module" Operating Instructions (order no. 8 010 001, German/English) respectively the "CDM 490 Connection Module" Operating Instructions (order no. 8 010 005, German/English). 10.4.5

Outdoor applications

If the CLV with integrated heater is used outdoors, it should be installed in a protective housing to prevent the front window from being damaged by rain, snow, or dust. The housing also acts as a wind protector.

8 009 993/O824/20-12-2004

© SICK AG · Division Auto Ident · Germany · All rights reserved

10-41

Page 208

Chapter 10

Appendix

Operating Instructions CLV 490 Bar Code Scanner

10.5

System messages

The CLV outputs all system messages in plain text on the terminal interface. The messages are in English and can be displayed in the Terminal Emulator in the "CLV-Setup" program. See also Chapter 6.6.1 Displaying messages, Page 6-60. 10.5.1

CLV without external parameter memory

Message

Meaning

"CLV 490 Soft.Vers. Vx.xx Aend. Nr. 0000 CLV 490/OTS"

The software version and revision is displayed when the power supply is switched on.

"internal parameters loaded"

When the CLV is started, it imports the parameter set that was last stored permanently in the internal EEPROM to the RAM

"laser safety timeout"

The CLV has shut off the laser diode 10 minutes (default setting) after a reading pulse has begun. The reading interval is still active, even though the CLV is no longer reading. The reading interval has to be terminated by a corresponding pulse signal. The laser diode is activated again with the next reading pulse.

"no code"

In Reading/Percentage Evaluation mode, the CLV indicates that no codes matching the parameterized evaluation conditions were detected during the reading interval (error status ST = 2 output on the host interface).

"No ext. parameter memory con- No external parameter memory was connected to the CLV when it nected" was started. Table 10-5: CLV system messages

10.5.2

CLV with external parameter memory connected

In addition to the messages above, the CLV also indicates whether it was able to access the external parameter memory Message

Meaning

"diff. SW version of parameters" When the CLV was started, the software version of the parameter set in the external memory was different from that in the CLV. The parameter set is still transferred. "external parameters loaded"

When the CLV was started, the external parameter set was imported successfully to the RAM.

"internal parameters copied to external memory"

When the CLV was started, the internal parameter set was successfully copied to the external parameter memory.

Table 10-6: Additional CLV system messages for the connected parameter memory

10-42

© SICK AG · Division Auto Ident · Germany · All rights reserved

8 009 993/O824/20-12-2004

Page 209

Appendix

Operating Instructions

Chapter 10

CLV 490 Bar Code Scanner

10.6

Installing and operating the "CLV-Setup" program

The scope of delivery of the CLV includes a CD-ROM ("Manuals & Software") containing the "CLV-Setup" program. The software can be installed on a PC with the following minimum configuration: 80486 processor, 66 MHz, and 16 MB RAM, CD drive, serial interface and mouse (recommended) as well as an operating system Windows 95TM /98TM, Windows NTTM, Windows 2000TM or Windows XPTM. Approximately 25 MB of hard disk space is required to install CLV-Setup (V 4.10) and CLV Assistant, approximately 20 MB to install CLV-Setup Help (V 4.10). Note

For the lastest version of the software, see www.sick.com. 10.6.1

Preparations

1.

Make sure you have the CD-ROM ("Manuals & Software") at hand.

2.

Connect the CLV to the CDB 420 or CDM 490 Connection Module (see Chapter 5.5.3 Connecting the supply voltage, Page 5-19).

3.

Connect the PC to the terminal interface on the CLV using a 3-core RS 232 data cable (null modem cable) (CDB 420 or CDM 490: connect the PC to the internal 9-pin "Aux" plug). See Chapter 5.5.6 Connecting the PC, Page 5-22.

4.

Switch on the power supply to the CDB 420 or CDM 490. Once the CLV has started, the "Device Ready" LED lights up.

5.

Switch on the PC and start Windows.

