Operating Instructions: CLV 41x Bar Code Scanner, Standard line: 6 Operation, pages 63 to 70

Pages 63 to 70 of 142, from Operating Instructions: CLV 41x Bar Code Scanner, Standard line

The original document is the PDF (2.2 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.

For the Sick CLV410, Sick CLV412 and Sick CLV414

Page 63

Operating Instructions

Operation

Chapter 6

CLV 41x bar code scanner

Action

Settings in CLV Setup

Set up reading pulse: Continous Read

Ö DEVICE CONFIGURATION tab Ö READING TRIGGER PARAMETERS button Start:

Ö READING TRIGGER SOURCE section – Continous Read Stop: Ö END OF READING INTERVALL section – Generated by trigger source or – Timer: Timeout (referred to the start of the reading pulse) Note: No laser timeout is available in this pulse mode (the laser diode is not switched off if the reading interval has not be terminated). Table 6-3: Guide: parameterizing reading pulse source and read result output time (contd.)

c) Parameterizing laser timeout Action

Settings

1. Select the duration for the laser timeout

Ö DEVICE CONFIGURATION tab Ö READING TRIGGER PARAMETERS button Ö LASER TIMEOUT section

– or –

– Duration 2. Deactivate the laser timeout (laser diode is always active when pulsing, regardless of the pulse duration)

Ö LASER TIMEOUT section – Click (deactivate) the ACTIVE checkbox

Table 6-4: Guide: parameterizing laser timeout d) Parameterizing evaluation characteristics •

Activate code types for evaluation Ö CODE CONFIGURATION tab Ö SYMBOLOGIES section and EDIT buttons

•

Activate code comparison Ö DEVICE CONFIGURATION tab Ö CODE COMPARISON section

•

Mode for TEACHING IN match code 1 via the "Sensor 2" switching input Ö DEVICE CONFIGURATION tab Ö TEACH-IN section

Separation of identical bar codes (same code type/identical contents) No. of bar codes per object

Line/raster scanner

1. Number n = 1

Ö CODE CONFIGURATION tab Ö CODE POSITION section – Deactivate COMPARE!

2. Number n > 1: – Same code type – Contents different or identical

Ö CODE POSITION section – Activate COMPARE!

Ö NUMBER OF CODES section – MINIMUM – MAXIMUM

3. Number n > 1: – Code type different – Contents different or identical

Ö CODE POSITION section – Deactivate COMPARE!

Ö NUMBER OF CODES section – MINIMUM – MAXIMUM

Table 6-5: Guide: settings to be made for the evaluation of identical bar codes

8 009 974/O079/01-09-2004

© SICK AG · Division Auto Ident · Germany · All rights reserved

6-9

Page 64

Chapter 6

Operation

Operating Instructions CLV 41x bar code scanner e) Parameterizing output characteristics Result status: •

Define function of result status output of "Result 1" to "Result 3" switching outputs Ö DEVICE CONFIGURATION tab Ö RESULT OUTPUT PARAMETERS button

•

Define function of result status output and volume of beeper Ö DEVICE CONFIGURATION tab Ö RESULT OUTPUT section and READING OUTPUT PARAMETERS button

Host interface: •

Arrangement in the data network Ö DEVICE CONFIGURATION tab Ö ARRANGEMENT section

•

Communication parameters Ö HOST INTERFACE tab Ö DATA FORMAT section

•

Protocol Ö HOST INTERFACE tab Ö INTERFACE PROTOCOL section

•

Sort reading result Ö DATA STRINGS tab Ö OUTPUT SEQUENCE SORT section

•

Masking of the reading result Ö DATA STRINGS tab Ö FORMAT MASK section

Data output string:

6-10

•

Select contents of Header, Separator and Terminator blocks Ö DATA STRINGS tab Ö OUTPUT FORMAT section

•

Position of separator in data output string Ö DATA STRINGS tab Ö SEPARATOR section

•

Set wrong read format Ö DATA STRINGS tab Ö WRONG READ FORMAT section and ERROR STRING section

•

If necessary, parameterize/activate test string Ö DATA STRINGS tab Ö TEST STRING section

© SICK AG · Division Auto Ident · Germany · All rights reserved

8 009 974/O079/01-09-2004

Page 65

Operating Instructions

Operation

Chapter 6

CLV 41x bar code scanner

6.4.4

Configuring the CLV with AutoSetup

AutoSetup enables the reading and code configurations of the CLV to be adapted automatically to the application at hand. The following conditions must be fulfilled with regard to this: • one bar code (code type, code length)

