For the Sick CLV490

Page 197

Appendix

Operating Instructions

Chapter 10

CLV 490 Bar Code Scanner ang le α Ape rture

Ap ngle re a u t r e

α=

=4 0°

40°

Dimensions in mm (in)

Reading distance

Reading conditions: see Table 10-1, Page 10-1

Characteristics field scanning frequency/reading distance: see Fig. 10-30, Page 10-32

–15° to +15° (typical values!)

–30° to +30°

Tilt

–45° to +45°

Low density: Reading performance data of line scanner with oscillating mirror

Reading field height

Line scanner with oscillating mirror (side reading window): CLV 490-7010/-7011

Resolution: 0.5 mm (19.7 mil)

10.2.8

Fig. 10-29: CLV 490-7010/-7011 (Low density): Reading field height as a function of the reading distance and tilt at a resolution of 0.5 mm (19.7 mil)

8 009 993/O824/20-12-2004

© SICK AG · Division Auto Ident · Germany · All rights reserved

10-31

Page 198

Appendix

Chapter 10

Operating Instructions CLV 490 Bar Code Scanner

Reading distance at a resolution of 0.5 mm (19.7 mil)/aperature angle 40° see Fig. 10-26, Page 10-28 Reading distance at a resolution of 0.5 mm (19.7 mil)/aperature angle 50° see Fig. 10-27, Page 10-29

Line scanner with oscillating mirror (side reading window): CLV 490-7010/-7011 Dimensions in mm (in)

Scanning frequency (Hz)

Radial reading distance Resolution: 0.50 mm (19.7 mil)

Reading conditions: see Table 10-1, Page 10-1

Fig. 10-30: Characteristics field CLV 490-7010/-7011 (Low density): Scanning frequency as a function of the radial reading distance and resolution

10-32

© SICK AG · Division Auto Ident · Germany · All rights reserved

8 009 993/O824/20-12-2004

Page 199

Appendix

Operating Instructions

Chapter 10

CLV 490 Bar Code Scanner

Line scanner with oscillating mirror (side reading window): CLV 490-7010/-7011

Deflection range

Dimensions in mm (in)

Defle a ction 40° max. ngle

Radial reading distance Tilt:

Resolution:

0.50 mm (19.7 mil)

–45° to +45° –30° to +30° –15° to +15°

Reading conditions: see Table 10-1, Page 10-1

Fig. 10-31: CLV 490-7010/-7011 (Low density): Deflection range as a function of radial reading distance, deflection angle and tilt at a resolution of 0.5 mm (19.7 mil)

8 009 993/O824/20-12-2004

© SICK AG · Division Auto Ident · Germany · All rights reserved

10-33

Page 200

Appendix

Chapter 10

Operating Instructions CLV 490 Bar Code Scanner

10.3

Installing and operating the external parameter memory (connector cover)

The external parameter memory is available as an accessory. It is located in a connector cover and comprises a permanent, rewriteable memory (EEPROM) with 4 kByte from CLV firmware V 2.0. The connector cover covers the two connections of the CLV in the enclosure rating IP 65. The cover is shown in Fig. 10-32. Depending on the type, two (one) connection cable(s) that are secured by means of conduit thread connections supply the signals of the CLV with identical pin assignment or limited pin assignment.

22 (0.86) 88.5 (3.48) 22 (0.86)

27 (1.06) 50 (1.79)

27 (1.06)

Fig. 10-32: External parameter memory, installed on the CLV

Note

If the CLV 490 is used to replace a CLV 265/CLV 295 equipped with a connector cover (parameter memory), the CLV 490 cannot store its extensive parameter set in this external parameter memory (too small). The parameter memory in the connector cover is available in different implementations:

Order no.

Design

CLV

Length

2 030 023

Connector cover, one connection cable with 15-pin D Sub HD plug. Pin assignment see Chapter 5.3 Connector pin assignment, Page 5-7.

Without heater

3m CLV to CDB 420 (9.84 ft)

2 033 126

Connector cover, one connection cable with 15-pin D Sub HD Without heater plug, one power supply cable with open end. Suitable for track chain use. Pin assignment see Chapter 5.3 Connector pin assignment, Page 5-7.