10.6.2

Installing the software

The "CLV-Setup" installation program creates a directory "CLV", a series of subdirectories, and the necessary links. A deinstaller is also installed automatically so that you can remove the program from the PC at any time. As of version V 2.6, CLV-Setup will only be available and further developed as a 32-bit application. Note

The following instructions describe how to install the software on Windows XPTM. These instructions may differ depending on the operating system you are using (for example, administrator rights are required to install the software on Windows XP). Initial installation User software CLV-Setup:

8 009 993/O824/20-12-2004

1.

Shut down all the applications running in Windows.

2.

Insert the CD-ROM into the CD-ROM drive.

3.

If the auto run function is enabled, the start page (table of contents) of the CD-ROM is automatically displayed in the your browser.

4.

Select the SOFTWARE folder on top left.

5.

Select CLV, ICR, OPS SETUP SOFTWARE in the listing. A table with CLV-Setup information is displayed.

6.

Select DOWNLOAD under SOFTWARE FILE.

7.

In the FILE DOWNLOAD dialog box select the option to install directly from the CD-ROM. Confirm with OK. The software is automatically saved in the “Programs\CLV“ directory on your hard disk. – or –

© SICK AG · Division Auto Ident · Germany · All rights reserved

10-43

Page 210

Chapter 10

Appendix

Operating Instructions CLV 490 Bar Code Scanner

Select the option to save the software locally on your local hard disk and confirm with OK. In the FILE DOWNLOAD dialog box select the desired target directory on your hard disk. The “CLVSetupxx.exe“ file is then saved at this location (xx = version number). Select OPEN in the dialog box to start the file. 8.

The installation program starts and guides you through the installation with screen messages. The program asks you for your user name and company name. This information appears as a header in printouts. “CLV-Setup“ and CLV Assistant are installed. The software will be entried in the Windows start menu under “Programs\CLVSetup“.

9.

Please read the Readme file, as it contains the latest information and problems regarding CLV-Setup.

10. Then confirm the final installation message by clicking “OK. CLV-Setup Help (Online): 1.

Return to the start page of the CD-ROM with the opened folder “Software“.

2.

Select CLV, CLV HELP SOFTWARE in the listing. A table with CLV Help information is displayed.

3.

Select DOWNLOAD under SOFTWARE FILE.

4.

In the FILE DOWNLOAD dialog box select the option to install directly from the CD-ROM. Confirm with OK. The software is automatically saved in the “Programs\CLV“ directory on your hard disk. – or – Select the option to save the software locally on your local hard disk and confirm with OK. In the FILE DOWNLOAD dialog box select the same target directory as selected for CLV-Setup on your hard disk. The “CLVHelp.exe“ file is then saved at this location.

5.

After downloading the software, select OPEN in the dialog box to start the software. The software is installed in the sub directory “Help“ of the main directory “CLV“. CLV Help will also be entried in the Windows start menu under “Programs\CLV-Setup“.

¾

Restart the PC. The Windows DLL files are possibly updated.

The “CLV-Setup“ program, the CLV Assistant and the online “CLV-Setup Help“ are installed and ready. The program logs all of the files that are installed to the “install.log“ file. This list is used by the uninstaller to remove CLV-Setup and must not be deleted if you want to use the uninstaller at some time in the future. Updating the program There are two ways of installing a new version of CLV-Setup:

10-44

•

You can install the new version and still keep the old version (parallel installation) – or –

•

You can install the new version over the old version (overwrite)