• fixed reading distance

• conveyor belt off (static reading)

The CLV determines the code type, code length and the most favorable scanning rate for the bar code presented in the maximum reading distance of the application. This takes place in Free Running mode. Once AutoSetup has been completed successfully, the CLV copies the parameter values determined to its parameter set. All other code types and code lengths are then disabled. If AutoSetup was not completed successfully, the parameter set is not changed. The CLV does not read a Pharmacode or any EAN/UPC types in AutoSetup. It does not respond to an external reading pulse. AutoSetup can be activated in three different ways in the CLV: • via the user interface of CLV Setup

• by presenting profile bar code no. 10 (on the card supplied no. 8 008 085) to the optical interface (reading window). See Chapter 10.4.1 Activating AutoSetup with a profile bar code, Page 10-16.

• via a command string (via the serial interface). See Chapter 10.5 Configuring the CLV with command strings, Page 10-22.

While AutoSetup is running, the CLV does not output any read result. Once AutoSetup has been completed, no error messages are output for diagnosed malfunctions in the event of a no read. Connecting a PC and using CLV Setup allows you to check the reading quality of the taught-in bar code in the Percentage Evaluation mode. Explanation

AutoSetup procedure 1.

Starting at 800 Hz, the CLV decreases its scanning frequency to 200 Hz in steps of 100 Hz. It evaluates the reading quality of 100 scans for each step.

2.

The CLV performs verification reads +/– 50 Hz around the scanning frequency with the highest reading quality.

3.

It then copies the optimum scanning frequency, the code type and code length to its parameter set.

Activating AutoSetup via the user interface of CLV Setup In CLV Setup, AutoSetup can be started via the View menu, via the terminal emulator or as an executable device function (interactively). Preparing AutoSetup:

8 009 974/O079/01-09-2004

1.

Switch on the power supply of the CDB 410 or CDM 410 connection module (of the CLV). After the CLV has started, it outputs a tone to indicate that the self-test was successful. After a short while, it outputs two consecutive tones to indicate that it has switched to reading mode. The "Device Ready" LED lights up.

2.

Upload the current parameter set from the CLV with CLV Setup (the parameter set initially contains the factory default settings). CLV Setup displays the current values on the tabs.

© SICK AG · Division Auto Ident · Germany · All rights reserved

6-11

Page 66

Operation

Chapter 6

Operating Instructions CLV 41x bar code scanner

3.

AutoSetup overwrites important values in the current parameter set of the CLV. Save the existing parameter set as a configuration file "*.scl" in CLV Setup before you change the parameter settings. See Chapter 6.4.1 Configuring CLV with the user interface of CLV Setup, Page 6-4.

Performing AutoSetup: 1.

Click in the toolbar. The terminal emulator is launched. The CLV is in reading mode.

2.

Present the application-specific bar code to the CLV at the maximum reading distance for the application. Note the limits of the CLV here. See reading field diagrams in Chapter 10.2 Specification diagrams, Page 10-1. To prevent surface reflections, rotate the bar code towards the CLV approximately 15° away from the vertical axis. See Fig. 4-7, Page 4-5.

3.

Click AUTO SETUP under DEVICE MODE. The CLV cancels reading mode. The "Device Ready" LED is extinguished. The CLV then starts AutoSetup. The red scan line appears when the "Laser On" LED lights up.

4.

The CLV confirms that AutoSetup has been completed successfully by outputting two consecutive tones via the beeper and briefly activating the "Result" LED. To begin with, it only overwrites the appropriate parameter values in the RAM. The CLV returns to reading mode. The "Device Ready" LED lights up. The CLV only evaluates the presented code type and length the next time a bar code is read.

5.

Perform an upload from the CLV. This is done by clicking in the toolbar. CLV loads the current parameter set from the RAM of the CLV and displays the new values in the tabs.

6.

Perform a download to the CLV. This is done by clicking The DOWNLOAD PARAMETER dialog box is displayed.

7.

Confirm the dialog box by selecting the PERMANENT save option. The CLV saves the changes made to the parameter set permanently in the EEPROM.

8.

Close the terminal emulator.

in the toolbar.

AutoSetup canceled by the CLV in the event of a no read: If the CLV was not able to execute AutoSetup successfully after approx. 35 s, it cancels the search mode after the last decrease made to the parameter values by outputting three consecutive tones via the beeper. Then the CLV starts AutoSetup again with each wrong read. ¾

Note

Close the terminal emulator and open it again. The CLV returns to reading mode. The "Device Ready" LED lights up. The current parameter set in the CLV does not change.

