Operating Instructions: CLV 41x Bar Code Scanner, Standard line, pages 77 to 86

Pages 77 to 86 of 142: 7 Maintenance; 8 Troubleshooting

The original document is the PDF (2.2 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.

For the Sick CLV410, Sick CLV412 and Sick CLV414

Page 77

Maintenance

Operating Instructions

Chapter 7

CLV 41x bar code scanner

7

Maintenance 7.1

Cleaning during operation

To ensure optimum reading performance, the reading window of the CLV must be clean at all times. For this reason, we recommend that you check the reading window at regular intervals, particularly if the device is operated under extreme conditions (dust, abrasion, humidity, finger prints). Laser radiation can seriously damage your eyesight The CLV uses a class 2 red-light laser. Looking directly at the laser beam can seriously damage the retina in your eyes. ¾

As with sunlight, never look directly into the laser beam.

¾

Switch off the device while it is being cleaned.

¾

Never look directly into the reading window while the device is in operation.

Damage to the reading window The reading window is made of glass. Scratches and streaks on the window will reduce the reading performance. ¾

Use a mild, non-scouring detergent.

¾

Never use scratching or scouring motions to clean the reading window.

Cleaning reading window ¾

Clean the reading window at regular intervals with a mild, non-scouring detergent, e.g. an antistatic glass cleaning fluid. Fig. 7-1 shows the areas that require cleaning.

Use a soft, lint-free cloth. ¾

If necessary, clean the LEDs on the rear of the device. Line/Raster scanner

Line/Raster scanner with 105° angle attachment clean here: Fig. 7-1:

Note

8 009 974/O079/01-09-2004

Cleaning the reading window

Don’t open the device. The producer warranty will be forfeited if the device is opened.

© SICK AG · Division Auto Ident · Germany · All rights reserved

7-1

Page 78

Chapter 7

Maintenance

Operating Instructions CLV 41x bar code scanner

Cleaning other optical surfaces: ¾

If an external reading pulse generator with a sensor (e.g. photoelectric reflex switch) is used, make sure to clean the relevant optical surfaces (Fig. 7-2). Contamination on these surfaces can cause switching errors.

¾

If the reading pulse is generated by reflector polling, make sure to clean the surface of the reflector.

clean here:

Fig. 7-2:

7.2

Cleaning the external optical sensor (reading pulse generator)

Maintenance

The CLV does not require any maintenance except the described cleaning measures. Its self-monitoring function ensures fault-free operation over long periods of time. The CLV outputs detected device faults by sending the error status ST= 3 via the host interface (if the error status has been enabled via the DATA STRINGS tab. Disabled in default setting). See also Chapter 8.4 Troubleshooting, Page 8-3.

7.3

Disposal

Irreparable devices or devices that are no longer required are to be disposed of in an environmentally-friendly manner: 1.

Always observe the applicable national waste disposal regulations.

2.

Remove the CLV housing.

3.

Remove the electronic modules and the connection cable.

4.

Remove the glass panel from the reading window and send it to be recycled.

5.

Send the chassis and cover (zinc die-cast) to be recycled.

6.

Send the electronic modules and connection cable for disposal as special waste.

At present, SICK AG does not accept any unusable or irreparable devices.

7-2

© SICK AG · Division Auto Ident · Germany · All rights reserved

8 009 974/O079/01-09-2004

Page 79

Operating Instructions

Troubleshooting

Chapter 8

CLV 41x bar code scanner

8

Troubleshooting 8.1

Overview of errors and malfunctions which could occur

8.1.1

Installation errors

•

CLV aligned incorrectly with the object carrying the bar code (e.g. masking)

•

Polling reflector or reading pulse sensor positioned incorrectly

8.1.2 •

Electrical connection errors

Interfaces on the CLV connected incorrectly (wiring error in CDB 410 or CDM 410)

8.1.3

Parameterization errors

•

Functions not adjusted to local conditions, e.g. communication parameters on the host interface are set incorrectly

•

Technical limits of the device exceeded, e.g. relationship between the scanning frequency and the reading distance

8.1.4

Malfunctions during operation

•

Laser timeout for laser diode elapsed

•

Device error (hardware/software)

8.2

Monitoring errors and malfunctions

The CLV is self-monitoring:

8 009 974/O079/01-09-2004

•

After the power supply has been switched on, the CLV automatically carries out a selftest, in which it checks important hardware components, before it is initialized (loading of parameter set and initializing of device functions). The self-test can be repeated at any time by canceling reading mode. For calling up, see Chapter 6.5.5 Self-test, Page 6-20.