© SICK AG · Division Auto Ident · Germany · All rights reserved

8 009 993/O824/20-12-2004

Page 211

Operating Instructions

Appendix

Chapter 10

CLV 490 Bar Code Scanner

Installing the new version while still keeping the old one: If you want to install the new version and still keep the old version of CLV-Setup, follow the procedure described under Initial installation. When the program asks you for the target directory, you must specify a new directory. Both versions of the software are then available but must not be started simultaneously. The configuration files "*.scl " (scanner configuration), "*.npj" (network project), "*.opj" (OTS project) or "*.rpj" (redundant OTS project) of the old version can be used in the new version and contain the CLV parameter sets as well as project data if serveal scanners had been organized logically by projects. In order to do so, copy the configuration files from the "data" directory of the old version to the "data" directory of the new version. Installing the new version over the old one: Before you can install the new version of the software, you must remove the files of the old version – with the exception of the configuration files "*.scl " (scanner configuration), "*.npj" (network project), "*.opj" (OTS project) or "*.rpj" (redundant OTS project). They contain the CLV parameter sets as well as project data if serveal scanners had been organized logically by projects. The deinstaller of the old version removes the program files. You can use the deinstaller to remove all of the files (with the exception of the configuration files) or only selected files. The default setting is a complete deinstallation (the configuration files are not deleted). With the custom deinstallation, the files are listed and only selected files are deinstalled. If you choose this deinstallation method, make sure that the main directory containing the configuration files "*.scl" is not deleted. We recommend that you move these files to a different directory and copy them back to the "data" directory later. 1.

From the Windows Start menu, choose PROGRAMS and the deinstaller for CLV-Setup. The deinstaller starts and guides you step by step through the deinstallation procedure.

2.

Choose the deinstallation method (complete or custom).

3.

Install the new version of CLV-Setup as described under Initial installation, making sure to choose the same directory.

The new version of CLV-Setup is installed. The configuration files of the old version can be used again.

8 009 993/O824/20-12-2004

© SICK AG · Division Auto Ident · Germany · All rights reserved

10-45

Page 212

Chapter 10

Appendix

Operating Instructions CLV 490 Bar Code Scanner

10.6.3

Starting CLV-Setup

The "CLV-Setup" program is launched with the following default settings: Communication

COM 1, 9 600 bd, 8 data bits, 1 stop bit, no parity

Transmit control character

Start: STX, Stop: ETX

Receive control character

Start: STX, Stop: ETX

Unit of measure

Metric (depending on the selected installation)

Browser

User dependent

Language

As selected for the installation

Company name

As entered for the installation

User name

As entered for the installation

Automatic connection to CLV

On startup: yes On scanner selection: yes When the Terminal Emulator is started/closed: yes

Initial screen

Yes

Storage confirmation prompt

Yes

Last type selected

CLV 41x

File storage location

"data" (configuration files for CLV)

Table 10-7: Default settings in CLV-Setup

1.

Switch on the PC and start Windows.

2.

Choose CLV-SETUP from the Start menu. The SICK logo is then followed by the initial screen.

3.

Confirm the initial screen. CLV-Setup then checks whether a CLV is connected to the COM 1 port on the PC (default setting in CLV-Setup) and whether the communication parameters on the PC match those on the CLV. If this is the case, the CONNECTED status is displayed with the CLV specification in the status bar right in the bottom of the screen. The CLV type (here: CLV 490) is displayed in the DEVICE drop-down list in the top right of the toolbar.

The program then loads the internal description of the CLV as well as the default parameter values and displays them on the tabs. Finally, CLV-Setup uploads the current parameter set from the RAM of the CLV and displays the values on the tabs instead of the default values. You can edit the current parameter set on the tabs. The default values are displayed when the device is used for the first time. Troubleshooting If the program cannot establish communication, it outputs a timeout warning and displays NO CONNECTION in the status field. This may be because the CLV is not connected or because the communication parameters on the CLV do not match those on the PC. In this case, CLVSetup enters the CLV type of the device it last communicated with in the DEVICE field on the toolbar. The default setting the first time the device is used is the CLV 41x. The software then loads the internal device description for this CLV type and displays the default values on the tabs.

10-46

© SICK AG · Division Auto Ident · Germany · All rights reserved

8 009 993/O824/20-12-2004

Page 213

Operating Instructions

Appendix

Chapter 10

CLV 490 Bar Code Scanner

1.

Connect the PC to the terminal interface on the CLV using a 3-core RS 232 data cable (null modem cable) (CDB 420 or CDM 490: connect the PC to the internal 9-pin "Aux" plug). See Chapter 5.5.6 Connecting the PC, Page 5-22.

