DP2000-P, DP9000-P Process Indicators, pages 51 to 60

Pages 51 to 60 of 60

The original document is the PDF (1.9 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.

For the Omega DP2000

Page 51 read from a scan by OCR; it may hold reading errors

21.5 ZERO OFFSET (Z01-4)

$2 R2 sees iN span (s), |_| \ ~_¥ q be

| $1 $2 | $3

B 8 . A B BE 8-4 ee an ide Boe 8:8 Bon BB in G 8 & K ot i | Step 1: | Remove all push-on jumpers not used in the desired configuration(s). Step 2: Select the desired configuration from the chart below, then install the push-on jumpers indicated. Zero Offset Configuration DP2000-P DP9000-P S1 ZO1 -3000/-1895 counts -15000/-9475 counts A ZO02 -1900/-600 counts -9500/-3000 counts C ZO3 -605/+870 counts -3025/+4350 counts B Z04 | +865/+2100 counts +4325/+10500 counts D

47

Page 52 read from a scan by OCR; it may hold reading errors

22.0 CALIBRATION FOR DP2000-P

NOTE: The numbers used below are derived from the selection Section 20.3.

22.1 VOLTMETER (VM1-15)

1.

3.

Apply a short to the input terminals and adjust the zero (Z) pot (R1) to make the display read O00.

Apply an Input Voltage equal to the largest (+) magnitude of the VM range selected and adjust the span (S) pot (R2) to make the display read the appropriate number.

Repeat above steps as required to set the display to within +1 count.

22.2 AMMETER (AM1-18)

1.

3.

Apply a short to the input terminals and adjust the zero (Z) pot (R1) to make the display read 000.

Apply an Input Current equal to the largest (+) magnitude of the AM range selected and adjust the span (S) pot (R2) to make the display read the appropriate number.

Repeat above steps as required to set the display to within +1 count.

22.3 VOLTAGE RECEIVER (VR1-24)

1.

2.

3.

Apply a short to the input terminals and adjust the zero (Z) pot (R1) to make the display read the Zero Offset Number (ZON) calculated.

Apply an Input Voltage equal to the Upper Input (UI) number used and adjust the span (S) pot (R2) to make the display read that number.

Repeat above steps as required to set the display readings to within +1 count.

22.4 CURRENT RECEIVER (CR1-24)

1.

Apply a short to the input terminals and adjust the zero (Z) pot (R1) to make the display read the Zero Offset Number (ZON) calculated.

Apply an Input Current equal to the Upper Input (UI) number used and adjust the span (S) pot (R2) to make the display read that number.

Repeat above steps as required to set the display readings to within +1 count. .

48

Page 53 read from a scan by OCR; it may hold reading errors

23.0 CALIBRATION FOR DP9000-P

NOTE: The numbers used below are derived from the selection Section 20.3.

23.1 VOLTMETER (VM1-15)

1.

Apply a short to the input terminals and adjust the zero (Z) pot (R1) to make the display read 0000.

2. Apply an Input Voltage equal to the largest (+) magnitude of the VM range selected and adjust the span (S) pot (R2) to make the display read the appropriate number.

3. Repeat above steps as required to set the display to within +1 count. 23.2 AMMETER (AM1-18)

1.

3.

Apply a short to the input terminals and adjust the zero (Z) pot (R1) to make the display read 0000.

. Apply an Input Current equal to the largest (+) magnitude of the AM range selected and adjust the span (S) pot (R2) to make the display read the appropriate number.

Repeat above steps as required to set the display to within +1 count.

23.3 VOLTAGE RECEIVER (VR1-24)

1.

Apply a short to the input terminals and adjust the zero (Z) pot (R1) to make the display read the Zero Offset Number (ZON) calculated.

Apply an Input Voltage equal to the Upper Input (Ul) number used and adjust the span (S) pot (R2) to make the display read that number.

Repeat above steps as required to set the display readings to within +1 count.

23.4 CURRENT RECEIVER (CR1-24)

1.

Apply a short to the input terminals and adjust the zero (Z) pot (R1) to make the display read the Zero Offset Number (ZON) calculated.

Apply an Input Current equal to the Upper Input (Ul) number used and adjust the span (S) pot (R2) to make the display read that number.

Repeat above steps as required to set the display readings to within +1 count.

49

Page 54 read from a scan by OCR; it may hold reading errors

50

$1 RR 3 4

RI

10288A1-01

R2 a?