10 m CLV to CDB 420 (32.8 ft)

2 033 127 as 2 033 126, but with longer cables

Without heater

10 m CLV to CDB 420 (32.8 ft)

2 020 307

Connector cover, two connection cables with 15-pin D Sub HD Without heater connections (1 x plug/1x socket). Pin assignment see Chapter 5.3 Connector pin assignment, Page 5-7.

3m CLV to CDM 490 (9.84 ft)

2 020 981

Connector cover, two connection cables with open ends. Pin assignment see Chapter 5.3 Connector pin assignment, Page 5-7.

2 030 065

Connector cover, two connection cables with 15-pin D Sub HD With heater connections (1 x plug/1x socket) and one power supply cable with open end. Suitable for track chain use. Pin assignment see Chapter 5.3 Connector pin assignment, Page 5-7.

Without heater

Connection

3m CLV to non-Sick Power (9.84 ft) pack, free wiring 3m CLV to CDM 490 (9.84 ft)

Table 10-4: Types of the external parameter memory in the connector cover

10-34

© SICK AG · Division Auto Ident · Germany · All rights reserved

8 009 993/O824/20-12-2004

Page 201

Appendix

Operating Instructions

Chapter 10

CLV 490 Bar Code Scanner

Order no.

Design

CLV

Length

Connection

2 031 034 as 2 030 065, but with longer cables

With heater

10 m CLV to CDM 490 (32.8 ft)

2 027 543

Connector cover, two connection cables ending on a second connector cover (with wires for CAN interface1 ). Pin assignment see Chapter 5.3 Connector pin assignment, Page 5-7.

With heater

3m CLV to AMV 100/200 (9.84 ft)

Table 10-4: Types of the external parameter memory in the connector cover (contd.)

10.3.1

Function of the external parameter memory

When it is started, the CLV is initialized with the data stored in its internal parameter set (EEPROM). If this parameter set has been copied to an external memory, the device can be replaced easily. The replacement device, therefore, does not have to be configured manually. Instead, it accesses the external parameter memory automatically (assuming that the correct start option has been selected). It then imports the parameter set to its RAM. This option ("Start with external parameters...") is set by default. The "Device Ready" and "Read Result" LEDs indicate whether the external memory was accessed successfully. The CLV also outputs messages for troubleshooting purposes on the terminal interface. Note

If the CLV and an external parameter memory in the connector cover are connected together to a CDB 420 or CDM 490 Connection Module containing also an external parameter memory provided by the CMC 400, the CLV will detected both external parameter memories. Before starting the reading mode, the CLV will operated as followed: •

If the memory of the CMC 400 comprises a parameter set compatible with the CLV, the CLV loads this parameter set into the RAM and into its internal EEPROM. The "Device Ready" LED of the CLV blinks 20 times and then lights up constantly. Then the CLV clears the parameter memory in the connector cover. In CLV-Setup, the external parameter memory in the connector cover is not longer available.

•

If the memory of the CMC 400 comprises a parameter set not compatible with the CLV, the CLV loads the internal parameter set from its in EEPROM into the RAM. The CLV clears the parameter memory in the connector cover. In CLV-Setup, the external parameter memory in the connector cover is not longer available.

•

If the memory of the CMC 400 is empty, the CLV copies the internal parameter set from its EEPROM via the RAM to the CMC 400 if write protection is disabled ("Write Protection" switch of the CMC 400 in top position). The "Device Ready" of the CLV blinks 20 times alternately with the "Read Result" LED, then the "Device Ready" LED ligths up constantly. If write protection for the CMC 400 is enabled, the CLV loads the internal parameter set from its in EEPROM into the RAM. The "Device Ready" and "Read Result" blink constantly together. In both cases, the CLV clears the parameter memory in the connector cover. In CLV-Setup, the external parameter memory in the connector cover is not longer available.

The CLV supports the CMC 400 from firmware V 3.30 on. If necessary, the memory of the CMC 400 can be manually deleted. To do so using CLV-Setup, choose CLEAR CMC 400 under CLV 490 in the menu bar.

