The text of a manual · part 4 of 6

DP2000-P, DP9000-P Process Indicators, pages 31 to 40

Pages 31 to 40 of 60

The original document is the PDF (1.9 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.

For the Omega DP2000

Page 31 read from a scan by OCR; it may hold reading errors

16.0 DRAWINGS

16.1 DIMENSIONS 8,0 . 48,0 96,0 a 0) | a7) |e (0. y - = L, b 77) , sn \/ “ SLIDE 139,8 a ~e MAX RETAINER (5.50) le u . 45,0, = CLAMP A TERMINAL BLOCK COVER (1.77) RING SIDE VIEW TOP VIEW

NOTE: Dimensions are in millimeters +0,25 mm and inches are in () +0.01 in

PANEL THICKNESS

6,4 (.25) MAX 92,0 0,8 (.03) MIN oo) 16R - | (08), pos ta - 92,0 + 0,8/-0,0 (3.622 + .032/-.000)

atte a

| =6. 45,0 +_:«0,6/-0,0 REAR VIEW (1.772 + .024/-.000) Terminal block cover and bezel not PANEL CUTOUT shown for clarity.

Clamp rings rotated and slide retainers removed as shown for installation.

27

Page 32 read from a scan by OCR; it may hold reading errors

28, _}- Fa

W6

Cy w4>

E38 E39

T41

DP9AD DP90 0 pP92 0 pp94 0 bpe96 0 pe9s O

CRS

4 5 cH

S41 se| A

S3

16.2 DP9000 MAIN ASSEMBLY LED

Page 33 read from a scan by OCR; it may hold reading errors

29

, ° €23

+4V, op, PIE

X00 pgw mr E22

46201 O6 Gt G2 45145 03 45145 Dea 45345 pp2 P1-8 XXX py a 7 €21 pez PI-7 € OOK pa wel esp

SIG MI Tea1-8

1 SIG HI ee SIG LO e- SIG LO oH 4

ANA GND [ - ANA GNO

E33 P1-15<4 PIS & 27

SPARE P3-13¢4 vena

ARE Pa-P SPARE P1-P €+ - oE28 2-3

OS2-a Oss . 9004375 ebcaef g abcaefg 7 te asa F 1 pas

P2-3 - >

SPARE

SPARE SPARE

SPARE ~-B.2v

TEST BUFF

OL

100KHZ osc +P

ANALOG P out

+30V

Oo PI-D

P1-A <+ - o £30 SIG | 40

1-18 Pe-8 +

J near

P26 <Kye-6

Pe - P2-3 Prieta Kee Sig 4 HI K - te we

-4 b] -REF E34 | “2. 7 Pav [ x 2 62-5 =a S

LL E42

Pi-2 € ¢

1-84

‘ ? E24 pit et] pez

Pa-F ? Ea3

P14 bi €12

Pa-B

Pe] S00

SIG HI O13

SIG LO com pi-aad pin ed_ = VV 7.

4-3 HOLD

Ja. 2 100KHZ <u4- 4 +POL

Ps-17€ e327 5 Ps-16€ -44¢4 6 yKv5-6 =) ANALOG

-T et) 93 Pi-T

J3-8 TO OVM

J3-9 SIG HI

Ees +5V rin elt 4 pit € E1360 | AT Fa

I+ cr2 J3-4 RELAY GND

+5V<- - - - + 3-7 +5V

2500pt | 250v cs

L2 2500pf aga

<u3-6 DIGITAL of gs

J3-5 +Bv aus

1, crs 47545 |B. al 47024

(a) pis et £% a £37 i) a “d 220

47024 we 4 . - cAa < 3-4 +300

4 | {v2 -8.2v

“‘T

U3 412 8

U2 247 16

24

4 ut | 7ia5}) 44 |, - DESIG| TYPE |+4.9V| -8.2V

GND

16.3 DP9000 SCHEMATIC DIAGRAM LED

Page 34 read from a scan by OCR; it may hold reading errors

17.0 SPECIFICATIONS: BSCP PROCESS SIGNAL CONDITIONER

17.1

17.2

GENERAL

This option board, identified as BSCP (DP2000-P or DP9000-P), provides extensive offset and scaling capability, permitting a wide selection of readout span for input current spans of 1 to 50 mA FS or for input voltage spans of 0.5 to 20 V FS. The DP2000 or DP9000 prefix is determined by the main assembly board used with the BSCP option board.

Selection of gain and offset ranges is made by push-on jumpers between 0.025" square pins on 0.1" centers. Six current ranges and six voltage ranges are provided, as well as push-on jumpers for offset and polarity to preserve the resolution and adjustability of the offset and gain potentiometers.

Formulas and computation procedures are supplied for calculating the proper model number based on the desired HI and LO input values and the top and bottom readout values chosen.