ZERO

OFFSET RANGE o

Sl

~ 215 /- 77 WV

~77/+54mv

+54/ +190mV

Baca BcHBoB

+165/ +295mV oO/Ojojice

SPAN RANGE

$2

@.8 8:8 Gc Bos

0/25 %

25/50 %

50/75 %

75/100%

@a/ojojo dialalo

$2

FULL SCALE CURRENT RANGE

$3

FULL SCALE

LOmA

VOLTAGE RANGE 0.5 .

2.0 mA tov

5.0 mA .

2.0V tOmA

5.0V

20 mA

10.0V

50 mA

VBIOLO/S/olo qj, Bl(cjujwjo

20.0V

Orr pa) et

24.0 DRAWINGS

24.1 BSCP ASSEMBLY DIAGRAM

Page 55 read from a scan by OCR; it may hold reading errors

51

+ SvOC

R38 is RIS + 4Sv0C

+ 30VAC sic wm Te1-6 >

R36 R26 Red 1.00" 49.9% 24.9%

R29 R2? R26 R30 R32 Rs 499 24.9 0.0 249 100 439

ANALOG GND x 4 ” rs) dt. HW

29 Q . + REF $20 $2¢ 3 ass fase $R33 R22 ¢ R2 wie 1S3E +530 Bh S57 See, sa2k $ 200K 2

2 --6--- _- -0- =~ 5 asp fi S3A] S3C ‘538 /S3F R24 R23 [s] J 49.9k | 24.9K wv nie 2 ote of room a3 38 $2A 280K S26 $2A ycs

Woe Vv Sov

TR-a >

- REF

TaI-§ > $16 Lo

28

13)

St HI OUT

‘v2a>+ ” f

1 i] t

'

!

|

]

> 100K \ ag 89P oa Fa ne Z| Ru ce, 0) | !

{ t

!

t

H

+

!

i]

]

( t n~ Nn 4 & ” & . J & at hoon 7] if C4 nz (7)

TT 1000pF ne as, - - 10 ¥ ve Sia ! 3.9K G 47505-06 ©

3 590% ass

| 332

-6.2V

SiG LO OUT ae a es Se ie eee

-82V a=

U3 441 7 4 U 2 Lc27TM2 8 4 ul 768 M05 3 -

DESIG TYPE |+ 5v|-6.2v

24.2 BSCP SCHEMATIC DIAGRAM

Page 56 read from a scan by OCR; it may hold reading errors

25.0 DIGITAL PANEL METER INSTALLATION INSTRUCTIONS

IMPORTANT:

For proper installation electrical connections must be made according to the model number on the meter label. Write the model number in the following space and use the appropriate instructions for your model number.

.-- Power requirement (Section 25.3) .-- Analog output (see Analog Output Manual)

.-- Control output (see Controller/Interface : Manual)

.--- Signal input (Section 26.0)

Model number DP2 Model number DP9

25.1 UNPACKING & INSPECTION

Your OMEGA digital panel meter was systematically inspected and tested, then carefully packed before shipment.

Unpack the instrument and inspect for obvious shipping damage. Notify the freight carrier immediately upon discovery of any shipping damage.

25.2 MECHANICAL INSTALLATION

1. 2.

Insure that the panel cutout dimensions are as shown on Figure 6.

Remove the lower printed circuit board edge connector, (if installed) J1, by pushing two molded plastic tabs away from the connector body and pulling the connector off the printed circuit board. Remove the printed circutt poard edge connector, J2, if upper board output option was ordered.

Loosen two clamp screws on the rear of the case enough to rotate the two slide clamps.

Slide the two slide retainers toward the rear of the case and remove them.

From the front of the panel, insert the meter into the panel cutout.

Slide the slide retainers back onto the case and push up tightly against the rear of the panel.

Rotate the slide clamps back into their original position and tighten enough to hold the case in place. Overtightening can break the clamps.

Install the lower printed circuit board edge connector, if supplied, by pushing it on to the printed circuit board connections. Install the upper printed circuit board edge connector, if used.

52

Page 57 read from a scan by OCR; it may hold reading errors

~ PANEL THICKNESS 6,4 (25) MAX 0,8 (.03) MIN 92,0

(06), pics

92,0 + 0,8/-0,0 (3.622 + .032/-.000) |

|. 45,0 + 0,6/-0,0 REAR VIEW

(1.772 + .024/-.000) Terminal block cover and bezel not PANEL CUTOUT shown for clarity.