•

If the CLV detects an error during the self-test or at any other time, it outputs the error status ST= 3 via the host interface. Prerequisite: the error status is enabled for transmission. By default, transfer is blocked.

•

While it is in operation, the CLV constantly monitors the laser diode and speed of the polygon mirror wheel. In addition, a watchdog circuit responds to error statuses.

•

A timer automatically deactivates the laser diode 10 minutes (default setting) after a continuous reading pulse is initiated in reading mode with switching input pulse modes "Sensor Input" and "Serial Interface". In this case, the CLV outputs outputs three consecutive tones via the beeper. The reading pulse must be terminated by resetting the pulse signal. The laser diode is activated again by the next reading pulse.

© SICK AG · Division Auto Ident · Germany · All rights reserved

8-1

Page 80

Troubleshooting

Chapter 8

Operating Instructions CLV 41x bar code scanner

8.3

Error messages

The CLV does not output any error messages except the ST error status. Value

Meaning

Possible cause

Remedy

0

"Good read"

– not applicable

1

Incorrect check digit

The check digit calculated by the CLV during the reading process does not match that printed in the bar code.

Check whether the check digits in the bar code which were generated by the print process are correct.

2

No bar code of the enabled code type • No bar code in the CLV's reading range • Synchronize the pulse of the CLV found according to the evaluation during the reading pulse. with the entry of a bar code into the criteria reading area. • Code type/length in the CLV not • Correct the code configuration enabled for decoding. parameters. • Reading window obstructed/dirty. • Check the reading window.

3

Device fault

Device fault

Call up the self-test! See Chapter 6.5.5 Self-test, Page 6-20. If result ≠ 15000: Contact the SICK Service department.

5

Required number of successful multiple reads for the bar code not reached

• Conveyor speed of the object is too high. • Scanning frequency is too low. • Ladder-type arrangement for bar code relative to the conveyor direction: code height (bar length) too poor. • Print quality too poor.

• Check the conveyor speed. • Adjust the scanning frequency. • Check the code height.

• Check the print quality.

6

Master: • The parameterized number of slaves • Check the entered number of Number of slave reading results does does not match the actual number of slaves. not match the parameterized number slaves available. of slaves • Timeout for the master is selected too • Check the master timeout setting. short. • Data connection between slaves and/ • Check the cables between the or the master has been interrupted. slaves and the master.

9

The scanned object does not have a bar The OUTPUT FILTER function is also activated for the code comparison. code that matches the specified match The CLV has detected valid bar codes. code. However, these do not match the active match code(s)

not applicable

A

The CHECK MAX. NUMBER OF CODES One object in a set of objects with a function is active. constant number of bar codes, for The number of valid bar codes example, contains more bar codes than detected by the CLV in reading mode defined in the application. exceeds that specified under NUMBER OF CODES: MAXIMUM. It repeatedly outputs the defined error string until the number specified under NUMBER OF CODES: MINIMUM is reached.

This message is used to indicate errors on the objects (e.g. check whether objects are homogeneous: mix with incorrect objects).

D

The "Code 32" evaluation option is activated for Code 39. The CLV is attempting to interpret 6-digit C39 bar codes as C32 bar codes (output as 9-digit decimal values).

The read 6-digit bar code is not a C32 bar not applicable code. The CLV outputs the defined error string instead.

Table 8-1: Meaning of the ST error status in the reading result

8-2

© SICK AG · Division Auto Ident · Germany · All rights reserved

8 009 974/O079/01-09-2004

Page 81

Operating Instructions

Troubleshooting

Chapter 8

CLV 41x bar code scanner

8.4

Troubleshooting

The following are required for troubleshooting purposes using the tables below: •

Tools

•

A tape measure (up to 1 m (39.4 in))

•

A digital measuring device (ammeter/voltmeter)

•

A PC running "CLV Setup"

•

A 3-core RS 232 data cable (null modem cable), pins 2 and 3 crossed, e.g. no. 2 014 054

8.4.1

General malfunction: CLV not ready

Malfunction

Possible cause

The "Device Ready" LED is not lit. The "Device Ready" switching output (default setting: "Good read") is not outputting a pulse.