The BSCP has a restricted common-mode range: the BSCS strain gauge signal conditioner is recommended for signals with sizeable common-mode levels.

DP2000-P & DP9000-P: PROCESS SIGNAL INPUT SPECIFICATIONS

Configuration Single-ended, meter ground common to input LO.

Polarity Bipolar

Span Ranges Internally selectable by push-on jumpers 0 to 25% 25 to 50% 50 to 75% 75 to 100%

Zero Offset Ranges Internally selectable by push-on jumpers -215 to -77 mV -77 to +54 mV +54 to +190 mV +165 to +295 mV

Fine Zero Offset 50% of full scale minimum (front panel adjustment by potentiometer)

30

Page 35 read from a scan by OCR; it may hold reading errors

VOLTAGE RECEIVER

Readout Range

-1999 to +1999 (DP2000-P)

Readout Range

-9999 to +9999 (DP9000-P)

Input Range 0.5 1.0 2.0 5.0 10.0 } 20.0 VFS Input Impedance 0.10 } 0.25 0.50 | 1.09 | 1.04 | 1.02 MOhms Bias Current 100 50 25 13 13 13 pA NMR at 50/60 Hz 60 56 54 50 50 50 dB Maximum Voltage: 250 Vp

DC CURRENT RECEIVER

Readout Range -1999 to +1999 (DP2000-P)

Readout Range -9999 to +9999 (DP9000-P)

Input Range 1.0 2.0 5.0 10.0 | 20.0 | 50.0 mA FS Maximum Current 16 22 35 50 70 | 112 mA Shunt Resistance 499 | 249 100 | 49.9 | 249 | 10.0 | ohms Full Scale

Voltage Drop 0.5 0.5 0.5 0.5 0.5 0.5 VFS NMR at 50/60 Hz 60 60 60 60 60 60 dB

Common Mode

Analog ground to ac power ground CMR at dc to 60 Hz CMV at dc to 60 Hz

Accuracy at 25 C

Maximum Errors DP2000-P DP9000-P

Reading Tempco

Zero Tempco

120 dB +1500 Vp per HV test +354 Vp per IEC spacing

+0.05% R +1 count +0.05% R +2 counts

+0.01% R/°C +0.01% R/°C

Warmup to rated accuracy Less than 10 minutes

31

Page 36 read from a scan by OCR; it may hold reading errors

18.0 SIGNAL INPUT CONNECTIONS (TB1) (SEE FIGURE 5)

The signal input connections for the BSCP Process Signal Conditioner are made at the standard 3-terminal barrier strip:

Terminal

Connection Signal Input 4 No Connection * 5, Signal LO * - 6 + Signal HI * +

[~ - TERMINAL SIGNAL

REAR TERMINAL VIEW FIGURE 5. SIGNAL INPUT CONNECTIONS

32

Page 37 read from a scan by OCR; it may hold reading errors

19.0

19.1

TESTS AND DIAGNOSTICS

The signal conditioner board BSCP is designed to function with a main assembly as a minimum configuration. There is no provision for testing a signal conditioner board alone.

Signal input requirements for your configuration are identified in the specifications for the BSCP signal conditioner.

Operating power and connections for your configuration are identified in the Main Assembly Sections of this manual.

NOTE: lf using Main Assembly DP2000, refer to Section DP20/DP29. lf using Main Assembly DP9000, refer to Section DP90/DP98.

Inspect the panel meter for physical damage. If damage is apparent, contact OMEGA Engineering Customer Service Department at 1-800-622-2378.

FUNCTIONAL ELECTRICAL TESTING

NOTE: Perform this test after your meter has been configured.

1. Short terminals 5 and 6 on barrier strip (TB1).

2. Apply proper power for your configuration to terminals 1, 2 and 3 on barrier strip (TB1). Display will read approximately the zero offset number (ZON) from Section 20.3.

33

Page 38 read from a scan by OCR; it may hold reading errors

20.0 CONFIGURATION PROCEDURE

20.1

20.2

GENERAL

Use this procedure to determine the configuration of the BSCP Process Signal Option.

Configure the unit using the push-on jumpers provided separately or already positioned on the pin forests. Pin forest designations are shown with each configuration chart.

