CN-2000 Series Programmable Temperature Controllers Operator's Manual, pages 51 to 60

Pages 51 to 60 of 75

The original document is the PDF (2.2 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.

For the Omega CN2001

Page 51 read from a scan by OCR; it may hold reading errors

STOP BIT(S) {PARITY! SEVEN DATA BITS | START | BIT BiT

: Lz fe¢[sf[«f3sT727 fi [ o | Zz MSB

Figure 5-12. Serial Data Format

5.8.2 Data Protocol

The Model CN-2000 Controller is always the passive (listener) link in this protocol. Transmission is always initiated by the host computer. The host sends a command or a command and data'to the Model CN-2000 Controller. The Model CN-2000 responds within 100 ms. either with system status or with system status and data.

The COMMAND sent by the host must be in the following format except that CKSUM is optional with the host. The CN-2000 will ignore a lack of CKSUM if none is provided by the host computer. The CN-2000 will always respond with a CKSUM even if it is not used. See examples on Page 41 and 42.

TO READ DATA:

* UNIT#: R PARAMETER : CKSUM CR TO WRITE DATA: * UNIT#: W PARAMETER / DATA : CKSUM CR

NOTE The controller will ignore LF if transmitted after CR.

The Model CN-2000 witl RESPOND within 100 ms. in the following format:

RESPONSE TO READ COMMAND: $ STATUS : DATA : CKSUM CR LF

RESPONSE TO WRITE COMMAND: $ STATUS : CKSUM CR LF

NOTE

Spacing is included above purely for clarity. Actual data transmission must not include spacings. NULL characters HEX ‘‘00"’ may be sent within the context of the message to the host from the CN-2000 and should be ignored, but must not be sent to the CN-2000.

46

Page 52 read from a scan by OCR; it may hold reading errors

5.8.3

Data Protocol Definitions See examples in Section 5.8.8 and 5.8.10 of this manual. *

~ Standard ASCII character usecl by the host computer to initiate a commancl.

- Standard ASCII character used as a field separator.

/, Standard ASCII character used as a separator between the PARAMETER mnemonic and the Tome nn data irra WRITE COMMAND,.~ -- eo

$, Standard ASCIl character used by the Model CN-2000 as the first character in a response.

UNIT #, ASCII O to 63. The unit address ID is established via binary weighted internal switches on the Model CN-2000 communications board. The unit address can be omitted in transmis- sions when only one unit is involved and address switches are set to all zeros.

R, READ command WwW, WRITE command

PARAMETER - Controller variable shown as a three-character ASClt mnemonic. See Section 5.8.4. ONLY UP- PER CASE CHARACTERS MAY BE USED.

CKSUM - Two ASCII coded hexadecimal characters representing an eight-bit checksum formed by adding each byte of the string preceding the checksum into an eight-bit accumulator and ignoring any overflow. The resulting eight-bit sum is sent as two hexadecimal characters. The high order four bits are the first character and the low order four bits are the last character. CKSUM is optional with the user.

CR ~ Carriage Return

LF, Line Feed

STATUS - Two ASCIl coded hexadecimal characters defining the system status. See paragraphs 5.8.5 and 5.8.6.

DATA - Astring of ASCII digits, including minus sign where appropriate, representing the numerical value of the PARAMETER specitied. Limited to four character spaces.

47

Page 53 read from a scan by OCR; it may hold reading errors

5.8.4 Parameter Designations

LIMITS: All values must be set in as

Function | MNEMONIC Definition whole numbers. See Notes below. Operator PvO Process Variable DO NOT WRITE PROCESS VARIABLE. Functions csp Control Set Point MUST BE WITHIN SYSTEM SPAN LIMITS.

OP1 Output 1, PRimary Output | READ only, see Note 1

OP2 Output 2, SECondary READ only, see Note ’

Output Primary DBI Dead Band 1 25, 50 or 100 for % of SPAN? Control cTt Cycle Time 1 1 to 60 for seconds of Cycle Time ji |. PBI. | Proportional Band 1___|__1.t0 200-for % of SPAN es re

OUTPUT 1 RE1 Reset 1. : - 1 to 2000 for-R'M or 0 for RESET OFF34.

