The text of a manual · part 12 of 26
NCR 7870 Scanner/Scale Release 1.0 User's Guide: Chapter 5: Calibration, pages 96 to 104
Pages 96 to 104 of 228, from NCR 7870 Scanner/Scale Release 1.0 User's Guide
The original document is the PDF (1.3 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.
For the NCR 7870
Page 96
5-10
Chapter 5: Calibration
Display
Add Weight
Ready C-5.00 lb (2.500 kg)
5.00 lb. (2.500 kg)
Ready C-15.00 lb (5.000 kg)
10.00 lb. (2.500 kg)
Remove Weight
Ready C-30.00 lb (10.000 kg) 15.00 lb. (5.00 kg) Ready C-0.00 lb (0.000 kg)
30.00 lb. (10.000 kg)
Ready 0.00 lb (0.000 kg)
Note: Use the same kilogram weights for both 9.995 kg and 13.995 kg scale units. 6. Press Reset/Scale Zero button. Display will indicate Ready C 5.00 lb (or 2.500 kg). 7. Place a 5.00 lb (or 2.500 kg) weight in center of Top Plate. 8. Press Reset/Scale Zero button. Display will indicate Ready C 15.00 lb (5.000 kg). 9. Add a 10.00 lb (2.500 kg) weight in center of Top Plate. Total weight on scale should be 15 lbs. (5.00 kg). 10. Press Reset/Scale Zero button. Display will indicate Ready C 30.00 lb (10.000 kg). 11. Add a 15.00 lb (5.000 kg) weight in center of Top Plate. Total weight on scale should be 30.00 lbs (10.000 kg). 12. Press Reset/Scale Zero button. Display will indicate Ready C 0.00 lb (0.000 kg). 13. Remove all weights. 14. Press Reset/Scale Zero button. Verify display indicates Scale Ready 0.00 lb (0.000 kg). 15. Press and holdReset/ Scale Zero button. Record Cal and Par values. 16. The Cal value should be incremented by one, indicating that a calibration has been performed. The PAr value should remain unchanged.
Page 97
Chapter 5: Calibration
5-11
Verifying Calibration This section describes how to verify that the 7870 has been properly calibrated. No error codes should appear during calibration. The codes are shown in the table below. If a code is displayed, the calibration is invalid and a corrective action must be undertaken before attempting recalibration.
Error Code
Problem
Suspect Component
1----
Calibration error A/D output below 000h at digital to analog converter
Scale Board Load Cell
2----
Calibration error A/D output below 280h at digital to analog converter
Scale Board Load Cell
3----
Calibration error Analog to digital converter output too high at 15 kg (30.0 lb.)
Scale Board Load Cell
4----
Vibration to scale during calibration
Motors in checkstand
9----
Interference detected during calibration
Top Plate is touching checkstand, scale was jarred, or overload screw is misadjusted.
If no error codes occur during calibration, four test procedures must be performed to verify calibration. These procedures are, in order, •
Increasing Load Test ,
•
Over-Capacity Test,
•
Decreasing Load Test, and
•
Shift Test.
Page 98
5-12
Chapter 5: Calibration
Procedure 1: Increasing Load Test This test checks the scale’s accuracy by incrementally adding weight to the center of the Top Plate. Use weights that correspond to the NCR 7870 weight feature. If this test fails, recalibrate scale.
Instructions 1. Select the appropriate weight feature from the table below. Step
Weight Feature
Add Weight
1
9.995 kg 13.995 kg 30.0 lb. 9.995 kg 13.995 kg 30.0 lb.
0.1 kg 0.1 kg 0.2 lb. 2.5 kg 2.5 kg 5.0 lb.
3
9.995 kg 13.995 kg 30.0 lb.
2.5 kg 4.5 kg 5.0 lb.
4
9.995 kg 13.995 kg 30.0 lb.
2.5 kg 3.0 kg 10.0 lb.
5
9.995 kg 13.995 kg 30.0 lb.
2.495 kg 3.995 kg 10.0 lb.
2
Remove Weight
0.01 kg 0.01 kg 0.2 lb.
Display
0.1 ± 0.00 kg 0.1 ± 0.00 kg 0.2 ± 0.00 lb. 2.5 ± 0.00 kg 2.5 ± 0.00 kg 5.0 ± 0.00 lb. 5.0 ± 0.005 kg 7.0 ± 0.005 kg 10.0 ± 0.01 lb. 7.5 ± 0.005 kg 10.0 ± 0.005 kg 20.0 ± 0.01 lb. 9.995 ± 0.005 kg 13.995 ± 0.005 kg 30.0 ± 0.01 lb.
2. Follow steps shown in table for the weight feature of your scale. 3. Check readout after completing each step and verify the measured value falls within the acceptable limits. 4. If a measured value does not fall within limits, recalibrate scale. If all measured values are acceptable, the scale has passed the Increasing Load Test. Go to Procedure 2: Over-Capacity Test. Note: Do NOT remove any weights from the Top Plate.
Page 99
Chapter 5: Calibration
5-13
Procedure 2: Over-Capacity Test This test checks for the proper indication from the scale when too much weight is placed on the Top Plate. An over-capacity is indicated by dashes appearing on the display. If a failure occurs during this test, the scale must be recalibrated. Note: This test must immediately follow the Increasing Load Test. Do not remove any of the weights prior to running this test.
