Operating Instructions CLP100 Bar Code Reader: 6 Startup and configuration

Pages 29 to 36 of 58, from Operating Instructions CLP100 Bar Code Reader

The original document is the PDF (1.3 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.

For the Sick CLP100

Page 29

Operating Instructions

Startup and configuration

Chapter 6

CLP100 Bar Code Reader

6

Startup and configuration Startup, adjustment, configuration and diagnosis are carried out via the CLP Setup configuration software. In normal reading operation the CLP100 operates fully automated.

6.1

Overview ot the startup procedure

•

Startup of the CLP100 with the default setting set by factory (Quick-Start).

•

Installing the CLP Setup software (Chapter 10.2, Page 46).

•

Connecting the PC to the CLP100 (Chapter 5.4.2, Page 25).

•

In order for optimising the functionality of the CLP100, adjusting the CLP100 and configuring the CLP100 for the application.

•

Checking correct functioning of CLP100 in automatic reading operation.

6.2

Quick-Start

1.

Carry out the electrical installation in accordance with Chapter 5, Page 23. A PC does not have to be connected if the CLP100 is operated with the default setting.

2.

Switch on the supply voltage. The LEDs of the CLP100 light up and are extinguished after the self-test has been completed successfully.

3.

Present the bar code sample from Fig. 6-1 to the CLP100 at a distance of approx. 35 mm.

4.

Start the reading pulse: Cover the optical path of the sensor (close the switch). Both LEDs on the rear light up.

5.

End the reading pulse: Unmask the optical path of the sensor (open the switch). If the reading has been successful, the red LED is extinguished, the green LED lights up. The CLP100 is ready to operate with the default setting of our works.

The CLP100 can be switched off without the configuration data being lost, since no changes were carried out to the parameter set.

0

Fig. 6-1:

1

2

3

4

1

2

3

4

5

Bar code sample (code 39, 0.35 mm (13.8 mil)), printing ratio 2:1)

The default setting shows Table 6-1, Page 35.

8008912/R289/2008-11-28

© SICK AG · Division Auto Ident · Germany · All rights reserved

29

Page 30

Chapter 6

Startup and configuration

Operating Instructions CLP100 Bar Code Reader

6.3

Operating modes and output of the read result

When configuring the CLP100, it is possible to choose between the following operating modes: •

Reading mode

•

Diagnostic mode

•

Adjusting mode

•

On-line test mode

In reading mode the CLP100 detects the presented bar codes and emits the read result at the end of the reading pulse parallel via the host and aux interface. In diagnosis mode the quality of the bar code readings is judged. These are brought statistically into the read field of the CLP100 (no transport movement). The output is emitted via the host and aux interface. For details on the diagnosis mode refer to Chapter 6.4.8, Page 33. In adjusting mode the CLP100 is adjusted to the bar code (see Chapter 4.4.2, Page 21). In the on-line test mode the CLP100 emits the read result only via the aux interface and not via the host interface. When outputting the read result, it is possible to choose between the normal reading function and the code comparison function (matchcode). The matchcode function allows the comparison of the read bar codes with a code which is entered at the CLP100. It is possible to select between complete or partial matching of the codes.

6.4

Configuration

The CLP Setup configuration software or command strings optimise the CLP100 to reading conditions on site. Starting point for this is the factory default setting which can be adjusted to optimise the CLP100. The CLP Setup configuration software is used to create an application-specific parameter set which can be loaded permanently into the CLP100 and saved/ archived as a configuration file (slp file) on the PC. CLP100

Download

PC with CLP Setup configuration software

Upload Parameter set in the main memory

Opened configuration file with current parameter set (after upload) RAM RAM

Permanently saved parameter set PROM Factory default setting

Saved configuration file (*.slp) with archived parameter set Data base (hard drive)

ROM Fig. 6-2:

Configuration with CLP Setup and storage of the parameter set

To this purpose a PC has to be connected and the CLP Setup has to be installed.

30

© SICK AG · Division Auto Ident · Germany · All rights reserved

8008912/R289/2008-11-28

Page 31

Operating Instructions

Startup and configuration

Chapter 6

CLP100 Bar Code Reader

6.4.1

Transferring the parameter set between the CLP Setup and CLP100

In order to edit the current parameter set of the CLP100 it must first be copied from the CLP100 to the CLP Setup. This process is designated as uploading. Changes to the parameter set carried out in the CLP Setup are not effective until they have been transferred to the CLP100. CLP Setup always copies the complete parameter set. Saving the parameters to the CLP100 is designated as downloading. Uploading the parameter set from the CLP100 ¾

Click on in the CLP Setup toolbar.

