Operating Instructions: CLV 41x Bar Code Scanner, Standard line: 6 Operation, pages 55 to 62

Pages 55 to 62 of 142, from Operating Instructions: CLV 41x Bar Code Scanner, Standard line

The original document is the PDF (2.2 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.

For the Sick CLV410, Sick CLV412 and Sick CLV414

Page 55

Operating Instructions

Operation

Chapter 6

CLV 41x bar code scanner

6

Operation 6.1

Overview of procedure for starting up the CLV

•

Start up the CLV with the factory default settings (Quick Start). With this configuration, the CLV can be operated without being connected to a PC.

•

Connect the PC. See Chapter 5.5.5 Connecting the PC to the host interface 2 (RS 232), Page 5-7 and Chapter 10.3 Installation and operating instructions for the PC-based "CLV Setup" program, Page 10-5.

•

Configure (parameterize) the CLV for the desired application.

6.2

Default settings

Table 6-1, Page 6-2 shows an overview of the factory default settings of the CLV. The default parameters are such that the CLV can be put into operation either straight away or following a few minor adjustments. A PC is not required for startup with the default settings. The default settings of the parameters are stored permanently both in the CLV (ROM) and in the CLV Setup database. They can be loaded into the RAM of the CLV or displayed on the tabs of CLV Setup at any time. Displaying and printing out complete default settings in CLV Setup: 1.

Choose FILE and SAVE AS, and enter a file name in the dialog window that appears. CLV Setup saves the current settings in a configuration file "*.scl".

2.

Click in the toolbar. CLV Setup loads the default settings from its internal database and displays them on the tabs.

3.

Click in the toolbar. The PRINT FILE dialog window is then displayed.

4.

If desired, a comment can be entered in the input field for the header of the printout. Confirm the dialog window with "OK". The PRINT dialog window for configuring the printer is then displayed.

5.

Edit the dialog window accordingly and confirm.

CLV Setup prints out all of the default settings in the form of a table. The header contains the company and user names that were entered during the CLV Setup installation routine.

8 009 974/O079/01-09-2004

© SICK AG · Division Auto Ident · Germany · All rights reserved

6-1

Page 56

Operation

Chapter 6

Operating Instructions CLV 41x bar code scanner

6.2.1

Default settings: line/raster scanner CLV 41x

Parameter

Default setting

Active code types

Codabar, Code 39, UPC, EAN, 2/5 Interleaved, Code 93, Code 128, EAN 128

Code length freely assignable (2/5 Interleaved: interval 4 to 50 characters)

Quiet zone

Start/stop ratio, automatic

Min./max. code position

0/100 CP

Multiple read

3

Min./max. no. codes

1

Scanning frequency

500 Hz

Resolution (min. bar width)

0.5 mm (19.7 mil)

Min. reading distance

50 mm (1.97 in)

Reading pulse source

Start: Reflector polling; end: Reflector polling

"Sensor 2" switching input

Teach-in match code

"Sensor 1" switching input not used, Start and end of reading pulse (level: active high), debouncing 10 to 20 ms

Switching outputs

Not inverted

Status output function

Result 1: "Device Ready" (static), Result 2: "Good read" (100 ms), Result 3: "No read" (100 ms)

Beeper

Status output function: "Good read", volume: low

Arrangement to host

Standalone

Device number

1

Host interfaces (type)

RS 422/485, RS 232

Protocol

NAK ; start character: STX, stop character: ETX

Transfer rate

9,600 bd

Data format

8 data bits; no parity; 1 stop bit

Output format

Good Read: Header: blank; Separator: blank; Terminator: blank; No Read: Separator and error string Error string: NOREAD

Output sorting in accordance with code position

Output time

Reading result: end of reading pulse

Test string not activated

Separator: after code

Table 6-1: Extract: Default parameter settings of the line/raster scanner CLV 41x

Note

If the current communication parameters in the CLV are different to the default setting, CLV Setup automatically resets these parameters temporarily to the default setting when starting the communication with the CLV. Only the host interface 2 (RS 232) remains active. This is useful when the CLV is integrated in a SICK network (RS 485). The CLV need not to be disconnected from the network. You also can manually set the parameters to the default setting by presenting the profile bar code no. 11 (Host interface: Temporary Default setting ON) from the card no. 8 008 085 during a reading pulse. Use the profile bar codes no. 12 (Host interface: Temporary Default setting OFF) to reset the CLV to the current parameters.

