Page 61

OaktonLog 10.

Chapter 1

Change the temperature units to change the temperature scale Click Toggle °C/°F button from Fahrenheit to Celsius and vice versa. Note: To change the units in the OaktonLog display, use the Setup dialog box (see page 62). 11.

Export data to Excel

Click Export to Excel to export the currently displayed data to an Excel spreadsheet. MicroLab will open a new Excel workbook displaying the data along with the OaktonLog info, including the OaktonLog name, serial number and alarm levels setup.

61

Page 62

Chapter 1

OaktonLog

12. Copy the graph as a picture You can copy the graph to the clipboard as a picture and then paste it to other Windows programs such as Word and PowerPoint: 1. On the Graph menu, click Copy graph. 2. Open the destination file. 3. In the destination file, right click and select paste. 1.4.5.

Program OaktonLog

1. Setup Use the Setup dialog box to view or to change the OaktonLog and the Cradle settings

62

Page 63

OaktonLog

Chapter 1

If you intend to setup more then one OaktonLog or Cradle use the Save setup tool (see page 72). •

Connect OaktonLog or the Cradle to the PC

•

Click Setup to open the Setup dialog box. The dialog box is made up of five sections: a.

OaktonLog info

Comment Click the edit box and type a name that will serve to identify the specific OaktonLog (e.g. its location).

63

Page 64

Chapter 1

OaktonLog

S/N Displays the OaktonLog’s serial number Battery Level If the indicator is in the red zone the battery should be replaced (see page 34). b.

Cradle info

Cradle ID The cradle’s identification number sets the cradle’s transmission time in daily download mode. Battery Level If the indicator is in the red zone the battery should be replaced (see page 171).

64

Page 65

OaktonLog c.

Chapter 1

Setup

Temperature Click the Temperature check box to activate the internal temperature sensor Humidity Click the Humidity check box to activate the internal humidity sensor External Click on the External check box to activate the external sensor option, and then select a sensor in the drop list. If you want to use a sensor that is not found in the drop list, you can define new sensor (see page 81). The sensors’ current readings will appear in the labels next to the sensor name once OaktonLog begins logging for the first time.

65

Page 66

Chapter 1

OaktonLog

Temperature unit To set the temperature unit in the OaktonLog display, select the option you want (Celsius or Fahrenheit). Interval Determines the logging interval, or the time interval between successive data recordings. The time format is hh:mm:ss. Set the time setting to select a recording time interval from between 10 seconds and 2 hours. For example, to set a time interval of one hour, five minutes and thirty seconds, click the hours (hh) and type 1 or use the arrows to select 01, then click the minutes (mm) and type 5 or use the arrows, and finally, click the seconds (ss) and type 30 or select 30 using the arrows.

66

Page 67

OaktonLog

Chapter 1

Timer run Click the Timer run check box if you want OaktonLog to start recording at a predetermined time. This option is convenient if you are using several OaktonLogs at once and want them to all begin logging at the same time. Use the time and date selectors to set the start time. Cyclic run In Cyclic run mode, OaktonLog overwrites the old measurements (starting with the oldest one) once the OaktonLog’s memory is full. Click the Cyclic run check box to operate in this mode. If the check box is clear, OaktonLog will operate in Normal run mode and will stop recording when the memory is full.

67

Page 68

Chapter 1

OaktonLog

Daily download Check the Daily download checkbox to enable automatic daily download (see page 86 for details). d.

Alarm levels Type in the desired minimum and maximum alarm levels. If OaktonLog records a reading that exceeds either of these levels, the OaktonLog’s LCD will start to flash and the cradle’s alarm will sound. To stop the LCD’s flashing, press either of its two keys. To stop the cradle’s alarm, press both the OaktonLog’s keys simultaneously.

68

Page 69

OaktonLog

Chapter 1

The default alarm levels are the lower and upper ends of the sensors. Click Cancel Alarm to restore the default levels. e.

Cradle time settings

Workday hours Use the up and down arrow buttons to set the daily period when you want the Alarm and/or the Daily Download to be active, or type the desired period in manually. Alarm delay Click the drop list to select the time delay between the time OaktonLog records a reading that exceeds the alarm levels and the time the alarm will sound.

69

Page 70

Chapter 1

OaktonLog

Alarm duration Click the drop list to select the time for which the alarm will sound before it stops automatically. Complete the setup Click Send setup to send the new settings to OaktonLog, which will complete the setup. Click Cancel if you don’t want to change the setup. Note: The Send setup command erases all existing data in the OaktonLog.

70