Instructions 1. Use the following chart to select the correct weight. 2. Follow steps on chart for adding and removing weights. Weight must be placed in center of Top Plate.
Step
Weight Feature
Add Weight
1
9.995 kg 13.995 kg 30.0 lb.
0.04 kg 0.04 kg 0.08 lb.
2
9.995 kg 13.995 kg 30.0 lb.
Remove Weight
Display
- -.- - -.- - -.- 0.04 kg 0.04 kg 0.08 lb.
9.995 ± 0.005 kg 13.995 ± 0.005 kg 30.0 ± 0.01 lb.
3. If a series of dashes are not shown, then the scale has failed and must be recalibrated. 4. The scale passes the test if a series of dashes is displayed and all measured values fall within the acceptable limits Continue to Procedure 3: Decreasing Load Test
Page 100
5-14
Chapter 5: Calibration
Procedure 3: Decreasing Load Test This test checks the scale’s accuracy when incrementally removing weight from the Top Plate. Use weights that correspond to the NCR 7870 weight feature. If a failure occurs during this test, you must calibrate the scale again. Note: This test must immediately follow the Over-Capacity Test; do not remove any of the weights prior to running this test.
Instructions 1. Select appropriate Weight Feature for your Scanner/Scale. 2. Follow the four steps shown in chart. Step
Weight Feature
1
Add Weight
Remove Weight
Display
9.995 kg 13.995 kg 30.0 lb.
2.495 kg 3.995 kg 10.0 lb.
7.5 ± 0.005 kg 10.0 ± 0.005 kg 20.0 ± 0.01 lb.
2
9.995 kg 13.995 kg 30.0 lb.
5.0 kg 7.5 kg 15.0 lb.
2.5 ± 0.00 kg 2.5 ± 0.00 kg 5.0 ± 0.00 lb.
3
9.995 kg 13.995 kg 30.0 lb.
2.5 kg 2.5 kg 5.0 lb.
0.1 ± 0.00 kg 0.1 ± 0.00 kg 0.2 ± 0.00 lb.
4
9.995 kg 13.995 kg 30.0 lb.
0.1 kg 0.1 kg 0.2 lb.
0.0 ± 0.00 kg 0.0 ± 0.00 kg 0.0 ± 0.00 lb.
0.1 kg 0.1 kg 0.2 lb.
4. If display does not read within limits shown, the scale has failed and must be recalibrated. 5. If the displayed measurements fall with the acceptable limits, the scale has passed. Go to Procedure 4: Shift Test
Page 101
Chapter 5: Calibration
5-15
Procedure 4: Shift Test This test consists of measuring a weight in the center and four positions on the Top Plate.
5
2 1
4
3
11033
Instructions 1. Place a 15.00 lb (or 5.000 kg) weight in Position 1 and record measured weight. 2. Move sequentially to Positions 2, 3, 4, and 5. Record measured weight for each position. 3. Move weight back to Position 1 and record measured value. 4. All measurements should fall within •
15.00 lb ± 0.01 lb (for 30.00 lb scales)
•
5.000 kg ± 0.005 kg (for 9.995 and 13.995 kg scales)
5. If any measured weight does not fall within acceptable range, scale has failed test and must be recalibrated. 6. If the scale passes all four tests, the calibration is successful. 7. Now the calibration seals, if required by local laws, can be placed and the Calibration Switch secured before reassembling the unit.
Page 102
5-16
Chapter 5: Calibration
Securing the Calibration Switch Most countries require that the Calibration Switch and Scale Board be secured before a unit can be put into operation. Some states and jurisdictions permit the use of the Audit Trail in place of the physical seals. Check with your local inspector or governmental agency about using the Audit Trail. This procedure begins with the unit already disassembled with Tower Cabinet in place and access to the Calibration Switch.
Prior Steps Calibration Switch accessible. See Section 2: Accessing the Calibration Switch.
Tools Needle-nosed Pliers Lead/Wire Seal Lead/Wire Seal Press Film/Paper Seal
P/N: 603-8001097 P/N: 603-9000157 Obtain locally – must meet local government requirements
Instructions 1. Install Scale Board Shield. a)
Position Shield so it encloses edge of Load Cell Shield, if present.
b)
Insert tabs of Shield into slots.
c)
Place Shield into position and attach thumbscrew. Tighten with pliers if necessary.
2. Attach seal and/or wire tag to control access to Calibration Switch.
Page 103
Chapter 5: Calibration
5-17
Note: When you perform a scale certification, someone must attach a seal to the Calibration Switch Security Cover. Depending on your local laws, Weights and Measures officials may be required to attach the seal. 3. Install Subplate Assembly. a)
Position Subplate with front end up to near vertical.
b)
Push back edge of Subplate under edge of Tower Cabinet.
c)
Gently press Subplate into position on unit chassis. Ensure Retaining Tab snaps into slot.
d)
Install Cover/Lift Strap in access hole. Tighten screw and snap cover into place.
4. Set Top Plate on rubber bumpers of Scale Hinge Assembly. Verify bumpers are seated in sockets. 5. Apply power to 7870. 6. Verify that display indicates Scale Ready 0.00 lb (0.000 kg). 7. Record the Cal and PAr values, shown on the display, on the Placed Into Service form if required by your local Weights and Measures regulations.
Page 104
5-18
Chapter 5: Calibration