6-2

© SICK AG · Division Auto Ident · Germany · All rights reserved

8 009 974/O079/01-09-2004

Page 57

Operating Instructions

Operation

Chapter 6

CLV 41x bar code scanner

6.3

Quick start

A PC need not be connected when the CLV is operated with the factory default settings. All bar code types are the default code types enabled except Pharmacode. 6.3.1

Starting up the line/raster scanner with the factory default settings

1.

Connect the CLV to the SICK CDB 410 or CDM 410 connection module.

2.

Switch on the power supply to the CDB 410 or CDM 410. Once the CLV has started, it outputs a tone to indicate that the self-test was successfully completed. Shortly afterwards, it outputs two tones to indicate that it has assumed reading mode. The "Device Ready" LED lights up. The "Result 1" ("Device Ready") output switches.

3.

Align the CLV to the provided white polling reflector (special code) with a reading distance of approx. 300 mm (11.82 in) (not for CLV 414).

4.

Start the reading pulse: block the polling reflector, e. g. with a sheet of paper. The "Laser On" LED lights up. The CLV switches the laser diode on and the red scan line appears.

5.

Present the bar code pattern from Fig. 6-1 to the CLV at the reading distance specified in Table 6-2.

6.

Align the bar code in such a way that the red scan line(s) is (are) almost at a 90° angle to the bars.

7.

End the reading pulse: unblock the polling reflector. The "Laser On" LED is extinguished. The CLV switches off the laser diode. The CLV confirms that the bar code was read successfully by outputting a tone via the beeper. The "Result" LED lights up until the end of the next reading pulse. The "Result 2" switching output (Good Read) becomes live for the duration of 100 ms.

The CLV can now be operated with the factory default settings. The device can be switched off without the configuration data being lost as no changes have been made to the parameter set.

U

Fig. 6-1:

V

W

X

Y

Z

Bar code pattern (code 39; module width 0.5 mm (19.7 mil); print ratio 2:1)

Device types

Scan procedure

Reading distance

CLV 410

Line/raster scanner approx. 200 mm (7.88 in)

CLV 412

Line/raster scanner approx. 200 mm (7.88 in)

CLV 414

Line/raster scanner

–

Table 6-2: Reading distances for module width 0.5 mm (19.7 mil), default settings

8 009 974/O079/01-09-2004

© SICK AG · Division Auto Ident · Germany · All rights reserved

6-3

Page 58

Chapter 6

Operation

Operating Instructions CLV 41x bar code scanner

6.4

Configuration (parameterizing)

The CLV is configured locally in accordance with the application at hand. The read, evaluation and output characteristics can be parameterized as required. The factory default settings or an application-specific parameter set of the CLV are in effect before changes are made. The CLV supports four different configuration methods: •

Configuration via the "CLV Setup" program and the CLV Assistant (parameter values are set via the serial interface).

•

Configuration via AutoSetup (reading and code configurations for a bar code are adapted automatically via the optical interface (reading window)), see Chapter 6.4.4 Configuring the CLV with AutoSetup, Page 6-11.

•

Configuration via profile bar codes (parameter values are set via the optical interface (reading window)), see Chapter 10.4 Configuring the CLV with profile bar codes, Page 10-16.

•

Configuration via command strings (parameter values are set via the serial interface), see Chapter 10.5 Configuring the CLV with command strings, Page 10-22.