If AutoSetup has been interactively started via DEVICE FUNCTION, the CLV automatically cancels the procedure after a few seconds. Terminating AutoSetup prematurely: ¾

6-12

To abort AutoSetup when once started via the terminal emulator proceed as describe above.

© SICK AG · Division Auto Ident · Germany · All rights reserved

8 009 974/O079/01-09-2004

Page 67

Operation

Operating Instructions

Chapter 6

CLV 41x bar code scanner

Running AutoSetup again: 1.

Click READING MODE under DEVICE MODE. The CLV switches to reading mode.

2.

Click AUTO SETUP under DEVICE MODE again. The CLV then starts AutoSetup.

3.

If AutoSetup is unsuccessful again, check whether the CLV is able to read the bar code correctly (reading distance, tilt, more than one bar code in the reading field).

Checking parameter values stored in the CLV (after parameters are saved):

Note

1.

Switch the power supply of the connection module (of the CLV) off and on again.

2.

Perform an upload from the CLV. This is done by clicking

3.

Check the values in the READING CONFIGURATION and CODE CONFIGURATION tabs. Only one code type may be active on the CODE CONFIGURATION tab.

in the toolbar.

The CLV does not enter a real value in the parameter set for the parameters "Minimum Bar Width" and "Minimum Reading Distance“, as it is not able to determine any absolute dimensions. This is why the value displayed on the READING CONFIGURATION tab does not match the actual conditions. Checking the stored read and code configurations: Once AutoSetup has been completed successfully, check the CLV settings as follows:

8 009 974/O079/01-09-2004

1.

Click in the toolbar. The Terminal Emulator is launched. The CLV is in reading mode.

2.

Click PERCENT EVALUATION under DEVICE MODE. The dialog box for the distance configuration is then displayed.

3.

Confirm the dialog box with "OK". The CLV starts the percentage evaluation and displays the reading quality of each set of 100 scans continuously. See Chapter 6.5.2 Percentage evaluation, Page 6-17.

4.

Present the application-specific bar code at the distance taught in (at the minimum and maximum reading distance if the bar code in the application is in a variable reading range).

5.

Check the reading quality in the terminal emulator. This should be between 70 and 100 %.

© SICK AG · Division Auto Ident · Germany · All rights reserved

6-13

Page 68

Chapter 6

Operation

Operating Instructions CLV 41x bar code scanner

6.5

Operating modes and outputting the reading result

The following CLV operating modes/functions can be selected in CLV Setup: Standard operating mode •

Reading mode

For startup •

Percentage evaluation

•

Read Rate Evaluation

For adapting device to application •

AutoSetup (automatic adaptation to reading conditions): See Chapter 6.4.4 Configuring the CLV with AutoSetup, Page 6-11.

•

Profile programming (configuration): See Chapter 10.4.2 Profile programming, Page 10-18.

•

Parameterization (configuration): See Chapter 6.4.1 Configuring CLV with the user interface of CLV Setup, Page 6-4.

•

Teach-in match code 1/activate match code comparison with "Sensor 2" switching input. See Chapter 10.8.1 Triggering the teach-in match code 1 via the "Sensor 2" switching input, Page 10-26.

For monitoring purposes/correcting faults •

Displaying and editing operating data

•

Self-test

6.5.1

Reading mode (standard operating mode)

The CLV performs a self-test after it has been switched on. It is then set to the reading mode and confirms this by outputting two consecutive tones via the beeper. In the default setting, the reflector polling is the trigger source of the internal reading interval. The reading result is output by the CLV at the end of the reading interval via both host interfaces. Depending on the configuration, the switching outputs "Result 1" to "Result 3" become live for the predefined pulse length when defined events occur during the reading process (e. g. Good Read). The reading result of the host interface 2 (RS 232) can be displayed in the CLV Setup terminal emulator. The structure of the represented reading result correspondents to the data output format defined on the DATA STRINGS tab. The READING MODE can be called up by choosing VIEW in the menu bar or via the terminal emulator. Displaying the reading result in the terminal emulator:

6-14

1.

Click in the toolbar. The TERMINAL EMULATOR dialog box is then displayed.

2.

Click READING MODE under DEVICE MODE.

3.

Align the CLV to the provided white polling reflector (special code) with a reading distance of approx. 300 mm (11.82 in) (not for CLV 414).

4.

Start the reading pulse: block the polling reflector, e. g. with a sheet of paper. The "Laser On" LED lights up. The CLV switches the laser diode on and the red scan line appears.