CLP Setup copies the parameter set last saved in the CLP100 from the CLP100 into its database and displays its values in the tab cards. If the CLP Setup software does not recognize the loaded parameters during uploading, a warning is emited. Unknown parameters can be edited in the terminal emulator under observance of the conventions for command strings. These parameters are also taken into consideration when the parameter set is saved. Downloading the parameter set to the CLP100 ¾

Click on in the CLP Setup toolbar.

The new parameter set is saved in the CLP100. Saving the parameter set in the CLP Setup 1.

Select the SAVE AS menu item in the FILE menu bar in order to save the modified parameter set as the new configuration file in the CLP Setup or to overwrite an existing file. The SAVE CLP FILE dialog box is opened.

2.

Enter the path and file name in the dialog box (extension of the file name: "*.slp") and click on OK. The new parameter set is saved in the CLP Setup.

6.4.2

Parameterize the CLP100 with command strings

The CLP100 can be parameterized for example by means of the terminal emulator by entering command strings. Parameters which are unknown to the CLP Setup can also be used. The command language of the CLP100 accesses the command interpreter of the device directly. Fig. 6-3, Page 32 shows a view of the terminal emulator. A0 is entered as an example in the framed text field for entering commands. This input is repeated in the large output window. For a full description of the command language refer to Table 10-4, Page 53.

Important

8008912/R289/2008-11-28

1.

Click on in the CLP Setup toolbar. The terminal emulator window opens.

2.

Enter the command in the framed text field and press the Return key. The command is transmitted to the CLP100.

The CLP100 does not transmit an answer, but reacts immediately to the command.

© SICK AG · Division Auto Ident · Germany · All rights reserved

31

Page 32

Chapter 6

Startup and configuration

Operating Instructions CLP100 Bar Code Reader

Fig. 6-3:

Terminal emulator with input of a command string

6.4.3

Selecting reading mode

After being switched on the CLP100 carries out a self-test and then changes over to the reading mode (default setting). Reading mode can be selected as follows if the CLP100 is in another operating mode: 1.

In the CLP Setup select the DEVICE CONFIGURATION tab card.

2.

Click in the OPERATING MODE field and select READING MODE.

3.

Carry out the download to CLP100. The CLP100 is in reading mode.

6.4.4

Adjusting the reading pulse

In the CLP100 default setting the reading pulse (trigger) is supplied by an external sensor. Debouncing of the sensor and the type of reading pulse end (through a time window or a sensor signal) can be set by means of the CLP Setup software. 1.

In the CLP Setup select the DEVICE CONFIGURATION tab card.

2.

Select the settings in the READING PULSE or READING PULSE END field.

3.

Carry out the download to CLP100.

6.4.5

Triggering the reading pulse

In its default setting the CLP100 receives the reading pulse (trigger) via an external sensor or a command string (software trigger). For test purposes, the reading pulse can also be triggered directly via the terminal emulator. 1.

32

Click on in the CLP Setup toolbar. The terminal emulator window opens. The CLP100 is in reading mode.

© SICK AG · Division Auto Ident · Germany · All rights reserved

8008912/R289/2008-11-28

Page 33

Operating Instructions

Startup and configuration

Chapter 6

CLP100 Bar Code Reader

2.

Click on START READING.

3.

Present the bar code.

4.

Click on STOP READING. The CLP100 outputs the read result in the window of the terminal emulator.

6.4.6

Setting the structure of the read result

In its default setting the CLP100 sends the data contents of the bar code to the host as the read result. In the default setting this data string is STX-/ETX-framed. The structure of the data string can be selected by means of the HOST INTERFACE and READING CONFIGURATION tab cards. 1.

In the CLP Setup select the READING CONFIGURATION tab card.

2.

Click in the SEND PREFIX list box and select a prefix from the list (A-Z, 1-0).

3.

Select the HOST INTERFACE tab card.

4.

Click in the SEPARATOR list field and select a separator.

5.

Carry out the download to CLP100. The CLP100 operates with the new settings.

6.4.7

Setting the matchcode function

The matchcode function allows the comparison of the read bar codes with a code which is entered manually at the CLP100 or trained in automatically (matchcode). 1.

In the CLP Setup select the DEVICE CONFIGURATION tab card.

2.

Click in the ACTIVE box in the COMPARE CODE field.

3.

Enter the code which is to be used for comparison in the MATCHCODE text field.

4.

If only a part of the code has to agree, enter the partial code in the PARTIAL COMPARISON OF THE MATCHCODE text field, and complete the field with question marks or @ characters.

5.

Selection the function of the switching output in the SWITCHING OUTPUT AT CODE COMPARISON field.

6.

Carry out the download to CLP100. The CLP100 outputs a result either at agreement of the read codes with the matchcode or at non-agreement.