2.

Click (AutoBaud detect) in the toolbar or choose OPTIONS, AUTOBAUD DETECT. CLV-Setup scans the serial interface by varying the communication parameters and sends a telegram to the CLV repeatedly. As soon as a response is registered from the CLV, CLV-Setup signals the detected communication parameters. Fig. 10-36 shows an example of the results of the AutoBaud detect function. The CONNECTED status is displayed right in the status bar in the bottom of the screen.

3.

Confirm the AUTO DETECT dialog box with OK. CLV-Setup displays the detected CLV type in a separate dialog box and asks you whether you want to upload the current parameter set from the CLV.

4.

Confirm the dialog box with Yes. CLV-Setup uploads the current parameter set from the RAM of the CLV to its database and displays the values on the tabs.

You can edit the current parameter set on the tabs.

Fig. 10-36: CLV-Setup: Result display of the AutoBaud Detect function

– or – 2.

Choose OPTIONS, SERIAL INTERFACE from the menu bar. CLV-Setup displays the current communication parameter settings on the PC in the COM PARAMETERS dialog box.

3.

Make sure that the communication parameters on the PC and CLV are identical (connected COM port, 9 600 bd, 8 data bits, 1 stop bit, no parity)

4.

Confirm the dialog box with OK. CLV-Setup attempts to communicate with the CLV again. If it is successful, it displays the CONNECTED status right in the status bar in the bottom of the screen.

5.

Click in the toolbar. CLV-Setup then uploads the current parameter set from the RAM of the CLV to its database and displays the values on the tabs.

You can edit the current parameter set on the tabs.

8 009 993/O824/20-12-2004

© SICK AG · Division Auto Ident · Germany · All rights reserved

10-47

Page 214

Chapter 10

Appendix

Operating Instructions CLV 490 Bar Code Scanner

10.6.4

User interface

The graphical user interface (GUI) of the CLV-Setup is largely self explanatory. The online CLV-Setup Help function provides a description of how to use the program under HELP TO CLV-SETUP in the help navigation tree. The GUI is shown in Fig. 10-37. The program window of the GUI comprises the following elements: •

Title bar that displays the program name, current configuration file, and its status (e. g. "No File" )

•

Menu bar with pull-down menus

•

Toolbar with buttons for triggering various functions

•

Drop-down list (top right) for selecting the device type

•

Frame for displaying the navigation tree of the tabs (on left-hand side)

•

Frame for displaying the tabs for CLV 490 (Reading Configuration, Device Configuration, etc.) (on right-hand side). The numbers of displayed tabs depends on the selected items in the navigation tree on the left side. The parameters on the tabs are grouped according to their function. Some of these parameters open further dialog boxes

•

Status bar (bottom) with two display fields for the communication between CLV-Setup and the CLV, the PC’s interface parameter display, error display field (system errors), device specification field and status display for the connection to the CLV.

Fig. 10-37: User interface of the "CLV-Setup" software

10-48

© SICK AG · Division Auto Ident · Germany · All rights reserved

8 009 993/O824/20-12-2004

Page 215

Operating Instructions

Appendix

Chapter 10

CLV 490 Bar Code Scanner

10.6.5

Functions

The "CLV-Setup" program (from version V 4.0) features the following functions to carry out: •

Uploading the default parameter set from the database of CLV-Setup

•

Automatic communication attempt with the CLV when the program is started

•

Automatic software compatibility check in the CLV

•

Uploading and displaying parameter set stored in the CLV

•

Changing the parameter/parameter values on the tabs

•

Downloading the changes to the parameter set in the CLV

•

Saving the changes to the parameter set as a configuration file "*.scl"

•

CLV Assistant for simplified parameterization (not for CLV 490)

•

Printing the configuration files

•

Exporting the configuration files in RTF format

•

Exporting the command strings contained in the profile bar codes in ASCII format

•

Printing the parameter set in the form of profile bar codes (not for CLV 490)