RAt Rate 1 1 to 500 for Minutes or O for Rate OFF3 Secondary ASP AUXiliary Set Point Secondary set point must be within SPAN. Control DB2 Dead Band 2 (AUX DB) 25, 50 or 100 for % of SPAN3 cT2 Cycle Time 2 1 to 60 for seconds of Cycle Time

PB2 Proportional Band 2 1 to 200 for % of SPAN OUTPUT 2 RE2 Reset 2 1 to 2000 for RM or O for RESET OFF3

RA2 Rate 2 1 to 500 for Minutes or O for Rate OFF? Alarms AL1 Alarm 1 Set Point Must be within system span limits (Option) AL2 Alarm 2 Set Point Must be within system span limits Timers* R1H Relay 1 Hours Oto 999 for hours (Option) RIM Relay 1 Minutes Oto 59 for minutes Note 2

R2H Relay 2 Hours 0 to 999 for hours (see Note 2}

R2M Relay 2 Minutes O to 59 for minutes Status PSW Process Status

CSW Control Status

“Timers can be set to operate in Hours/Minutes or Minutes/Seconds at time of initial calibration. For Minutes/Seconds operation R1H and R2H will read in minutes and R1M and R2M will read in seconds.

NOTES

1. Output power level is communicated as a digital value from O to 4095. To read % output, divide transmitted value by 40.95.

2. Set in hours and minutes without colon, i.e., 9959 for 99:59.

3. Set desired value as a whole number. The CN-2000 internal logic will multiply your number by 0.01 to yield correct decimal value.

48

Page 54 read from a scan by OCR; it may hold reading errors

5.8.5 Control Commands~ - Process Status (PSW)

The USER may READ or WRITE the Process Status (mnemonic PSW) as two ASCII HEX characters. Before WRITING a new command, you must first read the existing command data bits. All data bits you do not wish to change MUST be repeated in the new command including bits designated RESERVED. DATA BITS[ 7] 6] 5, 4]3 1211 ]f¢ 4] MSB LSB BIT 0 STATUS 1 STATUS _ 7 ____ Stop Start - 7

6 Manual Auto Control

5 Normal Input Open (READ only) 4 Normal Data Lost (READ only) 3 "REM" flashing "REM" full on

2 Reserved Reserved

1 ALM/Timer 1 OFF ALM/Timer 1 ON x Reserved Reserved

5.8.6 Control Commands- - Control Status (CSW)

The USER may READ or WRITE the Control Status (mnemonic CSW} as two ASCII HEX characters. Before WRITING a new command, you must first read the existing command data bits. All data bits you do not wish to change MUST be repeated in the new command including bits designated RESERVED.

pata Bits[7]e6 [5,4 [312 Tilf]

MSB LSB

BIT 0 STATUS 1 STATUS

7 Reserved Reserved

6 Reserved Reserved

5 Reserved Reserved

4 Normal Timer override (event alarm only) 3 Reserved Reserved

2 Reserved Reserved

1 Reserved Reserved

& Reserved Reserved

5.8.7 Status Response

Every command from the host computer causes the Model CN-2000 to generate a two ASCII HEX character status report as part of its response. Normal state (0) is assumed to be normal controller operation. pata BITSs[ T] Cc] ST DTP TR aTul MSB LSB BIT ABNORMAL CONDITION *1* T Transmission Error Cc CKSUM Error on received Data Ss Syntax Error or Illegal Input D Communications Turned OFF P Parity Error on received Data R INPUT OPEN A Alarm Condition (Relay 1) U Reserved

49

Page 55 read from a scan by OCR; it may hold reading errors

5.8.8 Protocol Example - Read

Acommand to READ the measurement of Unit Number 23 would be:

READ COMMAND : KR PV@ : 2B CR

* 23 To initiate command, ___J | | Unit Number 23 |

Separator R for READ command Mnemonic for data Separator ~CKSUM- a . _ a eS pees Carriage Return

The Model CN-2000 response to the READ command if the measurement were 542 degrees would be:

RESPONSE 542 : 93 CR LF

$ OO : To initiate response | | T Status, see paragraph 5.8.7 Separator Requested data Separator CKSUM

Carriage Return Line Feed

CKSUM Computation is as follows (for example in this section):

READ COMMAND RESPONSE REPLY CHAR HEX CHAR HEX * 2A $ 24 2 32 ") 3¢ 3 33 G 30 : 3A : 3A R 52 5 35 P 50 4 34 Vv 56 2 32 i) 39 : 3A : 3A TOTAL 22B TOTAL 193 BINARY 0010 0010 1011 BINARY 0001 1001 0011 DISCARD _l | DISCARD _| | | lst HEX = 2 lst HEX = 9. | 2nd HEX = B 2nd HEX = J

Note: Parity bits are not included in the CKSUM

50

Page 56 read from a scan by OCR; it may hold reading errors

5.8.9

WRITE COMMAND * 17 : W CSP/234 : OB CR = r

To initiate command is | 7

Unit Number 17

Separator

W for WRITE command Mnemonic for setpoint Separator

_ - - New setpoint value ~ . an pO, Separator CKSUM Carriage Return

Protocol Example - Write

A command to WRITE (enter new data) into Unit Number 17 to make the set point 234 degrees would be:

The Model CN-2000 response to the WRITE command would be:

RESPONSE $ 00 :: F8 CR LF

To initiate response | |

Status, see paragraph 5.8.7, _ - -~. Separator Separator

CKSUM

Carriage Return Line Feed

NOTE

The beginning of data and end of data (field) separators (:) will both be present even if status only (NO DATA) is being transmitted. 5.8.10 ASCII Table “ S S y & S CY s x YEE YTV II VIII IIS MSB 0 1 2 3 4 5 6 7 8 9 al|ajfec ) ej fr Nut | SOH | stTx | Erx | EoT |ENQ |ACK | BEL | @s | Ht | LE | vt | FF | cR | so | st 000 | 0 DLE | OC1 | OC2 {| C3 | OCA |NAK | SYN {ETB [CAN | Em jsue | esc | Fs | GS | AS | US 0001; 1 sey rt" be) S$ 1) HP wl Pe pe pep te] yy tod. de oor, 2 : ° 1 2 3 | 4 5 6 r | 8 9 ;[<]2e |] >]? oo; 3 T H a 8 c ° £ F G n ' J K L “ N ° oo} « | ©@ y , P Q R Ss T u v w x Y z { \ 1 A - O11) $ I \ a b c d e t 9g hn i i k ’ m n o O110) 6 . qa e s t u ’ w x y z { ' } ~ | OE oui) 7

BINARY, HEX, ASCH

51

Page 57 read from a scan by OCR; it may hold reading errors

SECTION 6 CALIBRATION

6.1

The CN2000 Series Controllers are delivered fully calibrated and ready to use. Recalibration is not normally required or recommended but it may be necessary in order to meet plant operating standards or to recover from extraordinary circumstances such as DATA LOST/PLEASE CAL.

The calibration procedure for the CN2000 Controllers include both a CONFIGURATION procedure and a REFERENCE procedure. The CON- FIGURATION procedure refers to those steps that must be performed to tell the microprocessor details of the application including input,

~output, control action,_alarms,display-units;-span-and-time. base. -.---

These details are to be keyed through the front keypad in a structured sequence (refer to Calibration Loop, Figure 6-1.) The REFERENCE procedure is the more traditional calibration using an external reference source. A structural program also exists for this calibration (see Calibration Loop, Figure 6-1). Please note that a type J thermocouple is required for the thermocouple reference calibration.

Controller CONFIGURATION calibration can be performed at the normal controller installation or on a bench. Only a properly protected (cover exposed terminals) ac power source and an appropriate input is necessary. If input is not convenient, a jumper between Terminals F and H is adequate for CONFIGURATION calibration. REFERENCE calibration requires a precise calibration source for the input signal.

COMPUTER DEVICES ARE NON-FORGIVING. KEY SE- QUENCES AND INSTRUCTIONS MUST BE FOLLOWED PRECISELY. DO NOT SKIP ANY STEP.

ENTERING CALIBRATION PROGRAM

Asecurity code protects the calibration procedure. To enter the calibration mode push the three keys as indicated in Figure 6-2 simultaneously. Note that on the CN2000 Series the upper left key is not shown on the keypad but it does exist under the front panel. A front display “CAL indicator will illuminate. If it does not, push the RETURN key and then push the 3 key code again.

- Figure 6-2. Entering Calibration Loop

52

Page 58 read from a scan by OCR; it may hold reading errors

ENTER CAL CONFIGURATION CALIBRATION LOOP HERE UNITS SELECTION TIME SELECTICN PN XX Xx [PC UNITS Uu c UNITS uU LPC UNITS uu UNITS uu XX XX XX TIME him TIME him time Hem EEC | tome wes y YES ¥ YES LO AND HI SPAN SELECTION PN ROUTINE SEE MANUAL FOR DETAILS Lo span PPC gs ur span PC OR XXXX UU XXXX UU CODE: DECIMAL POINT AND DISPLAY VALUE SELECTION ESyy DECIMAL LO VAL F-PC HI VAL PC OR, - -} - -, PLACES Xx -XXXX - _UU-$ fFRRX |---| - CODE. NO REFERENCE CALIBRATION ANALOG INPUT CALIBRATION REFERENCED TO LO AND HI VAL LEGEND INPUT Pc INPUT *% = NUMBER LO CAL HI CAL U_ = UNITS PC = PARAM KE’ THERMOCOUPLE CALIBRATION INPUT PC 3 T/C IN 50.000MV ICE BATH INTERNAL RTD CALIBRATION DECISION. INPUT PC input ES ACTUAL 100 ouMS 300 OHMS DISPLAY

ANALOG REMOTE SETPOINT CALIBRATION rc y

REFERENCED TO LO SPAN AND HI SPAN YES soreorar INPUT PC rnpur FY ETPO LO REMSP HI REMSP 0 CAL COMPLETE RETURN

OPERATOR LOOP

CAUTION: DO NOT TURN OFF POWER WHILE IN CAL LOOP.