GLOSSARY

The chart below explains various terms which appear throughout the following procedure:

Voltmeter Range Input Range, Full Scale Selection For DP2000-P & DP9000-P

VM1 +190/235 mV VM2 +233/350 mV VM3 +348/450 mV VM4 +440/580 mV VM5 +575/850 mV VM6 +800/1030 mV VM7 +1.0/1.5 V VM8 +1.4/1.8 V VMQ9 4+1.7/2.5 V VM10 +2.4/3.1 V VM11 +3.0/4.5 V VM12 +4.4/5.1 V VMi3 +5.0/6.5 V VM14 +6.4/9.6 V VM15 +9.5/11.0 V

Current Range Input Range, Full Scale

Selection For DP2000-P & DP9000-P AM1 4+370/470 uA AM2 +460/700 uA AM3 +690/800 uA AM4 +790/940 uA AM5 +935/1400 uA AM6 +1.39/1.91 mA AM7 +1.90/2.35 mA AM8 +2.33/3.50 mA AM9 +3.48/4.00 mA AM10 +3.90/4.70 mA AM11 +4.65/7.00 mA

34

Page 39 read from a scan by OCR; it may hold reading errors

AM12 +6.90/7.80 mA AM13 +7.70/9.40 mA AM14 4+9.35/14.0 mA AM15 +13.9/19.1 mA AM16 +19.0/23.5 mA AM17 4+23.3/35.0 mA AM18 +34.0/51.0 mA Zero Offset Output Count Offset Range Selection DP2000-P DP9000-P ZO1 -3000/-1895 -15000/-9475 Z02 -1900/-600 -9500/-3000 ZO3 -605/+870 -3025/+4350 ZO04 +865/+2100 +4325/+10500 Abbr Definition ZON Zero Offset Number LI Lower Input Number Ul Upper Input Number LD Lower Display Number UD Upper Display Number G Gain in Counts/Input 20.3 SELECTION lf the Input is to be configured as a:

- Voltmeter, proceed to Section 20.3.1

- Ammeter, proceed to Section 20.3.2

- Voltage Receiver, proceed to Section 20.3.3 - Current Receiver, proceed to Section 20.3.4

20.3.1 Voltmeter Range Selection VM1,2,3,4,5,6,7,8.9,10,11,12,13,14,15 *

1.

3.

Specify the magnitude of the largest + or - input voltage:

DP2000-P: VM = DP9000-P: VM =

Volts (For a 2000 count reading) Volts (For a 10000 count reading)

Select the lowest range where VM (from step 1) is equal to or less than the limit of that range.

VM1 = 0.235 V VM2 = 0.350 V VM3 = 0.450 V VM4 = 0.580 V VM5 = 0.850 V VM =

VM6 = 1.030 V VM7 = 1.500 V VM8 = 1.800_V VM9 = 2.500°V

VM10 = 3.100 V This selection will be use in configuring the

Proceed to Installation (Section 20.4).

VM11 = 4.500 V VM12 = 5.100 V

VM13 = 6.500 V VM14 = 9.600 V VM15 = 11.00 V meter (Section 21.1).

* Range Selection assumes that the meter has an offset range (ZO3) selected which allows a shorted input to read 000 on the DP2000-P display and 0000 on the DP9000-P display.

35

Page 40 read from a scan by OCR; it may hold reading errors

20.3.2 Ammeter Range Selection (AM1,2.3,4,5,6,7,8,9,10,11,12,13,14,15,16,17,18) *

20.3.3

1.

Specify the magnitude of the largest + or - input current.

DP2000-P = AM DP9000-P = AM

Select the highest current range where AM is equal to or less than the limit of that range.

mA (For a 2000 count reading) mA (For a 10000 count reading)

AM1 = 0.470 mA AM7 = 2.350 mA AM13 = 9.400 mA AM2 = 0.700 mA AM8 = 3.500 mA AM14 = 14.00 mA AM3 = 0.800 mA AM9 = 4.000 mA AM15 = 19.10 mA AM4 = 0.940 mA AM10 = 4.700 mA AM16 = 23.50 mA

AM5 = 1.400 mA AM11 = 7.000 mA AM17 = 35.00 mA AM6 = 1.910 mA AM12 = 7.800 mA AM18 = 51.00 mA

AM = This selection will be used in configuring the meter (Section 21.2).

3. Proceed to Installation (Section 20.4).

Voltage Receiver Selection P1 thru P7

1.

Using the Input Span Range and the required display readings, calculate the zero offset required (with polarity).

(L!) Lower Input = (UI) Upper Input = (LD) Lower Display = (UD) Upper Display =

When specifying the lower input (LI) and upper input (UI), the maximum display reading for a DP2000-P is +1999. The maximum display for a DP9000-P is +9999.

ZON = _(LDx Ul - (UDxLI) = (UI - LI)

Select a Zero Offset Range (ZO) from the appropriate OMEGA series where the Zero Offset Number (ZON) falls between the minimum and maximum numbers of that range.

DP2000-P DP9000-P ZO1 = -3000/-1895 ZO1 = -15000/-9475 ZO2 = -1900/-600 ZO2 = -9500/-3000 ZO3 = -605/+870 ZO3 = -3025/+4350 ZO4 = +865/+2100 ZO04 = +4325/+10500 ZO = This selection will be used in configuring the meter (Section 21.5).

36