After switching on the power supply (5 to 30 V DC): • CLV not supplied with operating voltage

Remedy

• Check wiring (power connector fitted securely to the CDB 410 or CDM 410 connection module). Measure voltage at pin 1 and pin 5. • The CLV has diagnosed a device error in • Switch the CLV off and on again. Does the LED light up? If not, contact the SICK the self-test Service department. During operation: • The CLV is not in "reading" mode

• Return to reading mode using the CLV Setup program See Chapter 6.5.1 Reading mode (standard operating mode), Page 6-14. • Terminate reading pulse. • The CLV has deactivated the laser diode Check the reading pulse function. 10 min (default setting) after the start of Generate new pulse or parameterize the current reading pulse (pulse mode: new laser timeout duration using the CLV sensor input/serial interface) Setup user interface. See Chapter 6.4.3 Parameterization guide, Page 6-7. Table 8-2: Troubleshooting: restoring operation (reading mode)

8 009 974/O079/01-09-2004

© SICK AG · Division Auto Ident · Germany · All rights reserved

8-3

Page 82

Troubleshooting

Chapter 8

Operating Instructions CLV 41x bar code scanner

8.4.2 Prerequisite

Malfunctions in reading mode: reading pulse errors

The "Device Ready" LED lights up. If not, see remedy in Table 8-2, Page 8-3.

Malfunction

Possible cause

Remedy

1. Pulse mode: Reflector polling

• Incorrect reading pulse source parameterized on the CLV

• In the CLV Setup program: Select the tab DEVICE CONFIGURATION and click the READING TRIGGER PARAMETERS button. READING TRIGGER SOURCE group: reflector polling selected? • Check: reading distance of reflector in specified reading range? • Check reflector type, Table 4-2, Page 4-9. • Check sequence: Read: 1. Block reflector, 2. Read bar code, 3. Unblock reflector.

The CLV cannot be pulsed: – the "Laser On" LED flickers but does not light up constantly – the red scan line is very faint

• Incorrect reading distance (polling reflector) • Incorrect reflector type used • Incorrect pulsing sequence (reflector is not blocked to start the reading interval on the CLV)

2. Pulse mode: • Incorrect reading pulse source sensor input parameterized on the CLV (external sensor). The CLV cannot be pulsed: – the "Laser On" LED does not light up • Polarity of switching input inverted – the red scan line does not appear

• In the CLV Setup program: Select the tab DEVICE CONFIGURATION and click the READING TRIGGER PARAMETERS button. READING TRIGGER SOURCE group: sensor input selected? • Active high: reading pulse starts when power applied. • Sensor not connected or connected to • Connect sensor to "Sensor 1" switching input. incorrect switching input Check sensor wiring (Fig. 5-5, Page 5-10). Measure output signal on sensor. • If sensor supplied by CLV: sensor not • Insert jumper between pin 5 and pin 15. connected to ground potential (CDB 410: switch S 3, CDM 410: • Photoelectric switch not aligned with the switch S 6) reflector (photoelectric switch/inductive • Align sensor. Check functioning. sensor not dampened)

3. Pulse mode: • Incorrect reading pulse source Serial Interface parameterized on the CLV The CLV cannot be pulsed: – the "Laser On" LED does not light up – the red scan line does not appear

• In the CLV Setup program: Select the tab DEVICE CONFIGURATION and click the READING TRIGGER PARAMETERS button. In the READING TRIGGER SOURCE group: is serial interface selected? Group SERIAL INTERFACE: is trigger type correct? • The CLV is not receiving any command • Check data connection to host. In the CLV Setup program: Select the string to start the reading interval on the HOST INTERFACE tab. In DATA FORMAT data interface group: are the correct interface type (hardware) and data format selected? Group: INTERFACE PROTOCOL: are the correct start and stop characters selected?

4.

CLV does not respond to an external signal that ends the reading pulse (sensor input, reflector polling, serial interface)

Incorrect end of reading pulse parameterized on the CLV

In the CLV Setup program: Check by choosing the DEVICE CONFIGURATION tab, click the READING TRIGGER PARAMETERS button. END OF READING INTERVAL group: is "Generated by Trigger Source" selected?