Clamp rings rotated and slide retainers removed as shown * for installation.

CASE, UPPER CLAMP SCREW CONTROL INTERFACE SLIDE CLAMP

EXCITATION VOLTAGE MODEL & POWER TAB SERIAL NO. : WN

7 IS ~~ SS SZ SLIDE CASE

RETAINER CONTROL INTERFACE

Seas BOARD OPTION | TERMINAL BLOCK COVER | | f law INPUT POWER

4 SIGNAL CONDITIONER es" BOARD

INTERCONNECT CABLE

ZERO ADJ.

ANALOG ns OUTPUT Si BOARD MAIN BOARD SPAN ADJ. . DECIMAL POINT SELECT DISPLAY BOARD CASE, LOWER

FIGURE 6. PANEL CUTOUT DIMENSIONS AND INSTALLATION

53

Page 58 read from a scan by OCR; it may hold reading errors

NOTES:

54

Page 59 read from a scan by OCR; it may hold reading errors

10288ML-98B Cover M1390 12/10/98 7:59 AM Page

[ omega. com ‘| MEOMEGA *%

OMEGAnet* On-Line Service Internet e-mail http://www.omega.com info@omega.com

Servicing North America:

USA: One Omega Drive, Box 4047 §SO 9001 Cartified Stamford, CT 06907-0047 Tel: (203) 359-1660 e-mail: info@omega.com Canada: 976 Bergar Laval (Quebec) H7L 5At Tel: (614) 856-6928 e-mail: canada@omega.com

For immediate technical or app

USA and Canada: Sales Service: 1-800-826-6342 / 1-800-TC-OMEGA™ Customer Service: 1-800-622-2378 / 1-800-622-BEST™ Engineering Service: 1-800-872-9436 / 1-800-USA-WHEN™ TELEX: 996404 EASYLINK: 62968934 CABLE: OMEGA

FAX: (203) 359-7700

FAX: (514) 856-6886

Mexico and Latin America: Tel: (95) 800-TC-OMEGA™

En Espafil: (203) 359-7803 Servicing Europe: Benelux: Postbus 8034, 1180 LA Amstelveen, The Netherlands Tel: (31) 20 6418405 FAX; (31) 20 6434643 Toll Free in Benelux: 06 0993344 e-mail; nl@omega.com Czech Republic: ul. Rude armady 1868, 733 01 Karvina - Hranice Tel: 420 (69) 6311899 FAX: 420 (69) 6311114 e-mail: czech@omega.com France: 9, rue Denis Papin, 78190 Trappes Tel: (33) 130-621-400 FAX; (33) 130-699-120 Toll Free in France: 0800-4-06342 e-mail; france@omega.com G y/Austria: Daim 26, D-75392 Deckenpfronn, Germany Tel: 49 (07056) 3017 FAX; 49 (07056) 8540 Toll Free in Germany: 0130 11 21 66 e-mail: germany@omega.com United Kingdom: 25 Swannington Road, PO. Box vA Omega Drive, (SO 9002 Certified ghton Astley, Lei hi Irlam, N LE9 6TU, England M44 SEX, England Tel: 44 (1455) 285520 Tel: 44 (161) 777-6611 FAX: 44 (1455) 283912 FAX: 44 (161) 777-6622 Toll Free in England: 0800-488-488 e-mail: uk@omega.com \tis the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon The information contained in this document is believed to be correct but OMEGA Engineering, Inc. accepts no liability for any errors it contains, and reserves the right to alter specifications without notice. ING: These products are not designed for use in, and should not be used for, patient connected applications. This device is marked with the international hazard symbol. It is important to read the Setup Guide before installing or commissioning this device as it contains important information relating to safety and EMC.

