Page 39

3 Controls, Connectors, & Indicators

CHAPTER 3

This chapter describes the physical layout, controls, connectors and indicators on the CiMAX 78xx Scanner, the Standard Interface Unit, and the Hex Interface Unit.

Scanner LED Display There are six status indicators on the side panel of the CiMAX 78xx scanner, as shown in Figure 3–1. FIGURE 3–1. LED Display yellow green

PWR/Laser ON I/O XMIT/RCV Decode Presence Setup

Setup The SETUP LED indicates that the scanner is in either the Setup or Diagnostics mode rather than in the normal operating mode. When this LED is on, the CiMAX 78xx scanner is not running application software.

Presence The PRESENCE LED shows the status of the scan zone presence detect photoeyes. If the scanner is configured to use presence, the LED lights when an object is in the scan zone.

Rev H, February 2002

CiMAX 78xx Technical Reference Manual

3-1

Controls, Connectors, & Indicators

3

Page 40

Chapter

3

Controls, Connectors, & Indicators

Decode The DECODE LED lights when the scanner is decoding a barcode.

Xmit/Rcv The XMIT/RCV LED will blink whenever the scanner is sending or receiving data. This includes messages to and from all sources that are sent via the Host, Setup and Starnode ports. The Ethernet port is not monitored. This LED helps to verify correct wiring and confirms that the scanner is receiving communications from any source.

I/O The I/O LED shows the state of all inputs and outputs (one through eight) combined. That is, if any of the inputs or outputs are activated, this LED will be on. To view the status of each input or output, the user must run the Inputs/Outputs test of the Diagnostics Program. The user may also observe which relays are active in the Interface Unit.

Pwr/Laser On (green) This LED indicates that power to the scanner is on. When power is first applied to the scanner, the scanning motor starts and the on-board computer performs internal diagnostics. After the diagnostics tests are completed, the laser is turned on and normal operation begins.

Power On Indicator This LED, located on the front of the scanner, lights when scanner power is ON, as shown in Figure 3–2. It is connected in parallel with the PWR/LASER ON LED.

3-2

CiMAX 78xx Technical Reference Manual

Rev H, February 2002

Page 41

Interface Units

FIGURE 3–2. Power ON Indicator

LASER ON WHEN LIT

Green LED

Controls, Connectors, & Indicators

3

AVOID EXPOSURE LASER LIGHT IS EMITTED FROM THIS APERTURE

DO NOT STARE INTO BEAM

Interface Units Standard Interface Unit The Standard Interface Unit allows you to make most CiMAX 78xx Scanner connections at a more convenient location than at the scanner itself, which may be mounted where it is not easily accessible. The Standard Interface Unit can usually be mounted at the side of the conveyor allowing easy access from the floor. Figure 3–3 illustrates how the scanner and the Standard Interface Unit are interconnected.

Rev H, February 2002

CiMAX 78xx Technical Reference Manual

3-3

Page 42

Chapter

3

Controls, Connectors, & Indicators

FIGURE 3–3. Standard Interface Unit - Interconnection Block Diagram

CiMAX 7800 CIX Scanner Terminal or PC DB-9 SET/STAR

7800

Connect to

DB-25 IN/OUT PWR DB-9 HOST Standard Interface Unit J3

DB-9 SET-UP TERMINAL

TB 1

TB 2 relays

AC line

RS-232 Host computer Presence photoeyes

Hex Interface Unit The Hex Interface Unit is a special box that allows more than one scanner to be used in tunnel type application. You can have up to six scanners attached to one interface unit. This allows most of the common signals to be connected to more than one scanner without additional wiring. It also allows for a quicker and more reliable installation. Because of the specialized nature of these applications, this box must be specific to the application. Starnode drop cables can be directly connected to TB7, and TB8. These terminal blocks can be an extension of any Starnode network branch. All Starnode rules still apply in layout of Starnode network. Refer to Starnode Installation & Site Planning Guide. This allows a common connect for the Starnode drop cables and eliminates the need for individual T-TAP boxes.

Standard Interface Connections The Standard Interface Unit is connected to the scanner by the cables provided with the system. The cables connect DC power to the scanner and all of the scanner input, output and communications signals, to the interface unit. You must connect the Starnode and Ethernet cables directly to the scanner.

3-4

CiMAX 78xx Technical Reference Manual

Rev H, February 2002

Page 43

Interface Units

You can connect a PC, running terminal software, to the SETUP port in the SET/STAR connector on the scanner. However, it may be more convenient to connect it to the SETUP connector inside the Standard Interface Unit. The SETUP connector in the Standard Interface Unit does not support the CiMAX 1400 HandHeld terminal.