Table 8-3: Troubleshooting: reading pulse errors in reading mode

8-4

© SICK AG · Division Auto Ident · Germany · All rights reserved

8 009 974/O079/01-09-2004

Page 83

Operating Instructions

Troubleshooting

Chapter 8

CLV 41x bar code scanner

8.4.3 ¾

Malfunction

Malfunctions in reading mode: result output errors

Check whether the "Device Ready" LED illuminates. If the LED does not illuminate, follow Table 8-2, Page 8-3. For the following, the number of bar codes to be read is 1. Possible cause

1. The "Result" LED (Good read) does not • Read not successful, since there was no light up at the end of the reading pulse. bar code in the reading field during the reading pulse – The "Result 2" switching output • Code comparison: the code read does (default setting: "Good read") is not not match the defined match code outputting a pulse – The CLV outputs the status ST = 2 in • Reading configuration for the presented bar code in CLV parameterized the reading result on the host incorrectly interface (if ST is enabled for the separator; this is not selected in the default setting) • Scan line positioned incorrectly

•

• • •

•

Remedy • Synchronize reading with an object in the reading field. See Table 8-3, Page 8-4. • not applicable

• In the CLV Setup program: Check by choosing the READING CONFIGURATION tab, READING PARAMETERS section. Are the min. reading distance and min. bar width values correct? • Align CLV. Is bar code at center of scan line? In the CLV Setup program: check reading quality. Choose PERCENTAGE EVALUATION. See Chapter 6.5.2 Percentage evaluation, Page 6-17. • In the CLV Setup program: Check by Evaluation range of the scan line is choosing the READING CONFIGURATION tab, limited improperly (CP values) CODE POSITION group: are the min. and max. code position values correct? Bar code presented at incorrect reading • Check: reading distance of bar code in specified reading range? distance • Align bar code accordingly. The bar code is tilted excessively with respect to the scan line The reading angles at which the bar code • In the CLV Setup program: Start PERCENTAGE EVALUATION. See appears to the CLV are too large Chapter 6.5.2 Percentage evaluation, Page 6-17. Monitor reading quality (reading quality >70 %!). Realign CLV if necessary. • Optimize CLV alignment (scan line should Bar code is in total reflection zone strike the bar code at an skew of approx. 15°). In the CLV Setup program: check reading quality. Start PERCENTAGE EVALUATION. See Chapter 6.5.2 Percentage evaluation, Page 6-17.

Table 8-4: Troubleshooting: result output errors in reading mode

8 009 974/O079/01-09-2004

© SICK AG · Division Auto Ident · Germany · All rights reserved

8-5

Page 84

Chapter 8

Troubleshooting

Operating Instructions CLV 41x bar code scanner

Malfunction

Possible cause

Remedy

1. The "Result" LED does not light up at the • Evaluation criteria for bar code incorrect • In the CLV Setup program: Choose the end of the reading pulse. CODE CONFIGURATION tab, SYMBOLOGIES (default setting: "Good read") group. Enable all code types (except Pharmacode), set code lengths to FREE. – continued – Download temporarily to CLV! Start PERCENTAGE EVALUATION. Code type and length are displayed in the reading result. Enable recorded code type in the SYMBOLOGIES group and disable all other types. Enter the code length under FIXED on the code type tab. Download to CLV! • Check: Are sufficient blank zones • Bar code quality not adequate present? Is the print contrast adequate? Are the specified print tolerances exceeded? In the CLV Setup program: Check by choosing READING CONFIGURATION tab, QUIET ZONE group. Is "Start/Stop" selected? Perform test read with a reference code in good condition. 2. The CLV is not transferring any reading results to the host.

• Reading pulse is not functioning correctly • See Table 8-3, Page 8-4. • Check wiring as shown in Fig. 5-2, • Wiring fault in data connection Page 5-5. • Measure voltages. • Voltage level incorrect • In the CLV Setup program: • Host interface in CLV parameterized Check by choosing HOST INTERFACE tab, incorrectly DATA FORMAT group. Is the correct data format selected? Check settings in INTERFACE PROTOCOL group.

3. The CLV transfers the status ST = 3 to The CLV has diagnosed a device error in the host in the reading result the self-test. (if ST is enabled for the separator; this is not selected by default).