6.4.1

Configuring CLV with the user interface of CLV Setup

To be able to use "CLV Setup", a PC has to be connected and the program installed on it. The procedure for connecting the PC to the CLV is described in Chapter 5.5.5 Connecting the PC to the host interface 2 (RS 232), Page 5-7. Installation and startup of "CLV Setup" and operation of the user interface are described in the appendix. See Chapter 10.3 Installation and operating instructions for the PC-based "CLV Setup" program, Page 10-5. Transferring parameter sets between CLV Setup and the CLV Upload When the CLV 41x is being parameterized, CLV Setup runs in the offline mode. To be able to modify the current parameter set of the CLV 41x, this first has to be loaded to CLV Setup from the CLV 41x. This procedure is referred to as an upload ("Upload from CLV 41x" in the CLV 41x menu or [F3] key) during which CLV Setup always loads a complete copy of the current CLV 41x parameter set. This parameter set remains unchanged until it is overwritten by CLV Setup. With the help of the context menu (right mouse button), only the parameter just edited can be loaded ("Upload parameter") or from version 3.6 all parameters of one tab or one dialog box ("Upload parameters of this view") from the CLV 41x’s memory (RAM). Download Changes made to the current parameter set in CLV Setup do not take effect until the parameter set has been transferred to the CLV 41x. CLV Setup always sends a copy of the complete parameter set to the CLV 41x, i. e. all of the existing parameter values in the CLV 41x are overwritten. The procedure for transferring and saving the parameter values in the CLV 41x is referred to as a download ("Download to CLV 41x" in the CLV 41x menu or [F4] key). With the help of the context menu (right mouse button), only the parameter just edited ("Download parameter") or from version 3.6 all parameters of one tab or one dialog box ("Download parameters of this view") will be temporarily loaded into the CLV 41x’s memory

6-4

© SICK AG · Division Auto Ident · Germany · All rights reserved

8 009 974/O079/01-09-2004

Page 59

Operating Instructions

Operation

Chapter 6

CLV 41x bar code scanner

(RAM). To finish the parametrization done by this way all parameters must be download again to the CLV with the "Permanent" option. Loading complete parameter set from the CLV (upload) ¾

Click in the toolbar or press [F3] key. CLV Setup loads the current CLV parameter set from the RAM of the CLV to its database and displays the values on the tabs. The "Device Ready" LED is extinguished during the upload.

If the "CLV Setup" program does not recognize the parameters transferred during the upload, it outputs a warning. Unknown parameters can be edited on the EXTRAS tab by following the conventions for command strings. When the parameter set is saved, these parameters are also taken into account. Transferring the complete parameter set to the CLV and saving it (download) 1.

2.

Click in the toolbar or press [F4] key. The "Device Ready" LED is extinguished. CLV Setup copies the parameter set to the RAM of the CLV. The DOWNLOAD PARAMETERS dialog window with the save options is then displayed. PERMANENT:

CLV Setup copies the parameter set to the RAM and to the EEPROM of the CLV.

TEMPORARY:

CLV Setup copies the parameter set to the RAM only. The changes are lost when the CLV power supply is switched off.

Confirm the dialog box with the desired save option.

The "Device Ready" LED lights up again. The new parameter set is saved in the CLV, permanentely or temporarily dependent on the selected option. Saving the parameter set in CLV Setup 1.

To save the modified parameter set as a new configuration file in CLV Setup, or to overwrite an existing file, choose FILE and SAVE AS. The SAVE AS dialog window is then displayed.

2.

Enter the file name in the dialog window (file extension "*.scl") and confirm the entry.

The new parameter set is now saved in CLV Setup in the subdirectory "data. 6.4.2

Function of tabs in CLV Setup (overview)

Reading Configuration This tab (and additional dialog windows) are used to set the following: •

Scanning frequency

•

Min. module width

•

Min. reading distance

•

Quiet zone

•

Limitation of the evaluation range of the scan line

Device Configuration This tab (and additional dialog windows) are used to set the following:

8 009 974/O079/01-09-2004

•

Device number

•

Source of the reading pulse

© SICK AG · Division Auto Ident · Germany · All rights reserved

6-5

Page 60

Operation

Chapter 6

Operating Instructions CLV 41x bar code scanner

•

Output time of the reading result referred to the start of the reading pulse

•

Laser timeout

•

Physical arrangement for data output via the host interface (standalone, master/slave)

•

Functional assignments of the "Result 1" to "Result 3" switching outputs

•

Functional assignment and volume of the beeper

•

Mode for programming match code 1 (teach-in)

•

Code comparison

Code Configuration This tab (and additional dialog boxes) are used to set the following:

Recommendation

•

Activation of evaluation routines for individual code types

•

Number of identical reads (multiple reads)

•

Minimum and maximum number of bar codes to be read/output

•

Activation of comparison of code position for the separation of identical bar codes

To enhance the reading reliability with fast applications, only activate those code types and code lengths that are actually relevant. Host Interface This tab (and additional dialog boxes) are used to set the following: •

Data format and transfer rate

•

Data transfer protocol

•

Start and stop characters of the interface protocol

Data Strings This tab (and additional dialog boxes) are used to set the following: •

Data output format

•

Constants and reading diagnosis data in the Header, Separator and Terminator blocks

•

Arrangement of the separators in the data string

•

Output format for no reads and contents of the error string

•

Test string function

•

Output sequence and sort criteria for reading more than one bar code per reading pulse

•

Activation and structure of format mask

Extras This tab is used to edit parameters that were not recognized by CLV Setup after an upload. Note

6-6

The online "CLV Setup Help" contains a detailed description of the functions of the parameters and their valid entries. See Chapter 10.3.8 Online help – CLV Setup Help, Page 10-13.