115/230 VAC (auto selected) ±15%, 50/60 Hz, 40 W max power is required by the Standard Interface Unit. This supplies DC power to itself, to the scanner, and provides AC voltage for relays controlling AC-operated devices. The AC Power Cable is fed into the interface through a punch-out hole in the case. The AC line connections are made at TB7, using the screws labeled NEUT and LINE. The ground wire must be connected directly to the case, as shown Figure 3–4. FIGURE 3–4. AC Power Connection

TB 7

LINE

NEUT

GROUND

CASE AC LINE IN

Toggle switch S1 on the circuit board is an AC ON/OFF switch. Because the cover of the interface will normally be closed, AC power should be supplied to the interface through an external, fused switch.

Hex Interface Connections The Hex Interface Unit requires its own AC power. There is a 4 A 250 VAC, 3AG fuse located in the fuse compartment above the power switch. To access the fuse, open the fuse compartment with a small screwdriver. If you position the Hex Interface Unit with the scanner cables exiting on your left side, scanner A is the cable closest to the slide Starnode door. The scanners are numbered A-F with scanner F located closest to the corner of the box. The Hex Interface Unit provides 6 setup connectors, 1 for each scanner.

Rev H, February 2002

CiMAX 78xx Technical Reference Manual

3-5

3

Controls, Connectors, & Indicators

AC Power

Page 44

Chapter

3

Controls, Connectors, & Indicators

In some applications, scanner A is acting as a master scanner using its host port to communicate over a sub-system RS422 multi drop connection to other scanners in the system. This master's setup port is used to communicate to a host computer using RS-232 on TB1 pins 39 and 40. If Switch 1 (S1) is in the normal position, scanner A setup port is also connected to a RS-232 to RS-422 converter. The RS-422 host connection will be made on TB9 pins 1-4. Switch 2 is a dipswitch that may be used to connect a number of scanners setup port receive data lines together. This may be used when data from the Programmable Interface Controller must be sent to all of the scanners. SICK Auto Ident, Inc. cable (A1-63397-1) will be connected between Programmable Interface Controller J1 and Hex Interface Box J18. Make sure that the cable end with pin 8 is connected to the Programmable Interface Controller. Switch 2 positions are as follows: •

If all scanners are to receive the same data on the Setup Port, S2 should be set (11111xxx), where: –

1=On

–

0=Off

– x= don’t care

•

Position 5 - F (BCDEFxxx) Position 5 must be set to ON to connect data to any other scanners. Scanner F gets data directly from J18.

•

Scanner A requires that S1 be in the setup position to receive data.

When the 78xx Scanner is connected to the Hex Interface Unit, ensure the correct switch is On for that scanner. Those scanners must use the SETUP port to receive data from the Programmable Interface Controller. If you wish to use this port for setup or diagnostics, set the applicable switch to Off temporarily, during testing.

3-6

CiMAX 78xx Technical Reference Manual

Rev H, February 2002

Page 45

Interface Units

FIGURE 3–5. S1 & S2 Connection Diagram

Normal S1

J3-8

(A) RXD

RS-232/422 Converter

3

Controls, Connectors, & Indicators

(A) TXD

J3-7 Setup S2

J6 J9 J12 J15 J18

Scanner (B) RXD Scanner (C) RXD Scanner (D) RXD Scanner (E) RXD Scanner (F) RXD

1 2 3 4 5 6 7 8

Another use of these switches is when you wish to update the operating system of more than one scanner at the same time. If you wish to upgrade O.S., communication to a scanner must be made to that scanners setup port through one of the six connectors in Hex Interface Unit. Each connector is used for each individual scanner. As an example, if you have two CiMAX 78xx Scanner attached to positions C and D, connect to computer to J9 or J12. Set S1 to normal and S2 to 01100000 (0=off, 1=on)

Control Input Connections Interface Unit Inputs All standard solid state control input connections to the Interface Unit(s) are available at terminal block TB1, as shown in Figure 3–8, “Standard Interface Unit Circuit Board” on page 3-11, and Figure 3–9, “Hex Interface Unit Circuit Board” on page 3-12. Connector assignments are listed in Table 3–1, “Interface Unit I/O Connections - TB1,” on page 3-16. Eight low-voltage DC solid state control inputs are provided and are, by default, assigned to accept inzone and outzone presence, tachometer and five height sensor signals, as specified in Table 3–1, “Interface Unit I/O Connections - TB1,” on page 3-16. The assignments are shown in parentheses in the table. A custom C program or the Setup parameters may used to override the default assignments. Rev H, February 2002

CiMAX 78xx Technical Reference Manual

3-7

Page 46

Chapter

3

Controls, Connectors, & Indicators

An optional DC-to-DC converter provides isolated DC power for photoeyes and tachometer. Isolated +12 and +5 volts DC and ground terminals are available on terminal block TB5. Using these voltages and the opto-isolated inputs will provide total isolation.