Switch the CLV off and on again. Does the "Device Ready" LED light up? If not, contact the SICK Service department.

4. The data content of the bar code output • The data format on the host interface is • In the CLV Setup program: Select the by the CLV is incorrect or incomplete. parameterized incorrectly in the CLV HOST INTERFACE tab, DATA FORMAT group. Are the values correct? • The CLV suppresses the last character in • In the CLV Setup program: Select the CODE CONFIGURATION tab. Click the the bar code. SYMBOLOGY PARAMETERS button. Choose the tab for the relevant code. Is Transmit Check Digit activated? Change if necessary. Download to CLV! Table 8-4: Troubleshooting: result output errors in reading mode (contd.)

8-6

© SICK AG · Division Auto Ident · Germany · All rights reserved

8 009 974/O079/01-09-2004

Page 85

Operating Instructions

Troubleshooting

Chapter 8

CLV 41x bar code scanner

8.4.4 Prerequisite

Malfunctions in reading mode: errors when outputting the result status

The "Device Ready" LED lights up. If not, see remedy in Table 8-2, Page 8-3.

Malfunction

Possible cause

Remedy

1. The switching outputs – "Result 2" (default setting: "Good Read") – "Result 3" (default setting: "No read") are not outputting any pulses.

• The event of the assigned function for • not applicable outputting the result status does not occur during the reading process • Incorrect switching mode parameterized • In the CLV Setup program: Check by in CLV choosing the DEVICE CONFIGURATION tab. Click the RESULT OUTPUT PARAMETERS button. Are the outputs set to "Invert Output"? Change if necessary. Download to CLV!

2. The beeper is not confirming the output of the reading result (default setting: "Good read").

Beeper deactivated

In the CLV Setup program: Check by choosing tab DEVICE CONFIGURATION, group RESULT OUTPUT/BEEPER VOLUME: is "Low" or "High" selected?

Table 8-5: Troubleshooting: errors in the result status output in reading mode

8.4.5 Prerequisite

Malfunctions: configuration errors (parameterization)

The "Device Ready" LED lights up. If not, see remedy in Table 8-2, Page 8-3.

Malfunction

Possible cause

1. AutoSetup: The CLV does not read the presented, application-specific bar code (no confirmation from beeper)

• CLV not in "AutoSetup" mode

•

• •

•

Remedy

• Check: Does the "Device Ready" LED light up? If so, the CLV is in "reading" mode again. In CLV Setup program: Restart AUTOSETUP from the VIEW pull-down menu. • Switch the CLV off and on again. If AutoSetup is started with profile bar Present profile bar code no. 10 within code no. 10: 5 s. wait time of 5 s after power-up has been Then present the application-specific bar exceeded code. Bar code presented at incorrect reading • Check: Is the reading distance of the bar code within the specified reading area? distance • Optimize bar code alignment (scan line Bar code is in total reflection zone should strike the bar code at an angle of approx. 15°). • See Malfunction 1, Table 8-4, Page 8-5 Bar code quality not adequate

2. Profile programming: • Wait time of 5 s after power-up has been • CLV does not read the presented profile exceeded bar code (no confirmation from beeper) • Wait time of 10 s after the first profile bar • code was presented has been exceeded • In reading mode: CLV not pulsed

Switch the CLV off and on again. Present profile bar code within 5 s. Once the first profile bar code has been read successfully, present the next bar code within 10 s, etc. • Clock the CLV accordingly and present the first profile bar code. Present the next bar code within 10 s, etc.

Table 8-6: Troubleshooting: configuration errors (parameterization)

8 009 974/O079/01-09-2004

© SICK AG · Division Auto Ident · Germany · All rights reserved

8-7

Page 86

Chapter 8

Troubleshooting

Operating Instructions CLV 41x bar code scanner

8.5

SICK Support

If the error status ST = 3 occurs repeatedly or if you cannot correct an error using the above measures, the CLV may be defective. The CLV contains no user-serviceable parts. Please contact our local SICK office or subsidary:

8-8

•

The telephone numbers and email addresses are listed on the back page of this manual. For postal addresses see also www.sick.com.

¾

Do not send the device to the SICK service without first contacting us.

© SICK AG · Division Auto Ident · Germany · All rights reserved

8 009 974/O079/01-09-2004