6.4.8

Setting the diagnosis mode

In diagnosis mode the quality of the bar code readings is judged. These are brought statistically into the read field of the CLP100 (no transport movement). The CLP100 carries out 100 readings after the reading pulse and evaluates them. The data string in the diagnosis mode is STX-/ETX-framed. The data string contains the bar code information (data) at the first place, followed by a slash and a three-digit identifier. The data string is followed by the combination of slash and identifier repeated thrice (a total of four identifiers separated by slashes): <STX>Data/www/xxx/yyy/zzz<ETX>. The maximum value of an identifier is 100. If the value is lower (one- or two-digit number), the leading blanks will be filled with spaces. If the DISPLAY CONTROL CODE function is activated, a space is displayed as <SPC>. The identifiers mean: www = Number of successful readings ("Good reads") xxx = Number of successful decoding processes (with misinterpretations)

8008912/R289/2008-11-28

© SICK AG · Division Auto Ident · Germany · All rights reserved

33

Page 34

Chapter 6

Startup and configuration

Operating Instructions CLP100 Bar Code Reader yyy zzz

= Number of successful indentifications of a start or stop character (depending on the scan direction) = Number of readings carried out (the CLP100 carries out 100 readings each in the diagnosis mode)

Error-free reading and decoding of a bar code would thus have the following form: <STX>Data/100/100/100/100<ETX>. A reading with some failed decodings has, for example, the following form: <STX>Data/<SPC>67/<SPC>67/100/100<ETX>. If the CLP100 has only recognized the start character, it transfers a question mark instead of the data string: <STX>?/<SPC><SPC>0/<SPC><SPC>0/100/100<ETX>. Fig. 6-4 shows further examples of reading results in the diagnosis mode:

Fig. 6-4:

Terminal emulator with reading results of the CLP100 in diagnosis mode

The diagnosis mode can be selected directly in the terminal emulation window or set via the tab cards:

34

¾

If the terminal emulator window is active: Click in the DIAGNOSTIC MODE box and present the bar code. The reading process is started automatically. The CLP100 scans the bar code once per second. - or -

1.

If the terminal emulator window is not active: In the CLP Setup select the DEVICE CONFIGURATION tab card.

2.

Click in the OPERATING MODE field and select DIAGNOSTIC MODE.

3.

Carry out the download to CLP100. The CLP100 is in diagnosis mode.

© SICK AG · Division Auto Ident · Germany · All rights reserved

8008912/R289/2008-11-28

Page 35

Operating Instructions

Startup and configuration

Chapter 6

CLP100 Bar Code Reader

4.

Click on in the CLP Setup toolbar. The terminal emulator window opens.

5.

Click on START READING.

6.

Present the bar code. The CLP100 outputs the read result in the window of the terminal emulator.

7.

In order to return to reading mode proceed as described in Chapter 6.4.3, Page 32.

6.5

Default setting

Table 6-1 shows an overview of the default setting of our works for the CLP100. The parameters in the default setting are selected so that the CLP100 can be used directly for many applications in this configuration. With the default setting no PC is required for starting up. Parameter

Default setting

Active code type

Code 39

Code length

Free

Multiple reading

2

Number of codes

1

Reading direction

Both

Send start/stop characters

Not active

Check digit

None

Reading pulse generation

Active low (unchangable)

Switching output

Not active

Host interface

RS 232 (unchangable)

Protocol

Start characters STX, Stop characters ETX

Data transmission rate

9,600 Bd

Data format

8 data bits, no parity, 1 stop bit

Output instant

Immediately

No Read message

BR

Table 6-1: Default setting of the parameters of the CLP100

If the CLP100 is connected to a PC, the current parameters can be changed by using the CLP Setup software. The parameter set of the default setting is saved in the CLP Setup and in the CLP100 and can be re-activated at any time. The complete default setting can be viewed by using the CLP Setup software and printed out, if required:

8008912/R289/2008-11-28

1.

In order to save the current settings: Select the SAVE AS menu item in the FILE menu bar and enter the file name.

2.

Click on in the toolbar. The default setting is loaded and is displayed in the tab cards.

3.

Click on in the toolbar. The FILE PRINT dialog box is opened.

4.

Enter the comment for the header of the output and click on OK. The dialog box for the print settings is opened.

5.

Enter the settings for printing and click on OK. CLP Setup prints out the default settings.

© SICK AG · Division Auto Ident · Germany · All rights reserved

35

Page 36

Chapter 6

Startup and configuration

Operating Instructions CLP100 Bar Code Reader

6.6

Switching off the CLP100

1.

If the parameter set was changed, download the parameter set to the CLP100 and save the parameter set as a configuration file in the CLP Setup.

2.

Switch off the supply voltage. The last parameter set saved in the CLP100 remains valid.

36