© SICK AG · Division Auto Ident · Germany · All rights reserved

8 009 974/O079/01-09-2004

Page 69

Operation

Operating Instructions

Chapter 6

CLV 41x bar code scanner

5.

Present the bar code pattern from Fig. 6-1, Page 6-3 to the CLV at the reading distance specified in Table 6-2, Page 6-3.

6.

End the reading pulse: unblock the polling reflector. The "Laser On" LED is extinguished. The CLV switches off the laser diode. The CLV outputs a tone via the beeper to indicate that the reading result is displayed in the window of the Terminal Emulator. The "Result" LED lights up until the end of the next reading pulse.

Fig. 6-2 shows two examples of reading result output: "good read" and "no read". The reading result comprises the data contents of the bar code(s) and the reading diagnosis data/constants parameterized to be output. In the default setting, the "Header", "Separator" und "Terminator" blocks do not contain any reading diagnosis data or constants. The CLV only reads and outputs here one bar code. The reading result for “Good Read“ only comprises the data content of the bar code, the reading result for “No Read“ comprises the parameterized wrong read format. Chapter 6.5.2 Percentage evaluation, Page 6-17 explains the structure and function of a part of the reading diagnosis data. Note

The CLV only outputs several bar codes in the reading result if the parameterized minimum and maximum number of bar codes is > 1, and several bar codes have been presented. The number of bar codes to be read/output for each reading interval can be selected on the CODE CONFIGURATION tab in the NUMBER OF CODES group.

with: UVWXYZ = NOREAD = Fig. 6-2:

Note

8 009 974/O079/01-09-2004

Good Read: data content of bar code No Read: wrong read format

CLV Setup: Reading mode: output of the reading result in the terminal emulator

In the reading mode, the CLV does not output the contents of the polling reflector.

© SICK AG · Division Auto Ident · Germany · All rights reserved

6-15

Page 70

Operation

Chapter 6

Operating Instructions CLV 41x bar code scanner

Tip

The structure of the data output string can be configured via the DATA STRINGS tab: •

Up to 10 elements, consisting of constants (control characters, letters, digits) and/or reading diagnosis data, can be entered in the "Header", "Separator" und "Terminator" blocks.

•

The position of the "Separator" block can be choosen before or after the code.

1.

Select the DATA STRINGS tab.

2.

Click in the HEADER input field. The dialog window EDIT PARAMETER: TFH is displayed.

3.

Click the desired constants or placeholders for the reading diagnosis data. The selected elements appear in the text box at the top of the dialog window in the sequence in which they were selected.

4.

Confirm the dialog box with "OK".

5.

Repeat this procedure for the "Separator" and "Terminator" blocks.

6.

Perform a download to the CLV. This is done by clicking The DOWNLOAD PARAMETER dialog box is displayed.

7.

Confirm the dialog box by selecting the PERMANENT save option. The CLV outputs the selected elements in the data output string of the host interface with the next reading result.

in the toolbar.

Triggering the reading pulse via the terminal emulator: In the default setting, the reflector polling is the trigger source of the reading pulse. For test purposes the reading pulse can also be triggered directly via the terminal emulator of CLV Setup. To do so, a different trigger source must be selected temporarily in the CLV.

6-16

1.

Choose the DEVICE CONFIGURATION tab.

2.

Click the READING TRIGGER PARAMETERS button. The READING TRIGGER PARAMETERS dialog box is then displayed.

3.

Click SERIAL INTERFACE under READING TRIGGER SOURCE.

4.

Confirm the dialog box with "OK".

5.

Perform a download to the CLV. To do so, click the SERIAL INTERFACE option again with the right mouse button and choose DOWNLOAD PARAMETER in the context menu. CLV Setup copies the parameter to the CLV temporarily. The serial interface has been activated as the trigger source of the reading pulse and is ready to receive appropriate commands (until the CLV is switched off).

6.

Click in the toolbar. The TERMINAL EMULATOR dialog box is then displayed. The CLV is in reading mode.

7.

Click the SW-TRIGGER ON button or press [F7] key. CLV Setup sends a start command to the CLV. The "Laser On" LED lights up. The red scan line then appears.

8.

Present the bar code sample from Fig. 6-1, Page 6-3.

9.

Click the SW-TRIGGER OFF button or press [F8] key. CLV Setup sends a stop command to the CLV. The "Laser On" LED is extinguished. The CLV switches off the laser diode. The CLV outputs a tone via the beeper to indicate that the reading result is displayed in the terminal emulator. The "Result" LED lights up until the end of the next reading pulse.