Operating Instructions CLP100 Bar Code Reader: 5 Electrical installation

Pages 23 to 28 of 58, from Operating Instructions CLP100 Bar Code Reader

The original document is the PDF (1.3 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.

For the Sick CLP100

Page 23

Operating Instructions

Electrical installation

Chapter 5

CLP100 Bar Code Reader

5

Electrical installation 5.1 Important

Overview of installation sequences

Electrical installation should only be carried out by qualified staff. The following list provides an overview of a typical installation sequence: •

Connecting the supply voltage

•

Wiring the host interface (Connecting the PC)

•

Wiring the "Sensor" switching input

•

Wiring the "Result" switching output

Once electrical installation has been completed, the CLP100 is started up and configured (see Chapter 6 Startup and configuration, Page 29).

5.2

Planning the electrical installation

The following general requirements should be observed for the electrical installation: •

Supply voltage 5 V DC ± 5 % (functional extra-low voltage in accordance with IEC 60364-4-41). The power supply unit must output a continuous output of 2 W and be provided by the customer using e.g. a supply power pack.

•

With external reading pulsing: Suitable reading pulse sensor (start/stop), e.g. photoelectric reflex switch, for registering an object in the field of view

•

Host computer with RS RS 232 data interface for further processing the reading data

The following toots and resources are required for electrical installation:

8008912/R289/2008-11-28

•

Tool

•

Digital measuring device (current/voltage measurement)

© SICK AG · Division Auto Ident · Germany · All rights reserved

23

Page 24

Chapter 5

Electrical installation

Operating Instructions CLP100 Bar Code Reader

5.3

Electrical connections and cables

The electrical connection of the CLP100 is carried out either via a 9-pin D-Sub cable plug or the open cable end. The cable length is 0.9 m (2.95 ft). 5.3.1

Pin assignment of the connection plug or wire colors of the cable end 1

5

6

9

Pin

Signal

1

Sensor

2

RxD (Host)

3

TxD (Host)

4

Result "GO/NG"

5

GND

6

Not assigned

7

RxD (Terminal), TTL

8

TxD (Terminal), TTL

9

+5 V

Table 5-1: Pin assignment of the D Sub plug

Wire color

Signal

Pink

Sensor

Brown

RxD (Host)

Gray

TxD (Host)

White

Result "GO/NG"

Black

GND

Yellow

RxD (Terminal), TTL

Orange

TxD (Terminal), TTL

Red

+5 V

Blue

RTS

Green

CTS

Table 5-2: Assignment of the wire colors of the open cable end

24

© SICK AG · Division Auto Ident · Germany · All rights reserved

8008912/R289/2008-11-28

Page 25

Operating Instructions

Electrical installation

Chapter 5

CLP100 Bar Code Reader

5.4

Carry out the electrical installation

NOTICE Claims under the warranty rendered void! Do not open the housing of the CLP100. If the device is opened, the SICK AG warranty shall not apply. •

Connect or disconnect current linkages only under de-energised conditions.

•

Wire cross sections and their correct protection have to be selected and implemented according to valid engineering standards.

•

For data connection use only shielded data cables (twisted pair).

•

In order to avoid interferences, do not install data cable parallel to the power- supply and motor cables over long distances (for example in cable ducts).

•

Lay all cables so that there is no possibility of people tripping over them and so that they are protected against damage.

5.4.1

Supply voltage

For operation the CLP100 requires a supply voltage of 5 V DC ± 5 % in accordance with IEC 60364-4-41 (functional extra-low voltage). The functional extra-low voltage can be generated by a power supply unit using a safety transformer . The power supply unit must output a continuous output of 2 W. Important

The output circuit of the power supply unit must be electrically insulated from the input circuit. This is usually created by means of a safety transformer in accordance with IEC 742. To ensure that the incoming supply cables are protected against short-circuits/overloads, the wire cross-sections must be dimensioned and protected in accordance with valid standards. Connecting the supply voltage: ¾

Connect the supply voltage to Pin 5 and Pin 9 of the 9-pin D Sub plug (see Table 5-1, Page 24) respectively to the wires of the open cable end (see Table 5-2, Page 24).

5.4.2

Wiring the host interface (Connecting the PC)

The CLP100 is operated and configured with the CLP Setup software. For this purpose it must be connected temporarily to the PC via the host interface. The host interface of the CLP100 is an RS 232 interface. For the pin assignment refer to Table 5-1, Page 24 or Table 5-2, Page 24. Table 5-3 shows the recommended maximum cable length as a function of the data transmission rate. .

Interface type

Transmission rate

Max. distance to the host

RS 232

Up to 19,200 Bd

12 m (39.37 ft)

Table 5-3: Maximum cable length between the CLP100 and the host computer

When the PC is connected, the communications parameters are set as described in Chapter 6.4.1 Transferring the parameter set between the CLP Setup and CLP100, Page 31.