© SICK AG · Division Auto Ident · Germany · All rights reserved

8 009 974/O079/01-09-2004

Page 61

Operating Instructions

Operation

Chapter 6

CLV 41x bar code scanner

6.4.3

Parameterization guide

Overview of parameterization procedure •

Parameterize optical reading characteristics of the scanner

•

Set up reading pulse

•

Parameterize the laser timeout

•

Parameterize the evaluation characteristics of the decoder

•

Parameterize output characteristics (data, result status)

When the CLV is started up for the first time, the factory default settings are in effect. The following parameters must then be set: a) Parameterizing optical reading characteristics •

Scanning frequency Ö READING CONFIGURATION tab Ö SCANNING FREQUENCY section

•

Minimum bar width Ö READING CONFIGURATION tab Ö MINIMUM BAR WIDTH section

•

Minimum reading distance Ö READING CONFIGURATION tab Ö MINIMUM READING DISTANCE section

•

Rest zone Ö READING CONFIGURATION tab Ö QUIET ZONE section

•

Limitation of evalulation range of the scan line Ö READING CONFIGURATION tab Ö CODE POSITION READ ZONE section b) Parameterizing reading pulse source Action

Settings in CLV Setup

Set up reading pulse: External sensor

Ö DEVICE CONFIGURATION tab Ö READING TRIGGER PARAMETERS button

– or –

Ö READING TRIGGER SOURCE section

Start: – Sensor input (Aktiv High) or – Sensor input (Aktiv Low) Ö DEBOUNCING section – Standard or – Fast Ö FIRST TRIGGER section – Edge or – Level Stop: Ö END OF READING INTERVAL section – Generated by trigger source or – Timer: Timeout (referred to the start of the reading pulse) Read result output referred to the start of the reading pulse: Ö OUTPUT ON GOOD READ section – End of reading gate or – Immediate Note: Connect the external sensor to the “Sensor 1“ switching input.

Table 6-3: Guide: parameterizing reading pulse source and read result output time

8 009 974/O079/01-09-2004

© SICK AG · Division Auto Ident · Germany · All rights reserved

6-7

Page 62

Chapter 6

Operation

Operating Instructions CLV 41x bar code scanner

Action

Settings in CLV Setup

Set up reading pulse: Command strings

Ö DEVICE CONFIGURATION tab Ö READING TRIGGER PARAMETERS button

– or –

Ö READING TRIGGER SOURCE section

Start: – Serial interface Stop: Ö END OF READING INTERVAL section – Generated by trigger source or – Timer: Timeout (referred to the start of the reading pulse) Trigger characters: Ö SERIAL INTERFACE section – Standard Trigger or – Single characters Ö for single characters: start and stop characters Read result output referred to the start of the reading pulse: Ö OUTPUT ON GOOD READ section – End of reading gate or – Immediate

Set up reading pulse: Free-running

Ö DEVICE CONFIGURATION tab Ö READING TRIGGER PARAMETERS button

– or –

Ö READING TRIGGER SOURCE section

Start: – Free-running with timeout Stop (by trigger source): Ö FREE RUNNING section – Timeout Read result output referred to the start of the reading pulse: Ö OUTPUT ON GOOD READ section – End of reading gate or – Immediate Note: No laser timeout is available in this pulse mode (the laser diode is not switched off if the reading interval has not be terminated).

Set up reading pulse: Reflector polling

Ö DEVICE CONFIGURATION tab Ö READING TRIGGER PARAMETERS button

– or –

Ö READING TRIGGER SOURCE section

Start/Stop: – Reflector polling Read result output referred to the start of the reading pulse: Ö OUTPUT ON GOOD READ section – End of reading gate or – Immediate

Table 6-3: Guide: parameterizing reading pulse source and read result output time (contd.)

6-8

© SICK AG · Division Auto Ident · Germany · All rights reserved

8 009 974/O079/01-09-2004