FAX: (95) 203-359-7807 e-mail: espanol@omega.com

WARRANTY/DISCLAIMER #8385 AAMAS

OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materiais and workmanship for a period of 13 months from date of purchase. OMEGA Warranty adds an additional one (1) month grace period to the normal one {1) year product warranty to cover handling and shipping time. This ensures that OMEGA’s customers receive maximum coverage on each product. If the unit should malfunction, it must be returned to the factory for evaluation. OMEGA’s Customer Service Department will issue an Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA, if the unit is found to be defective it will be repaired or replaced at no charge. OMEGA’s WARRANTY does not apply to defects resulting from any action of the purchaser, including but not limited to dling, interfacing, outside of design limits, improper repair, or unauthorized modification. This WARRANTY is VOID if the unit shows evidence of having been tampered with or shows evidence of being damaged as a result of excessive corrosion; or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating conditions outside of OMEGA’s control. Components which wear are not warranted, including but not limited to contact points, fuses, and triacs. OMEGA is pk d to offer on the use of its various products. However, OMEGA neither assumes responsibility for any omissions or errors nor assumes liability for any Gamages that result the use of its products in accordance with information In provided by OMEGA. either verbal or written. OMEGA warrants only that the manufactured by it will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES ‘OR REPRESENTATIONS OF ANY KIND WHATSOEVER,

EXPRESSED OR IMPLIED, EXCEPT TITLE, AND ALL IMI [ARRANTIES INCLUDING ANY "WARRANTY OF MERCHANTABILITY AND I FITNESS FORA ARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF LIABILITY: The Hees purchaser set forth herein are exclusive ane the total liebitity of OMEGA with respect to this order, whether based on war ranty nification, strict liability or otherwise, shall not exceed the purchase price of the component upon which ciability jis based. In no event shall MEGA b be liable for F special CONDITIONS: Equipment sold b OMEGA i is not intended to be used, nor shall it be used: (1) as a “Basic Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any Product(s) be used in or with any nuclear installation or activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility as set forth in our basic WARRANTY/ DISCLAIMER language, and additionally, purchaser will indemnify OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the Product(s) in such a manner.

* RETURN REQUESTS / INQUIRIES SAA Direct all warranty and repair fequests/inguiries to the OMEGA Customer Service Department. BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR} NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return package and on any correspondence. The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit.

FOR WABBANTY RETURNS, please have FOR NON-WARRANTY REPAIRS, consult the BEFORE | OMEGA for current repair charges. Have contacting OMEGA: the following information available 1. P.O, number under which the product BEFORE contacting OMEGA: was PURCHASED, 1, P.O. number to cover the cOsT of the repair, 2. Model and serial number of product, and 3. Repair instructions and/or specific 3. Repair instructions and/or specific problems relative to the product. problems relative to the product. OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our the latest in and

2. Model and serial number of the product under warranty, and

OMEGA is a regis of OMEGA i. INC. © Copyright 1998 OMEGA ENGINEERING, INC. vy rights reserved. This document may not be copied, photocopied, reproduced, translated, or reduced to a medium or hii dable form, in whole or in part, without prior written consent of OMEGA ENGINEERING, INC.

Page 60 read from a scan by OCR; it may hold reading errors

10288ML-98B Cover M1390 12/10/98 8:18 AM Page

Where Do | Find Everything | Need for

Process Measurement and Control? C € Made in the USA Regs OMEGA...Of Course! vomplant TEMPERATURE

[4 Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies

& Wire: Thermocouple, RTD & Thermistor

& Calibrators & Ice Point References

[ Recorders, C Ilers & Process Monit

Infrared Pyrometers

PRESSURE, STRAIN AND FORCE Transducers & Strain Gauges

& Load Cells & Pressure Gauges

& Displacement Transducers Instrumentation & Accessories

FLOW/LEVEL

G Rotameters, Gas Mass Flowmeters & Flow Computers & Air Velocity Indicators

{ Turbine /Paddlewheel Systems

(& Totalizers & Batch Controllers pH/CONDUCTIVITY

& pH Electrodes, Testers & Accessories

(& Benchtop/Laboratory Meters

{ Controllers, Calibrators, Simulators & Pumps Industrial pH & Conductivity Equipment

DATA ACQUISITION

Data Acquisition & Engineering Software

( Communications-Based Acquisition Systems # Plug-in Cards for Apple, IBM & Compatibles i Datalogging Systems

[# Recorders, Printers & Plotters

HEATERS

( Heating Cable

4 Cartridge & Strip Heaters ( Immersion & Band Heaters [ Flexible Heaters

( Laboratory Heaters

MONITORING AND CONTROL D P2 P, (& Metering & Control instrumentation DP9000-P a Refractometers.

umps & Tubin; le fF Air, Soil & Water Monitors Process Indicators

( Industrial Water & Wastewater Treatment z l pH, Conductivity & Dissolved Oxygen Instruments

10288ML-98 Rev. B M1390/1298 http.//www.omega.com e-mail: infe@omega.com