Operating Instructions CLP100 Bar Code Reader: 8 Troubleshooting

Pages 40 to 42 of 58, from Operating Instructions CLP100 Bar Code Reader

The original document is the PDF (1.3 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.

For the Sick CLP100

Page 40

Chapter 8

Troubleshooting

Operating Instructions CLP100 Bar Code Reader

8

Troubleshooting This chapter describes how errors at the CLP100 can be recognised and eliminated.

NOTICE Claims under the warranty rendered void! Do not open the housing of the CLP100. If the device is opened, the SICK AG warranty shall not apply.

8.1

Possible errors and faults

8.1.1

Installation faults

•

CLP100 aligned unfavorably to the bar code (for example, reflections).

•

Reading-pulse sensor positioned incorrectly

8.1.2 •

Interfaces of the CLP100 wired incorrectly

8.1.3 •

40

Parameterization errors

Functions not adapted to the local conditions, for example, communication parameters of the host interface set incorrectly

8.1.4 •

Faults during the electrical installation

Faults during operation

Device fault (hardware/software)

© SICK AG · Division Auto Ident · Germany · All rights reserved

8008912/R289/2008-11-28

Page 41

Operating Instructions

Troubleshooting

Chapter 8

CLP100 Bar Code Reader

8.2

Troubleshooting table

Aids during troubleshooting: •

These operating instructions

•

Multimeter for measuring voltage and current

Fault

Possible cause

Check

Remedy

1.

When the supply voltage is applied to the CLP100: The LEDs do not light up (LEDs must light up for three seconds after voltage is applied if the self-test is successful)

1.1 CLP100 without supply voltage

1.1 Check the supply voltage

1.1 Check the wiring and measure the supply voltage (DC 5 V ± 5 %)

1.2 CLP100 has diagnosed a device fault during the self-test

1.2 Switch the device off and on again. Do the LEDs light up?

1.2 If the LEDs do not light up, contact the SICK service (refer to Chapter 8.3, Page 42)

CLP100 cannot be started via the external sensor, Scan field of the CLP100 (red light) is not displayed or LEDs do not light up after an executed signal for reading

2.1 External sensor connected incorrectly

2.1 Check the wiring of the external sensor

2.1 Measure the output signal of the external sensor and the input signal at the CLP100 (active low)

2.2 Photoelectric switch not aligned to the reflector or Optical path covered (light scanner or inductive sensor is not dampened)

2.2 Check the external sensor for functionality and a free optical path

2.2 Re-align the external sensor

2.3 CLP100 does not receive command string for pulsing via the host interface

2.3 Check the data connection between the host and the CLP100. Check the command strings of the host computer

2.3 Carry out the connection in accordance with Chapter 5.4, Page 25 and observe the pin assigment. Carry out an upload in the CLP Setup. On the HOST INTERFACE tab card set the data format, protocol, start and stop characters, baud rate correctly. Carry out the download to the CLP100

3.1 No bar code in the reading area during the reading process

3.1 Check the interval between the reading pulse and presence of the bar code in the reading area

3.1 Refer to Item 2

3.2 Scan field does not contact the bar code

3.2 Check the adjustment. Is the bar code in the middle of the scan field?

3.2 Readjust the CLP100 (refer to Chapter 4.4.2, Page 21)

3.3 Reading distance incorrect

3.3 Check the reading distance

3.3 Set the correct reading distance, but take the module width into consideration (refer to Chapter 10.3, Page 51)

3.4 Evaluation criteria for the code type set incorrectly

3.4 Check the code type and code length

3.4 Enter correct code type in CLP Setup. Carry out download to CLP100

3.5 Reading angle under which the bar code appears is too large

3.5 Check the reading angle

3.5 Readjust the CLP100 while ensuring that the maximum reading angle is observed (refer to Fig. 4-4, Page 19)

2.

3.

Green LED does not light up after the end of the reading pulse (reading not successful)

Table 8-1: Troubleshooting table

8008912/R289/2008-11-28

© SICK AG · Division Auto Ident · Germany · All rights reserved

41

Page 42

Chapter 8

Troubleshooting

Operating Instructions CLP100 Bar Code Reader

Fault –continued– 3. Green LED does not light up after the end of the reading pulse (reading not successful)

4.

CLP100 does not transfer read results after the end of the reading pulse

Possible cause

Check

Remedy

3.6 Fault due to reflection of the light at the bar code

3.6 Check the angle

3.6 Readjust the CLP100 and tilt it relative to the vertical (refer to Fig. 4-5, Page 20)

3.7 Bar code quality is insufficient

3.7 Check the printing contrast, idle zones and print tolerances of the bar code

3.7 Have the bar code checked (for example by the SICK service (refer to Chapter 8.3, Page 42))

4.1 No reading pulse occurred

4.1 Do the LEDs light up after the reading pulse?

4.1 Refer to Item 2

4.2 Data connection cable wired incorrectly

4.2 Check the wiring

4.2 Carry out the wiring in accordance with the pin assignment and the connection scheme

4.3 Voltage level incorrect

4.3 Check the voltage level

4.3 Correct the voltage level

4.4 Host interface parameterized incorrectly

4.4 Check the data format, protocol, start and stop characters

4.4 Carry out an upload in the CLP Setup. On the HOST INTERFACE tab card set the data format, protocol, start and stop characters, baud rate correctly. Carry out the download to the CLP100

Table 8-1: Troubleshooting table (contd.)

8.3

SICK support

NOTICE Repairs to the CLP100 should only be carried out by qualified and authorised SICK AG service staff. If an error cannot be eliminated using the above measures, it is possible that the CLP100 is defective. The CLP100 cannot be repaired by the user, meaning that it is not possible to re-estabilsh functions after a failure. However, the CLP100 can be rapidly replaced by the user. See Chapter 7.3 Replacing a CLP100, Page 38. If an error occurs which cannot be eliminated, please contact SICK Service: •

International: Competent SICK branch office or SICK subsidary – For telephon numbers and e-mail addresses please see the back page of these operating instructions – For the postal address please visit www.sick.com.

¾

42

Only return devices after consultation with the SICK Service.

© SICK AG · Division Auto Ident · Germany · All rights reserved

8008912/R289/2008-11-28