8 009 993/O824/20-12-2004

© SICK AG · Division Auto Ident · Germany · All rights reserved

10-35

Page 202

Chapter 10

Appendix

Operating Instructions CLV 490 Bar Code Scanner

The application of the CMC 400, the changed function of the "Device Ready" and "Read Result" LEDs of the CLV as well as the remedy for errror messages describes the CMC 400 Operating Instructions (order no. 8 010 002, German/English). 10.3.2

Installation and electrical connection

The pin assignments/wire colors for the connection cables of the parameter memory are shown in the following tables: –

Parameter memory no. 2 030 023: Table 5-5, Page 5-8

–

Parameter memory no. 2 020 307: Table 5-6 and Table 5-7, Page 5-9

–

Parameter memory no. 2 033 126/no. 2 033 127: Table 5-8 and Table 5-9, Page 5-10

–

Parameter memory no. 2 030 065/no. 2 031 034: Table 5-10 and Table 5-11, Page 5-11

–

Parameter memory no. 2 027 543: Table 5-13 and Table 5-14, Page 5-13

–

Parameter memory no. 2 020 981: Table 5-17 and Table 5-18, Page 5-15

–

Parameter memory no. 2 021 267: Table 5-19 and Table 5-20, Page 5-16

¾

Mount the parameter memory on the CLV as shown in Chapter 5.5.3 Connecting the supply voltage, Page 5-19 and connect the cables to the CBD 420 or CDM 490 Connection Module or to the non-SICK power supply.

10.3.3

Operation

The parameter memory in the connector cover connected to the CLV is accessed • using the selected CLV start option

• after the parameters have been downloaded to the CLV with the permanent storage function and the START WITH EXTERNAL PARAMETERS option

The start option is set at the bottom right of the DEVICE CONFIGURATION tab as shown in Fig. 10-32 (grey frame). The START WITH EXTERNAL PARAMETERS option is selected by default.

Fig. 10-33: CLV-Setup: "Device configuration" tab with the CLV start options

10-36

© SICK AG · Division Auto Ident · Germany · All rights reserved

8 009 993/O824/20-12-2004

Page 203

Operating Instructions

Appendix

Chapter 10

CLV 490 Bar Code Scanner

Start with internal parameters When it starts, the CLV loads the internal parameter set to its RAM. The "Device Ready" LED lights up constantly. Since the data is stored permanently, it is not lost when the device is switched off. Start with external parameters (default setting) When it starts, the CLV loads the external parameter set to its RAM. The "Device Ready" LED blinks for approx. 10 s and then lights up constantly. The data in the CLV is lost when the device is switched off. Start with copy internal -> external When it starts, the CLV loads the internal parameter set to its RAM. It then copies the parameter set to the external memory in the connector cover and resets the start option to "Start with External Parameters". The two "Device Ready" and "Read Result" LEDs blink alternately for approx. 10 s. The "Device Ready" LED then lights up constantly. The data is not lost when the device is switched off since it is stored permanently in the CLV and is identical to the internal parameter set. 10.3.4

Switching on the device for the first time

Since the external parameter memory in the connector cover is empty (e. g. on delivery) after the CLV has started, the CLV copies its current parameter set to the external parameter memory if no CMC 400 parameter memory module is available in the CDB 420 or CDM 490 Connection Module. Then the "Device Ready“ LED lights up. The CLV is ready for operation. If an external parameter memory containing a corresponding parameter set is connected to the CLV, the "Device Ready" LED blinks for 10 s when the device starts up correctly and then lights up constantly. 10.3.5

Adjusting the parameter set in the external parameter memory (connector cover) after it has been downloaded to the CLV

If the CLV is started with the START WITH EXTERNAL PARAMETERS option with an external parameter memory in the connector cover and if the parameter set is downloaded from CLVSetup to the CLV with the permanent storage option, CLV-Setup asks you whether the new parameter set is also to be copied to the external parameter memory (Fig. 10-34, Page 10-38). This function prevents the CLV from reloading the old parameter set the next time it starts.

8 009 993/O824/20-12-2004

© SICK AG · Division Auto Ident · Germany · All rights reserved

10-37

Page 204

Chapter 10

Appendix

Operating Instructions CLV 490 Bar Code Scanner

Fig. 10-34: CLV-Setup: dialog box for adjusting the external parameter memory

¾

Adjust the parameter set in the external parameter memory as described in Chapter 6.4.1 Configuring the CLV via the user interface of CLV-Setup, Page 6-5.

