Page 81
OaktonLog
Chapter 1 g. Click Calibrate. To restore default calibration, click Default. 6. Defining a custom sensor You can use OaktonLog with any sensor that has a 0 – 20mA current output or a 0 – 10V voltage output. a.
Click the Logger menu and then select Define new sensors to open the Define New Sensor dialog box.
b.
Click Add to add a new sensor to the list.
c.
In the Based on drop list, select an external sensor that matches your sensor’s output d.
Type the sensor’s name in the Sensor Name edit box.
81
Page 82
Chapter 1
82
OaktonLog e.
Type the sensor’s unit in the Sensor Unit edit box.
f.
In the Calibration Values section, enter two values of your sensor that correspond to the base sensor values. For example, see the definition for a 0 – 100mbar pressure sensor whose output is 0 – 20mA:
Page 83
OaktonLog
Chapter 1
Figure 4: Define New Sensor g. Click OK. The new sensor will then appear in the External Sensor list in the Setup dialog box.
83
Page 84
Chapter 1
OaktonLog
7. Communication setup Communication between the PC and an OaktonLog takes place automatically whenever you send a command to OaktonLog. However, the Communication Setup dialog box can be used for more advanced communication options. Click on the Logger menu and then select Com setup to open the Communication Setup dialog box: If you are not using a receiver and cradle, uncheck the Search for Receiver and Cradle checkbox in order to speed up the search.
84
Page 85
OaktonLog
Chapter 1
Figure 5: MicroLab’s Communication Setup dialog box The checked COM ports are available for communication. Click Try to connect to establish communication. Click Work offline to work with saved files.
85
Page 86
Chapter 1
OaktonLog
1.4.6. Automatic Daily Download You can program all the OaktonLogs on your line to automatically transmit their data to a PC every day, at a preset time. To work with this option you will need to mount the OaktonLogs onto wireless Cradles (see page 166) and connect a receiver to the PC (see page 178). 1.
Prepare the OaktonLog.
To prepare a Cradle with OaktonLog system for daily download:
86 a.
Connect the cradle to the PC (see page 168).
b.
Setup the OaktonLog and the cradle (see page 62).
Page 87
OaktonLog
Chapter 1
Notes: 1. Enter an identification number for the cradle in the Unit ID edit box. The number should be an integer between 1 and 200 and every cradle should have a different ID number. (We recommend attaching a label with the ID number to every cradle.) 2. The recording interval must be equal to or greater than 1 minute for recording data from one sensor, 2 minutes for 2 sensors and 3 minutes for 3 sensors. c.
Check the Daily download checkbox.
d.
To set the download time and file location click Daily download at the bottom of the setup dialog to open the Daily download dialog box.
87
Page 88
Chapter 1
OaktonLog e.
Enter the desired download time in the Download start time box. MicroLab automatically calculates and sets the cradles’ transmission times according to their ID numbers, so that the cradles will transmit the data successively.
88 f.
MicroLab stores the data files in the following folder by default: C:\Program Files\Fourier Systems\MicroLab\OaktonLog Data\Daily download. If you want to store your data in another location click Browse and navigate to the desired folder.
g.
Click OK to return to the setup dialog.
h.
Click Send setup.
i.
Repeat this procedure (except for steps d to g which
Page 89
OaktonLog
Chapter 1 should be performed only once) with every cradle. Remember to assign a different ID number to each cradle. j.
After you have finished setting up all your cradles, connect the RF receiver to the PC (see page 178).
k.
Click Logger on the menu bar, then click Com setup to open the Communication setup dialog.
l.
Make sure that the Search for Receiver and Cradle checkbox is checked.
m. Click Try to connect to establish communication with the receiver.
89
Page 90
Chapter 1 n.
OaktonLog
Place the cradles at the desired data logging locations.
2. Automatic download At the download start time the cradles should be within 300m of the receiver and in its line of sight. The cradles will automatically transmit the data one after the other according to their ID number. The time interval between successive transmissions is two minutes. MicroLab automatically saves the data after every transmission under the name: daily download [&date]. After the daily download has been completed, the file will automatically be closed. Every 24 hours MicroLab creates new data file.
90