•

Online access to the CLV via a Terminal Emulator (select operating mode, start operating functions, log file for communication between the CLV and CLV-Setup, extended Terminal mode, e. g. for displaying CLV messages)

•

AutoBaud Detect for automatic communication setup with the CLV

•

Selecting the COM port of the PC for adapting the data transfer parameters

•

Monitoring the data communication of the CLV host interface

•

Selecting language setting

•

Selecting units of measure

•

Displaying the last 10 configuration files that were processed

•

Saving the CLV type that was selected last

•

Automatic storage confirmation prompt when changes are made to the configuration file

•

Wizards for individual functions

•

Clearing the optional CMC 400 parameter memory in the CDB/CDM Connection Module

•

CLV 490 only: Aligment check to calculate the required setting for angles

•

CLV 490 only: Background analysis

•

CLV 490 only: I/O monitoring

•

Setting switches for starting the program (in the "CLVmain.ini" file)

•

Hot keys [F keys] for important functions (using the program without a mouse)

•

Starting program with SCL file as argument

•

Macro player for recording and playing a macro file

•

Context-sensitive help via the [F1] key

10.6.6

8 009 993/O824/20-12-2004

Hot keys

[F1]

Start "CLV-Setup Help“ online

[F2]

Open dialog box to execute device functions

[F3]

Upload complete set of parameters from CLV

[F4]

Download complete set of parameters to CLV

[F5]

Open configuration file

[F6]

Save configuration file

© SICK AG · Division Auto Ident · Germany · All rights reserved

10-49

Page 216

Chapter 10

Appendix

Operating Instructions CLV 490 Bar Code Scanner

[F7]

Load default setting of CLV from CLV-Setup database

[F8]

Start AutoBaud Detect

[F9]

Load operating data from CLV to be displayed and reset

[F10] Open Terminal Emulator 10.6.7

Opening and closing tabs

After starting the user interface, CLV-Setup displays those tabs in the frame on right-hand side whose names are framed in the navigation tree on left-hand side. The first tab covers the other tabs in the sequence but their marks are still visible on the top of the tabs. If you click some buttons on a tab in the fore-ground, several sub dialogs will also be opened as tabs and added to the first tab. In the frame, all tabs on the right will then be shifted more to the right side. To scroll the tabs, use the sliding control elements at the left corner on top of the frame which are displayed by CLV-Setup. You can open and close the tabs in the following way: In the navigation tree: •

Opening: Double-click the desired entry or put the mouse cursor on the entry and select “Show dialog“ in the context menu using the right mouse button. To display several entries under a branch, click the corresponding “+“ symbol. To hide displayed sub entries, click the corresponding “–“ symbol.

•

Closing: Put the mouse cursor on the desired entry and select “Hide dialog“ in the context menu using the right mouse button.

In the frame for displaying tabs: •

Opening: Click on the desired mark on the top of the tab. If neccessary, scroll the tabs.

•

Closing: To close the tab in the fore-ground, click on the red cross at the left corner on the top.

Pressing the left mouse button, the left frame with the navigation tree can be move from the program window to a seperate position on the screen. If the navigation tree has been closed, it can be reopened using the VIEW menu. 10.6.8

Online help – CLV-Setup Help

The procedure for setting the CLV parameters is supported by the online help function, which is displayed in an HTML browser, such as Internet ExplorerTM. You can call up contextsensitive help on each parameter as well as a description of its function.

10-50

1.

Press the [F1] key. The browser is launched automatically and displays the help topic.

2.

If the program cannot find a browser, it asks you to specify the storage location on the hard disk. Enter the name of the executable file and the path in the dialog box.

3.

Use the [ALT]+[TAB] keys to switch between the CLV-Setup and CLV-Setup Help applications and prevent several browser windows from being opened.

4.

To display an overview of the Help function, Choose HELP, CONTENTS from the CLV-Setup menu bar.

© SICK AG · Division Auto Ident · Germany · All rights reserved

8 009 993/O824/20-12-2004