10.3.6

Meaning of the LEDs

The function of the two "Device Ready" and "Read Result" LEDs, as defined by the selected start option, is shown in Table 3-3, Page 3-12. Their function in the event of an error accessing the external memory in the connector cover as well as troubleshooting measures are described in Table 8-2, Page 8-3. 10.3.7

Error messages

If errors occur accessing the external parameter memory in the connector cover, the CLV outputs error messages on the terminal interface. When the CLV is started, these messages can be displayed using the "CLV-Setup" program. 1.

Connect the PC to the terminal interface on the CLV using a 3-core RS 232 data cable (null modem cable) (CDB 420 or CDM 490: connect the PC to the internal 9-pin "Aux" plug). See Chapter 5.5.6 Connecting the PC, Page 5-22.

2.

Switch on the PC and start Windows.

3.

Start the "CLV-Setup" program and open the Terminal Emulator.

4.

Switch on the power supply for the CDB 420 or CDM 490. The CLV starts and loads the parameter set from the external memory.

5.

Monitor the messages in the Terminal Emulator.

Table 8-3, Page 8-5 lists the error messages with the appropriate troubleshooting measures.

10-38

© SICK AG · Division Auto Ident · Germany · All rights reserved

8 009 993/O824/20-12-2004

Page 205

Operating Instructions

Appendix

Chapter 10

CLV 490 Bar Code Scanner

10.3.8

Replacing a CLV

The device is replaced by starting the new CLV with the START WITH EXTERNAL PARAMETERS option (default). This is the case when the device is put into service for the first time, e. g. when it is taken from the warehouse. The device must be replaced with a device of the same type. You do not need to connect a PC to replace the device. ¾

Replace the device as described in Chapter 10.12.2 Importing the parameter set from the external memory, Page 10-74.

The CLV imports the external parameter set into the RAM. This is sufficient to ensure that the device operates correctly. If the parameter set is to be stored permanently in the CLV (EEPROM), a PC with the "CLV-Setup" program must be connected to upload the parameter set from the CLV to CLV-Setup and then download it again to the CLV with the PERMANENT storage option. When asked to do so, adjust the external memory again.

8 009 993/O824/20-12-2004

© SICK AG · Division Auto Ident · Germany · All rights reserved

10-39

Page 206

Chapter 10

Appendix

Operating Instructions CLV 490 Bar Code Scanner

10.4

Optional heating

10.4.1

Features

•

Integrated heater (permanently installed)

•

Wider range of applications for CLV: up to max. –35 °C (–31 °F)

•

Supply voltage 24 V DC +20 %/–10 %

•

CLV enabled via internal temperature switch (power-up delay approx. 35 to 40 min. at 24 V DC and min. ambient temperature of –35 °C (–31 °F))

•

Required cable cross-section (power supply): min. 0.75 mm2 (approx. 20 AWG)

10.4.2

Design

The heating system comprises two parts: • the front window heater

• the housing heater

The optional heating system is installed and tested at the factory. The user cannot install it on site. 10.4.3

Function

When the 24 V DC supply voltage is applied, a temperature switch first supplies electric current to the heater only. This switch enables the supply voltage for the CLV when the internal temperature rises above 7 °C (+44.6 °F) during the warm-up phase (approx. 35 min). The device then performs a self-test and switches to Reading mode. The "Device Ready" LED lights up to indicate that the device is ready. When the internal temperature has reached approx. 25 °C (+77 °F), a further temperature switch deactivates the housing heater and reactivates it if necessary. Reading mode is not interrupted as a result. The front window heater remains active constantly. The temperature curve inside the housing is shown in Fig. 10-35. Temperature inside housing [°C] +40

104 [°F]

+30

86

+20

68

+10

50

0

32

-10

14

-20

-4

-30

-22

-40

-40

The temperature enables the operating voltage for the CLV internally

Time (min) 40

80

Fig. 10-35: CLV with heater: temperature curve inside the housing

10-40

© SICK AG · Division Auto Ident · Germany · All rights reserved

8 009 993/O824/20-12-2004