Figure 6-1. Calibration Loop

53

Page 59 read from a scan by OCR; it may hold reading errors

6.2

PART NUMBER

PN 11.23 | The PART NUMBER is the first CAL loop display If this 445677 display does not appear, push the LAST key until it is on the display. This 10 digit PART NUMBER fully defines the controller. PART ] r 7 7 | NUMBER, - ! | i - | | \ -d Piven d DOo © S © -.4.-- - -- Medel oe, ; A Input Eo 2000 No communications ° -J_ thermocouple ~ E1 2000 Remote Analog Setpoint 01 K_ thermocouple E2 2000 R$232C non-isolated 02 R thermocouple E3 2000 RS232C isolated 03 $ thermocouple E4 2000 RS422 non-isolated 04 T thermocouple E5 2000 RS422isolated oS 5 Cerro E6 i mi 2000 20 mA Current (isolated) 08 PLATINELINTIC EA 2000* Nocommunications 03 Ni/N' 18% Moly T/C, €B 2000* Remote Analog Setpoint 10 WE%Re/W26%Re TC EC 2000* RS232Cnon-isolated 11 W3%Re/W25%Re T/C ED 2000* RS232C isolated 12) WIW26%Re TIC EE 2000* RS422non-isolated 20 RTD 100 ohm pt. EF 2000* RS422isolated 0 aie wo ohm pt. (0.1 res} * A isolated -5 volts D EG 2000 20 mA Current (isolated) 60 4.20mADG * With AT (SELF TUNE) option. ; . 5 Alarm/Timer 1 2 Primary Output 0 noalarm O NOPRIMARY 1 Hi process 1. PRI. HEAT SSR ON/OFF 2 LOprocess 2 PRI. COOL SSR ON/OFF 3 Ht deviation 3 PRI. HEAT SSR PROP. 4 LO deviation 4 PRI, COOL SSR PROP. 5 ON Timer 5 PRI. HEAT 4-20 mA DC 6 OFF Tlmer 6 PRI. COOL 4-20 mA DC 6 Alarm/Ti 2 7 PRI. HEAT 0-5 VDC 0 onla imer 8 PRI. COOLO-5 VDC no alarm 1 HI process 3 Secondary Output 2 LOprocess QO NOSECONDARY 3. Hideviation 1 SEC. HEAT SSR ON/OFF 4 lOdleviation 2 SEC. COOL SSR ON/OFF 5 ONTimer 4 SEC. COOL SSR PROP. 6 OFF Timer 6 SEC. COOL 4-20 mA DC : 7 SEC. COOLO-5 VDC 7 Options 00 Nooptions XX Any special

Note: Units with “Auto/Manual” option (reptacing the “Start/Stop” option) have code “Z” (instead of code “E”) in the part number. .

54

Page 60 read from a scan by OCR; it may hold reading errors

6.3

6.4

The CN2000 Controllers have been manufactured with specific hardware that determines input and output capability. Tne PART NUMBER defines this capability. The PART NUMBER also defines some items that can be changed by a keypad instruction without any hardware change. For example: In a controller provided with thermocouple input, the type of thermocouple can be changed by a keypad instruction. However an input cannot be changed to an RTD or process input unless a hardware change is also made. CONFIGURATION calibration deals specifically with those items that can be changed strictly by a keypad instruction.

PART NUMBER CHANGES... Snes

DIGIT 1: Any number change requires a hardware change.

DIGIT 2: Numbers 1, 2, 3, 4 are interchangeable by keypad instructions only.

Number 5 and 6 are interchangeable by keypad instructions only.

Number 7 and 8 are interchangeable by keypad instructions only.

Any other change requires a hardware change too.

DIGIT 3: Numbers 1, 2 and 4 are interchangeable by keypad instructions only. Any other change requires a hardware change too.

DIGIT 4: Numbers 00 thru 12 are interchangeable by keypad instructions only. Any other change requires a hardware change too.

DIGIT 5: All numbers except O are interchangeable by keypad instruction only.

DIGIT 6: All numbers except O are interchangeable by keypad in- .

struction only.

DIGIT 7: This number has no effect on the microprocessor commands. It is used to identify specials and should only be changed by authorized personnel.

Most recalibration requirements will use the original part number found on the instrument labels. If you make a change in the PART NUMBER, be sure you write the new number down on the back cover of this manual for future reference.

If you do not need to reset or change the part number, push the PARAM DISPLAY key to advance the display and proceed to paragraph 6.5

If you do want to change the part number, proceed with paragraph 6.4.

SETTING THE PART NUMBER

An additional security code protects the part number. To change the part number push the two keys indicated in Figure 6.3 simultaneously.

55