Page 71

OaktonLog

Chapter 1

If you set it to Timer run mode, OaktonLog will wait in standby mode displaying time.

, and will begin recording at the specified

71

Page 72

Chapter 1

OaktonLog

2. Saving setup When setting up multiple OaktonLogs or Cradles use the Save Setup option in the Setup dialog. 1. After you have finished the settings selection of the first OaktonLog or Cradle and before sending the setup command, click Save Setup to save all the setup settings. 2. Connect another OaktonLog or Cradle to the PC, click

3. 4. 5.

72 on the main toolbar, then click Load Setup Setup to load your setup settings. Change the Cradle ID. Change the comment (optional). Click Send Setup.

Page 73

OaktonLog 3.

Chapter 1

Start recording

Click Run every time you want to start a new recording. The Run command erases all previous data in the OaktonLog’s memory and begins recording. Note: If you setup the OaktonLog using the Setup command, it will automatically begin recording, and you don’t have to click Run

73

Page 74

Chapter 1 4.

OaktonLog

Stop recording

Click Stop to stop recording. In Stop mode, OaktonLog keeps all recorded data but does not record new data. Use this mode to save battery power. 5. Sensor calibration A new OaktonLog comes fully calibrated. After a period of long use, however, you may want to recalibrate the humidity or the temperature sensors. The humidity calibration affects both OaktonLog and MicroLab and should be carried out while OaktonLog is connected to the PC. The temperature calibration affects only the MicroLab software and can be carried out regardless of whether OaktonLog is connected to the PC.

74

Page 75

OaktonLog

Chapter 1

Calibration password To prevent accidental change of the calibration, the calibration procedure is protected by a password. The default password is: 1234. To change the password: a.

Click Logger on the menu toolbar, then click Calibration.

b.

Click Change Password to open the Change Password dialog box.

c.

Enter the current password in the Current Password edit box.

d.

Enter the new password in the New Password edit box.

75

Page 76

Chapter 1

OaktonLog e.

Enter the new password a second time in the Confirm New Password edit box to confirm your new password.

f.

Click OK.

Note: The password must include at least 4 characters and is case sensitive. Humidity calibration To calibrate the humidity sensor you will need a humidity chamber.

76 a.

Connect OaktonLog to the PC b.

Set up OaktonLog to record every 10 seconds (see page

Page 77

OaktonLog

Chapter 1

62). c.

Click Logger on the menu bar, and then click Calibration.

d.

Enter the calibration password, and then click OK.

e.

Select Humidity in the Choose sensor drop list

77

Page 78

Chapter 1

78

OaktonLog f.

Click Default to restore the original values g.

Disconnect OaktonLog from the PC and insert it into

Page 79

OaktonLog

Chapter 1 the humidity chamber. h.

Set the humidity chamber to the first reference value. Wait until the humidity level is stabilized and write down OaktonLog’s reading.

i.

Repeat the last step with the second reference value.

j.

Connect OaktonLog to the PC.

k.

Enter the two OaktonLog values into the OaktonLog value edit boxes.

l.

Enter the two reference values into the Reference value edit boxes.

m. Click Calibrate. To restore default calibration, click Default.

79

Page 80

Chapter 1

OaktonLog

Temperature calibration

80 a.

Click Logger on the menu bar, and then click Calibration.

b.

Enter the calibration password, and then click OK.

c.

Select Temperature or Ext. Temperature in the Choose sensor drop list.

d.

Click the Default button to restore the original values.

e.

Enter the two OaktonLog values into the OaktonLog value edit boxes.

f.

Enter the two reference values into the Reference value edit boxes.