Programmable Interface Controller Inputs Standard inputs 4-8 are used for height sensing. When used, the least significant height is connected to input 4 of the Interface Unit(s). Up to seven additional low-voltage DC inputs are available at terminal block J2 on an optional Programmable Interface Controller, as shown in Figure 3–10, “Programmable Interface Controller Board” on page 3-13. The inzone photoeye is connected to input 8 of the Programmable Interface Controller. The extra height photoeyes are connected to inputs 1 - 7 of the Programmable Interface Controller with lowest input connected to input 1 of Programmable Interface Controller. On the leading edge of inzone, the peak-detected state of the 7 inputs, the highest number of photoeyes blocked since the last inzone, are sent serially to the scanner. They permit additional height sensing, for better resolution of height and corresponding better accuracy for x, y positioning to allow smaller gaps between boxes.

Optional Input Relay Modules As many as eight optional solid-state input relay modules can be used to accommodate AC input signals. Refer to “Relays in the Interface Units” on page 3-13 for more information.

Control Output Connections Eight solid-state open-collector control outputs are available at terminal block TB1 on the main printed circuit board, to operate relays, alarms, diverters, etc. Connector assignments are listed in Table 3–1, “Interface Unit I/O Connections - TB1,” on page 3-16. Refer to Figure 3–6, “Typical Solid State Output Circuit” on page 3-9. Three of the outputs are assigned, by default, to output NO READ, MATCH and NO MATCH signals, as specified Table 3–1, “Interface Unit I/O Connections - TB1,” on page 3-16. The assignments are shown in parentheses in the table. A custom C program or the Setup parameters may be used to override the default assignments. As many as seven optional solid-state output relay modules, and one form C (SPDT) electromechanical relay, are available to switch AC voltages. Refer to “Relays in the Interface Units” on page 3-13 for more information.

Special Inputs & Outputs Laser Control A digital low or switch closure to ground on this input will turn the laser On or Off, depending upon how a jumper inside the scanner was configured at the factory. The default position of jumper JH3 is between pins 1 and 2. This is referred to as a normal laser control. With no connection, or logical high on the input laser control, the laser will be On. A ground or logical low on laser control input causes the laser to turn Off. Moving the jumper to position 2 to 3 causes this function to be reversed and referred to as an inverted laser control.

3-8

CiMAX 78xx Technical Reference Manual

Rev H, February 2002

Page 47

Interface Units

Beeper The Beeper output is an open collector output providing an audio tone to signal the operator each time that the scanner has read a label. By default, Output 4 is used to drive the beeper. Jumper JH6 in the Standard Interface Unit is used to connect Output 4 to the beeper. For beeper operation, position JH6 across pins 2 and 3. Refer to Figure 3–8, “Standard Interface Unit Circuit Board” on page 3-11 which shows the location of JH6.

1.

Place the JH6 jumper across pins 1 and 2.

2.

Set parameter 003 IOMODE to +16 to disable the beeper. Refer to the Scanner Parameters Reference Guide for more information.

Communications Inputs & Outputs Host and SETUP port connections are available on terminal block TB1. Terminal assignments for these ports are shown in Table 3–1, “Interface Unit I/O Connections - TB1,” on page 3-16.

DC Power Outputs Terminal block TB1 also provides several non-isolated low voltage DC outputs for powering external sensors, as well as associated ground connections. The ratings of the DC outputs are shown in Table 3–1, “Interface Unit I/O Connections - TB1,” on page 3-16.

Solid State Input & Output Circuits Figure 3–6 and Figure 3–7 show typical Solid State Input and Output Circuits. FIGURE 3–6. Typical Solid State Output Circuit

Rev H, February 2002

CiMAX 78xx Technical Reference Manual

3-9

Controls, Connectors, & Indicators

3

To use Output 4 as a general purpose output:

Page 48

Chapter

3

Controls, Connectors, & Indicators

FIGURE 3–7. Typical Solid State Input Circuit

+12 VDC 10K Ohms

3-10

CiMAX 78xx Technical Reference Manual

Rev H, February 2002