8008912/R289/2008-11-28

© SICK AG · Division Auto Ident · Germany · All rights reserved

25

Page 26

Electrical installation

Chapter 5

Operating Instructions CLP100 Bar Code Reader

Fig. 5-1 shows the wiring of the host interface.

*) 9-pin internal D-Sub plug to PC

Fig. 5-1:

Wiring the host interface (RS 232)

NOTICE Damage to the interface module in the CLP100! Electronic components can be damaged if the host interface is wired incorrectly.

Recommendation

¾

Wire the host interface correctly in accordance with Fig. 5-1.

¾

Check the wiring before switching on the CLP100.

¾

Connect the host interface of the CLP100 EMC-compatibly to the host by means of shielded cables. Observe the maximum cable lengths (see Table 5-3, Page 25).

¾

Connect the shielding at one end.

1.

Switch off the PC and the supply voltage of the CLP100.

2.

Connect the desired port "COM x" of the PC to the CLP100. To this purpose use an RS-232 data connection cable (null modem, RxD and TxD transposed).

3.

Switch on the PC and the supply voltage of the CLP100.

4.

Set the communication parameters of the PC (see Chapter 10.2.3, Page 48). In the default setting the CLP100 communicates via the RS 232 interface with the host using the values listed in Table 5-4.

Parameter

Host interface value

Aux interface value

Data transmission rate

9,600 Bd

9,600 Bd

Data bit

8

8

Parity

None

None

Stop bit

1

1

Protocol

SICK: Sending start characters: STX Sending stop characters: ETX Repeat request: None Timeout: 50 ms

SICK: Sending start characters: STX Sending stop characters: ETX Repeat request: None Timeout: 50 ms

Table 5-4: Communication parameters of the host interface (default setting)

26

© SICK AG · Division Auto Ident · Germany · All rights reserved

8008912/R289/2008-11-28

Page 27

Electrical installation

Operating Instructions

Chapter 5

CLP100 Bar Code Reader

5.4.3

Wiring the auxiliary (terminal) interface

As an alternative the PC may be connected to the Aux interface which operates parallel to the host interface with the same communication parameter values. The Aux interface can be use e.g. to monitor the data traffic on the host interface. Fig. 5-2 shows the wiring of the Aux interface. Important

The terminal interface is a TTL level interface.

7

3*

8

2*

5

5*

*) 9-pin internal D-Sub plug to PC

Fig. 5-2:

Wiring the Aux interface (RS 232 TTL)

5.4.4

Wiring the "Sensor" switching input

If a reading process is to be triggered via an external sensor, connect the sensor to the "Sensor" switching input. This triggering type is selected as the default setting of the CLP100. Fig. 5-3 shows the wiring of the "Sensor" switching input. Table 5-5 lists the characteristic data of this switching input.

Fig. 5-3:

¾

Wiring of the "Sensor" switching input

Connect the sensor as shown in Fig. 5-3.

Switching behaviour

Connecting GND to the input starts the read port of the CLP100 (low)

Properties

Not electrically separated

Electrical values

Low (Reading gate active): 0 V ≤ VS ≤ 0.8 V I ≤ 1.5 V µA

High (Reading gate inactive): 2.4 V ≤ VS ≤ 5 V I ≤ 1.5 V µA

Table 5-5: Characteristic data of the "Sensor" switching input

TIP

8008912/R289/2008-11-28

The DEVICE CONFIGURATION tab card of the CLP Setup software can be used to modify the response time of the "Sensor" input (debouncing).

© SICK AG · Division Auto Ident · Germany · All rights reserved

27

Page 28

Electrical installation

Chapter 5

Operating Instructions CLP100 Bar Code Reader

5.4.5

Wiring the "Result" switching output

The switching output can have various result functions assigned to it. If the corresponding event occurs during a reading process, the switching output becomes live for the selected pulse duration. Fig. 5-4 shows the wiring of the "Result" switching output. Table 5-6 lists the characteristic data of the switching output.

Fig. 5-4:

¾

Wiring of the "Result" switching output (NPN)

Wire the switching output as described in Figure 5-4.

Switching behaviour

NPN switching against GND

Properties

Not electrically separated

Electrical values

Vout = 5 V ≤ V ≤ 30 V Iout = ≤ 50 mA

Table 5-6: Characteristic data of the "Result" switching output

Recommendation

TIP

28

¾

In order to check the switching functions wire the switching output to a load and measure the voltage with a high-voltage digital voltmeter. This avoids the display of incorrect voltage values and switching states.

The DEVICE CONFIGURATION tab card of the CLP Setup software can be used to modify the function assignment, the pulse duration and the output instant of the signal.

© SICK AG · Division Auto Ident · Germany · All rights reserved

8008912/R289/2008-11-28